K121997 · Imris, Inc. · LNH · Aug 10, 2012 · Radiology
Device Facts
Record ID
K121997
Device Name
IMRIS ONCOLOGY PACKAGE
Applicant
Imris, Inc.
Product Code
LNH · Radiology
Decision Date
Aug 10, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The IMRIS Oncology package uses the IMRISneuro family of Intra-operative MRI systems/Siemens MAGNETOM MR systems indicated for use as a diagnostic imaging device that produce transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and / or spectra, and that display the internal structure and / or function of the entire body, including, but not limited to head, neck, and pelvis regions. Depending on the region of interest, contrast agents may be used. These images and / or spectra when interpreted by a trained physician yield information that may assist in diagnosis. The IMRIS Oncology Package provides an additional patient tabletop for IMRIS MR systems /Siemens MAGNETOM that allows patients to be imaged on a flat surface. The flat patient surface enables IMRIS MR system/Siemens MAGNETOM to acquire images in patient positions similar to other modalities that also utilize a flat patient surface such as X-ray, CT, PET and radiation therapy. The IMRIS MR systems/Siemens MAGNETOM MR systems with IMRIS Oncology Package may also be used with MR safe/MR conditional patient positioning and immobilization accessories to assist in obtaining consistent patient positions throughout multiple imaging sessions.
Device Story
The IMRIS Oncology Package is an accessory kit for IMRISneuro and Siemens MAGNETOM 1.5T MRI systems, designed to facilitate radiation therapy planning. It consists of a flat patient tabletop (MR SIMS table top) and specialized RF coils (HNC150 head/neck coil and PCC150 pelvic coil). The flat tabletop allows patient positioning consistent with other modalities like CT, PET, and X-ray. The device is used in hospital diagnostic rooms by trained clinicians. The system acquires MR images which are interpreted by physicians to assist in diagnosis and treatment planning. The package supports the use of MR-safe/conditional immobilization accessories to ensure consistent patient positioning across multiple sessions. The device benefits patients by enabling high-quality MRI imaging in a geometry compatible with radiation oncology workflows.
Clinical Evidence
Bench testing only. Testing included IEC 60601-1 and 60601-2-33 compliance, MRI compatibility (artifacts, heating), surface heating, and workflow/loading tests. Coil performance (SNR and image non-uniformity) was validated per NEMA MS 9-2008 and NEMA MS 6-2008 standards. Clinical images were provided to demonstrate clinical effectiveness.
Technological Characteristics
Components: Acetal tabletop, 18-channel head/neck coil (HNC150), 11-channel pelvic coil (PCC150). Sensing: Phased array RF coils. Connectivity: Plugs into MRI system interface. Safety: Active/passive decoupling, RF fuse. Standards: IEC 60601-1, IEC 60601-2-33, UL 94, NEMA MS 9-2008. Form factor: Flat surface tabletop for 1.5T MRI systems.
Indications for Use
Indicated for use as a diagnostic imaging device to produce cross-sectional images (transverse, sagittal, coronal, oblique) and spectra of the entire body, including head, neck, and pelvis, for diagnostic purposes. Used in conjunction with IMRISneuro or Siemens MAGNETOM 1.5T MRI systems. No specific age or gender restrictions; contraindications include standard MRI contraindications.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K121997
## IMRIS
AUG 1 0 2012
510(k) Summary for Oncology Package
### 510(k) SUMMARY FOR IMRIS ONCOLOGY PACKAGE
(As required by 21 CFR 807.92)
#### 1. GENERAL INFORMATION Establishment: IMRIS Inc. Address: 100-1370 Sony Place Winnipeg, Manitoba Canada, R3T 1N5 3003807210 Registration Number: Mr. Sanjay Shah Contact Person: QA and Regulatory Engineer Email: sshah@imris.com Phone: 1-204-480-7070 Fax: 1-204-480-7071 Date of Summary July 23, 2012 Preparation: Device Name / Trade name IMRIS Oncology Package Classification Name: Magnetic resonance diagnostic device. Classification Panel: Radiology Classification (CFR section): 21 CFR 892.1000 Class: Class II Product Code: LNH
### 2. PREDICATE DEVICES
The IMRIS Oncology Package is substantially equivalent to the following predicate medical devices.
