The ULTRASONIC PROBE UM-S20-17S has been designed to be used with an endoscopic ultrasound center, a probe driving unit, and an endoscope for intraluminal ultrasonic imaging of the upper airways and tracheobronchial tree. The ULTRASONIC PROBE UM-S20-20R has been designed to be used with an endoscopic ultrasound center, a probe driving unit, and an endoscope for intraluminal ultrasonic imaging of the upper airways and tracheobronchial tree.
Device Story
Ultrasonic probes (UM-S20-17S, UM-S20-20R) designed for intraluminal imaging of upper airways and tracheobronchial tree; used with endoscope, probe driving unit, and ultrasound center. Probes inserted through endoscope channel; 20MHz piezoelectric transducer at tip rotates mechanically to produce 360-degree radial B-mode sonograms. Transducer converts electrical signals to ultrasound waves; receives reflected waves; converts back to electrical signals for image generation by ultrasound center. Used in healthcare facilities by trained professionals. Output displayed on ultrasound center for clinical visualization; aids in diagnosis and observation of airway structures. Benefits include high-resolution intraluminal imaging for clinical decision-making.
Clinical Evidence
No clinical data were collected to support the performance of the subject devices. Substantial equivalence is supported by non-clinical bench testing, including acoustic output, durability, measurement accuracy, human factors, biocompatibility, and electrical safety/EMC testing.
Technological Characteristics
20MHz piezoelectric transducer; mechanical radial scanning; 360-degree field of view; B-mode. Materials: Liquid paraffin (High White 70) ultrasonic medium. Sterilization: ETO. Electrical safety: IEC 60601-1, IEC 60601-2-18, IEC 60601-2-37. EMC: IEC 60601-1-2. Biocompatibility: ISO 10993 series. Reprocessing: ISO 17664-1, AAMI TIR12, ANSI AAMI ST98/ST58.
Indications for Use
Indicated for intraluminal ultrasonic imaging of the upper airways and tracheobronchial tree in adult patients. Used with endoscopic ultrasound center, probe driving unit, and endoscope. Prescription use only.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 11, 2025
Olympus Medical Systems Corporation
% Brenda Geary
Senior Manager Regulatory Affairs
Olympus Surgical Technologies of the Americas
800 West Park Drive
WESTBOROUGH MA 01581
Re: K250762
Trade/Device Name: Ultrasonic Probe UM-S20-17S (UM-S20-17S); Ultrasonic Probe UM-S20-20R (UM-S20-20R)
Regulation Number: 21 CFR 892.1570
Regulation Name: Diagnostic Ultrasonic Transducer
Regulatory Class: Class II
Product Code: ITX, EOQ
Dated: June 11, 2025
Received: June 11, 2025
Dear Brenda Geary:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250762 - Brenda Geary
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250762 - Brenda Geary
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MARJAN NABILI -S for
Yanna Kang, Ph.D.
Assistant Director
Mammography and Ultrasound Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250762
Device Name
ULTRASONIC PROBE UM-S20-17S (UM-S20-17S)
ULTRASONIC PROBE UM-S20-20R (UM-S20-20R)
Indications for Use (Describe)
The ULTRASONIC PROBE UM-S20-17S has been designed to be used with an endoscopic ultrasound center, a probe driving unit, and an endoscope for intraluminal ultrasonic imaging of the upper airways and tracheobronchial tree.
Modes of operation:
The Mode of Operation is B mode.
Operator qualifications:
Appropriately trained healthcare professionals.
Device use settings:
Healthcare facility
The ULTRASONIC PROBE UM-S20-20R has been designed to be used with an endoscopic ultrasound center, a probe driving unit, and an endoscope for intraluminal ultrasonic imaging of the upper airways and tracheobronchial tree.
Modes of operation:
The Mode of Operation is B mode.
Operator qualifications:
Appropriately trained healthcare professionals.
Device use settings:
Healthcare facility
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}
OLYMPUS
TRADITIONAL 510(k)
ULTRASONIC PROBES UM-S20-17S and UM-S20-20R
510(k) Summary
# 510(k) Summary: K250762
## 1. COMPANY INFORMATION
### Applicant
OLYMPUS MEDICAL SYSTEMS CORPORATION
2951 Ishikawa-cho, Hachioji-shi
Tokyo 192-8507, Japan
FDA Establishment Registration #: 8010047
### Official Correspondent
Wendy Perreault
c/o Olympus Surgical Technologies of the Americas
800 West Park Drive
Westborough, MA 01581
Cell: 404-542-5854
Email: wendy.perreault@olympus.com
### Manufacturing Site
SHIRAKAWA OLYMPUS CO., LTD.
