EMMA 1.5T MRI System

K183621 · Time Medical Limited · LNH · Feb 22, 2019 · Radiology

Device Facts

Record IDK183621
Device NameEMMA 1.5T MRI System
ApplicantTime Medical Limited
Product CodeLNH · Radiology
Decision DateFeb 22, 2019
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 892.1000
Device ClassClass 2

Indications for Use

The EMMA 1.5T MRI System is indicated for use as a magnetic resonance diagnostic device (MRDD) which produces transverse, sagittal, coronal and oblique cross-sectional images, and those display the internal structure and/or function of the head, body, or extremities. Depending on the region of interest, contrast agents may be used. These images when interpreted by trained physician yield information that may assist medical diagnosis.

Device Story

The EMMA 1.5T MRI System is a 1.5T superconducting magnet MRI scanner used in clinical settings to produce cross-sectional images of the human body. It consists of a magnet, gradient coils, RF transmission/receive coils, patient table, and a Windows-based workstation running Prodiva software. The system operates by placing the patient in a strong magnetic field, exciting nuclei with RF signals, and acquiring relaxation signals via receiver coils. The spectrometer controls pulse sequences, while the workstation reconstructs raw data into images. Operators (technicians/radiologists) use the system to configure scans and manage data. The resulting images are interpreted by physicians to assist in medical diagnosis, providing anatomical and functional information to benefit patient care.

Clinical Evidence

No formal clinical trials were conducted. Evidence consists of bench testing (biocompatibility per ISO 10993-1, electrical safety/EMC per IEC 60601-1/2-33, acoustic noise per NEMA MS 4-2010, and performance per NEMA MS-1, 2, 3, 5, 6, 8) and a comparison of sample clinical images between the subject and predicate devices to demonstrate consistent diagnostic image quality.

Technological Characteristics

1.5T superconducting magnet system; materials include FRP and ABS. Components: gradient coils, RF transmission/receive coils, spectrometer, and multi-core PC workstation. Connectivity: DICOM compliant. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-33, NEMA MS series. Software: Prodiva (Windows-based).

Indications for Use

Indicated for use as a magnetic resonance diagnostic device (MRDD) to produce transverse, sagittal, coronal, and oblique cross-sectional images of the head, body, or extremities for diagnostic purposes. May be used with contrast agents. For use by trained physicians.

