Bevel Up Holder

K193074 · Nsp Tech Pte, Ltd. · JKA · Apr 21, 2020 · Clinical Chemistry

Device Facts

Record IDK193074
Device NameBevel Up Holder
ApplicantNsp Tech Pte, Ltd.
Product CodeJKA · Clinical Chemistry
Decision DateApr 21, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1675
Device ClassClass 2

Intended Use

The Bevel Up Holder is used for the collection of blood collection tubes in routine venipuncture procedures. It is used with blood collection sets, needles and tubes. The device is to be used by trained healthcare professional only.

Device Story

Bevel Up Holder is a single-use, non-invasive, clear polypropylene tube holder for venipuncture. Device interfaces with blood collection needles and sets (e.g., Becton Dickinson Eclipse, Vacutainer Multi Sample, Safety Lok) at one end and blood collection tubes at the other. Operated by trained healthcare professionals in clinical settings. Connecting end allows rotation to adjust venous access device position. Facilitates secure connection between needle/set and collection tube; ensures proper fit and prevents leakage during blood draw. Benefits include standardized, reliable blood specimen collection.

Clinical Evidence

Bench testing only. Simulated blood collection performed with 20 units per interface device type (Eclipse Needle, Vacutainer Multi Sample, Safety Lok). Evaluated for fit, leakage, and function; all met acceptance criteria. Additional testing included deformation (5kg force), insertion/evacuation force, connectivity, and accelerated aging (ASTM F1980-2007). Biocompatibility leveraged from K181743.

Technological Characteristics

Material: Polypropylene. Form factor: Clear plastic holder. Connectivity: Mechanical interface for blood collection needles/sets and tubes. Sterilization: Not sterilized. Standards: ASTM D4169-16 (shipping), ISO 14971:2007 (risk management), ASTM F88/F883-15 (seal strength), ASTM F1980-2007 (shelf life).

Indications for Use

Indicated for use by trained healthcare professionals for blood collection into tubes during routine venipuncture procedures using blood collection sets and needles.