3-1 Okamiyama, Odakura
NISHIGO-MURA
NISHISHIRAKAWA-GUN Fukushima, JAPAN 961-8061
- Date Prepared: 9 July 2025
## 2. PRODUCT INFORMATION
Trade Name: ULTRASONIC PROBE UM-S20-17S (UM-S20-17S); ULTRASONIC PROBE UM-S20-20R (UM-S20-20R)
Common Name: Diagnostic Ultrasound Transducer
Classification Number: 21 CFR 892.1570, 21 CFR 874.4680
Product Code Names: Transducer, Ultrasonic Diagnostics; Bronchoscope (Flexible or Rigid) and Accessories
Product Codes: ITX, EOQ
{5}
OLYMPUS
TRADITIONAL 510(k)
ULTRASONIC PROBES UM-S20-17S and UM-S20-20R
510(k) Summary
Regulatory Class: II
Classification Panel: Radiology
## 3. PREDICATE DEVICE
The Subject devices are equivalent to the Predicate device listed below in Table 1.
Table 1: Predicate device for ULTRASONIC PROBE UM-S20-17S and UM-S20-20R
| Device name | 510(k) Submitter | 510(k) No. |
| --- | --- | --- |
| FUJIFILM Endoscopic Ultrasonic Probe (P2612S-L); Endoscopic Ultrasonic Probe (P2620S-L) | Fujifilm Healthcare Americas Corporation | K231666 |
The Predicate device has not been subject to a design-related recall.
The Reference device supporting the performance of the Subject devices is listed below in Table 2.
Table 2: Reference device for ULTRASONIC PROBE UM-S20-17S and UM-S20-20R
| Device name | 510(k) Submitter | 510(k) No. |
| --- | --- | --- |
| UM-3R Ultrasonic Probe | OLYMPUS MEDICAL SYSTEMS CORPORATION | K063683 |
## 4. DEVICE DESCRIPTION
The Ultrasonic Probes have been designed to be used with an Olympus Endoscopic Ultrasound Center, a Probe Driving Unit, other ancillary equipment and an endoscope for intraluminal ultrasonic imaging of the upper airways and tracheobronchial tree and surrounding organs.
Ultrasonic Probes UM-S20-17S and UM-S20-20R are designed to be used in conjunction with bronchoscopes. The Probes are inserted into the patient through a channel of the endoscope.
The Ultrasonic Probes consist of an insertion tube and a connector section. The connector section is connected to the Probe Driving Unit and the Probe Driving Unit is connected to the Ultrasound Center.
The Ultrasonic Probe sends and receives electrical signals to and from the Ultrasound Center through the Probe Driving Unit. The Probes use a 20MHz frequency piezoelectric transducer and produce B-mode scan. They produce 360-degree mechanical/radial sonograms.
Page 2 of 8
{6}
OLYMPUS
TRADITIONAL 510(k)
ULTRASONIC PROBES UM-S20-17S and UM-S20-20R
510(k) Summary
The transducer is built into the insertion tube at the tip of the Probe. The transducer is rotated by the Probe Driving Unit within the insertion tube.
The transducer converts the electrical signal to the ultrasound wave, sends it to the object, receives the reflected wave from the object and converts it to the electrical signal. The electrical signal is input to the Endoscopic Ultrasound Center and the ultrasound image is generated by the Endoscopic Ultrasound Center.
UM-S20-17S and UM-S20-20R use direct contact method only.
The Subject devices submitted for clearance each include one (1) major component: the Ultrasonic Probe, which is packaged with a Probe Holder (MH-245) and a Water-resistant Cap (MH-244).
The target population for this device is Adults.
# 5. INDICATIONS FOR USE
- The ULTRASONIC PROBE UM-S20-17S has been designed to be used with an endoscopic ultrasound center, a probe driving unit, and an endoscope for intraluminal ultrasonic imaging of the upper airways and tracheobronchial tree.
[Modes of operation]
The Mode of Operation is B mode.
[Operator qualifications]
Appropriately trained healthcare professionals.
[Device use settings]
Healthcare facility
- The ULTRASONIC PROBE UM-S20-20R has been designed to be used with an endoscopic ultrasound center, a probe driving unit, and an endoscope for intraluminal ultrasonic imaging of the upper airways and tracheobronchial tree.
[Modes of operation]
The Mode of Operation is B mode.
[Operator qualifications]
Page 3 of 8
{7}
OLYMPUS
TRADITIONAL 510(k)
ULTRASONIC PROBES UM-S20-17S and UM-S20-20R
510(k) Summary
Appropriately trained healthcare professionals.
