Diagnostic ultrasound imaging or fluid flow analysis of the human body as tollows: Intraluminal ultrasound for upper airways and tracheobronchial tree.
Device Story
The Olympus BF Type UM40 is an ultrasonic bronchofiberscope used for endoscopic ultrasonic imaging of the upper airways and tracheobronchial tree. It functions as an ultrasonic transducer inserted into the body cavity to perform mechanical scanning of the targeted site. The device transmits ultrasound waves, receives reflected echoes, and converts them into electrical signals for processing and display as images on an external endoscopic ultrasound center. It is intended for use by trained clinicians in a clinical setting, typically in conjunction with an ultrasound center, light source, and display monitor. The device provides real-time visualization of internal structures, aiding clinicians in diagnostic assessments of the airway and bronchial tree, potentially improving diagnostic accuracy and patient management.
Clinical Evidence
Bench testing only. The device was evaluated for compliance with IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-2-18, and CISPR11. Biocompatibility testing was performed on the new insertion tube material to ensure safety.
Technological Characteristics
Mechanical scanning ultrasonic transducer; operates via ultrasound pulse-echo principle. Insertion tube outer surface uses a new biocompatible material. Complies with IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-2-18, and CISPR11. System is designed for integration with external endoscopic ultrasound centers, light sources, and display monitors.
Indications for Use
Indicated for intraluminal ultrasound imaging of the upper airways and tracheobronchial tree in adult patients.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
EUS EXERA Ultrasonic Gastrovideoscope OLYMPUS GF TYPE UM160 (K011886)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
# MAY 1 7 2017
Olympus America, Inc. c/o Mr. Donald James Sherratt Medical Stream Director Intertek Testing Services 70 Cadman Hill Road Boxborough, MA 01719
Re: K021204
Trade/Device Name: Olympus BF Type UM40 Ultrasonic Bronchofiberscope Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: PSV, ITX, ODG Dated: April 15, 2002 Received: April 16, 2002
Dear Mr. Sherratt,
This letter corrects our substantially equivalent (SE) letter of May 1, 2002 and our subsequent corrected SE letter of July 27, 2015.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
## Page 2 - Mr. Donald James Sherratt
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet
address http://www.fda.gov/MedicalDevices/ResourcesforYowIndustry/default.htm. Also. please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go
to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet
address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Eric A. Mann -S 2017.05.17 14:12:01 -04'00'
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### 4.3.1 Indications for Use Form for Ultrasonic Bronchofiberscope OLYMPUS BF TYPE UM40
#### Diagnostic Ultrasound Indications for Use Form
| Clinical Application | Mode of Operation | | | | | | | | | |
|--------------------------|-------------------|---|---|-----|-----|---------------|-------------------|------------------------|--------------------|-----------------|
| | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (specify) | Other (specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative | | | | | | | | | | |
| Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | | |
| Conventional | | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | | |
| Superficial | | | | | | | | | | |
| Other (specify) | | N | | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as tollows:
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
Note 1: Specification for "Other"
Intraluminal ultrasound for upper airways and tranche obronchial tree.
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED] Concurrence of CDRH, Office of Device Evaluation (ODE)
.
Prescription Use Per 21 CFR 801.109)
Rate Pulla
(Division Sign-Off) Division of Reproductive, A and Radiological Devic 5 10(k) Number
Section 4.3 Page 2
{3}------------------------------------------------
## 4.3.1 Indications for Use Form For
# Indications FU-M60 EUS EXERA ENDOSCOPIC ULTRASOUND CENTER With Olympus GF Type UM40 Ultrasonic Endoscopic Transducer
## Diagnostic Ultrasound Indications for Use Form
Intended Use of new transducer: Intraluminal ultrasound for upper airways and tracheobronchial tree
| | Mode of Operation | | | | | | | | | |
|-------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | P | | | | | | | | |
| Transrectal | | P | | | | | | | | |
| Transvaginal | | P | | | | | | | | |
| Transurethral | | P | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | |
| Other (specify) | | P | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: Olympus EU-M60 EUS EXERA Endoscopic Ultrasound Center Previously cleared for use for
Intraluminal ultrasound for upper airways and tracheobronchial tree under K011886
Specification for "Other": Gastrointestinal tract, biliary, pancreatic duct and surrounding organs,
Intraluminal ultrasound for upper alrways and tracheobronchial tree, urinary tract,
female reproductive tract.
