Endoscopic Ultrasonic Probe (P2612S-L); Endoscopic Ultrasonic Probe (P2620S-L)
K231666 · FUJIFILM Healthcare Americas Corporation · ITX · Dec 13, 2023 · Radiology
Device Facts
| Record ID | K231666 |
| Device Name | Endoscopic Ultrasonic Probe (P2612S-L); Endoscopic Ultrasonic Probe (P2620S-L) |
| Applicant | FUJIFILM Healthcare Americas Corporation |
| Product Code | ITX · Radiology |
| Decision Date | Dec 13, 2023 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1570 |
| Device Class | Class 2 |
Intended Use
This product is a medical ultrasonic probe. It is intended for the observation and diagnosis of the gastrointestinal tract and surrounding organs under the management of physicians at medical facilities. This product is intended for adults. Modes of Operation: B-mode Never use this product for any other purposes.
Device Story
Endoscopic ultrasonic probe (P2612S-L, P2620S-L) used for intraluminal imaging of gastrointestinal tract and surrounding organs. Device connects to ultrasonic observation system (SP-900); converts electrical signals into ultrasonic waves via mechanical radial transducer; receives reflected waves; converts back to electrical signals for image generation. Operated by physicians in medical facilities. Provides real-time B-mode ultrasound images to assist in clinical diagnosis and observation. Benefits include minimally invasive visualization of GI structures.
Clinical Evidence
No clinical data. Bench testing only, including electrical safety (ANSI/AAMI ES60601-1, IEC 60601-2-37), electromagnetic compatibility (IEC 60601-1-2), and biocompatibility (ISO 10993-1, 5, 10, 23).
Technological Characteristics
Mechanical radial ultrasonic probe. Materials: Polyethylene (distal cap), stainless steel (distal ring), epoxy resin (bonded part), PTFE (sheath). Frequencies: 10MHz (P2612S-L) and 17MHz (P2620S-L). Single-element transducer. Connectivity: Connects to SP-900 processor. Reprocessing: Manual cleaning/disinfection, EOG sterilization. Standards: IEC 60601-2-37, ISO 10993.
Indications for Use
Indicated for observation and diagnosis of the gastrointestinal tract and surrounding organs in adult patients.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
Predicate Devices
- OLYMPUS UM-S30-25R ULTRASONIC PROBE AND ASSOCIATED ANCILLARY EQUIPMENT . FOR GASTROINTESTINAL TRACT (K994103)
Reference Devices
- EUB-525 Diagnostic Ultrasound Scanner (K012239)
- EUB-2000 Diagnostic Ultrasound Scanner (K012239)
- Sp-711 Sonoprobe System (K012239)
- PL2226-12 (K012239)
- P2226-20 (K012239)
Related Devices
- K214089 — FUJIFILM Ultrasonic Processor SP-900 and FUJIFILM Ultrasonic Probe PB2020-M2 · Fujifilm Corporation · Jun 3, 2022
- K063847 — FUJINON ULTRASONIC ENDOSCOPE AND PROCESSOR, MODELS EG-530UR, EG-530UT AND SU-700 · Fujinon, Inc. · Mar 1, 2007
- K994103 — OLYMPUS UM-S30-25R ULTRASONIC PROBE AND ASSOCIATED ANCILLARY EQUIPMENT FOR GASTROINTESTINAL TRACT · Olympus America, Inc. · Sep 5, 2000
- K221952 — Endoscope Model EG-580UT and Endoscope Model EG-580UR · Fujifilm Corporation · Aug 4, 2022
- K171207 — FUJIFILM Ultrasonic Processor SP-900 and FUJIFILM Ultrasonic Probe PB2020-M · Fujifilm Medical Systems U.S.A, Inc. · Jan 11, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
December 13, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Fujifilm Healthcare Americas Corporation % Chaitrali Kulkarni Sr. Regulatory Affairs Specialist 798 MIYANODAI KAISEI-MACHI ASHIGARAKAMI-GUN, KANAGAWA 258-8538 JAPAN
# Re: K231666
Trade/Device Name: Endoscopic Ultrasonic Probe (P2612S-L); Endoscopic Ultrasonic Probe (P2620S-L) Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: Class II Product Code: ITX Dated: November 7, 2023 Received: November 7, 2023
# Dear Chaitrali Kulkarni:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Yanna S. Kang -S
Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
Submission Number (if known)
K231666
Device Name
Endoscopic Ultrasonic Probe (P2612S-L); Endoscopic Ultrasonic Probe (P2620S-L)
Indications for Use (Describe)
This product is a medical ultrasonic probe. It is intended for the observation and diagnosis of the gastrointestinal tract and surrounding organs under the management of physicians at medical facilities. This product is intended for adults.