[Device use settings]
Healthcare facility
## 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
Compared to the Predicate and Reference devices, the Subject devices offer similar functions. A detailed comparison of the technological characteristics of the devices is provided in Table 3 below. These differences in technological characteristics do not raise new questions of safety or effectiveness.
Page 4 of 8
{8}
OLYMPUS
TRADITIONAL 510(k)
ULTRASONIC PROBES UM-S20-17S and UM-S20-20R
510(k) Summary
Table 3: Subject, Predicate and Reference Device Comparison of Technological Characteristics
| Item | Subject Devices (SD) | | Predicate Device (PD) | Reference Device (RD) |
| --- | --- | --- | --- | --- |
| | UM-S20-17S | UM-S20-20R | FUJIFILM Endoscopic Ultrasonic Probe (P2612S-L); Endoscopic Ultrasonic Probe (P2620S-L) | UM-3R |
| Manufacturer | OLYMPUS MEDICAL SYSTEMS CORPORATION | OLYMPUS MEDICAL SYSTEMS CORPORATION | Fujifilm Healthcare Americas Corporation | OLYMPUS MEDICAL SYSTEMS CORPORATION |
| 510(k) # | N/A | N/A | K231666 | K063683 |
| Classification Product Code | ITX | ITX | ITX | IYN |
| Subsequent Product Code(s) | EOQ | EOQ | N/A | ITX, IYO, ODG |
| Indications for Use | The ULTRASONIC PROBE UM-S20-17S has been designed to be used with an endoscopic ultrasound center, a probe driving unit, and an endoscope for intraluminal ultrasonic imaging of the upper airways and tracheobronchial tree. [Modes of operation] The Mode of Operation is B mode. [Operator qualifications] Appropriately trained healthcare professionals. [Device use settings] Healthcare facility | The ULTRASONIC PROBE UM-S20-20R has been designed to be used with an endoscopic ultrasound center, a probe driving unit, and an endoscope for intraluminal ultrasonic imaging of the upper airways and tracheobronchial tree. [Modes of operation] The Mode of Operation is B mode. [Operator qualifications] Appropriately trained healthcare professionals. [Device use settings] Healthcare facility | This product is a medical ultrasonic probe. It is intended for the observation and diagnosis of the gastrointestinal tract and surrounding organs under the management of physicians at medical facilities. This product is intended for adults. Modes of Operation: B-mode Never use this product for any other purposes. | These instruments have been designed to be used with an Olympus Endoscopic Ultrasound Center, a Water Supply Unit (For Ultrasonic Endoscope), a Probe Driving Unit, a Probe/Irrigation Plug and an endoscope for intralumenal ultrasonic imaging of the gastrointestinal tract and surrounding organs, the upper airways and tracheobronchial tree and the urinary organs. |
| Frequency | 20 MHz | 20 MHz | P2612S-L: 10MHz±15% P2620S-L: 17MHz±15% | 20 MHz |
Page 5 of 8
{9}
OLYMPUS
TRADITIONAL 510(k)
ULTRASONIC PROBES UM-S20-17S and UM-S20-20R
510(k) Summary
| Item | Subject Devices (SD) | | Predicate Device (PD) | | Reference Device (RD) | |
| --- | --- | --- | --- | --- | --- | --- |
| | UM-S20-17S | UM-S20-20R | FUJIFILM Endoscopic Ultrasonic Probe (P2612S-L); Endoscopic Ultrasonic Probe (P2620S-L) | | UM-3R | |
| Display mode | B-mode | B-mode | B-mode | | B-mode | |
| Compatible Endoscopic Ultrasound System | OLYMPUS : EU-ME2
EU-ME2 Premier Plus
EU-ME3 | OLYMPUS : EU-ME2
EU-ME2 Premier Plus
EU-ME3 | FUJIFILM Ultrasonic Processor SP-900 | | OLYMPUS XEU-M60A | |
| Number of Transducers | 1 | 1 | - | | 1 | |
| Shape of Transducer | Rectangle | Rectangle | - | | Rectangle | |
| Size of Transducer | 0.75 × 2.3mm | 1.0 × 2.3mm | - | | 1.6 × 2.4mm | |
| Scanning direction | Perpendicular to the direction of insertion | Perpendicular to the direction of insertion | - | | Perpendicular to the direction of insertion | |
| Scanning field of view (Scanning area) | 360° | 360° | - | | 360° | |
| Frame rate | 6.67rps | 6.67rps | - | | 6.67rps | |
| Axial resolution | 2mm or less | 2mm or less | 1mm or less | | 2mm or less | |
