DC-3/DC-3T DIAGNOSTIC ULTRASOUND SYSTEM
K091941 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · ITX · Jul 9, 2009 · Radiology
Device Facts
| Record ID | K091941 |
| Device Name | DC-3/DC-3T DIAGNOSTIC ULTRASOUND SYSTEM |
| Applicant | Shenzhen Mindray Bio-Medical Electronics Co., Ltd. |
| Product Code | ITX · Radiology |
| Decision Date | Jul 9, 2009 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1570 |
| Device Class | Class 2 |
| Attributes | Pediatric, 3rd-Party Reviewed |
Intended Use
The device is intended for use by a qualified physician for ultrasound evaluation of gynecology, obstetrics, abdominal, pediatric, small parts (breast, thyroid, testicle, etc), neonatal cephalic, transcranial, cardiac transvaginal, transrectal, peripheral vascular, intraoperative, urology, orthopedial, and musculoskeletal, perficial vascular, exams exams.
Device Story
The DC-3/DC-3T is a mobile, software-controlled diagnostic ultrasound system. It utilizes linear, convex, and phased array probes (2.0–12.0 MHz) to acquire and display images in B-Mode, M-Mode, Color, PW, CW, Power, and DirPower modes. Operated by qualified physicians in clinical settings, the system processes acoustic signals to generate real-time diagnostic images. It includes advanced imaging features such as Tissue Harmonic Imaging, Smart3D, iScape, and iBeam. The output is displayed for the clinician to perform diagnostic evaluations across various anatomical regions. The system aids in clinical decision-making by providing visual anatomical and physiological data, potentially improving diagnostic accuracy and patient management.
Clinical Evidence
Bench testing only. The device was evaluated per the FDA guidance 'Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers' (2008). Acoustic output was measured and calculated according to NEMA UD-2 and NEMA UD-3 standards. Safety and electromagnetic compatibility were verified against IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-2-37, and ISO 10993-1 standards.
Technological Characteristics
Mobile ultrasound system; frequency range 2.0–12.0 MHz; supports linear, convex, and phased array transducers. Imaging modes: B, M, Color, PW, CW, Power, DirPower, and combined modes. Features: Tissue Harmonic Imaging, Smart3D, iScape, iBeam. Conforms to IEC 60601-1, IEC 60601-2-37, and ISO 10993-1. Software-controlled architecture.
Indications for Use
Indicated for use by qualified physicians for ultrasound imaging in gynecology, obstetrics, abdominal, pediatric, small parts (breast, thyroid, testes), neonatal/adult cephalic, transcranial, cardiac, transvaginal, transrectal, transurethral, peripheral vascular, intraoperative, urology, orthopedics, and musculoskeletal applications.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
Predicate Devices
- DC-3/DC-3T Diagnostic Ultrasound System (K083505)
- Mindray M5 (K080640)
- Mindray DC-6 (K072164)
- GE Voluson 730 (K041688)
Related Devices
- K081320 — DC-3, DC-3T DIAGNOSTIC ULTRASOUND SYSTEM · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · May 27, 2008
- K102865 — DC-7 DIAGNOSTIC ULTRASOUND SYSTEM, DC-3/DC-3T DIAGNOSTIC ULTRASOUND SYSTEM · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Nov 3, 2010
- K083505 — DC-3/DC-3T DIAGNOSTIC ULTRASOUND SYSTEM · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Dec 12, 2008
- K072164 — DC-6 DIAGNOSTIC ULTRASOUND SYSTEM · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Sep 5, 2007
- K110199 — DC-T6 DIAGNOSTIC ULTRASOUND SYSTEM · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Feb 8, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
JUL - 9 2009
K091941
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §801 13 Donny
