uMR 780, uMR 790
K191157 · Shanghai United Imaging Healthcare Co., Ltd. · LNH · May 29, 2019 · Radiology
Device Facts
| Record ID | K191157 |
| Device Name | uMR 780, uMR 790 |
| Applicant | Shanghai United Imaging Healthcare Co., Ltd. |
| Product Code | LNH · Radiology |
| Decision Date | May 29, 2019 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 892.1000 |
| Device Class | Class 2 |
Intended Use
The uMR 780/uMR790 system is indicated for use as a magnetic device (MRDD) that produces sagittal, transverse, coronal, and oblique cross sectional images, and that display internal anatomical structure and/or function of the head, body and extremities. These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist the diagnosis. Contrast agents may be used depending on the region of the scan.
Device Story
uMR 780 (65cm bore) and uMR 790 (60cm bore) are 3.0T superconducting magnetic resonance diagnostic devices. Systems comprise magnet, RF power amplifier, RF coils, gradient power amplifier, gradient coils, patient table, spectrometer, computer, and vital signal module. Devices acquire MR signals to produce sagittal, transverse, coronal, and oblique cross-sectional and spectroscopic images. Operated by trained clinicians in clinical settings to visualize internal anatomy/function and derive physical parameters for diagnostic assistance. Contrast agents may be administered. Modification involves updated magnet model and addition of seven previously cleared RF coils (K183186).
Clinical Evidence
No clinical data. Bench testing only, including SNR, geometric distortion, image uniformity, magnetic field homogeneity, and magnetic field decay, performed in accordance with IEC 60601-2-33 Ed. 3.2.
Technological Characteristics
3.0T superconducting MRI system. Components: magnet, RF power amplifier, RF coils, gradient power amplifier, gradient coils, patient table, spectrometer, computer. Bore sizes: 65cm (uMR 780), 60cm (uMR 790). Conforms to NEMA and DICOM standards. Safety testing per IEC 60601-2-33 Ed. 3.2.
Indications for Use
Indicated for patients requiring diagnostic imaging of head, body, and extremities. Used to produce cross-sectional and spectroscopic images of internal anatomical structures and functions. Contraindications include standard MRI safety exclusions (e.g., incompatible implants).
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
- uMR 780 (K181370)
- uMR 790 (K181371)
Reference Devices
- Head Coil - 12, Head Coil - 32, Carotid Coil - 8, Temporomandibular Joint Coil - 4, Infant Coil - 24, Cardiac Coil - 24, Foot & Ankle Coil - 24 (K183186)
Related Devices
- K181370 — uMR 780 · Shanghai United Imaging Healthcare Co., Ltd. · Oct 3, 2018
- K181371 — uMR 790 · Shanghai United Imaging Healthcare Co., Ltd. · Oct 3, 2018
- K183137 — uMR 580, uMR 588 · Shanghai United Imaging Healthcare Co., Ltd. · Jan 11, 2019
- K240540 — uMR Omega · Shanghai United Imaging Healthcare Co., Ltd. · Mar 22, 2024
- K180926 — uMR 560 · Shanghai United Imaging Healthcare Co., Ltd. · Jun 6, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the Department of Health & Human Services - USA seal. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. The logo is simple and professional, and it is easily recognizable.
Shanghai United Imaging Healthcare Co., Ltd. % Xin Gao Regulatory Affairs Specialist No. 2258 Chengbei Road Shanghai, Shanghai 201807 CHINA
Re: K191157
Trade/Device Name: uMR 780, uMR 790 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH Dated: April 30, 2019 Received: May 1, 2019
Dear Xin Gao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
May 29, 2019
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191157
Device Name uMR 780, uMR 790
#### Indications for Use (Describe)
The uMR 780/uMR790 system is indicated for use as a magnetic device (MRDD) that produces sagittal, transverse, coronal, and oblique cross sectional images, and that display internal anatomical structure and/or function of the head, body and extremities.
These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist the diagnosis. Contrast agents may be used depending on the region of the scan.
Type of Use (Select one or both, as applicable)
| <span style="font-size:100%;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------------------------------------------|-----------------------------------------------|
|-------------------------------------------------------------------------------------|-----------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized "U" shape that also resembles a shield with a vertical line running through the center. The color of the logo is a dark teal.
