K122687 · Greiner Bio-One North America, Inc. · JKA · Jan 10, 2013 · Clinical Chemistry
Device Facts
Record ID
K122687
Device Name
BLOOD CULTURE HOLDER
Applicant
Greiner Bio-One North America, Inc.
Product Code
JKA · Clinical Chemistry
Decision Date
Jan 10, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1675
Device Class
Class 2
Indications for Use
The Blood Culture Holder is used to fill blood into blood culture bottles and tubes in routine venipuncture procedures. This device is to be used by properly trained healthcare professionals only in accordance with these instructions. The SAFETY Blood Collection Set and/or Blood Collection Set + Blood Culture Holder are used in routine venipuncture procedures. The winged needle of the SAFETY Blood Collection Set is designed with a safety shield, which can be activated to cover the needle immediately following blood collection to aid in the protection against accidental needlestick injury. The Blood Culture Holder is used to fill blood into blood culture bottles and tubes.
Device Story
Device is a single-use, non-sterile plastic cylindrical holder; designed to interface with male luer adapters and blood collection sets. Operates as an external housing for VACUETTE Multiple Sample Luer Adapters; allows connection of blood collection needles to evacuated blood culture bottles or tubes. Used in clinical settings by trained healthcare professionals for routine venipuncture. Does not contact blood directly; fluid path is contained within the attached luer adapter and collection set. Provides mechanical support for blood collection containers; aids in safe, closed-system blood transfer. When supplied as a system with a collection set, the entire assembly is sterilized to maintain fluid path integrity.
Clinical Evidence
Bench testing only. Simulated blood draw studies were conducted using 20 units of the Blood Culture Holder connected to Safety Blood Collection Sets. Testing evaluated compatibility with various blood culture bottles, vials, and evacuated tubes (13x75mm to 16x100mm). Endpoints included proper fit, leakage, and functional performance during blood collection. All samples met acceptance criteria.
Technological Characteristics
Plastic cylindrical holder; threaded for compatibility with male luer adapters. Non-sterile (unless supplied as part of a system). Tested for pyrogenicity per ISO-10993-11 (rabbit pyrogen test) and endotoxin limits per LAL Turbidimetric Method. Mechanical interface only; no energy source or software.
Indications for Use
Indicated for use by trained healthcare professionals during routine venipuncture procedures to facilitate the collection of blood into blood culture bottles and evacuated tubes.
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
Predicate Devices
Smiths Medical ASD Inc. Saf-T Closed Blood Collection System® (K081229)
Reference Devices
Greiner Safety Blood Collection Set with Luer Adapter (K011786)
Submission Summary (Full Text)
{0}------------------------------------------------
# K122687 510(K) SUMMARY
### December 7. 2012
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is:
CONTACT:
Manfred Abel Greiner Bio-One North America, Inc. P.O Box 1026 Monroe, NC 28111
NAME OF DEVICE: Trade Name:
Common Names/Descriptions: Classification Name:
Product Code:
Class:
Blood Culture Holder; Safety Blood Collection Set + Luer Adapter with preattached Blood Culture Holder
Tube Holder
21 CFR 862.1675 Blood specimen collection device
JKA
ll
PREDICATE DEVICE:
Smiths Medical ASD Inc. Saf-T Closed Blood Collection System® (K081229)
# DEVICE DESCRIPTION: INTENDED USE:
Blood Culture Holder:
The Blood Culture Holder is used to fill blood into blood culture bottles and tubes in routine venipuncture procedures. This device is to be used by properly trained healthcare professionals only in accordance with these instructions.
Safety Blood Collection Set and/or Blood Collection Set + Blood Culture Holder:
The SAFETY Blood Collection Set and/or Blood Collection Set + Blood Culture Holder are used in routine venipuncture procedures. The winged needle of the SAFETY Blood Collection Set is designed with a safety shield, which can be activated to cover the needle immediately
{1}------------------------------------------------
following blood collection to aid in the protection against accidental needlestick injury. The Blood Culture Holder is used to fill blood into blood culture bottles and tubes.
### PRODUCT DESCRIPTION:
The Blood Culture Holder is a single-use, non-sterile, plastic holder with threads which are compatible with male luer adapters. The holder can be used with all VACUETTE® Multiple Sample Luer Adapters and Safety Blood Collection Sets for routine venipuncture into various manufacturers' evacuated blood culture bottles and tubes. It is supplied alone or with the Greiner sterile Safety Blood Collection Set with Luer Adapter.
The Blood Culture Holder does not come into contact with the blood specimen. It is an external holder for the Safety Blood Collection Set and the VACUETTE® Multiple Sample Luer Adapter. The Safety Blood Collection Set consists of tubing with a blood collection needle at one end and a female luer at the other. The VACUETTE® Multiple Sample Luer Adapter consists of a male luer at one end and a cannula covered with a sheath. The luer adapter is threaded into the Blood Culture Holder, with the male luer end on the outside and the covered cannula on the inside of the holder barrel. The blood flows through the blood collection needle and tubing of the Safety Blood Collection Set, into the male luer of the VACUETTE® Multiple Sample Luer Adapter, through the luer into the cannula, and into the blood culture tube or holder when punctured by the cannula. The fluid path consists of the Safety Blood Collection Set and the VACUETTE® Multiple Sample Luer Adapter. The Blood Culture Holder never comes into contact with the blood specimen during conditions of use.