The assigned 510(k) number is: _______________________________________________________________________________________________________________________________________________
#### 1. Submitter:
Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen,
518057, D. Cu 518057, P. R. China
Tel: +86 755 2658 2888 Fax: +86 755 2658 2680
#### Contact Person:
Tan Chuanbin Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
Date Prepared: May 26, 2009
2. Device Name: DC-3/DC-3T Diagnostic Ultrasound System
#### Classification
Regulatory Class: II
Review Category: Tier II
21 CFR 892.1550 Ultrasonic Pulsed Doppler Imaging System (90-TYN)
21 CFR 892.1560 Ultrasonic Pulsed Echo Imaging System (90-11
21 CFR 892.1570 Diagnostic Ultrasound Transducer (90-ITX)
## 3. Marketed Device:
The subject device is substantially equivalent in its technologies and functionality to the original DC-3/DC-3T Diagnostic Ultrasound System that is already cleared under premarket notification number K083505, and the other predicate devices are listed below: Mindray M5 (K080640), Mindray DC-6 (K072164),GE Voluson 730(K041688),
{1}------------------------------------------------
## 4. Device Description:
The DC-3/DC-3T Diagnostic Ultrasound System is a general purpose, mobile, software controlled, ultrasound diagnostic system. It function is to acquire and display ultrasound images in B-Mode, M-Mode, Color mode, PW mode, CW mode, Power mode, DirPower mode, DirPower mode or the combined mode (i.e. B/M Mode). This system is a Track 3 device that employs an array of probes that include linear array, convex array and phased array with a frequency range of approximately 2.0 MHz to 12.0 MHz.
## 5. Intended Use:
The device is intended for use by a qualified physician for ultrasound evaluation of gynecology, obstetrics, abdominal, pediatric, small parts (breast, thyroid, testicle, etc), neonatal cephalic, transcranial, cardiac transvaginal, transrectal, peripheral vascular, intraoperative, urology, orthopedial, and musculoskeletal, perficial vascular,
exams exams.
## 6. Safety Considerations:
The DC-3/DC-3T Diagnostic Ultrasound System has been tested as Track 3 Device per the FDA Guidance document "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers' issued in September 2008. The acoustic output is measured and calculated per NEMA UD 2 Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment: 2004 and NEMA UD 3 Standards, such, as JEC 6004. The device conforms to applicable medical device safety standards, such as IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-2, IEC 60601-2-37 and ISO 10993-1.
## Conclusion:
The conclusions drawn from testing of the DC-3/DC-3T Diagnostic Ultrasound System demonstrate that the device is as safe and effective as the legally marketed predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular in shape, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is a stylized image of what appears to be an abstract symbol, possibly representing a person or a bird.
## JUL = 9 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Shenzhen Mindray Bio-Medical Electronics Co., Ltd. % Mr. Robert Mosenkis President CITECH 5200 Butler Pike Plymouth Meeting, PA 19462-1298
Re: K091941
Trade/Device Name: DC-3/DC-3T Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYO, IYN, and ITX Dated: June 24, 2009 Received: June 30, 2009
Dear Mr. Mosenkis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the DC-3/DC-3T Diagnostic Ultrasound System, as described in your premarket notification:
Transducer Model Number
3C5A ୧୯.୮୮.୧ 7L4A 716 10L4 6C2 6LE7
{3}------------------------------------------------
| 6LB7 |
|------|
| 3C1 |
| 2P2 |
| 7L5 |
| 7LT4 |
| D6-2 |
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
If you have any questions regarding the content of this letter, please contact Paul Hardy at (240) 276-3666.