| Lateral resolution | 2mm or less | 2mm or less | 3mm or less | | 2mm or less | |
| Scanning method | Mechanical, radial scanning | Mechanical, radial scanning | Mechanical, radial scanning | | Mechanical, radial scanning | |
| Contact method | Direct contact method | Direct contact method | - | | Direct contact method | |
| | N/A | N/A | - | | Sterile De-aerated water immersion method | |
| | N/A | N/A | - | | Balloon method (Balloon sheath [MH-246R] is necessary). | |
| Insertion tube maximum outer diameter (mm) | φ 1.45 (Distal end side 1085mm)
φ 1.8 (Connector side) | φ 1.75 (Distal end side 850mm)
φ 2.05 (Connector side) | 2.6 | | φ 2.5 | |
| Insertion tube outer diameter (mm) | φ 1.4 (Distal end side 1085mm)
φ 1.7 (Connector side) | φ 1.7 (Distal end side 850mm)
φ 2.0 (Connector side) | 2.5 | | φ 2.4 | |
Page 6 of 8
{10}
OLYMPUS
TRADITIONAL 510(k)
ULTRASONIC PROBES UM-S20-17S and UM-S20-20R
510(k) Summary
| Item | Subject Devices (SD) | | Predicate Device (PD) | | Reference Device (RD) | |
| --- | --- | --- | --- | --- | --- | --- |
| | UM-S20-17S | UM-S20-20R | FUJIFILM Endoscopic Ultrasonic Probe (P2612S-L); Endoscopic Ultrasonic Probe (P2620S-L) | | UM-3R | |
| Ultrasonic medium | Liquid paraffin (High White 70) | Liquid paraffin (High White 70) | - | | Liquid paraffin (High White 70) | |
| Working length (mm) | 2150 | 2050 | 2620 | | 2050 | |
| Total length (mm) | 2225 | 2140 | - | | 2140 | |
| Sterilization method for reprocessing | ETO | ETO | ETO | | N/A | |
| Compatible Olympus Reprocessor for cleaning and disinfection | OER-Pro
OER-Elite | OER-Pro
OER-Elite | N/A | | N/A | |
| Electrical safety | Compliance to IEC 60601-1, IEC 60601-2-18, IEC 60601-2-37 | Compliance to IEC 60601-1, IEC 60601-2-18, IEC 60601-2-37 | Compliance to IEC 60601-1, IEC 60601-2-37 | | Compliance to IEC 60601-1, IEC 60601-2-18 | |
| EMC | Compliance to IEC 60601-1-2 | Compliance to IEC 60601-1-2 | Compliance to IEC 60601-1-2 | | Compliance to IEC 60601-1-2 | |
Page 7 of 8
{11}
OLYMPUS
TRADITIONAL 510(k)
ULTRASONIC PROBES UM-S20-17S and UM-S20-20R
510(k) Summary
# 7. SUMMARY OF NON-CLINICAL PERFORMANCE DATA
Results of the following testing support the safety and performance of the Subject devices and demonstrate their equivalence to the Predicate device and Reference device; all testing passed/met the acceptance criteria, demonstrated compliance with the cited standards and FDA Guidance shown below, and supported equivalent safety and performance to the Predicate and Reference devices:
- Performance Testing – Bench (including Acoustic Output in compliance with IEC 60601-2-37, IEC 62359, IEC 62127-1, and FDA Guidance: Marketing Clearance of Diagnostic Ultrasound Systems and Transducers - Guidance for Industry and Food and Drug Administration Staff (February 21, 2023); Durability, Measurement Accuracy; and Human Use Factors following FDA Guidance Documents Applying Human Factors and Usability Engineering to Medical Devices: Guidance for Industry and Food and Drug Administration Staff (issued February 3, 2016) and Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling Guidance for Industry and Food and Drug Administration Staff (issued March 17, 2015).
- Biocompatibility Testing (ISO 10993-1; ISO 10993-5; ISO 10993-10; ISO 10993-12; ISO 10993-17; ISO 10993-18; ISO 10993-23)
- Reprocessing Validation (ISO 17664-1; AAMI TIR12; ANSI AAMI ST98; ANSI AAMI ST58; ISO 11135; ISO 11138-2)
- Electrical Safety/EMC Testing (IEC 60601-1; IEC ES60601-1; IEC 60601-1-2; IEC 60601-2-18; IEC 60601-2-37; IEC TR 60601-4-2)
# 8. SUMMARY OF CLINICAL PERFORMANCE DATA
No clinical data were collected to support the performance of the Subject devices.
# 9. CONCLUSION
The results of non-clinical performance testing demonstrate that the Subject devices are as safe and effective as the Predicate device to support a substantial equivalence determination.
Page 8 of 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.