SyMRI

K162943 · Syntheticmr AB · LNH · Aug 29, 2017 · Radiology

Device Facts

Record IDK162943
Device NameSyMRI
ApplicantSyntheticmr AB
Product CodeLNH · Radiology
Decision DateAug 29, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1000
Device ClassClass 2
AttributesSoftware as a Medical Device

Intended Use

SyMRI is a post-processing software medical device intended for use in visualization of the brain. SyMRI analyzes input data from MR imaging systems. SyMRI utilizes data from a multi-delay, multi-echo acquisition (MDME) to generate parametric maps of R1, R2 relaxation rates, and proton density (PD). SyMRI can generate multiple image contrasts from the parametric maps. SyMRI enables post-acquisition image contrast adjustment. SyMRI is indicated for head imaging. When interpreted by a trained physician, SyMRI images can provide information useful in determining diagnosis. SyMRI should always be used in combination with at least one other, conventional MR acquisition (e.g. T2-FLAIR).

Device Story

SyMRI is a post-processing software device for brain MRI. It takes multi-delay, multi-echo (MDME) acquisition data as input. The software transforms this data into parametric maps of R1, R2 relaxation rates and proton density (PD). From these maps, it generates various synthetic image contrasts (e.g., T1, T2, PD, FLAIR, IR) by allowing users to adjust echo time (TE), repetition time (TR), and inversion delay (TI) post-acquisition. Used in clinical settings by physicians to aid diagnosis; the output provides enhanced slice registration due to the single-scan source. Benefits include post-acquisition contrast flexibility and reduced need for multiple scans, improving workflow and diagnostic utility.

Clinical Evidence

Bench testing only. Verification and validation performed per internal processes. Phantom head-to-head comparison between SyMRI and MAGiC evaluated R1, R2, and PD parametric maps and major synthetic image contrasts (T1w, T2w, T2 FLAIR). Results showed no difference between the subject and predicate devices.

Technological Characteristics

Post-processing software for MRI. Inputs: MDME sequence data. Outputs: Parametric maps (R1, R2, PD) and synthetic MR images. Connectivity: DICOM (NEMA PS 3.1-3.20). Standards: AAMI/ANSI 62304 (software lifecycle), ISO 13485 (quality), ISO 14971 (risk management), IEC 62366-1 (usability), IEC 82304-1 (health software safety).

Indications for Use

Indicated for head imaging in patients requiring brain visualization. Must be used by a trained physician in conjunction with conventional MR acquisitions (e.g., T2-FLAIR) to assist in diagnosis.