| NAME OF THE DEVICE | 510(K) NUMBER | DATE OF<br>CLEARANCE | MANUFACTURER |
|-------------------------------|---------------|----------------------|---------------------------|
| MR Radiation Oncology Options | K102155 | August 31, 2010 | GE Medical System,<br>LLC |
{1}------------------------------------------------
## IMBIS
### 510(k) Summary for Oncology Package
### 3. DEVICE DESCRIPTION
The IMRIS Oncology Package provides an additional patient tabletop (MR SIMS table top) that can be used with IMRIS Neuro 1.5T / Siemens MAGNETOM 1.5T MRI scanners for radiation therapy planning. The MR SIMS table top allows patients to be imaged on a flat surface that matches the imaging patient position. The oncology package includes Head and Neck coil (HNC150) and Pelvic coil (PCC 150). Used in the IMRISneuro 1.5Tsystem/Siemens MAGNETOM 1.5 Tesla, it is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal, and oblique images of the internal structures of the head, neck and pelvic regions. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis. The IMRIS HNC150 and PCC150 coils can also be used as standard diagnostic head coils and pelvic coil respectively for diagnostic examinations. The Oncology package may be used with MR-conditional/ MR safe patient immobilization accessories to assist in consistent positions throughout multiple imaging sessions.
### 4. INDICATION FOR USE
The IMRIS Oncology package uses the IMRISneuro family of Intra-operative MRI systems /Siemens MAGNETOM MR systems indicated for use as a diagnostic imaging device that produce transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and / or spectra, and that display the internal structure and / or function of the entire body, including, but not limited to head, neck, and pelvis regions. Depending on the region of interest, contrast agents may be used. These images and / or spectra when interpreted by a trained physician vield information that may assist in diagnosis.
The IMRIS Oncology Package provides an additional patient tabletop for IMRIS MR systems /Siemens MAGNETOM that allows patients to be imaged on a flat surface. The flat patient surface enables IMRIS MR system/Siemens MAGNETOM to acquire images in patient positions similar to other modalities that also utilize a flat patient surface such as X-ray. CT, PET and radiation therapy.
The IMRIS MR systems/Siemens MAGNETOM MR systems with IMRIS Oncology Package may also be used with MR safe/MR conditional patient positioning and immobilization accessories to assist in obtaining consistent patient positions throughout multiple imaging sessions.
| Characteristic | GE MR Radiation Oncology Options | IMRIS Oncology Package |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FDA 510(k) # | K102155 | Current Submission |
| Manufactured<br>by | GE Medical Systems, LLC | IMRIS Inc. |
| Intended use<br>/Indications for<br>use | The MR Radiation Oncology Options<br>are a patient positioning package<br>intended for use with the GE wide bore<br>MR Scanners. The MR Radiation<br>Oncology Options when used with a<br>GE wide bore MR scanner is a whole<br>body magnetic resonance scanner<br>designed to support high resolution, | The IMRIS Oncology package<br>uses the IMRISneuro family of<br>Intra-operative MRI<br>systems/Siemens MAGNETOM<br>MR systems indicated for use as<br>a diagnostic imaging device that<br>produce transverse, sagittal,<br>coronal and oblique cross |
| | | 510(k) Summary for Oncology Package |
| | high signal-to-noise ratio and short<br>scan times. The GE wide bore MR<br>scanners with MR Radiation Oncology<br>Options are indicated for use as a<br>diagnostic imaging device to provide<br>axial, sagittal, coronal, and oblique<br>images, spectroscopic images, and/or<br>spectra, dynamic images of the internal<br>structures and organs of the entire<br>body, including but not limited to head,<br>neck, TMJ, spine, breast, heart,<br>abdomen, pelvis joints, prostate, blood<br>vessels, and musculoskeletal regions<br>of the body. The images produced by<br>the GE wide bore scanners with MR<br>Radiation Oncology Options reflect the<br>spatial distribution of protons (hydrogen<br>nuclei) exhibiting magnetic resonance.<br>These images and/or spectra when<br>interpreted by a trained physician yield<br>information that may assist in<br>diagnosis.<br>The MR Radiation Oncology Options<br>include a removable patient table insert<br>for GE wide bore MR scanners that<br>allow patients to be imaged on a flat<br>surface. The flat patient surface allows<br>image acquisition in patient positions<br>similar to other modalities that also use<br>a flat patient surface such as X-Ray,<br>CT, PET, and radiation therapy.