Regulatory Classification

Identification

A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a larger font size than the rest of the name. The words "U.S. FOOD & DRUG" are stacked on top of the word "ADMINISTRATION". April 21, 2020 NSP Tech Pte Ltd HC Tan Operations Manager 10 Admiralty Street, Northlink Building, #02-06 757695 SINGAPORE Re: K193074 Trade/Device Name: Bevel Up Holder Regulation Number: 21 CFR 862.1675 Regulation Name: Blood Specimen Collection Device Regulatory Class: Class II Product Code: JKA Dated: March 2, 2020 Received: March 4, 2020 ### Dear HC Tan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. for Tina Kiang Ph.D. Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K193074 Device Name BEVEL UP HOLDER Indications for Use (Describe) The Bevel Up Holder is used for the collection of blood collection tubes in routine venipuncture procedures. It is used with blood collection sets, needles and tubes. The device is to be used by trained healthcare professional only. Type of Use (Select one or both, as applicable) | <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | |------------------------------------------------------------------------------------| | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### l. SUBMITTER: ## Applicant Name: NSP Tech Pte Ltd 10 Admiralty Street, Northlink Building #02-06, Singapore 757695 Phone: (65) 67478177 ## Contact Person: Name: HC Tan Phone: (65) 98180450 Fax: (65) 67476533 Email: hockchoon.tan@nsptech.com.sg ## Date Prepared: 20th February 2020 #### II. DEVICE | Name | : | Bevel Up Holder | |---------------------|---|----------------------------------| | Trade Name | : | BUH | | Common Name | : | Blood Collection Tube Holder | | Classification Name | : | Blood Specimen Collection Device | | Regulatory code | : | 21 CFR 862.1675 | | Regulatory Class | : | Class II | | Product Code | : | JKA | #### lll. PREDICATE DEVICE: | Trade Name | : | VACUETTE Blood Culture Holder | |---------------|---|-------------------------------| | 510(K) Number | : | K122687 | | Common Name | : | Blood Collection Tube Holder | {4}------------------------------------------------ | Classification Name | Blood Specimen Collection Device | |---------------------|----------------------------------| | Regulatory code | 21 CFR 862.1675 | | Regulatory Class | Class II | | Product Code | JKA | #### DEVICE DESCRIPTION IV. The Bevel Up Holder device is a single use and non-invasive device to be used in venipuncture procedures for the collection of blood specimens. It is to be used as a Tube Holder with blood collection tubes such as the Becton Dickinson Vacutainer® blood collection tube. The Bevel Up Holder device is also used with blood collection needles and blood collection sets such as the Becton Dickinson Eclipse Needle set, Becton Dickinson Vacutainer Multi Sample and Becton Dickinson Safety Lok Blood Collection Set for the collection of blood specimen into the blood collection tubes. The Bevel Up Holder is to be used by trained healthcare professional only and is intended for prescription use. The Bevel Up Holder has a clear plastic holder body with an opening on one end for the insertion of blood collection tube and an interface device connecting feature on the other end for venous access device for connecting to blood collection needle and blood collection set. The connecting end of the venous access device of the Bevel Up Holder (BUH) device allows the adjusting of the position of the venous access device with rotating it. #### V. INDICATIONS FOR USE STATEMENT The Bevel Up Holder is used for the collection of blood collection tubes in routine venipuncture procedures. It is used with blood collection sets, needles and tubes. The device is to be used by trained healthcare professional only. #### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE Information was submitted to demonstrate that there are no significant differences in technological characteristics between the Bevel Up Holder (BUH) and the predicate device. The following comparison chart shows that the subject device and the predicate device are substantially equivalent: ### TABLE OF COMPARISON BETWEEN SUBJECT DEVICE AND PREDICATE DEVICE {5}------------------------------------------------ | Device Name | NSP Tech Pte Ltd<br>Bevel Up Holder (BUH) | Greiner Bio One<br>VACUETTE® Blood Culture Holder | |----------------|-------------------------------------------|---------------------------------------------------| | 510(k) Number | K193074 | K122687 | | Classification | Class II | Class II | | Product Code | JKA | JKA | | Device Name | NSP Tech Pte Ltd<br>Bevel Up Holder<br>(BUH) | Greiner Bio One<br>VACUETTE® Blood<br>Culture Holder | Comparison | |--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications<br>For Use | The Bevel Up<br>Holder is used for<br>the collection of<br>blood into blood<br>collection tubes in<br>routine<br>venipuncture<br>procedures. It is<br>used with blood<br>collection sets,<br>needles and<br>tubes. The device<br>is to be used by<br>trained healthcare<br>professional only. | The Blood Culture<br>Holder is used to fill<br>blood into blood<br>culture bottles and<br>tubes in routine<br>venipuncture<br>procedures. This<br>device is to be used<br>by properly trained<br>healthcare<br>professionals only<br>in accordance with<br>these instructions. | Both the Bevel<br>Up Holder and<br>the predicate<br>device are used<br>for collection of<br>blood into<br>collection tubes<br>carried out in<br>routine<br>venipuncture<br>procedures by<br>trained<br>healthcare<br>professional.<br>Hence, they are<br>similar in the<br>Indication For<br>Use. | | Manufacturer | NSP Tech Pte Ltd | Greiner Bio One | N/A | | Product<br>Configuration | Clear plastic<br>Holder | Clear plastic Holder | Same | | Material | Polypropylene | Polypropylene | Same | | Sterility | Not sterilized | Not sterilized | Same | {6}------------------------------------------------ | Packaging | Bulk pack in poly<br>bag | Bulk pack in poly<br>bag. | Same | |----------------------|---------------------------------------------------------------------------|---------------------------|-----------| | Use | Single-Use Only | Single-Use Only | Same | | Bio<br>Compatibility | Leveraged from<br>Blood Transfer<br>Accessory<br>(K181743)10993<br>series | ISO 10993 series | Same Same | | Shelf Life | 3 years | 3 years | Same | ### Similarities: The Bevel Up Holder device has similarities with the predicate device in areas such as Indications For Use, Product configuration, Material used, Principle of operations of safety feature and technological characteristics. The safety and effectiveness of the two devices are similar. ### Differences: The Bevel Up Holder device differs from the predicate device in the design where the latter has a bluish appearance for the color of the device. #### PERFORMANCE DATA VII. Simulated blood collection was performed on the Bevel Up Holder using the respective interface devices each time which is the - . Becton Dickinson Eclipse Needle Set - . Becton Dickinson Vacutainer Multi Sample - . Becton Dickinson Safety Lok Blood Collection Set For each of the above interface device, the test involves using a new Bevel Up Holder connected to a new set of interface device each time and the blood collection process carried out. This is done with 20 sets of the subject device and interface device. The results were evaluated for proper fit, leakage and function of the devices. All Bevel Up Holders met the acceptance criteria for fit, leakage and function when used for blood collection. The device meets the functional and performance requirements, underwent testing in accordance with the following recognized consensus standards: - ASTM D4169-16: Standard practice for performance testing of shipping . containers and systems. - ISO 14971;2007: Medical Devices-Application of Risk management of medical ● devices - ASTM F88/F883-15: Standard test method for seal strength of flexible barrier . materials {7}------------------------------------------------ In addition, the following performance testing was carried out on the Bevel Up Holder. ### · Deformation Test The device was tested by holding it in place with a fixture and a 5 kg force is applied on the device for a period of 5 minutes. Any permanent change in outer diameter of the device before and after compression by the applied force is measured and recorded. This test assesses the quality of the device. Result of the test shows no permanent change in the outer diameter of the Bevel Up Holder (BUH). ### • Insertion and Evacuation Force Test The device is held in place by a fixture and the force taken to insert or evacuate a blood collection tube into the device is recorded and result shows the device meeting the defined acceptance criteria in the insertion and evacuation of the blood collection tube into the Bevel Up Holder (BUH) for blood collection operation. ### • Connectivity Test The device was tested for connectivity by connecting them to the interface devices such as Becton Dickinson Eclipse Needle Set, Becton Dickinson Vacutainer Multi Sample and Becton Dickinson Safety Lok Blood Collection Set for the test with each of the respective devices. The blood collection operation is then simulated and result shows no leakage with good fit and function for the connectivity of the Bevel Up Holder (BUH) to the interface devices. ### · Test: Shelf Life An Accelerated Aging Test is done with reference to ASTM F1980-2007 is conducted for the device to simulate the defined years of shelf life. Result shows that the Bevel Up Holder (BUH) meeting the defined shelf life with device still functioning well. ### · Bio Compatibility The Bevel Up Holder device in its finished form is identical to Blood Transfer Accessory (K181743) in formulation (material), processing, and geometry and no other chemicals have been added (e.g., plasticizers, fillers, additives, cleaning agents, mold release agents).Leveraging on the bio compatibility result of this similar identical device, the Bevel Up Holder (BUH) is deemed to have met the biocompatibility requirements for Cytotoxicity, Sensitization and Irritation Reactivity. ### · Sterilization The device is not required to be sterilized. #### SUBSTANTIAL EQUIVALENCE CONCLUSION VIII. The result of the non-clinical testing demonstrates that the Bevel Up Holder (BUH) is substantially equivalent to the predicate device. VACUETTE® Blood Culture Holder. Hence, we conclude that based on the information above, the Bevel Up Holder (BUH) is substantially equivalent to the predicate device, VACUETTE® Blood Culture Holder.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...