Sincerely yours,
Heidi Remer
Janine M. Morris
Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{4}------------------------------------------------
× |
Diagnostic Ultrasound Indications for Use Form
Transducer
DC-3/DC-3T
System Model:
510(k) Number(s)
| Clinical Application | Mode of Operation | | | | | | | |
|-------------------------------|-------------------|---|-----|-----|---------------|-------------------|--------------------|-----------------|
| | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Combined (specify) | Other (specify) |
| Ophthalmic | | | | | | | | |
| Fetal | P | P | P | P | P | P | P | Note 1, 2, 3, 4 |
| Abdominal | P | P | P | P | P | P | P | Note 1, 2, 3, 4 |
| Intraoperative (specify)* | P | P | P | P | P | P | P | Note 2, 3, 4 |
| Intraoperative (Neuro) | | | | | | | | |
| Laparoscopic | | | | | | | | |
| Pediatric | P | P | P | P | P | P | P | Note 1, 2, 3, 4 |
| Small organ(specify)** | P | P | P | P | P | P | P | Note 2, 3, 4 |
| Neonatal Cephalic | P | P | P | P | P | P | P | Note 1, 2, 3, 4 |
| Adult Cephalic | P | P | P | P | P | P | P | Note 1,2, 3 |
| Trans-rectal | P | P | P | P | P | P | P | Note 2, 3, 4 |
| Trans-vaginal | P | P | P | P | P | P | P | Note 2, 3 |
| Trans-urethral | | | | | | | | Note 2, 3 |
| Trans-esoph.(non-Card.) | | | | | | | | |
| Musculo-skeletal Conventional | P | P | P | | P | P | P | Note 2, 3, 4 |
| Musculo-skeletal Superficial | P | P | P | | P | P | P | Note 2, 3, 4 |
| Intravascular | | | | | | | | |
| Cardiac Adult | P | P | P | P | P | P | P | Note 1,2, 3 |
| Cardiac Pediatric | P | P | P | P | P | P | P | Note 1,2, 3 |
| Intravascular (Cardiac) | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | |
| Intra-Cardiac | | | | | | | | |
| Peripheral Vascular | | | | | | | | |
| Other (specify)*** | P | P | P | | P | P | P | Note 2, 3, 4 |
| | P | P | P | | P | P | P | Note1, 2, 3, 4 |
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number _
0085
{5}------------------------------------------------
## Mindray Co. Ltd.- DC-3/DC-3T Diagnostic Ultrasound System
Diagnostic Ultrasound Indications for Use Form
3CSA
Transducer
×
System Model: 510(k) Number(s)
| | Mode of Operation | | | | | | | |
|----------------------------------------------------------------------------------------------------|------------------------------------------------|---|------|------------|------------------|----------------------|-----------------------|-----------------|
| Clinical Application | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Combined<br>(specify) | Other (specify) |
| Ophthalmic | | | | | | | | |
| Fetal | P | P | P | | P | P | P | Note 1, 2, 3 |
| Abdominal | P | P | P | | P | P | P | Note 1, 2, 3 |
| Intraoperative (specify)* | | | | | | | | |
| Intraoperative (Neuro) | | | | | | | | |
| Laparoscopic | | | | | | | | |
| Pediatric | P | P | P | | P | P | P | Note 1, 2, 3 |
| Small organ(specify) ** | | | | | | | | Note 1, 2, 3 |
| Neonatal Cephalic | | | | | | | | |
| Adult Cephalic | | | | | | | | |
| Trans-rectal | | | | | | | | |
| Trans-vaginal | | | | | | | | |
| Trans-urethral | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | |
| Intravascular | | | | | | | | |
| Cardiac Adult | | | | | | | | |
| Cardiac Pediatric | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | |
| Intra-Cardiac | | | | | | | | |
| Peripheral Vascular | P | P | P | | P | P | P | Note 1, 2, 3 |
| Other (specify) *** | P | P | P | | P | P | P | Note 1, 2, 3 |
| N=new indication; P=previously cleared by FDA; E=added under Appendix E | | | | | | | | |
| Additional comments: Combined modes: B+M, PW+B, Color + B, Power + B, PW +Color+ B, Power + PW +B. | | | | | | | | |
| * Intraoperative includes abdominal, thoracic, and vascular etc. | | | | | | | | |
| ** Small organ-breast, thyroid, testes, etc. | | | | | | | | |
| *** Other use includes Urology. | | | | | | | | |
| Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents. | | | | | | | | |
