OLYMPUS UM 2R/3R ULTRASONIC PROBES AND ASSOCIATED ANCILLARY EQUIPMENT, FOR BRONCHIAL USE

K982323 · Olympus America, Inc. · ITX · Dec 11, 1998 · Radiology

Device Facts

Record IDK982323
Device NameOLYMPUS UM 2R/3R ULTRASONIC PROBES AND ASSOCIATED ANCILLARY EQUIPMENT, FOR BRONCHIAL USE
ApplicantOlympus America, Inc.
Product CodeITX · Radiology
Decision DateDec 11, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1570
Device ClassClass 2

Intended Use

The Olympus UM-2R and UM-3R Ultrasonic Probes have been designed for use in combination with Olympus Endoscopic Ultrasound System for intraluminal sonographic imaging of the upper airways and tracheobronchial tree.

Device Story

Ultrasonic probes (UM-2R/3R) provide transendoscopic access to upper airways and tracheobronchial tree; used with bronchoscopes (min 2.8 mm channel). Probes utilize de-aerated water immersion method for B-mode, 360-degree mechanical/radial scanning. Probe driving unit controls transducer rotation. UM-2R operates at 12 MHz; UM-3R operates at 20 MHz. Used by physicians in clinical settings to perform intensive examination, observation, and diagnosis of airways. Output displayed via compatible Olympus Endoscopic Ultrasound Systems (EU-M30, EU-M20, EU-M3). Enables visualization of tissue beyond conventional bronchoscope limits, facilitating navigation through strictures and anatomical ducts; aids clinical decision-making regarding airway pathology.

Clinical Evidence

Bench testing only. Compliance with IEC 60601-1 and FDA's 1993/1985 Diagnostic Ultrasound Guidance confirmed safety and effectiveness. No clinical data provided.

Technological Characteristics

Mechanical/radial scanning ultrasound transducer; 2.4 mm outer diameter; 2050 mm length. UM-2R (12 MHz) and UM-3R (20 MHz). De-aerated water immersion coupling. Non-sterile, reprocessable. Compatible with Olympus EU-M30, EU-M20, and EU-M3 systems. Complies with IEC 60601-1.

Indications for Use

Indicated for intraluminal sonographic imaging of the upper airways and tracheobronchial tree in patients requiring endoscopic ultrasound examination.