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is blue. Time Medical Limited % Mr. Nick Tse Compliance Manager G/F, Bio-Informatics Center No., 2 Science Park West Avenue Shatin, N.T., Hong Kong CHINA February 22, 2019 Re: K183621 Trade/Device Name: EMMA 1.5T MRI System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH Dated: December 12, 2018 Received: December 26, 2018 Dear Mr. Tse: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Michael D. O'Hara For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### 2.2 INDICATIONS FOR USE FDA Indications for Use Form (FDA Form 3881) is shown below and at Annex K. | DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration<br>Indications for Use | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020<br>See PRA Statement below. | |------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------| | 510(k) Number (if known) | | | K183621 | | | Device Name | | EMMA 1.5T MRI System #### Indications for Use (Describe) The EMMA 1.5T MR System is a magnetic resonance diagnostic device (MRDD) which produces transverse, sagittal, coronal and oblique cross-sectional images, and those display the internal structure and/or function of the head, body, or extremities. Depending on the region of interest, contrast agents may be used. These images when interpreted by trained physician yield information that may assist medical diagnosis. | Type of Use (Select one or both, as applicable) | | | | | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|--|--|--| | 2 Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | | | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. | | | | | | *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .* | | | | | | The burden time for this collection of information is estimated to average 79 hours per response, including the<br>time to review instructions, search existing data sources, gather and maintain the data needed and complete<br>and review the collection of information. Send comments regarding this burden estimate or any other aspect<br>of this information collection, including suggestions for reducing this burden, to: | | | | | | Department of Health and Human Services<br>Food and Drug Administration<br>Office of Chief Information Officer<br>Paperwork Reduction Act (PRA) Staff<br>PRAStaff@fda.hhs.gov | | | | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of<br>information unless it displays a currently valid OMB number." | | | | | FORM FDA 3881 (7/17) Page 1 of 1 POC Fultidaing Services (301) 443-4345 EF {3}------------------------------------------------ K183621 #### 2.1 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS This 510(k) Summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. | Date of Submission: | February, 9, 2018 | |-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Sponsor: | Time Medical Limited<br>G/F Bio-Informatics Centre, No. 2 Science Park West Avenue,<br>Hong Kong Science Park, Shatin, New Territories, Hong Kong, China<br>Contact Person: Nick Tse - Compliance Manager<br>Telephone number: +852 21561711 Fax.: +852 2156 0908 | | Correspondent: | Time Medical Limited<br>G/F Bio-Informatics Centre, No. 2 Science Park West Avenue,<br>Hong Kong Science Park, Shatin, New Territories, Hong Kong, China<br>Contact Person: Nick Tse - Compliance Manager<br>Telephone number: +852 21561711 Fax.: +852 2156 0908 | | Filing Device: | EMMA 1.5T MRI System | | Classification Name: | System. Nuclear Magnetic Resonance Imaging | | Regulatory Description: | Magnetic Resonance Diagnostic Device (MRDD) | | Classification Class | Class II / LNH / 21 CFR 892. 1000 | | Intended Use: | The EMMA 1.5T MRI System is indicated for use as a magnetic resonance<br>diagnostic device (MRDD) which produces transverse, sagittal, coronal and<br>oblique cross-sectional images, and those display the internal structure<br>and/or function of the head, body, or extremities. Depending on the region of<br>interest, contrast agents may be used. These images when interpreted by<br>trained physician yield information that may assist medical diagnosis. | | Device Description: | The EMMA 1.5T MRI System is a 1.5T superconducting magnet MRI<br>system which produces transverse, sagittal, coronal and oblique<br>cross-sectional images, and those display the internal structure and/or<br>function of the head, body, or extremities. It is composed of Magnet, Magnet<br>Enclosure, Patient Table, Gradient Coil, Transmission Coil, Receiver Coil,<br>Client PC, and Imaging Cabinet. The system software, Prodiva, a<br>Windows-based software, is an interactive program with user friendly<br>interface. The device is conformed to IEC and DICOM standards. | #### General Comparison to Predicate Device: | Comparison Element | Filing Device: | Predicate Device: | |---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Number | N/A | K161567 | | Model | 15000-01 | SIGNA Voyager 1.5T | | Device Name | EMMA 1.5T MRI System | SIGNA Voyager | | Applicant | Time Medical Limited | GE Healthcare | | Applicant Address | Unit G01-G03, 05 & 08, G/F,<br>Bio-Informatics Centre, Hong Kong<br>Science Park, Shatin, N.T | 3200 N. Grandview Blvd.