SUPERSTAR 0.35T MRI SYSTEM

K071154 · Neusoft Medical Systems Co., Ltd. · LNH · May 10, 2007 · Radiology

Device Facts

Record IDK071154
Device NameSUPERSTAR 0.35T MRI SYSTEM
ApplicantNeusoft Medical Systems Co., Ltd.
Product CodeLNH · Radiology
Decision DateMay 10, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1000
Device ClassClass 2
Attributes3rd-Party Reviewed

Intended Use

The Superstar 0.35T is an imaging device, and is intended to provide the physician with physiological and clinical information obtained non-invasively and without the use of ionizing radiation. The MRI system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by MRI system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.

Device Story

Superstar 0.35T is a 0.35T permanent magnet MRI system; magnet constructed of NdFeB material. System utilizes gradient subsystem for controlled magnetic fields and RF subsystem for signal transmission/reception. Inputs: proton magnetic resonance signals. Outputs: 2D/3D cross-sectional images (transverse, coronal, sagittal, oblique, curved). Operated by trained clinicians in clinical settings. Software provides scanning control, image reconstruction, and archive management. Physicians interpret images to assess internal structures (head, body, extremities) for diagnostic purposes. Benefits include non-invasive visualization of anatomy without ionizing radiation.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

0.35T permanent magnet MRI system; magnet material: NdFeB. Standards: IEC60601-1, IEC60601-2-33, IEC60825-1, DICOM 3.0, NEMA MS Series (MS1-MS8). Connectivity: DICOM 3.0. Software: Windows-based interactive program for scanning, reconstruction, and archiving.

Indications for Use

Indicated for patients requiring diagnostic imaging of head, neck, shoulder, breast, wrist, ankle, body, spine, and extremities. Utilizes proton excitation for T1, T2, proton density, diffusion-weighted imaging, and MR angiography. No specific age or gender contraindications listed.

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K071154 # Neusoft 510(k) MAY 1 0 2007 ## Attachment 1 # Summary of Safety and Effectiveness Page 1 of 2 This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h) ### General Information: | Trade Name: | Superstar 0.35T | |---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product Model: | Superstar 0.35T | | Common Name of Device | MRI System | | CFR Section: | 21 CFR Part 892.1000<br>Magnetic resonance diagnostic device | | Classification Name: | System, Nuclear Magnetic Resonance Imaging | | Product Code: | LNH | | Device Class: | Class II | | Applicable Standard: | IEC60601-1, Medical electrical equipment - Part 1: General<br>Requirements for Safety<br>IEC60601-2-33, Medical electrical equipment -- Part 2-33: Particular<br>requirements for the safety of magnetic resonance equipment for<br>medical diagnosis<br>21 CFR Subchapter J, Radiological Health<br>IEC60825-1, Safety of laser products-Part1:Equipment classification,<br>requirement and user's guide<br>DICOM 3.0<br>NEMA MS Series (MS1 - MS8) | | Manufacture and<br>Distributor: | Neusoft Medical Systems Co., Ltd.<br>No.3-11, Wenhua Road, Heping District,<br>Shenyang, China<br>Post Code : 110004 | | Submitter: | Contact : Tian Yanfang<br>Title : Manager of Quality Management Department<br>Tel : 86-24-83660649<br>Fax : 86-24-83780480<br>E-Mail : Tianyanfang@neusoft.com | {1}------------------------------------------------ #### Safety and Effectiveness information Intended Uses: The Superstar 0.35T is intended to produce images that reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis. Device Description: The Superstar 0.35T is a 0.35T permanent magnet MRJ system. The magnet is mainly made of NdFeB material. The system software based on Windows (TM) is an interactive program with user-friendly interface. Its functions cover scanning control, image reconstruction and image/archive management and maintenance. Predicated Device: K030918 : Superopen 0.35T K024042 : Panorama Enhancement #### Statement of Substantial Equivalence: The Superstar 0.35T system is comparable and substantially equivalent to the Superopen 0.35T MRI system (K030918, another product of Neusoft ) and Panorama Enhancements (K024042) in that they are similar in technology and intended uses. Both of these systems are open-magnet MR Imaging System, use Gradient Subsystem to provide controlled and uniform gradient magnet fields in the X, Y and Z planes, and use RF Subsystem to complete the function of RF signal transmitting/receiving and processing. Image reconstruction is controlled by console's computer that has an interactive user interface, and the system produces 2D and 3D image that can be filmed or electronically stored for future review. Both of these systems have the traditional MRI unit. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the department's name around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized caduceus or a representation of human services. Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850 Neusoft Medical Systems Co., Ltd. % Mr. Tamas Borsai Division Manager, Medical Division TUV Rheinland of North America 12 Commerce Road NEWTOWN CT 06470 MAY 1 0 2007 Re: K071154 Trade/Device Name: Superstar 0.35T Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: April 16, 2007 Received: April 25, 2007 ### Dear Mr. Borsai: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Image /page/2/Picture/10 description: The image is a black and white circular logo. The logo has the letters "PA" in a bold, sans-serif font in the center. Above the letters "PA" is the text "1906-1986". Below the letters "PA" is the word "Centennial" in a cursive font. Below the word "Centennial" are three stars. The logo is surrounded by a dotted border. Protecting and Promoting Public Health {3}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |----------------|----------------------------------|--------------| | 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 894.xxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html. Sincerely yours, Nancy C. Hogdon Nancy C. Brogdon Director. Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ K071154 510(k) ## Attachment 2 ## Indications for Use The Superstar 0.35T is an imaging device, and is intended to provide the physician with physiological and clinical information obtained non-invasively and without the use of ionizing radiation. The MRI system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by MRI system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination. The indications for use are as follows: Anatomical Region: Head, Neck, Shoulder, Breast, Wrist, Ankle Body, Spine, Extremities Nucleus excited: Proton Diagnostic uses: T1,T2, proton density weighted imaging Diffusion weighted imaging MR Angiography Imaging processing Imaging capabilities: 2D, 3D Spin Echo( SE ) Turbo spin echo (TSE) Short time inversion recovery (STIR); Fast STIR, IRFFE, IRSE, Fast IR, IR TSE Fluid attenuated inversion recovery (FLAIR); Fast FLAIR 2D,3D Fast Field Echo (FFE) T1/T2/N-Fast field echo (FFE); B-Fast filed echo (B-FFE);N/B-Fast filed echo 3D; T1-Fast field echo 3D; Dual echo (DE);DTSE; DSE; MR angiography FE, FFE 3D; Echo Planar Imaging (EPI) -SE-EPI -Diffusion DW-EPI Nancy C Brogdon (Division Sign-O Division of Reproductive, Ab and Radiological Devi 510(k) Number > cription Use (Per 21 CFR 801.109)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...