Modes of Operation: B-mode
Never use this product for any other purposes,
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# K231666
#### FUJIFILM Corporation
## FUJIFILM Ultrasonic Probe P2612S-L/P2620S-L
Date: December 13, 2023
#### Submitter's Information:
FUJIFILM Corporation 798 MIYANODAI KAISEI-MACHI ASHIGARAKAMI-GUN, KANAGAWA 258-8538 JAPAN
#### Contact Person:
Chaitrali Kulkarni Senior Regulatory Affairs Specialist E-Mail: chaitrali.kulkarni@fujifilm.com Telephone: (704) 517 4886
#### Identification of the Subject Device:
| Device Name | Endoscopic Ultrasonic<br>Probe (P2612S-L);<br>Endoscopic Ultrasonic<br>Probe (P2620S-L) |
|------------------------|-----------------------------------------------------------------------------------------|
| Common Name | Ultrasonic Probe |
| Product Code | • ITX |
| Device Class | Class 2 |
| Regulation Number | • 892.1570 |
| Regulation Description | • Diagnostic ultrasonic<br>transducer |
| Review Panel | • Radiology |
#### Predicate Device:
- OLYMPUS UM-S30-25R ULTRASONIC PROBE AND ASSOCIATED ANCILLARY EQUIPMENT . FOR GASTROINTESTINAL TRACT (K994103)
#### Reference Device:
- Modifications to EUB-525 Diagnostic Ultrasound Scanner; EUB-2000 Diagnostic Ultrasound ● Scanner; Sp-711 Sonoprobe System (K012239)
#### Intended Use / Indications for Use:
This product is a medical ultrasonic probe. It is intended for the observation and diagnosis of the gastrointestinal tract and surrounding organs under the management of physicians at medical facilities. This product is intended for adults.
Modes of Operation: B-mode
Never use this product for any other purposes.
{4}------------------------------------------------
# Device Description:
FUJIFILM Endoscopic Ultrasonic Probe converts the electrical signal from the ultrasonic observation device connected via the probe scanner into ultrasonic waves by the ultrasonic transducer placed and emits it and receives reflected waves from the human body. The received wave is converted into an electric signal by transducer and sent to an ultrasonic device to create an ultrasonic image. The scanning method is a mechanical radial method by using a probe scanner.
Based on the guidance, Marketing Clearance of Diagnostic Ultrasound Systems and Transducers (Feb. 2023), this device follows the track 3 designation, as it does conform to IEC 60601-2- 37. The transducer model designation and type are a mechanical radial scan, with a size and spacing of elements of 1.9×2.0×0.6mm(M×L×T). There is one element in the array, with array dimensions of one and the maximum number of active elements for a single pulse is one. The nominal ultrasonic frequency of the P2612S-L is 10MHz and for P2620S-L is 17MHz, both in combination with SP-900.