| System | Diagnostic Ultrasound Indications for Use Form | | | | | | | |
| Model: | | | 6CVI | Transducer | X | | | |
| 510(k) Number(s) | | | | | | | | |
| | | | | | | | | |
| Clinical Application | Mode of Operation | | | | | | | |
| | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Combined (specify) | Other (specify) |
| Ophthalmic | | | | | | | | |
| Fetal | P | P | P | | P | P | P | Note 2, 3 |
| Abdominal | | | | | | | | |
| Intraoperative (specify)* | | | | | | | | |
| Intraoperative (Neuro) | | | | | | | | |
| Laparoscopic | | | | | | | | |
| Pediatric | | | | | | | | |
| Small organ(specify)** | | | | | | | | |
| Neonatal Cephalic | | | | | | | | |
| Adult Cephalic | | | | | | | | |
| Trans-rectal | P | P | P | | P | P | P | Note 2, 3 |
| Trans-vaginal | P | P | P | | P | P | P | Note 2, 3 |
| Trans-urethral | | | | | | | | |
| Trans-esoph.(non-Card.) | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | |
| Intravascular | | | | | | | | |
| Cardiac Adult | | | | | | | | |
| Cardiac Pediatric | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | |
| Trans-esoph.(Cardiac) | | | | | | | | |
| Intra-Cardiac | | | | | | | | |
| Peripheral Vascular | | | | | | | | |
| Other (specify) *** | P | P | P | | P | P | P | Note 2, 3 |
| N=new indication; P=previously cleared by FDA; E=added under Appendix E | | | | | | | | |
| Additional comments:Combined modes: B+M, PW+B, Color + B, Power + B, PW +Color+ B, Power + PW +B. | | | | | | | | |
| *Intraoperative includes abdominal, thoracic, and vascular etc. | | | | | | | | |
| ** Small organ-breast, thyroid, testes, etc. | | | | | | | | |
| *** Other use includes Urology. | | | | | | | | |
| Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents. | | | | | | | | |
| Note 2: Smart3D | | | | | | | | |
| | | | | | | | | |
| Note 3: iScape | | | | | | | | |
| Clinical Application | Mode of Operation | | | | | | | |
| | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Combined<br>(specify) | Other (specify) |
| Ophthalmic | | | | | | | | |
| Fetal | | | | | | | | |
| Abdominal | P | P | P | | P | P | P | |
| Intraoperative (specify)* | | | | | P | P | P | Note 2, 3, 4 |
| Intraoperative (Neuro) | | | | | | | | |
| Laparoscopic | | | | | | | | |
| Pediatric | P | P | P | | P | P | P | |
| Small organ(specify)** | P | P | P | | P | P | P | Note 2, 3, 4 |
| Neonatal Cephalic | P | P | P | | P | P | P | Note 2, 3, 4 |
| Adult Cephalic | | | | | | P | P | Note 2, 3, 4 |
| Trans-rectal | | | | | | | | |
| Trans-vaginal | | | | | | | | |
| Trans-urethral | | | | | | | | |
| Trans-esoph.(non-Card.) | | | | | | | | |
| Musculo-skeletal Conventional | P | P | P | | P | P | P | |
| Musculo-skeletal Superficial | P | P | P | | P | P | P | Note 2, 3, 4 |
| Intravascular | | | | | P | P | P | Note 2, 3, 4 |
| Cardiac Adult | | | | | | | | |
| Cardiac Pediatric | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | |
| Trans-esoph.(Cardiac) | | | | | | | | |
| Intra-Cardiac | | | | | | | | |
| Peripheral Vascular | P | P | P | | P | P | P | |
| Other (specify)*** | | | | | P | P | P | Note 2, 3, 4 |
| N=new indication; P=previously cleared by FDA; E=added under Appendix E | | | | | | | | |
| Additional comments: Combined modes: B+M, PW+B, Color + B, Power + B, PW +Color+ B, Power + PW +B. | | | | | | | | |
| * Intraoperative includes abdominal, thoracic, and vascular etc. | | | | | | | | |
| ** Small organ-breast, thyroid, testes, etc. | | | | | | | | |
| ***Other use includes Urology. | | | | | | | | |
| Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents. | | | | | | | | |
| Note 2: Smart3D | | | | | | | | |
| Note 3: iScape | | | | | | | | |
| Note 4: iBeam | | | | | | | | |
Note 2: Smart3D Note 3: iScape
Note 4: iBeam
Hula leum
(Division Sign-Dff) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
0086
{6}------------------------------------------------
Diagnostic Ultrasound Indications for Use Form
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number __
:
.