When the Blood Culture Holder is supplied attached to the Safety Blood Collection Set and the VACUETTE® Multiple Sample Luer Adapter, it undergoes sterilization as part of the entire system because the fluid path must be sterilized. However, when the Blood Culture Holder is provided alone, it is not sterilized and the user attaches their own sterile blood collection needle and luer adapter.
The Safety Blood Collection Set with Blood Culture Holder is non-pyrogenic and tested according to ISO-10993-11 rabbit pyrogen test. The devices also are below the limit for endotoxin and tested according to ISO-10993-11 LAL Turbidimetric Method.
Caution: US Federal law restricts this device for sale by or on the order of a physician.
{2}------------------------------------------------
# SUBSTANTIAL EQUIVALENCE:
The Blood Culture Holder is substantially equivalent to the Smiths Medical ASD Inc. Saf-T Closed Blood Collection System® (K081229) in intended use and design. The similarities and differences are listed below.
| | Greiner GBO<br>Blood Culture Holder | Smiths Medical ASD Inc.<br>Saf-T Closed Blood Collection System®<br>(predicate device) |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | Blood Culture Holder: K122687<br>Safety Blood Collection Set with<br>Luer Adapter: K011786 | K081229 |
| Intended Use | The Blood Culture Holder is used to<br>fill blood into blood culture bottles<br>and tubes in routine venipuncture<br>procedures. This device is to be<br>used by properly trained healthcare<br>professionals only in accordance<br>with these instructions. | The Saf-T Closed Blood<br>Collection System® device is<br>intended for use as a direct<br>blood draw device into a<br>vacuum tube or to allow a<br>syringe blood draw and<br>transfer to fill vacuum tubes. |
| Device<br>Description | Blood culture holder | Blood culture device |
| Device Design | Cylindrical holder for vacuum bottle<br>and tube placement : cylinder with<br>threads which are compatible with<br>male luer adapters<br><br>The design of the holder allows<br>collection of blood into blood culture<br>bottles and blood collection tubes. | Cylindrical holder for vacuum<br>bottle and tube placement :<br>cylinder with threads which<br>are compatible with blood<br>collection needles and male<br>luer adapters |
| Cylinder | Plastic | Plastic |
## PRODUCT PERFORMANCE:
Simulated blood draw testing was performed on the Safety Blood Collection Set with Luer Adapter and the Blood Culture Holder using a variety of blood culture bottles and evacuated tubes representing the range of bottle and tube sizes. The devices used in the study were:
- Blood collection devices: Greiner Safety Blood Collection Set with Luer . Adapter; Blood Culture Holders;
- Blood culture bottles and vials (2 manufacturers); .
- Blood collection tubes: VVACUETTE® EDTA K2, 4 mL, 13x75mm
(smallest diameter cap and tube size) and VACUETTE® Z Serum .
{3}------------------------------------------------
Separator and Clot Activator, 7 mL, 16x100mm (largest diameter cap and tube size).
Twenty Blood Culture Holders were connected to 20 Safety Blood Collection . Sets with Luer Adapters. The sets were tested in simulated blood draws with blood culture bottles, blood culture vials, and the two tube types.
The results were evaluated for proper fit of the Blood Culture Holders to the Safety Blood Collection Sets, leakage and function of the blood culture containers, and leakage and function of the blood collection tubes. All Blood Culture Holders met the acceptance criteria for fit, leakage, and function when connected to the Safety Blood Collection Sets and when used for blood collection into blood culture bottles and tubes.
Conclusion: The Blood Culture Holder can be used with blood culture bottles, blood culture vials, and the range of blood collection tube sizes from 13x75mm to 16x100mm.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, a traditional symbol associated with medicine and healthcare.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### January 10, 2013
Greiner Bio-One North America, Incorporated C/O Ms. Judith Smith Principal Judi Smith, Limited Liability Company P.O. Box 103 BALDWIN MD 21013
Re: K122687
Trade/Device Name: Blood Culture Holder Regulation Number: 21 CFR 862.1675 Regulation Name: Blood Specimen Collection Device Regulatory Class: II Product Code: JKA Dated: December 17, 2012 Received: December 20, 2012
#### Dear Ms. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Smith
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony Lo Re
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### INDICATIONS FOR USE
510(k) Number (if known): K122687
Device Name: Blood Culture Holder
Indication For Use:
The Blood Culture Holder is used to fill blood into blood culture bottles and tubes in routine venipuncture procedures. This device is to be used by properly trained healthcare professionals only in accordance with these instructions.
The SAFETY Blood Collection Set and/or Blood Collection Set + Blood Culture Holder are used in routine venipuncture procedures. The winged needle of the SAFETY Blood Collection Set is designed with a while shield, which can be activated to cover the needle immediately following blood collection to aid in the protection against accidental needlestick injury. The Blood Culture Holder is used to fill blood into blood culture bottles and tubes.
Prescription Use × (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Weihong Gu -S c=US, o=U.S. Government, ou=HHS, ou=FDA, ou=People, cn=Weihong Gu -S, 0.9.2342.19200300.100.1.1=2000380818 2013.01.11 11:49:54 -05'00'
Division Sign-Off Office of Device Evaluation
510(k) K122687
Page 1 of 1
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.