#### Comparison of Technological Characteristics:
Comparisons of technological characteristics between the subject devices are provided in the tables below:
#### Table 1
| Device name | Predicate Device | Reference Device: | Proposed Device | Equivalence |
|---------------|--------------------------|-------------------|-----------------------|-------------|
| | Olympus Ultrasonic Probe | | Endoscopic Ultrasonic | |
| | | | Probe | |
| | UM-S30-25R | PL2226-12 | P2612S-L/P2620S-L | |
| | | P2226-20 | | |
| 510(k) number | K994103 | K012239 | K231666 | |
{5}------------------------------------------------
| Indications for use (IFU) | The Olympus UM-S30-25R<br>Ultrasonic Probe has been<br>designed to be used with an<br>Olympus Endoscopic Ultrasound<br>center, Probe Driving Unit and an<br>endoscope for<br>intraluminal ultrasonic imaging of<br>the gastrointestinal tract and<br>surrounding organs. Modes of<br>Operation: B-mode | The intended use of the<br>subject device is for<br>endoscopic observation<br>of the gastrointestinal<br>tract (esophagus,<br>stomach, duodenum,<br>large intestine) and<br>biliary system (pancreato-<br>biliary ducts). The<br>Ultrasound Device<br>Indications Statements<br>for each application and<br>mode of the<br>system/transducers are<br>included with this<br>document. | This product is a medical<br>ultrasonic probe. It is<br>intended for the<br>observation and diagnosis<br>of the gastrointestinal<br>tract and surrounding<br>organs under the<br>management of<br>physicians at medical<br>facilities. This product is<br>intended for adults.<br>Modes of Operation: B-<br>mode<br>Never use this product for<br>any other purposes. | Same | |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Probe specification | | | | | |
| Diameter of insertion portion | 2.4mm | 2.6mm | 2.5mm | The proposed device<br>and the predicate<br>device have different<br>insertion portion<br>diameters, however the<br>proposed device has a<br>smaller insertion<br>portion diameter than<br>the reference device,<br>therefore there is no<br>new concern for safety<br>and efficacy. | |
| Maximum diameter of insertion<br>portion | | 2.7mm | 2.6mm | The proposed device<br>and the reference<br>device have different<br>maximum diameter of | |
| | | | | | |
| | | | | | insertion portion, however the proposed device has a smaller insertion portion diameter than the reference device, therefore there is no new concern for safety and efficacy. |
| Working length | | 2050mm | 2180mm | 2620mm | Although the length of the proposed device is longer, the working length of the proposed device is compatible with the working length of its compatible endoscope, therefore there are no new concerns for the safety and efficacy of the proposed device. |
| Applicable reprocess method | Manual cleaning | | Applicable | Applicable | Same as reference device |
| | Manual disinfection | | Applicable | Applicable | Same as reference device |
| | Endoscope automatic cleaning and disinfection | | Not Applicable | Applicable | The proposed device does have automatic cleaning and disinfection although the reference device does not. The difference in applicable reprocess methods does not affect the safety and efficacy of the device and cleaning instructions have been provided in the operation manual. |
| | EOG | | Not Applicable | Applicable | EOG sterilization is |
| | Sterilization | | | | applicable to the<br>proposed device but<br>not the reference<br>device. The difference<br>in applicable reprocess<br>methods does not affect<br>the safety and efficacy<br>of the device and<br>cleaning instructions<br>have been provided in<br>the operation manual. |
| | Stellad<br>Sterilization | | Not applicable | Not applicable | Same as reference<br>device |
| Applicable system | | Olympus Endoscopic Ultrasound<br>center, Probe Driving Unit and an<br>endoscope for<br>intraluminal ultrasonic imaging of<br>the gastrointestinal tract and<br>surrounding organs. | SP-711 | SP-900 | The difference in<br>applicable systems<br>between the subject<br>device and predicate<br>device do not affect the<br>safety or efficacy of the<br>subject device. The<br>subject device has been<br>tested with the SP-900<br>processor and the<br>applicable system is<br>listed in the operation<br>manual. |
