ASU-1000 DIAGNOSTIC ULTRASOUND TRANSDUCER

K974544 · Aloka Co., Ltd. · ITX · Mar 18, 1998 · Radiology

Device Facts

Record IDK974544
Device NameASU-1000 DIAGNOSTIC ULTRASOUND TRANSDUCER
ApplicantAloka Co., Ltd.
Product CodeITX · Radiology
Decision DateMar 18, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1570
Device ClassClass 2

Intended Use

The ASU-1000 and Volume mode software, like other marketed diagnostic ultrasound systems and 3-D rendering packages, is indicated for imaging body structures in two and three dimensions to aid in the diagnosis of disease or abnormality.

Device Story

The Aloka ASU-1000 is a piezo-electric diagnostic ultrasound transducer used with Aloka SSD-1700 and SSD-1400 systems. It transmits high-frequency sound waves into the body; reflected energy, based on tissue acoustic impedance, is processed into 2D slice images. The associated 'Volume Mode' software transforms these slices into 3-D surface renderings of target tissues. Used in clinical settings by healthcare professionals to aid in the diagnosis of disease or abnormality. Output is visualized on the ultrasound system display, assisting clinicians in anatomical assessment and diagnostic decision-making. Benefits include enhanced visualization of fetal and maternal structures through 3-D rendering.

Clinical Evidence

Bench testing only. No clinical data provided. Substantial equivalence is based on technological comparison, compliance with electrical/physical safety standards, and evaluation of patient contact materials.

Technological Characteristics

Piezo-electric transducer; 2D slice image acquisition; 3-D surface rendering software; compatible with Aloka SSD-1700 and SSD-1400 systems. Operates via standard ultrasound pulse-echo principles. Complies with electrical and physical safety standards.

Indications for Use

Indicated for diagnostic ultrasound imaging and Doppler analysis of fetal and maternal structures in obstetrics and abdominal applications. Includes 3-D surface rendering of target tissues. Prescription use only.