{7}------------------------------------------------
Diagnostic Ultrasound Indications for Use Form System Transducer Model: 7L4A 510(k) Number(s) ·
×
Hechd Lewn
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number _
{8}------------------------------------------------
System Transducer × Model: 716 510(k) Number(s) Mode of Operation Clinical Application Color Amplitude B M Combined PWD CWD Other (specify) Doppler Doppler (specify) Ophthalmic Fetal Abdominal P P P P P P Intraoperative (specify)* Note 2, 3, 4 Intraoperative (Neuro) Laparoscopic Pediatric P P P P P Small organ(specify)** P Note 2, 3, 4 P P P P P P Neonatal Cephalic Note 2, 3, 4 P P P P p Adult Cephalic P Note 2, 3, 4 Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card.) Musculo-skeletal Conventional P P P P P Musculo-skeletal Superficial P Note 2, 3, 4 P P P P P Intravascular P Note 2, 3, 4 Cardiac Adult Cardiac Pediatric Intravascular (Cardiac) Trans-esoph. (Cardiac) Intra-Cardiac Peripheral Vascular P P P P P P Other (specify) ** * Note 2, 3, 4 N=new indication; P=previously cleared by FDA; E=added under Appendix E Additional comments: Combined modes: B+M, PW+B, Color + B, PW +Color+ B, PW +Color+ B, Power + PW +B, * Intraoperative includes abdominal, thoracic, and vascular etc. ** Small organ-breast, thyroid, testes, etc. ***Other use includes Urology. Note I: Tissue Harmonic Imaging. The feature does not use contrast agents. Note 2: Smart3D Note 3: iScape Note 4: iBeam
Hes Rem
livision Sign Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
## Diagnostic Ultrasound Indications for Use Form
0089
{9}------------------------------------------------
Diagnostic Ultrasound Indications for Use Form System Transducer × Model: I OLA 510(k) Number(s) Mode of Operation Clinical Application Color-Amplitude B Combined M PWD CWD Other (specify) Doppler Ophthalmic Doppler (specify) Fetal Abdominal p P P P Intraoperative (specify)* P P Note 2, 3, 4 Intraoperative (Neuro) Laparoscopic Pediatric P P P P P Small organ(specify) ** P Note 2, 3, 4 P P P P P Neonatal Cephalic P Note 2, 3, 4 P P P P Adult Cephalic P P Note 2, 3, 4 Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card.) Musculo-skeletal Conventional P P P P Musculo-skeletal Superficial P P P Note 2, 3, 4 P p P Intravascular P P Note 2, 3, 4 Cardiac Adult Cardiac Pediatric Intravascular (Cardiac) Trans-esoph (Cardiac) Intra-Cardiac Peripheral Vascular p P P P P Other (specify) ** P Note 2, 3, 4 N=new indication; P=previously cleared by FDA; E=added under Appendix E Additional comments: Combined modes: B+M, PW+B, Color + B, PW +Color+ B, PW +Color+ B, PW +Color+ B, Power + PW +B, *Intraoperative includes abdominal, thoracic, and vascular etc. ** Small organ-breast, thyroid, testes, etc. ***Other use includes Urology. Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents. Note 2: Smart3D
Helda Perm
ision Sign Off) on of Reproductive, Abdominal a Radiological Devices 510(k) Number
Note 3: iScape Note 4: iBeam
{10}------------------------------------------------
and the comments of the count
.
.
...
.