# 510 (k) SUMMARY
K191157
# 1. Date of Prepared
April 30, 2019
## 2. Sponsor Identification
Shanghai United Imaging Healthcare Co.,Ltd. No.2258 Chengbei Rd. Jiading District, 201807, Shanghai, China
Contact Person: Xin GAO Position: Regulatory Affairs Specialist Tel: +86-021-67076888-5386 Fax: +86-021-67076889 Email: xin.gao@united-imaging.com
### 3. Identification of Proposed Device(s)
Trade Name: uMR 780, uMR 790 Common Name: Magnetic Resonance Diagnostic Device Model(s): uMR 780, uMR 790 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: II Product Code: LNH
## 4. Identification of Predicate Device(s)
Predicate Device #1 510(k) Number: K181370 Device Name: uMR 780 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: II Product Code: LNH
Predicate Device #2 510(k) Number: K181371 Device Name: uMR 790 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: II Product Code: LNH
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for United Imaging. The words "UNITED IMAGING" are stacked on top of each other in a bold, sans-serif font. To the right of the words is a stylized "U" shape, which is also in a bold font. The logo is simple and modern, and the colors are muted.
### Reference Device
510(k) Number: K183186 Device Name: Head Coil - 12, Head Coil - 32, Carotid Coil - 8, Temporomandibular Joint Coil - 4, Infant Coil - 24, Cardiac Coil - 24, Foot & Ankle Coil - 24 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: II Product Code: MOS
### 5. Device Description
The uMR 780 is a 3.0T superconducting magnetic resonance diagnostic device with a 65cm size patient bore. It consists of components such as magnet, RF power amplifier, RF coils, gradient power amplifier, gradient coils, patient table, spectrometer, computer, equipment cabinets, power distribution system, internal communication system, and vital signal module etc. The uMR 780 Magnetic Resonance Diagnostic Device is designed to conform to NEMA and DICOM standards.
The uMR 790 is a 3.0T superconducting magnetic resonance diagnostic device with a 60cm size patient bore. It consists of components such as magnet, RF power amplifier, RF coils, gradient power amplifier, gradient coils, patient table, spectrometer, computer, equipment cabinets, power distribution system, internal communication system, and vital signal module etc. The uMR 790 Magnetic Resonance Diagnostic Device is designed to conform to NEMA and DICOM standards.
The modification performed on the previously cleared devices (K181370 for uMR 780 and K181371 for uMR 790) is due to the modification of magnet. The modification, which do not affect the intended use or alter the fundamental scientific technology of the device, is following:
A modify magnet with a new model number;
In addition, seven coils which have been previously cleared by FDA via K183186 (Head Coil - 12, Head Coil - 32, Carotid Coil - 8, Temporomandibular Joint Coil - 4, Infant Coil - 24, Cardiac Coil - 24, Foot & Ankle Coil - 24) are added in this submission.
### 6. Indications for Use
The uMR 780/uMR 790 system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces sagittal, transverse, coronal, and oblique
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in a bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" shape, which is also in a bold font. The color of the text and the "U" shape is a dark teal.
cross sectional images, and spectroscopic images, and that display internal anatomical structure and/or function of the head, body and extremities. These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist the diagnosis. Contrast agents may be used depending on the region of interest of the scan.
## 7. Technological Characteristic
The technology characteristics of the modified uMR 780 and uMR 790, reflected in this 510(k) submission, do not alter the scientific technology of the devices and are substantially equivalent to those of the predicate devices.
# 8. Non-Clinical Tests
The following testing was conducted on the proposed devices:
- A Signal to Noise Ratio
- Geometric Distortion A
- A Image Uniformity
- A Magnetic Field Homogeneity
- A Magnetic Field Decay
- > IEC 60601-2-33 Ed. 3.2 B:2015, Medical Electrical Equipment - Part 2-33: Particular Requirements For The Basic Safety And Essential Performance Of Magnetic Resonance Equipment For Medical Diagnostic
The test results demonstrated that the device performs as expected and thus, it is substantially equivalent to the predicate devices to which it has been compared.
# 9. Clinical Tests
No clinical testing was conducted on the proposed devices.
## 10. Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, we concludes that the modified uMR 780 and uMR 790 Magnetic Resonance Diagnostic Device are substantially equivalent to the predicate devices. It does not introduce new indications for use, and has the same technological characteristics and does not introduce new potential hazards or safety risks.