(PLEASE DO NOT WRITE BELOW THIS LINE . CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Ra. a. Pally
Division Sign-Off
Division Sign-Off Division of Reproduc and Radiological Device 510(k) Number
Section 4.3 Page 3
{4}------------------------------------------------
## 4.3.1 Indications for Use Form For Olympus EU-M30 ENDOSCOPIC ULTRASOUND CENTER With Olympus GF Type UM40 Ultrasonic Endoscopic Transducer
# Diagnostic Ultrasound Indications for Use Form
Intended Use of new transducer: Intraluminal ultrasound for upper airways and tracheobronchial tree
| | Mode of Operation | | | | | | | | | |
|-------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | P | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | |
| Other (specify) | | P | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
Intraluminal ultrasound for upper airways and tracheobronchial tree with transducer UM 2R/3R under K902323
Specification for "Other": Gastrointestinal tract, biliary, pancreatic duct and surrounding organs. Cleared under
K982323
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Ra-C. Phillips
Sian-Off Division of Reproductive, and Radiological Devices 510(k) Number
Section 4.3 Page 4
{5}------------------------------------------------
K021204
# MAY 01 2002
## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21CFR, Part 807, Subpart E, Section 807.92.
## A. Submitter's name, address, telephone number, initial importer, contact person
#### 1. Manufacturer of the subject device
Name & Address of Manufacturer; Olympus Optical Co,. Ltd.
Registration Number : Address, Phone and Fax of R & D Department Endoscope Division
2-3-1 Shinjukuku Monolis Nishi-Shinjuku Shinjuku-ku, Tokyo, 163-0914 Japan 810047 2951 Ishikawa-cho Hachioji-shi, Tokyo 192-8507 Japan TEL 81-426-42-2891 FAX 81-426-46-5613
#### 2. Initial Importer
Name: Address: Olympus America Inc. Two Corporate Center Drive Melville, NY 11747-3157 TEL 516-844-5688 FAX 516-844-5416
#### 3. Name of Contact Person
Name:
Tsuyoshi Yanai Regulatory Affairs Manager, Olympus Optical Co., Ltd.
Address, Phone and Fax:
2951 Ishikawa-cho Hachioji-shi, Tokyo 192-8507 Japan TEL 81-426-42-2891 FAX 81-426-46-5613
{6}------------------------------------------------
#### B. Device Name, Common Name
- 1. Common/Usual Name
Ultrasonic endoscope
#### 2. Device Name
· Ultrasonic Bronchofiberscope OLYMPUS BF TYPE UM40
#### 3.Classification Name
| | FR Number | Product Code | Class |
|----------------------------------|-----------|--------------|-------|
| Endoscope and accessories | 876.1500 | KOG | II |
| Diagnostic Ultrasound Transducer | 892.1570 | ITX | II |
#### C. Identification of the predicate or legally marketed device
The following devices information demonstrates that this device is substantially equivalent to a legally marketed, predicate medical device.
| Device Name | #K |
|-----------------------------------------------------------------------------------------------|---------|
| EUS EXERA Ultrasonic Gastrovideoscope OLYMPUS GF<br>TYPE UM160 | K011886 |
| Olympus UM-2R/ 3R Ultrasonic Probes and associated<br>ancillary equipment (for bronchial use) | K982323 |
| BF-240/P240/1T240 Bronchovideoscope & Accessories | K963033 |
| Bronchoscope BF-N20 | K910423 |
### D. Device Description
#### 1. Summary
This subject device has been designed to be used with an OLYMPUS endoscopic ultrasound center, light source, documentation equipment, display monitor, endotherapy accessories and other ancillary equipment for endoscopic ultrasonic imaging of the upper airways and trancheobronchial tree.
#### 2. Design
This subject device is designed to comply with the standards listed below.
| IEC 60601-1 |
|----------------|
| IEC 60601-1-1 |
| IEC 60601-1-2 |
| IEC 60601-2-18 |
| CISPR11 |
{7}------------------------------------------------
#### 3. Materials
The material for Insertion Tube Outer Surface of this subject device has a new patientr no metional aterial. The biocompatibility test reports of the new material show that the new material is safe for its intended use.
#### E. Intended Use:
The intended use of this subject device, as defined by FDA guidance documents, is: Other
1)Intraluminal ultrasound for upper airways and tracheobronchial tree
#### F. Technological Characteristics:
This device operates identically to the predicate devices in that the transducer of the endoscope that is inserted into the body cavity mechanically scans the targeted site. Chaooops that is interial in the transducer is used as an ultrasound source to transmit sound waves. Sound waves are reflected back to the transducer and converted to electrical signals that are processed and displayed as images.
Technological Characteristics of this device is identical to the predicated devices identified in item C.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.