、
| System | |
|------------------|-----|
| Model: | 6C2 |
| 510(k) Number(s) | |
Diagnostic Ultrasound Indications for Use Form
| Transducer | X |
|------------|---|
|------------|---|
| Mode of Operation | | | | | | | | |
|-------------------------------|---|---|-----|-----|------------------|----------------------|-----------------------|-----------------|
| Clinical Application | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Combined<br>(specify) | Other (specify) |
| Ophthalmic | | | | | | | | |
| Fetal | | | | | | | | |
| Abdominal | P | P | P | | P | P | P | Note 2, 3 |
| Intraoperative (specify)* | | | | | | | | |
| Intraoperative (Neuro) | | | | | | | | |
| Laparoscopic | | | | | | | | |
| Pediatric | P | P | P | | P | P | P | Note 2, 3 |
| Small organ(specify)** | | | | | | | | |
| Neonatal Cephalic | P | P | P | | P | P | P | Note 2, 3 |
| Adult Cephalic | P | P | P | | P | P | P | Note 2, 3 |
| Trans-rectal | | | | | | | | |
| Trans-vaginal | | | | | | | | |
| Trans-urethral | | | | | | | | |
| Trans-esoph.(non-Card.) | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | |
| Intravascular | | | | | | | | |
| Cardiac Adult | P | P | P | | P | P | P | Note 2, 3 |
| Cardiac Pediatric | P | P | P | | P | P | P | Note 2, 3 |
| Intravascular (Cardiac) | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | |
| Intra-Cardiac | | | | | | | | |
| Peripheral Vascular | | | | | | | | |
| Other (specify)*** | P | P | P | | P | P | P | Note 2, 3. |
N=new indication; P=previously cleared by FDA; E=added under Appendix E
Additional comments:Combined modes: B+M, PW+B, Color + B, Power + B, PW +Color+ B, Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular etc.
**Small organ-breast, thyroid, testes, etc.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3: iScape
Note 4: iBeam
(Division Sign-Off)
Division of Reproductive, Abdominal and
Radiological Devices
| 510(k) Number | K091941 |
|---------------|---------|
|---------------|---------|
0091
.
{11}------------------------------------------------
| System | | | | Transducer | | X | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|---|------|---------|------------|------------------|----------------------|-----------------------|-----------------|
| Model: | | 6LE7 | | | | | | |
| 510(k) Number(s) | | | | | | | | |
| | | | | | | | | |
| Clinical Application | | | | | | Mode of Operation | | |
| | B | M | PW<br>D | CWD | Color<br>Doppler | Amplitude<br>Doppler | Combined<br>(specify) | Other (specify) |
| Ophthalmic | | | | | | | | |
| Fetal | P | P | P | | P | P | P | Note 2, 3, 4 |
| Abdominal | | | | | | | | |
| Intraoperative (specify)* | | | | | | | | |
| Intraoperative (Neuro) | | | | | | | | |
| Laparoscopic | | | | | | | | |
| Pediatric | | | | | | | | |
| Small organ(specify)** | | | | | | | | |
| Neonatal Cephalic | | | | | | | | |
| Adult Cephalic | | | | | | | | |
| Trans-rectal | P | P | P | | | | | |
| Trans-vaginal | P | P | P | | P | P | P | Note 2, 3, 4 |
| Trans-urethral | | | | | | | | |
| Trans-esoph.(non-Card.) | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | |
| Intravascular | | | | | | | | |
| Cardiac Adult | | | | | | | | |
| Cardiac Pediatric | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | |
| Intra-Cardiac | | | | | | | | |
| Peripheral Vascular | | | | | | | | |
| Other (specify)*** | P | P | P | P | P | P | P | Note2, 3, 4 |
| N=new indication; P=previously cleared by FDA; E=added under Appendix E | | | | | | | | |
| Additional comments:Combined modes: B+M, PW+B, Color + B, Power + B, PW +Color+ B, Power + PW +B.<br>* Intraoperative includes abdominal, thoracic, and vascular etc. | | | | | | | | |
| ** Small organ-breast, thyroid, testes, etc. | | | | | | | | |