Regulatory Classification

Identification

A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### 510(k) SUMMARY 1. MAR 1 8 1998 This summary statement complies with 21 CFR, section 807.92 as amended March 14, l ddd i This premarket notification has been submitted by Aloka Co., Ltd. and covers the Aloka ASU-1000 transducer and associated Volume Mode 3-D rendering software. The address is: 10 Fairfield Boulevard Wallingford, CT. 06492 The contact person is Paul D. Smolenski, Manager, Quality and Regulatory Affairs. The proprietary name for the transducer is the Aloka ASU-1000 diagnostic ultrasound transducer. The 3-D rendering software is "Volume Mode". The common name for this type of device is a diagnostic ultrasound transducer and associated accessories. The items in this submission are covered under the following classifications: 90 ITX - Transducer, Ultrasonic, Diagnostic 90 IYN - Ultrasonic pulsed Doppler system and accessories The above as stated in 21 CFR, part 892.1570, and 892.1550 have been classified as regulatory Class II. The Aloka ASU-1000 and Volume mode software are substantially equivalent to several previously marketed 3-D rendering systems for ultrasound including the TomTek Echo-Scan system. The ASU-1000 and Volume mode software can be used on the Aloka SSD-1700 and SSD-1400 diagnostic ultrasound systems. These systems function in the same manner as other diagnostic ultrasound devices. High frequency sound waves are transmitted into the body by a piezo-electric transducer. In the body, differences in the acoustic impedance of different tissues reflect a certain amount of the ultrasound energy back to the transducer, where it is processed into slice images. The slice images are processed by the Volume Mode software into a 3-D surface rendering of the target tissues. The ASU-1000 and Volume mode software, like other marketed diagnostic ultrasound systems and 3-D rendering packages, is indicated for imaging body structures in two and three dimensions to aid in the diagnosis of disease or abnormality. {1}------------------------------------------------ - The ASU-1000 and Volume mode software are indicated for the same diagnostic . ultrasound applications as other products currently marketed. - The ASU-1000 and Volume mode software have the same gray-scale and Doppler . abilities as other products currently offered by Aloka and others. - The ASU-1000 and Volume mode software have the same 3-D rendering abilities as . products currently available. - The ASU-1000 and Volume mode software uses essentially the same technologies for . imaging, Doppler functions and signal processing as other products currently marketed by Aloka and others. - The ASU-1000 and Volume mode have similar methods of use as other products . currently marketed by Aloka and others. - The ASU-1000 acoustic power output levels are below the maximum levels allowed . by the FDA. - The ASU-1000 and Volume mode software are subjected to the same Quality . Assurance systems in development and production as other products currently marketed by Aloka. - The patient contact material used in the ASU-1000 has been evaluated and found to be ● safe for this application. - The ASU-1000 and Volume mode software complies with the same electrical and . physical safety standards as other products currently marketed by Aloka. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure, composed of three curved lines that suggest the head, body, and tail feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird symbol. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAR 1 8 1998 Paul D. Smolenski Manager, Ouality and Regulatory Affairs Aloka Co., Ltd. 10 Fairfield Blvd. Wallingford, CT 06492 K974544 Re: Aloka ASU-1000 Diagnostic Ultrasound Transducer Dated: March 5, 1997 Received: March 6 1997 Regulatory Class: II 21 CFR 892.1570/Procode: 90 ITX Dear Mr. Smolenski: We have reviewed your section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug and Cosmetic Act. You may, therefore, market the device, subject to the general controls provisions Act (Act). The general controls provisions of the Act include requirements for registration, listing of devices, good manufacturing practices, labeling, and prohibitions against misbranding and adulteration. This determination of substantial equivalence applies to the Aloka ASU-1000 Diagnostic Ultrasound Transducers intended for use with the following diagnostic ultrasound systems, as described in your premarket notification: #### System Model Number #### Aloka SSD-1700 Aloka SSD-1400 If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic QS inspections, the FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification does not affect any obligation you may have under sections 53 t and 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. Please be advised that the determination above is based on the fact that no medical devices have been demonstrated to be safe and effective for in vitro fertilization or percutaneous umbilical blood sampling, nor have any devices been marketed for these uses in interstate commerce prior to May 28, 1976, or reclassified into class I (General Controls) or class II (Special Controls). FDA considers devices specifically intended for in vitro fertilization and percutaneous umbilical blood sampling to be investigational, and subject to the provision of the investigational device exemptions (IDE) regulations, 21 CFR, Part 812. Therefore, your product labeling must be consistent with FDA's position on this use. {3}------------------------------------------------ Page 2 - Paul D. Smolenski This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premaket notification" (2) CPR 807.97). Other general information on your responsibilities un Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". If you have any questions regarding the content of this letter, please contact Paul Gammell, Ph.D. at (301) 594-1212. Sincerely yours, Yurid L. Lymm Lillian Yin, Ph.D. Lillian Yin. Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health {4}------------------------------------------------ K974544 ## Ultrasound Device Indications Statement Page 1 of 2 | 510(k) Number (if known): | unknown at submission | |---------------------------|----------------------------------------| | Device Name: | ASU-1000 with Volume Mode 3-D Software | | System: | SSD-1400 | ### Fill our one form for each ultrasound system and transducer Diagnostic Ultrasound imaging and Doppler analysis of the human body as Indications for Use: follows: | | MODES of OPERATION | | | | | | | | | | |----------------------|--------------------|---|---|------|-----|---------------|-------------------|-----|-----------|-----------| | Clinical Application | A | B | M | PW D | CWD | Color Doppler | Amplitude Doppler | CVI | Combined | Other | | Opthalmic | | | | | | | | | | | | Fetal | | ✓ | ✓ | ✓ | | | | | See Below | See Below | | Abdominal | | ✓ | ✓ | ✓ | | | | | See Below | See Below | | Intra-Operative | | | | | | | | | | | | Intra-Operative | | | | | | | | | | | | Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organ | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac Adult | | | | | | | | | | | | Cardiac Pediatric | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transuretheral | | | | | | | | | | | | Intraluminal | | | | | | | | | | | | Peripheral Vessel | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | Combined Modes: B/M, B/PWD Other Indications or Modes:3-D rendering of fetal and maternal structures in Obstetrics. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED.) Concurrence of CDRH, Office of Device Evaluatition (ODE) Daniel Leyson Prescription Use (Per 21 CFR 801.109) (Division Sign-Off) Division of Reproductive, Abdominal, ENT and Radiological I 510(k) Number {5}------------------------------------------------ K974544 # Ultrasound Device Indications Statement Page 2 of 2 unknown at submission 510(k) Number (if known): ASU-1000 with Volume Mode 3-D Software Device Name: SSD-1700 System: ### Fill our one form for each ultrasound system and transducer Diagnostic Ultrasound imaging and Doppler analysis of the human body as Indications for Use: follows: | | MODES of OPERATION | | | | | | | | | | |---------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|-----|-----------|-----------| | Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | CVI | Combined | Other | | Opthalmic | | | | | | | | | | | | Fetal | | ✓ | ✓ | ✓ | | ✓ | ✓ | | See Below | See Below | | Abdominal | | ✓ | ✓ | ✓ | | ✓ | ✓ | | See Below | See Below | | Intra-Operative | | | | | | | | | | | | Intra-Operative<br>Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organ | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac Adult | | | | | | | | | | | | Cardiac Pediatric | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transuretheral | | | | | | | | | | | | Intraluminal | | | | | | | | | | | | Peripheral Vessel | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | Combined Modes: B/M, B/PWD, M/CD, B/PWD/CD Other Indications or Modes:3-D rendering of fetal and maternal structures in Obstetrics. > (PLEASE DQ NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IE NEEDED.) > ---- -- -Concurrence of CDRH, Office of Device Evaluatition (ODE) David le Lynn Division Sign Off ivision Sign-Off) Division of Reproductive, Abdominal, ENT and Radiological De 510(k) Number KG974544 Prescription Use (Per 21 CFR 801.109)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...