Regulatory Classification

Identification

A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### play 982323 DEC 1 1 1998 # 510(k) Summary Olympus UM-2R/3R Ultrasonic Probe | Device Name: | Olympus UM-2R/3R Ultrasonic Probe and ancillary<br>equipment for bronchial use | |------------------------------------|---------------------------------------------------------------------------------------------------------------------------| | Common/Usual Name: | Olympus Ultrasonic Probes | | Classification Number<br>and Name: | Class II, 21 CFR 892.1570<br>Diagnostic Ultrasound Transducer<br>Class II 21 CFR 874.4680<br>Bronchoscope and accessories | | Predicate Devices: | Olympus EU-M30 (K951994)<br>Olympus UM-2R/3R (K944610)<br>Olympus EU-M20 (K926514)<br>Olympus EU-M3 (K882061) | | Submitted by: | Laura Storms-Tyler<br>Olympus America Inc.<br>Two Corporate Center Drive<br>Melville, NY 11042-1179<br>(516) 844-5688 | | Summary Preparation<br>Date: | June 26, 1998 | ### Statement of Intended Use: The Olympus UM-2R/3R Ultrasonic Probes have been cleared for use within the gastrointestional tract in 510(k) #K944610. The Olympus UM-2R and UM-3R Ultrasonic Probes have been designed for use in combination with Olympus Endoscopic Ultrasound System for intraluminal sonographic imaging of the upper airways and tracheobronchial tree. ## Device Description In routine examination of the upper airways and tracheobronchial tree, there are situations where the physician prefers to perform an intensive examination, observation, and diagnosis of the upper airways and tracheobronchial tree. The conventional type bronchoscopes limit the physician's ability to access certain areas of interest. The UM-2R/UM-3R Ultrasonic Probes, when used with an endoscope offer transendoscopic access {1}------------------------------------------------ -K982323 p. 242 to the upper airways and tracheobronchial tree. The 2.4 mm insertion tube of these probes can be advanced through strictures and anatomical ducts. The Olympus Ultrasonic Probes to be used in conjunction with bronchoscopes with a minimum channel size of 2.8 mm. The rotation of the transducer is controlled by a probe driving unit. The UM-2R and UM-3R probes produce a B-mode scans using the de-aerated water immersion method and offer 360 degree mechanical/radial scanning of the tissue under observation. The outer diameter of the insertion tube is 2.4 mm and the length is 2050 mm. Both probes incorporate similar design, construction, intended use, and method of operation. The only difference between these two probes is that the UM-2R probe operates at 12 MHz and is compatible with both Olympus EU-M30, EU-M20 and EU-M3 Endoscopic Ultrasound Systems, while the UM-3R probe operates at 20 MHz and is compatible with the EU-M30 and the EU-M20 Endoscopic Ultrasound System. The Olympus EU-M30 Endoscopic Ultrasound Center was cleared for marketing in 510(k) # K951994. The Olympus EU-M20 Endoscopic Ultrasound System was cleared for marketing in 510(k) # K926514 and EU-M3 Endoscopic Ultrasound System was cleared for marketing in the 510(k) # K882061. All components and associated equipment of the UM-2R / UM-3R Ultrasonic Probes will be marketed non-sterile and can be reprocessed as described in the Instruction Manual. ### Safety The Olympus UM-2R and UM-3R Ultrasound Probes are designed, manufactured, and tested in compliance with International Standard IEC 60601-1. The ultrasound characteristics of Olympus UM-2R and UM-3R Ultrasound Probes meet the requirements of the FDA's 510(k) Diagnostic Ultrasound Guidance for 1993 and 1985. When compared to the predicate devices listed in the "Regulatory History" portion of this section, except for intended use, neither ultrasound probe incorporates any significant change in method of operation, material, or design that could affect safety or effectiveness. {2}------------------------------------------------ Image /page/2/Picture/2 description: The image shows a logo for the Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, stacked on top of each other. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC 1 1 1998 Laura Storms-Tyler Director, Regulatory Affairs Olympus America, Inc. 2 Corporate Center Melville, NY 11747-5416 K982323 Re: Trade Name: Olympus UM-3R Ultrasonic Probes and associated ancillary equipment, for bronchial use Requlatory Class: II/21 CFR 892.1570 Product Code: 90 ITX Dated: June 30, 1998 Received: July 2, 1998 Dear Ms. Storms-Tyler: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration. This determination of substantial equivalence applies to the following Ultrasound Systems intended for use with the Olympus UM-2R and UM-3R Ultrasonic Probes, as described in your premarket notification: System Model Number | EU-M30 | |--------| | EU-M20 | | EU-M3 | If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic QS inspections, the FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification does not affect any obligation you may have under sections 531 and 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ Page 2 - Laura Storms-Tyler Please be advised that the determination above is based on the fact that no medical devices have been demonstrated to be safe and effective for in vitro fertilization or percutaneous umbilical blood sampling, nor have any devices been marketed for these uses in interstate commerce prior to May 28, 1976, or reclassified into class I (General Controls) or class II (Special Controls). FDA considers devices specifically intended for in vitro fertilization and percutaneous umbilical blood sampling to be investigational, and subject to the provision of the investigational device exemptions (IDE) regulations, 21 Therefore, your product labeling must be consistent with FDA's CFR, Part 812. position on this use. This letter will allow you to begin marketing your device as described in your The FDA finding of substantial equivalence of your premarket notification. device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation "Misbranding by reference to premarket notification" (21 CFR entitled, Other general information on your responsibilities under the Act may 807.971. be obtained from the Division of Small Manufacturers Assistance at its tollfree number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". If you have any questions regarding the content of this letter, please contact Paul Gammell, Ph.D. at (301) 594-1212. Sincerely yours, Voril A. Klymm Lillian Yin, Ph.D. irector, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ KVDLJLJ Appendix F ﺮ # Diagnostic Ultrasound Indications for Use Form # Diagnostic Ultrasound system and each transducer. Fill out one form for each ultrasound system and each transducer. Fill out one form for each ultrasound system and each as Fill out one form for each ultrasounce of every of the human body as follows: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use: Diagnostic ultrasound imaging of fluid | Clinical Application | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (specify) | Other (specify) | |------------------------------|---|---|-----|-----|-----|---------------|-------------------|------------------------|--------------------|-----------------| | Ophthalmic | | | | | | | | | | | | Fetal | | | | | | | | | | | | Abdominal | | | | | | | | | | | | Intraoperative (specify) | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organ (specify) | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal | | | | | | | | | | | | Conventional | | | | | | | | | | | | Musculo-skeletal Superficial | | | N/P | | | | | | | | | Other (specify) | | | | | | | | | | | N = new indication; P= previously cleared by FDA; E= added under Appendix E Additional Comments: New indications = Intraluminal ultrasound for upper airways and tracheobronchial tree. Previously cleared indications = Intraluminal ultrasound for gastrointestinal tract. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices Prescription Use (Per 21 CFR 801.109) F-3 510(k) Number: K982323 જ્યન {5}------------------------------------------------ K982323 510(k) Number (if known): Device Name: Olympus UM-2R / UM-3R Ultrasonic Probes and associated Ancillary Equipment (for bronchial use) Indications for Use: Olympus UM-2R and UM-3R Ultrasonic Probes have been designed for use in conjunction with Olympus Endoscopic Ultrasound System for intraluminal sonographic imaging of the upper airways and tracheobronchial tree. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) | Concurrence of CDRH, Office of Device Evaluation (ODE) | | |-----------------------------------------------------------------------|--------------------------| | Prescription Use<br>(Per 21 CFR 801.109) | <span></span> | | | OR | | | Over-The-Counter Use | | (Division Sign-Off) | | | Division of Reproductive, Abdominal, ENT,<br>and Radiological Devices | | | 510(k) Number | K982323 | | | (Optional Format 1-2-96) |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...