| *** Other use includes Urology. | | | | | | | | |
| Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents. | | | | | | | | |
| Note 2: Smart3D | | | | | | | | |
| Note 3: iScape | | | | | | | | |
| Note 4: iBeam | | | | | | | | |
-
Diagnostic Ultrasound Indications for Use Form
Hele Keen
(Division Sign off)
Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
{12}------------------------------------------------
.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
. *-
: : : : : : : : : : : : : : : : : : : : : : : : :
System
Transducer X
| Model: | 6LB7 |
|------------------|------|
| 510(k) Number(s) | |
| Clinical Application | B | M | PW D | CWD | Color Doppler | Amplitude Doppler | Combined (specify) | Other (specify) |
|-------------------------------|---|---|------|-----|---------------|-------------------|--------------------|-----------------|
| Ophthalmic | | | | | | | | |
| Fetal | | | | | | | | |
| Abdominal | | | | | | | | |
| Intraoperative (specify)* | | | | | | | | |
| Intraoperative (Neuro) | | | | | | | | |
| Laparoscopic | | | | | | | | |
| Pediatric | | | | | | | | |
| Small organ(specify)** | | | | | | | | |
| Neonatal Cephalic | | | | | | | | |
| Adult Cephalic | | | | | | | | |
| Trans-rectal | P | P | P | | P | P | P | Note 2, 3, 4 |
| Trans-vaginal | | | | | | | | |
| Trans-urethral | | | | | | | | |
| Trans-esoph.(non-Card.) | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | |
| Intravascular | | | | | | | | |
| Cardiac Adult | | | | | | | | |
| Cardiac Pediatric | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | |
| Trans-esoph.(Cardiac) | | | | | | | | |
| Intra-Cardiac | | | | | | | | |
| Peripheral Vascular | | | | | | | | |
| Other (specify)*** | P | P | P | | P | P | P | Note2, 3, 4 |
N=new indication; P=previously cleared by FDA; E=added under Appendix E
Additional comments:Combined modes: B+M, PW+B, Color + B, Power + B, PW +Color+ B, Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular etc.
**Small organ-breast, thyroid, testes, etc.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3: iScape
Note 4: iBeam
(Division Sign-Off)
Division of Reproductive, Abdominal and Radiological Devices
| 510(k) Number | K091941 |
|---------------|---------|
|---------------|---------|
## Diagnostic Ultrasound Indications for Use Fe
0093
{13}------------------------------------------------
System Transducer × Model: 3C1 510(k) Number(s) Mode of Operation Clinical Application PW Color Amplitude Combined B M CWD Other (specify) D Doppler Doppler (specify) Ophthalmic Fetal P P P p P P Note 1, 2, 3 Abdominal P P P P P P Note 1, 2, 3 Intraoperative (specify)* Intraoperative (Neuro) Laparoscopic Pediatric P P P P p P Note 1, 2, 3 Small organ(specify)** Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph. (non-Card.) Musculo-skeletal Conventional Musculo-skeletal Superficial Intravascular Cardiac Adult P P P P P P Note 1, 2, 3 Cardiac Pediatric P P P P p P Note 1, 2, 3 Intravascular (Cardiac) Trans-esoph.(Cardiac) Intra-Cardiac Peripheral Vascular Other (specify)*** N=new indication; P=previously cleared by FDA; E=added under Appendix B Additional comments:Combined modes: B+M, PW+B, Color + B, Power + B, PW +Color+ B, Power + PW +B, * Intraoperative includes abdominal, thoracic, and vascular etc. ** Small organ-breast, thyroid, testes, etc. ***Other use includes Urology. Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents. Note 2: Smart3D Note 3: iScape Note 4: iBeam (Division Sign-Offy Division of Reproductive, Abdon
0094
Radiological Devices 510(k) Number
Diagnostic Ultrasound Indications for Use Form
{14}------------------------------------------------
Mindray Co. Ltd. - DC-3/DC
:
.
.
:
.
.
:
:40 : 0 : 0 : 0 : 0 : 0 :
.
:
.
