MAGNETOM TRIO SYSTEM

K013586 · Siemens Medical Systems, Inc. · LNH · Dec 28, 2001 · Radiology

Device Facts

Record IDK013586
Device NameMAGNETOM TRIO SYSTEM
ApplicantSiemens Medical Systems, Inc.
Product CodeLNH · Radiology
Decision DateDec 28, 2001
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1000
Device ClassClass 2

Intended Use

The MAGNETOM Trio system is a whole body scanner designed to support higher resolution imaging and shorter scan times. The Trio system is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures of the head and whole body. The images produced by the Trio system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin echo time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining diagnosis.

Device Story

MAGNETOM Trio is a whole-body MRI scanner; utilizes magnetic resonance to map proton density, T1, T2, and flow properties; produces cross-sectional images (transverse, sagittal, coronal, oblique) of head and body. Operated by trained clinicians in clinical settings; images interpreted by physicians to assist in diagnosis. System designed for higher resolution and faster scan times compared to predecessors. Safety parameters (static field, RF power, acoustic noise) and performance metrics (SNR, uniformity, distortion) evaluated to ensure equivalence to existing 1.5T and 3.0T systems.

Clinical Evidence

Bench testing only. No clinical data provided. Safety and performance were validated through evaluation of static magnetic field, rate of change of magnetic field, RF power deposition, acoustic noise, signal-to-noise ratio, image uniformity, geometric distortion, and slice profile/thickness/gap against FDA guidance for MR Diagnostic Devices.

Technological Characteristics

Whole-body MRI scanner; 3.0 Tesla static magnetic field. Sensing based on nuclear magnetic resonance (NMR) of hydrogen nuclei. Evaluated for safety parameters including RF power deposition and acoustic noise, and performance metrics including SNR, geometric distortion, and slice profile. Connectivity and software details not specified.

Indications for Use

Indicated for diagnostic imaging of head and whole body internal structures in patients requiring MRI. Produces transverse, sagittal, coronal, and oblique images based on proton density, T1, T2, and flow properties. For use by trained physicians.

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 1101 3586 # 11 Appendix: 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92. | I. | General Information. Establishment: | | | | | | |----|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--| | | • Address: | Siemens Medical Systems, Inc.<br>186 Wood Avenue South<br>Iselin, N.J. 08830 | | | | | | | Registration Number: | 2240869 | | | | | | | Contact Person: | Mr. Jamie Yieh<br>Technical Specialist, Regulatory Affairs<br>Phone: (732) 321-4625<br>Fax: (732) 321-4841<br>Email: jamie.yieh@siemens.com | | | | | | | Date of Summary Preparation: | 10/29/01 | | | | | | | Device Name: | | | | | | | | •Trade Name: | MAGNETOM Trio System | | | | | | | •Classification Name: | Magnetic Resonance Diagnostic Device, CFR § | | | | | | | •Classification: | Class II | | | | | | | •Performance Standards: | None established under Section 514 the Food, Drug, and Act. | | | | | 892.1000 Cosmetic {1}------------------------------------------------ ## II. Safety and Effectiveness Information Supporting Substantial Equivalence. ## · Device Description: ### • Intended Use The MAGNETOM Trio system is a whole body scanner designed to support higher resolution imaging and shorter scan times. The Trio system is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures of the head and whole body. The images produced by the Trio system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin echo time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining diagnosis. ## · Technological Characteristics The MAGNETOM Trio System is substantially equivalent to both the MAGNETOM 3.0 T Allegra Dedicated Head only System and GE's Body Imaging Coil for 3.0T Signa VH/I system. ## · General Safety and Effectiveness Concerns: Operation of the MAGNETOM Trio System is substantially equivalent to the commercially available MAGNETOM 1.5 T Sonata System and 3.0 T Dedicated Head Only Allegra System. The following are the safety parameter with action levels: - Maximum Static Field ● - Rate of Change of Magnetic Field ● - RF Power Deposition . - . Acoustic Noise Levels and performance levels - Specification Volume . - Signal to Noise ● - Image Uniformity . - Geometric Distortion . - Slice Profile, Thickness and Gap . - High Contrast Spatial Resolution . specified by the FDA guidance document for MR Diagnostic Devices were evaluated within this notification. The increase in static magnetic field of the system will affect the maximum static field and acoustic noise level, as well as certain performance levels with the system. However, the new levels are not significantly changed and, in the case of safety, parameters remain below the level of concern. {2}------------------------------------------------ #### • Substantial Equivalence: Laboratory testing were performed to support this claim of substantial equivalence and to show that the technological differences do not raise any new questions pertaining to safety and effectiveness. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image is a black and white seal for the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of three overlapping human profiles facing to the right. The profiles are arranged in a way that suggests unity and collaboration. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 #### DEC 2 8 2001 Mr. Jamie Yieh Technical Specialist, Regulatory Affairs Siemens Medical Systems, Inc. 186 Wood Avenue South ISELIN NJ 08830 Re: K013586 Trade/Device Name: Magnetom Trio System Magnetic resonance imaging system Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: 90 LNH Dated: October 29, 2001 Received: October 30, 2001 Dear Mr. Yieh: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ 510(k) Number (if known) 人のノ 35 86 MAGNETOM Trio System Device Name: #### Indications for Use: The MAGNETOM Trio system is a whole body scanner designed to support higher resolution imaging and shorter scan times. The Trio system is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures of the head and whole body. The images produced by the Trio system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin echo time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining diagnosis. (please do not write below this line- continue on another page if needed) | Concurrence of CDRH, Office of Device Evaluation | | |--------------------------------------------------|----| | Prescription Use | ✓ | | | OR | | Over-The-Counter Use | | cun broadon (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices Ko13586 510(k) Number.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...