BD VACUTAINER RAPID SERUM TUBE PLUS BLOOD COLLECTION TUBE, MODEL 368771

K091292 · Becton, Dickinson and Company (BD) · JKA · Aug 25, 2009 · Clinical Chemistry

Device Facts

Record IDK091292
Device NameBD VACUTAINER RAPID SERUM TUBE PLUS BLOOD COLLECTION TUBE, MODEL 368771
ApplicantBecton, Dickinson and Company (BD)
Product CodeJKA · Clinical Chemistry
Decision DateAug 25, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1675
Device ClassClass 2

Intended Use

The BD Vacutainer® Rapid Serum Tube Plus Blood Collection Tube (BD RST) is a single use tube used to collect, transport, separate, and process venous blood specimens to obtain serum for chemistry determinations for in vitro diagnostic use. It is used in settings where a venous blood sample is collected by a trained healthcare worker. The BD RST is not recommended for patients on heparin therapy, direct thrombin inhibitor therapy or with Factor I deficiency.

Device Story

BD Vacutainer® Rapid Serum Tube Plus (BD RST) is a sterile, single-use, evacuated plastic blood collection tube with rubber stopper. Designed for rapid serum preparation; clots blood specimens in 5 minutes to facilitate faster turnaround times for chemistry, serology, and immunology testing. Used by trained healthcare workers in clinical settings for venous blood collection. Operates as a closed system for specimen collection, transport, and processing. Healthcare providers use the resulting serum for diagnostic determinations. Benefits include reduced processing time compared to standard serum tubes.

Clinical Evidence

Clinical testing compared blood collected in BD RST versus the predicate BD SST™ tube. Testing included a battery of routine and special chemistry analytes, as well as selected serology (anti-CMV IgG, anti-CMV-IgM, CRP) and immunology (C3, C4, IgG1, IgM, Rf) analytes. Results demonstrated substantial equivalence in performance. The 5-minute clotting time was successfully verified.

Technological Characteristics

Sterile, single-use, evacuated plastic tube with rubber stopper. Closed system design. Additive-based rapid clotting mechanism (5 minutes).

Indications for Use

Indicated for venous blood specimen collection, transport, separation, and processing to obtain serum for chemistry determinations in patients requiring diagnostic testing. Not for patients on heparin therapy, direct thrombin inhibitor therapy, or with Factor I deficiency.

Regulatory Classification

Identification

A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(K) SUMMARY OF SAFETY AND EFFECTI #### 1. General Information This Summary of Safety and Effectiveness information is being submitted in accordance with the requirements of the SMDA of 1990 and 21 § 807.92 #### Establishment: - Address: . - Registration Number: . Contact Person: . BD Diagnostics, Preanalytical Systems 1 Becton Drive, MC300 Franklin Lakes, NJ 07417 1024879 291297 Mary Ann Alsberge Regulatory Affairs Specialist BD Diagnostics, Preanalytical Systems Telephone No .: (201) 847-3103 Fax No. (201) 847-4858 Date of Summary: . #### Device . - BD Vacutainer® Rapid Serum Tube Plus Trade Name: Blood Collection Tube May 1, 2009 - JKA (Tubes, vials, systems, serum separators, Classification Name: . blood collection) - Class II Classification: . - None Established under 514 of the Food, Drug Performance Standards: ● and Cosmetic Act Safety and Effectiveness Information Supporting the Substantial Equivalence Determination 2. > Device Description: BD Vacutainer® Tubes are sterile, single-use, evacuated plastic blood collection tubes with rubber stoppers that provide a means of collecting, transporting, separating, and processing blood in a closed tube. The BD Vacutainer® Rapid Serum Tube Plus Blood Collection Tube (BD RST) clots blood specimens much faster than other additives (5 minutes), and therefore is desirable when a fast turn-around-time is necessary. {1}------------------------------------------------ #### > Intended Use: The BD RST is a single used to collect, separate, transport and process venous blood specimens to obtain serum for chemistry determinations for in vitro diagnostic use. It is used in settings where a venous blood sample is collected by a trained healthcare worker. The BD RST is not recommended for patients on heparin therapy, direct thrombin inhibitor therapy or with Factor I deficiency. #### > Claims: - BD RST tube clots blood in 5 minutes. . - BD RST tube demonstrates substantially equivalent performance to the predicate . BD Vacutainer® SST™ Plus Blood Collection tube (BD SST™) for the collection, separation, transport, and processing of venous blood specimens for chemistry determinations requiring serum for in vitro diagnostic use. - . BD RST tube demonstrates substantially equivalent performance to the predicate BD SST™ tube for the following selected serology analytes: anti-CMV IgG, anti-CMV-IgM, and CRP. - BD RST tube demonstrates substantially equivalent performance to the predicate . BD SST™ tube for the following selected immunology analytes: C3, C4, IgG1 IgM, and Rf. - BD RST tube demonstrates substantially equivalent performance to the predicate . BD SST™ tube for 24 hr stability for analytes evaluated except LDL (19 hours) and Triglycerides (6 hours). #### > Synopsis of Test Methods and Results Clinical testing was performed on blood collected in both the evaluation and predicate tubes for a battery of routine and special chemistry analytes, and selected serology and immunology analytes. Clotting of blood within 5 minutes was demonstrated in the BD RST. The test results demonstrated that the evaluation device's performance was substantially equivalent to the legally marketed predicate device for tested analytes. Based on a comparison of the device features, materials, and intended use, the BD RST is substantially equivalent to the commercially available predicate device, the BD SST™ tube. Mary Ann Alsberge Regulatory Affairs Specialist BD Diagnostics, Preanalytical Systems Date {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus. Becton Dickinson & Company c/o Ms. Mary Ann Alsberge Regulatory Affairs Specialist 1 Becton Drive, MC 300 Franklin Lakes, NJ 07417 AUG 2 5 2009 Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002 Re: k091292 Trade Name: BD Vacutainer® Rapid Serum Tube Plus Blood Collection Tube Regulation Number: 21 CFR §862.1675 Regulation Name: Blood specimen collection devices. Regulatory Class: Class II Product Codes: JKA Dated: July 21, 2009 Received: July 22, 2009 Dear Ms. Alsberge: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820). {3}------------------------------------------------ Page 2 If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. G.C.H. Courtney C. Harper, Ph.D. Acting Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### Indication for Use ### 510(k) Number (if known): # k091292 Device Name: BD Vacutainer® Rapid Serum Tube Plus Blood Collection Tube Indication For Use: The BD Vacutainer® Rapid Serum Tube Plus Blood Collection Tube (BD RST) is a single use tube used to collect, separate, transport and process venous blood specimens to obtain serum for chemistry determinations for in vitro diagnostic use. It is used in settings where a venous blood sample is collected by a trained healthcare worker. The BD RST is not recommended for patients on heparin therapy, direct thrombin inhibitor therapy or with Factor I deficiency. Prescription Use X (21 CFR Part 801 Subpart D) And/Or Over the Counter Use (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD) Carol Benson Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K091292
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...