| Applicable scope | | Flexible endoscope: inner channel<br>is 2.8mm, 3.2mm,4.2mm,6mm and<br>5.Smm.<br>Rigid endoscopes: minimum<br>capacity is 9Fr. | Direct-viewing type<br>endoscopes with a<br>forceps channel diameter<br>of 2.8 mm or more and a<br>working<br>length of 1700 mm or<br>less | FUJIFILM endoscope<br>-Upper gastrointestinal<br>endoscope<br>-Lower gastrointestinal<br>endoscope<br>*1 The instrument<br>channel diameter of each<br>endoscope must be 2.8<br>mm or more and its<br>working<br>length 2200 mm or less.<br>*2 This product is not<br>applied in combination | The differences in the<br>applicable scope do not<br>affect the safety or<br>efficacy of the subject<br>device. The subject<br>device has been tested<br>with its applicable<br>scope and no new<br>concerns for safety or<br>efficacy was seen. |
| | | | with Duodenoscope and<br>Ultrasonic endoscope. | | |
| Ultrasound specification | | | | | |
| Scanning method | Mechanical radial | Mechanical radial | Mechanical radial | same | |
| Sound operating frequency | 30MHz | PL2226-12 :<br>12MHz±20%<br>PL2226-20 :<br>18MHz±20% | P2612S-L :<br>10MHz±15%<br>P2620S-L :<br>17MHz±15% | The differences in the<br>ultrasound<br>specification from the<br>predicate device and<br>the proposed device<br>have been tested<br>according to 60601-2-<br>37 | |
| Resolution | | Axial: 1.5mm or less<br>Lateral: 3.0mm or less | Axial: 1.0mm or less<br>Lateral: 3.0mm or less | The differences in the<br>ultrasound<br>specification from the<br>predicate device and<br>the reference device<br>have been tested<br>according to 60601-2-<br>37. | |
| Penetration | | PL2226-12 : 15mm<br>PL2226-20 : 5mm | P2612S-L : 20mm<br>P2620S-L: 5mm | The differences in the<br>ultrasound<br>specification from the<br>predicate device and<br>the proposed device<br>have been tested<br>according to 60601-2- | |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
| | | | 37. |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Patient contact materials | | | |
| Insertion portion | Distal end cap :<br>Polyethylene<br>• Distal ring : Stainless<br>steel rod<br>• Bonded part : Epoxy<br>resin<br>• Sheath :<br>Polytetrafluoroethylene | Distal end cap :<br>Polyethylene<br>• Distal ring : Stainless<br>steel rod<br>• Bonded part : Epoxy<br>resin<br>• Sheath :<br>Polytetrafluoroethylene | Same as reference<br>device |
{10}------------------------------------------------
## Performance Data:
FUJIFILM Ultrasonic P2612S-L/P2620S-L is supplied non-sterile and must be properly reprocessed prior to each use in accordance with its reprocessing instructions. The cleaning, disinfection, and sterilization were validated on P2612S-L/P2620S-L.
The biocompatibility was evaluated in accordance with FDA's quidance. Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", issued 2018. The cytotoxicity was evaluated according to ISO 10993-5. The sensitization study was conducted according to 10993-10 and irritation testing was conducted according to ISO 10993-23.
The subject device P2612S-L/P2620S-L contains electronic components. The testing was conducted to ensure the electrical safety of the subject device P2612S-L/P2620S-L according to ANSI/AAMI ES60601-1 and IEC 60601-2-37:2007. The subject device P2612S-L/P2620S-L along with the applicable system (SP-900) was evaluated for the electromagnetic compatibility according to IEC 60601-1-2:2014.
# Clinical Testing:
Clinical testing is not applicable to this submission.
# Conclusions:
The subject device FUJIFILM Ultrasonic Probe P2612S-L/P2620S-L and predicate device, Olympus UM-S30-25R Ultrasonic Probe and Associated Ancillary Equipment for Gastrointestinal Tract, have slight differences in the technological characteristics. Although there are differences in technological characteristics those differences do not raise any new concerns for safety and efficacy.
The subject device FUJIFILM Ultrasonic Probe P2612S-L/P2620S-L is substantially equivalent to the respective predicate device Olympus UM-S30-25R Ultrasonic Probe and Associated Ancillary Equipment for Gastrointestinal Tract (K994103)