<br>Waukesha, WI 53188, USA | | Classification Name | System, Nuclear Magnetic<br>Resonance Imaging | Magnetic Resonance Diagnostic<br>Device | | Product Code | LNH | LNH, MOS | | Regulation Number | 21 CFR 892.1000 | | | Panel | Radiology | | | Class | Class II | | | Indications for Use | The EMMA 1.5T MRI System is<br>indicated for use as a magnetic<br>resonance diagnostic device<br>(MRDD) which produces<br>transverse, sagittal, coronal and<br>oblique cross-sectional images,<br>and those display the internal<br>structure and/or function of the | The SIGNA Voyager is a whole body<br>magnetic resonance scanner<br>designed to support high resolution,<br>high signal-to-noise ratio, and short<br>scan time imaging. The SIGNA<br>Voyager is indicated for use as a<br>diagnostic imaging device to produce<br>axial, sagittal, coronal, and oblique | {4}------------------------------------------------ The filing device has the same US FDA classification including classification name, product code, regulation number, panel, and class with the predicate device. The filing device has the same major indications for use as the predicate device produces transverse, sagittal, coronal and oblique cross-sectional images, and those display the internal structure and/or function of the head, body, or extremities. Depending on the region of interest, contrast agents may be used. These images when interpreted by trained physician yield information that may assist medical diagnosis. The minor differences in indications for use do not constitute any safety and effectiveness issue, as indicated in performance data provided. | General | Filing Device | Predicate Device | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|--------------------------------------------------| | Design | Cylindrical | Cylindrical | | Materials | FRP, ABS | | | Power Source | Isolated power supply | | | Performance Data | Comply with:<br>- NEMA MS standards<br>- IEC standards: IEC 60601-1,<br>IEC60601-1-2,<br>IEC60601-2-33 | | | Magnet | Filing Device | Predicate Device | | Type | Superconducting | | | Field Strength | 1.5T | | | Weight | 3,200 kg | 3,900 kg | | Overall Dimensions | 1497 x 1880 x 2421 mm<br>(L x W x H) | 1730 x 2070 x 2360 mm<br>(L x W x H) | | Cryogen Type | Liquid Helium | | | Spatial Homogeneity | 1.45 ppm at 40 cm DSV | 0.40 ppm at 40 cm DSV | | Fringe Field | 5 gauss line < 2.5m radial x 4m<br>axial | 2.5 m radial x 4.0 m axial | | Magnet Room Lighting | For sufficient and even lighting, 16<br>overhead lamps (100 lumen each)<br>recommended | Available | | Room Ventilation | HVAC is installed for ventilation. | | | Gradient System | Filing Device | Predicate Device | | Amplitude | 43.6 mT/m | 36 mT/m | | Rise time | 0.17 ms | 0.24 ms | | Slew rate | 257 T/m/s | 150 T/m/s | | Patient Table | Filing Device | Predicate Device | | Dimensions (L x W x H) | 92.1 x 24.6 x 34.0 inch<br>(2340 x 624 x 864 mm) | 96.5 x 27.6 x 36.7 inch<br>(2450 x 700 x 932 mm) | | Patient positioning | Laser Localizer assisted automatic<br>patient table positioning.<br>Manual Position mode | Laser Localizer Available | | Max Patient Weight | 200kg | 250 kg | | Patient-control room<br>communication | Microphones and speaker for<br>two-way communication between<br>patient and radiologist | Magnacoustic Communication &<br>Music System | | Workstation | Filing Device | Predicate Device | | Model | Multi-Processing, Multi-Core PC<br>with 24-inch monitor | Quad-Core Intel Xeon with 24-inch<br>monitor | | Memory and Capacity | 64GB of RAM, 250GB SSD Storage | 32GB of RAM, 2 x 400 GB SSD<br>Storage | | Operating System | Microsoft Windows | Linux OS | | Receive Coils | Filing Device | Predicate Device | | Multi-Channel Head Coil | ✓ | ✓ | | Multi-Channel Head and<br>neck Coil | ✓ | ✓ | | Multi-Channel Shoulder<br>Coil | ✓ | ✓ | | Multi-Channel Spine Coil | ✓ | ✓ | | Multi-Channel Body Coil | ✓ | ✓ | | Multi-Channel Closed<br>Breast Coil | ✓ | ✓ | | Multi-Channel Knee Coil | ✓ | ✓ | | Multi-Channel Ankle Coil | ✓ | ✓ | | Multi-Channel Wrist Coil | ✓ | ✓ | | Multi-Channel<br>Multi-purpose Coils | ✓ | ✓ | | Pulse Sequences | Filing Device | Predicate Device | | T1 Spin Echo | ✓ | ✓ | | T1, T2 Fast Spin Echo,<br>HASTE | ✓ | ✓ | | T1, T2* Gradient Echo | ✓ | ✓ | | FLAIR | ✓ | ✓ | | TOF Angiography | ✓ | ✓ | | Inversion Recovery | ✓ | ✓ | | Fat-Water Separation | ✓ | ✓ | | Chemical Fat Sat | ✓ | ✓ | | MRCP | ✓ | ✓ | | DWI EPI | ✓ | ✓ | | CMGE | ✓ | ✓ | | SWI | ✓ | ✓ | | 2D Image Processing<br>Features | Filing Device | Predicate Device | | Image panning, zooming,<br>windowing | ✓ | ✓ | | Image flipping, inverting<br>and rotation | ✓ | ✓ | | Statistical Analysis | ✓ | ✓ | | Geometric Measurement | ✓ | ✓ | | Phase Image | ✓ | ✓ | | CINE | ✓ | ✓ | | Shutter | ✓ | ✓ | | Image Filtering | ✓ | ✓ | | Image Subtraction | ✓ | ✓ | | Slice Profile | ✓ | ✓ | | Histogram | ✓ | ✓ | | 3D Image Processing<br>(MPR/Max.IP) | Filing Device | Predicate Device | | Image panning, zooming,<br>windowing | ✓ | ✓ | | Cutting Operation Output<br>Stack | ✓ | ✓ | | Image export | ✓ | ✓ | | The filing device basically utilizes same technologies as predicate device, and can give consistent<br>results when using the images in medical diagnosis. The minor differences in technological<br>characteristics do not constitute any safety and effectiveness issue, as indicated in performance data<br>provided. | | | #### Comparison of Technological Characteristics with the Predicate Device: {5}------------------------------------------------ {6}------------------------------------------------ {7}------------------------------------------------ | Technological<br>Characteristic | EMMA 1.5T MRI system is based on the principle that certain atomic nuclei<br>present in the human body, when placed in a strong magnetic field and<br>excited by a radio frequency signal at the precession frequency, will emit a<br>relaxation signal. The emitted relaxation signals are acquired by signal<br>receiving coils and are analyzed by the system. Human-interpretable<br>images are reconstructed by and stored in the computer and displayed on<br>the screen. | |---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | The principle components of EMMA 1.5T MRI system includes | | | Superconducting magnet with corresponding cooling devices Gradient system RF transmission and receive system Cooling system for different components of the system Patient table and support components Spectrometer Server/Client PC and application software | | | The superconducting magnet is made of coils of superconducting wires<br>and is cooled to and maintained at cryogenic temperature by an enclosed<br>dedicated cooling system. When energized, current flows continuously<br>inside the superconducting wires as a closed loop; thus producing a<br>constant magnetic field. This has the same major technological<br>characteristics as the predicate device. | | | The gradient system consists of a gradient coil and the gradient amplifier.<br>The gradient coil produces varying linear gradient fields in the three<br>orthogonal spatial axes with currents given by the gradient amplifier. The<br>gradient coil can also produce various non-linear gradients fields to<br>improve the homogeneity of the main magnetic field. This has the same<br>major technological characteristics as the predicate device. | | | The RF transmission and receive system consists of the RF amplifier, the<br>RF transmission coil and various RF receiver coils for receiving the<br>relaxation signals emitted from the subject after RF transmission. They are<br>coils made of copper wires tuned to transmit and receive around the<br>designed frequency corresponding to the magnetic field strength. This has<br>the same major technological characteristics as the predicate device. | | | The cooling system maintains suitable operating temperatures for<br>amplifiers and other system devices. This has the same major<br>technological characteristics as the predicate device. | | | The patient table and support components provide a comfortable platform<br>for the patient and allows the operator to easily position the patient to the<br>location for imaging. This has the same major technological characteristics<br>as the predicate device. | | | The spectrometer controls the execution of the pulse sequence for RF<br>transmit and receive and gradient output. This has the same major<br>technological characteristics as the predicate device. | | | The server/client PC and the application software provides the operator a<br>platform to configure scans and manage scanned data. Raw data received<br>from the spectrometer is also reconstructed here. This has the same major<br>technological characteristics as the predicate device. | | | technological characteristics as the predicate device. | | | Therefore, the same scientific technology theory is applied to both the<br>EMMA 1.5T MRI system and predicate device. | | | Any minor differences in physical attributes do not constitute any safety and<br>effectiveness issue, as indicated in performance data provided. | {8}------------------------------------------------ {9}------------------------------------------------ | Nonclinical Testing<br>Summary: | Like the predicate device, the following performance data were provided in<br>support of the substantial equivalence determination. | |---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Biocompatibility testing<br>The biocompatibility evaluation for the filing device was conducted in<br>accordance with ISO 10993-1 "Biological Evaluation of Medical Devices –<br>Part 1: Evaluation and Testing Within a Risk Management Process" as<br>recognized by FDA. | | | Electrical safety and electromagnetic compatibility (EMC)<br>Electrical safety and EMC testing were conducted in accordance with AAMI<br>/ ANSI ES60601-1:2005/(R)2012 And A1:2012 "Medical Electrical<br>Equipment - Part 1: General Requirements For Basic Safety And Essential<br>Performance", IEC 60601-2-33:2010+A1:2013 "Medical Electrical<br>Equipment - Part 2-33: Particular Requirements For The Basic Safety And<br>Essential Performance Of Magnetic Resonance Equipment For Medical<br>Diagnostic for safety" and IEC 60601-1-2:2014 Edition 4 "Medical Electrical<br>Equipment - Part 1-2: General Requirements For Basic Safety And<br>Essential Performance - Collateral Standard: Electromagnetic<br>Disturbances - Requirements And Tests" for EMC as recognized by FDA | | | Surface Heating of RF Receive Coils<br>Surface heating was tested with both coil-plugged and coil-unplugged<br>configurations according to AAMI/ANSI ES60601-1. The measured<br>temperature of the surface of the coil never exceeds the maximum limit of<br>41°C in either configuration. | | | Software Verification and Validation Testing<br>Software verification and validation testing were conducted and<br>documentation was provided as recommended by FDA's Guidance for<br>Industry and FDA Staff, "Guidance for the Content of Premarket<br>Submissions for Software Contained in Medical Devices." | | | Acoustic testing<br>Acoustic testing were conducted in accordance with NEMA MS 4-2010<br>"Acoustic Noise Measurement Procedure for Diagnosing Magnetic<br>Resonance Imaging Devices" | | | Performance testing - Bench<br>The filing device complies with the following NEMA MS standard as does the<br>predicate device:<br>• NEMA MS-1-2008 (R2014) - Determination of Signal-to-Noise Ratio<br>(SNR) in Diagnostic Magnetic Resonance Images<br>• NEMA MS 2-2008 (R2014) - Determination of Two-Dimensional<br>Geometric Distortion in Diagnostic Magnetic Resonance Images<br>• NEMA MS 3-2008 (R2014) - Determination of Image Uniformity in<br>Diagnostic Magnetic Resonance Images<br>• NEMA MS 5-2010 - Determination of Slice Thickness in Diagnostic<br>Magnetic Resonance Imaging<br>• NEMA MS 6-2008 (R2014) - Determination of Signal-to-Noise Ratio<br>and Image Uniformity for Single-Channel Non-Volume Coils in<br>Diagnostic MR Imaging<br>• NEMA MS 8-2016 - Characterization Of The Specific Absorption<br>Rate For Magnetic Resonance Imaging Systems<br>The performance testing results demonstrate the safety and performance<br>as expected. Therefore, the filing device is substantially equivalent to the<br>legally marketed predicate device. | {10}------------------------------------------------ | Clinical Testing<br>Summary: | Sample clinical images are provided to verify the claim of filing device's<br>capability in generating images for diagnostic purposes. Sample clinical<br>image sets from filing device and predicate device on same pulse<br>sequences are provided to demonstrate the substantial equivalence of filing<br>device to legally marketed predicate device. | |-------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Conclusions Drawn<br>from Non-Clinical<br>Information and<br>Clinical Images: | The non-clinical information (safety and performance tests), combined with<br>clinical images (both from filing device and predicate device), demonstrate<br>that the filing device is as safe, as effective, and performs as well as the<br>predicate device. | | SE Conclusion: | | The filing device has the same intended use as the predicate device. The<br>minor differences in indication for use do not constitute any safety and<br>effectiveness issue, as indicated in performance data provided. | |----------------|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | | The filing device utilizes same technologies as predicate device, and gives<br>consistent results (images) in medical diagnosis. Any minor differences in<br>physical attributes do not constitute any safety and effectiveness issue, as<br>indicated in performance data [non clinical/clinical] provided. | | | | Therefore the filing device is demonstrated, for being substantially<br>equivalent to the legally marketed predicate device. |
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...