EME SELF MONITORING BLOOD GLUCOSE SYSTEM, EME PRO SELF MONITORING BLOOD GLUCOSE SYSTEM

K112901 · Eps Bio Technology Corp. · LFR · Dec 23, 2011 · Clinical Chemistry

Device Facts

Record IDK112901
Device NameEME SELF MONITORING BLOOD GLUCOSE SYSTEM, EME PRO SELF MONITORING BLOOD GLUCOSE SYSTEM
ApplicantEps Bio Technology Corp.
Product CodeLFR · Clinical Chemistry
Decision DateDec 23, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Indications for Use

The EME Self Monitoring Blood Glucose System is intended for the quantitative measurement of glucose in fresh capillary whole blood from fingertip, palm, or forearm. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by a single patient with diabetes and should not be shared, as an aid to monitor the effectiveness of diabetes control. The system is not to be used on neonates, nor for the diagnosis of, or screening for diabetes mellitus. Alternative site testing can be only used during steady-state blood glucose conditions. The system consists of the EME meter and the EME test strips. The EME meter only is used with the EME test strips to quantitatively measure glucose in fresh capillary whole blood from fingertip, palm, or forearm. The EME Glucose Control Solution For use with EME Blood Glucose Self Monitoring System as a quality control check to verify the accuracy of blood glucose test results. The EME Pro Self Monitoring Blood Glucose System is intended for the quantitative measurement of glucose in venous whole blood or fresh capillary whole blood from fingertip. Testing is done outside the body (In Vitro diagnostic use). It is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control. The system is only used with single-use lancing devices. The system is not to be used on neonates, nor for the diagnosis of, or screening for diabetes mellitus. The system consists of the EME Pro meter and the EME Pro test strips. The EME Pro meter only is used with the EME Pro test strips to quantitatively measure glucose in venous whole blood or fresh capillary whole blood from fingertip. The EME Glucose Control Solution For use with the EME Pro Blood Glucose Self Monitoring System as a quality control check to verify the accuracy of blood glucose test results.

Device Story

The EME and EME Pro systems are blood glucose monitoring devices. They utilize test strips to measure glucose concentrations in fresh capillary (fingertip, palm, forearm) or venous whole blood. The system consists of a meter and test strips, supported by a control solution for accuracy verification. The EME system is designed for single-patient home use, while the EME Pro is intended for multiple-patient use in professional clinical settings. Users apply a blood sample to the test strip, which the meter analyzes to provide a quantitative glucose reading. This output assists healthcare providers and patients in monitoring diabetes control. The devices are not intended for neonatal use or diabetes diagnosis.

Clinical Evidence

No clinical data. Bench testing only. Disinfection efficacy studies demonstrated complete inactivation of Hepatitis B Virus (HBV) using PDI Super SANI-CLOTH Germicidal wipes. Robustness testing confirmed no performance degradation after 20,000 disinfection cycles for the meter and 14,600 cycles for the lancing device.

Technological Characteristics

Glucose test system (21 CFR 862.1345). Consists of a meter, test strips, and control solution. Measures glucose in capillary or venous whole blood. Intended for in vitro diagnostic use.

Indications for Use

Indicated for single patient home use (EME Self Monitoring Blood Glucose System) and multiple patient use in a professional healthcare setting (EME Pro Self Monitoring Blood Glucose System) for blood glucose monitoring.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Submission Summary (Full Text)

{1} SPECIAL 510(k): Device Modification OIVD Review Memorandum (Decision Making Document is Attached) To: THE FILE RE: DOCUMENT NUMBER: k112901 This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary): 1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.) EG V1(BL) Self Monitoring Glucose Test System and EG V1 Pro Self Monitoring Glucose Test System k101037. 2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use). 3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. This change was for: A. The Physical Appearance of the meter has changed. The dimensions of the meter were modified from $3.5 \times 2.1 \times 0.9$ inches to $3.8 \times 2.0 \times 0.6$ inches. B. The trade name of the meter has changed from EG V1(BL) Self Monitoring Glucose Test System and EG V1 Pro Self Monitoring Glucose Test System to EME Self Monitoring Blood Glucose System and EME Pro Self Monitoring Blood Glucose System respectively. C. Power type changed from $2 \times$ CR2032 Lithium to $2 \times$ AAA Lithium Battery. D. Addition of a strip eject button on the side of the meter. 4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, analytes and performance characteristics. 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis. b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied c) A declaration of conformity with design controls. The declaration of conformity should include: i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review. 6. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure (and Class III Summary for Class III devices). The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device. {2} 2 # Reviewer's Comments: The device is intended for single patient home use (EME Self Monitoring Blood Glucose System) and multiple patient use in a professional healthcare setting (EME Pro Self Monitoring Blood Glucose System). Disinfection efficacy studies were performed on the materials comprising the meter and lancing device by an outside commercial laboratory testing services demonstrating complete inactivation of hepatitis B Virus (HBV) with PDI® Super SANI-CLOTH® Germicidal disposable wipes (EPA Reg. No: 9480-4). The sponsor also conducted robustness studies and demonstrated that there was no change in performance or in the external materials of the meter and lancing device (for single patient use only) after 20,000 disinfecting cycles for the meter and 14,600 disinfection cycles for the lancing device which represents 4 years of cleaning after each use (up to 9 times per day) and disinfection once a week by lay users. The meter (EME Pro Self Monitoring Blood Glucose System) was validated for multiple-patient use with 20,000 disinfecting cycles (each cycle tested consisted of one pre-clean wipe and one disinfecting wipe) to represent 3 years of cleaning and disinfection after each use.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...