:
.
| System | | | | Transducer | | X | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|---|---|---------|------------|------------------|----------------------|-----------------------|-----------------|
| Model: | | | 2P2 | | | | | |
| 510(k) Number(s) | | | | | | | | |
| | | | | | | | | |
| Clinical Application | | | | | | Mode of Operation | | |
| | B | M | PW<br>D | CWD | Color<br>Doppler | Amplitude<br>Doppler | Combined<br>(specify) | Other (specify) |
| Ophthalmic | | | | | | | | |
| Fetal | | | | | | | | |
| Abdominal | P | P | P | P | P | P | P | Note 1, |
| Intraoperative (specify)* | | | | | | | | |
| Intraoperative (Neuro) | | | | | | | | |
| Laparoscopic | | | | | | | | |
| Pediatric | P | P | P | P | P | P | P | Note 1, |
| Small organ(specify)** | | | | | | | | |
| Neonatal Cephalic | P | P | P | P | P | P | P | Note 1, |
| Adult Cephalic | P | P | P | P | P | P | P | Note 1, |
| Trans-rectal | | | | | | | | |
| Trans-vaginal | | | | | | | | |
| Trans-urethral | | | | | | | | |
| Trans-esoph.(non-Card.) | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | |
| Intravascular | | | | | | | | |
| Cardiac Adult | P | P | P | P | P | P | P | Note 1, 2 |
| Cardiac Pediatric | P | P | P | P | P | P | P | Note 1, 2 |
| Intravascular (Cardiac) | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | |
| Intra-Cardiac | | | | | | | | |
| Peripheral Vascular | | | | | | | | |
| Other (specify)*** | | | | | | | | |
| N=new indication; P=previously cleared by FDA; E=added under Appendix E | | | | | | | | |
| | | | | | | | | |
| Additional comments:Combined modes: B+M, PW+B, Color + B, Power + B, PW +Color+ B, Power + PW +B.<br>*Intraoperative includes abdominal, thoracic, and vascular etc. | | | | | | | | |
| ** Small organ-breast, thyroid, testes, etc. | | | | | | | | |
| ***Other use includes Urology. | | | | | | | | |
| Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents. | | | | | | | | |
| Note 2: Smart3D | | | | | | | | |
| Note 3: iScape | | | | | | | | |
| Note 4: iBeam | | | | | | | | |
Diagnostic III+
:
:
.
:
-
: . .
.
.
:
:
.
{15}------------------------------------------------
### Mindray Co. Ltd.- DC-3/DC-3T Diagnostic Ultrasound System
Diagnostic Ultrasound Indications for Use Form
| System | |
|------------------|-----|
| Model: | 7L5 |
| 510(k) Number(s) | |
| Transducer | X |
| | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Combined (specify) | Other (specify) |
|-------------------------------|---|---|-----|-----|---------------|-------------------|--------------------|-----------------|
| Clinical Application | | | | | | | | |
| Ophthalmic | | | | | | | | |
| Fetal | | | | | | | | |
| Abdominal | P | P | P | | P | P | P | Note 2, 3, 4 |
| Intraoperative (specify)* | | | | | | | | |
| Intraoperative (Neuro) | | | | | | | | |
| Laparoscopic | | | | | | | | |
| Pediatric | P | P | P | | P | P | P | Note 2, 3, 4 |
| Small organ(specify)** | P | P | P | | P | P | P | Note 2, 3, 4 |
| Neonatal Cephalic | P | P | P | | P | P | P | Note 2, 3, 4 |
| Adult Cephalic | | | | | | | | |
| Trans-rectal | | | | | | | | |
| Trans-vaginal | | | | | | | | |
| Trans-urethral | | | | | | | | |
| Trans-esoph.(non-Card.) | | | | | | | | |
| Musculo-skeletal Conventional | P | P | P | | P | P | P | Note 2, 3, 4 |
| Musculo-skeletal Superficial | P | P | P | | P | P | P | Note 2, 3, 4 |
| Intravascular | | | | | | | | |
| Cardiac Adult | | | | | | | | |
| Cardiac Pediatric | | | | |…