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other, with a wing-like shape extending from the top profile. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 29, 2017 SyntheticMR AB % Mr. Raymond Kelly Consultant Licensale, Inc. 68 Southwoods Terrrace Southbury CT 06488 Re: K162943 Trade/Device Name: SyMRI Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: July 22, 2017 Received: August 21, 2017 Dear Mr. Kelly: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, Michael D. O'Hara For Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K162943 Device Name SyMRI Indications for Use (Describe) SyMRI is a post-processing software medical device intended for use in visualization of the brain. SyMRI analyzes input data from MR imaging systems. SyMRI utilizes data from a multi-delay, multi-echo acquisition (MDME) to generate parametric maps of R1, R2 relaxation rates, and proton density (PD). SyMRI can generate multiple image contrasts from the parametric maps. SyMRI enables post-acquisition image contrast adjustment. SyMRI is indicated for head imaging. When interpreted by a trained physician, SyMRI images can provide information useful in determining diagnosis. SyMRI should always be used in combination with at least one other, conventional MR acquisition (e.g. T2-FLAIR). | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| | <span style="unicode-bidi:isolate; direction:ltr;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | |----------------------------------------------------------------------------------------------------------| | <span style="unicode-bidi:isolate; direction:ltr;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 003 510k Summary #### 510(k) Summary This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. Date Prepared: August 14, 2017 # Applicant: SyntheticMR AB Storgatan 11 SE-58223 Linköping, Sweden # Contact Person: Raymond Kelly 68 Southwoods Terrace Southbury, CT 06488 USA Phone: (203) 400-7566 # Device Information: Trade Name: SyMRI® 10 Common Name: Nuclear Magnetic Resonance Imaging System Classification: 21CFR 892.1000 Product Code: LNH Regulatory Class: II ### Predicate Device: MAGiC K161397 (LNH) ### Indications for Use SyMRI is a post-processing software medical device intended for use in visualization of the brain. SyMRI analyzes input data from MR imaging systems. SyMRI utilizes data from a multidelay, multi-echo acquisition (MDME) to generate parametric maps of R1, R2 relaxation rates, and proton density (PD). SyMRI can generate multiple image contrasts from the parametric maps. SyMRI enables post-acquisition image contrast adjustment. SyMRI is indicated for head imaging. When interpreted by a trained physician, SyMRI images can provide information useful in determining diagnosis. SyMRI should always be used in combination with at least one other, conventional MR acquisition (e.g. T2-FLAIR). ### Device Description SyMRI allows the user to generate multiple image contrasts from a single acquisition scan. This is accomplished by post-processing a multi-delay, multi-echo acquisition (MDME) into parametric maps. The parametric maps are R1, R2 relaxation rates, and proton density (PD). The inverse relaxation parameters, T1 relaxation time (1/R1), and T2 relaxation time (1/R2) are also provided as parametric maps. The parametric maps can be visualized as contrast weighted MR images, such as T1, T2, PD, and Inversion Recovery (IR) weighted images (including T1-FLAIR, T2-FLAIR, STIR, Double IR, and PSIR weighted images). SyMRI calculates the pixel signal intensity as a function of R1, R2, PD, and desired MR scanner settings, such as echo time (TE), repetition time (TR), and inversion delay time (TI). A number of default settings for TE. TR, and TI are provided, but the user has the ability to change the contrast of the images. SyMRI generates all the different image contrasts from the same parametric maps, derived from the same acquisition. This leads to enhanced image slice registration, owing to the absence of interacquisition patient movement. SyMRI provides the user the ability to change the contrast of the images after the acquisition. This is performed by adjusting the TE, TR, and/or TI parameters post-acquisition, to generate the specific contrast desired. SyMRI also provides image processing tools to extract the values of the parametric maps per individual pixel, per region of interest, or the entire imaging volume. {4}------------------------------------------------ # 003 510k Summary SyMRI is intended to be used on MDME sequence data from GE MAGiC. # Comparison of Technological Characteristics SyMRI and MAGiC are the same algorithm for post processing, MAGiC also includes acquisition. Input data to SyMRI and MAGiC is MDME acquisition. SyMRI and MAGiC both uses a single scan to acquire a multi-delay, multi- echo acquisition (MDME) data to generate parametric maps of R1, R2 relaxation rates, and proton density (PD) This data is used to generate images, which are possible to adjust, even post acquisition. In SyMRI the parametric maps are used to produce the synthetic images - as in MAGiC. SyMRI is the same as MAGiC post processing. By virtue of its intended use and physical and technological characteristics, SyMRI is substantially equivalent to the predicate devise that has been cleared for marketing in the US. The performance data shows that SyMRI is as safe and effective as the predicate device. ### Performance Data Quality and Safety: SyMRI was designed in compliance with following standards: - AAMI/ANSI 62304- Medical Device software Software life cycle processes ● - NEMA PS 3.1 3.20 (2016) DICOM Format ● - ISO 13485:2012 Quality management systems Requirements for regulatory purposes ● - ISO 14971:2012 Application of risk management to medical devices ● - IEC 62366-1:2015 Application of usability engineering to medical devices ● - IEC82304-1:2016 Health Software Part 1: General requirements for product safety ● ### Testing to Support of Substantial Equivalence Determination Verification and validation was performed according to SyntheticMR processes. Additional phantom head to head comparison of R1, R2 and PD parametric maps, which included one contrast of each major synthetic image (T1w, T2w, T2 FLAIR), were performed to compare SyMRI to MAGiC. There was no difference between SyMRI and MAGiC. ### Conclusion By virtue of its intended use and physical and technological characteristics, SyMRI is substantially equivalent to the predicate device that has been cleared for marketing in the US.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...