GREINER VACUTTE BLOOD COLLECTION TUBE W/ CLOT ACTIVATOR & GEL SEPARATOR
Device Facts
| Record ID | K960858 |
|---|---|
| Device Name | GREINER VACUTTE BLOOD COLLECTION TUBE W/ CLOT ACTIVATOR & GEL SEPARATOR |
| Applicant | Greiner America, Inc. |
| Product Code | JKA · Clinical Chemistry |
| Decision Date | Aug 30, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1675 |
| Device Class | Class 2 |
Intended Use
The product is intended for use in holding and separating blood serum from the cellular components of blood.
Device Story
Evacuated blood collection tube; contains silicon dioxide clot activator and inert polymeric gel separator; used to collect venous blood; centrifugation causes gel to form physical barrier between serum and cellular components; enables serum isolation for laboratory analysis; used by phlebotomists and clinical laboratory personnel; facilitates diagnostic testing by providing stable serum samples.
Clinical Evidence
Bench testing only. Comparative study of 41 analytes using paired samples collected in Greiner Vacuette™ and Becton Dickinson Vacutainer® tubes. Results demonstrated good correlation with coefficients ranging from 0.864 to 1.000.
Technological Characteristics
Evacuated blood collection tube; clear plastic tube body; silicon dioxide clot activator; inert polymeric gel separator; stopper material consistent with predicate.
Indications for Use
Indicated for use in clinical settings for the collection of blood samples and the subsequent separation of serum from cellular components.
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
Predicate Devices
- Becton Dickinson Vacutainer® SST tubes (K921806)
Related Devices
- K023331 — BD VACUTAINER SST II PLUS TUBES · Becton, Dickinson & CO · Nov 25, 2002
- K093910 — IMPROVACUTER GEL & CLOT ACTIVATOR TUBE · Guangzhou Improve Medical Instruments Co., Ltd. · Jul 12, 2010