<br>The GE wide bore MR scanners with<br>the MR Radiation Oncology Options<br>may also be used with MR- compatible<br>patient positioning and immobilization<br>accessories to assist in obtaining<br>consistent patient positions throughout<br>multiple imaging sessions. | sectional images, spectroscopic<br>images and / or spectra, and<br>that display the internal<br>structure and / or function of the<br>entire body, including, but not<br>limited to head, neck, and pelvis<br>regions. Depending on the<br>region of interest, contrast<br>agents may be used. These<br>images and / or spectra when<br>interpreted by a trained<br>physician yield information that<br>may assist in diagnosis.<br>The IMRIS Oncology Package<br>provides an additional patient<br>tabletop for IMRIS MR<br>systems/Siemens MAGNETOM<br>that allows patients to be<br>imaged on a flat surface. The<br>flat patient surface enables<br>IMRIS MR system/Siemens<br>MAGNETOM to acquire images<br>in patient positions similar to<br>other modalities that also utilize<br>a flat patient surface such as X-<br>ray, CT, PET and radiation<br>therapy.<br>The IMRIS MR<br>systems/Siemens MAGNETOM<br>MR systems with IMRIS<br>Oncology Package may also be<br>used with MR safe/MR<br>conditional patient positioning<br>and immobilization accessories<br>to assist in obtaining consistent<br>patient positions throughout<br>multiple imaging sessions. |
| Where used | Hospital Diagnostic room | Hospital Diagnostic room |
| Anatomical<br>sites | Entire body, including but not limited to<br>head, neck, TMJ, spine, breast, heart,<br>abdomen, pelvis joints, prostate, blood<br>vessels, and musculoskeletal regions<br>of the body | Entire body, including, but not<br>limited to head, neck, and pelvis<br>regions |
### 5. COMPARISION TO PREDICATE DEVICES
{2}------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
K121997
.
:
{3}------------------------------------------------
K12/997
## IMRIS
### 510(k) Summary for Oncology Package
6. SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE: (807.92 (A) (6))
.
| Characteristic | GE MR Radiation Oncology<br>Options | IMRIS Oncology Package | Comparison |
|--------------------------------------------------|------------------------------------------------------------------------------------|----------------------------------------------------------------|--------------------------------------------------|
| FDA 510(k) # | K102155 | Current Submission | |
| | MR-SIM table Top and S Frame mask adaptor | | |
| | MR radiation oncology<br>position insert | MR-SIM table top and S-<br>frame adaptor | |
| MRI system<br>Compatibility | GE 1.5T MRI scanners | Siemens MAGNETOM 1.5T<br>MRI systems IMRIS 1.5T<br>MRI systems | Same field<br>Strength<br>1.5Tesla |
| Surface | Flat Surface | Flat Surface | Same |
| | Mobile patient table top with<br>embedded high-density,<br>posterior RF coil array | Table top with cavity allows<br>mounting of posterior coils | Similar<br>design |
| Material | Fiber glass with foam core | Acetal | Different |
| MRI safety | MR compatible | MR safe | Same |
| S-Frame and mask<br>connecting with<br>table top | Yes | Yes | Same |
| IEC 60601-1<br>compliance | Yes | Yes | Same |
| Head and Neck Coil (HNC 150) | | | |
| | GE head and neck coil (6-<br>channel flex coil) | IMRIS Head and neck<br>coil (HNC150) | Comparison |
| MRI system<br>Compatibility | GE 1.5T MRI Scanners | Siemens 1.5T MRI<br>scanners<br>IMRIS 1.5T MRI system | Same field<br>strength -<br>1.5T MRI<br>Scanners |
| Coil Type | Receive-only six channel<br>phased array coil | Receive-only eighteen<br>channel phased array coil | Similar |
| RF Cable Interface | Interface Cable with<br>Insulated Cable Traps | Interface Cable with<br>Insulated Cable Traps | Same |
| Tune and Match | No tune, no match | No tune, no match | Same |
| Safety features | Active and Passive<br>Decoupling | Active and Passive<br>Decoupling | Same |
| | | RF Fuse | New |
| Coil Enclosure<br>Material and design | Vinyl coating closed cell<br>foam | Vinyl coating closed cell<br>foam | Same |
.
.
{4}------------------------------------------------
## K121997
# IMRIS
510(k) Summary for Oncology Package
| Characteristic | GE MR Radiation Oncology Options | IMRIS Oncology Package | Comparison |
|---------------------------------------|-------------------------------------------------------|-------------------------------------------------------|--------------------------------------------------|
| FDA 510(k) # | K102155 | Current Submission | |
| | Flexible | Flexible | Same |
| IEC 60601-1 Compliance | Yes | Yes | Same IEC<br>standards |
| Pelvic Coil (PCC 150) | | | |
| | GEM anterior array coil<br>GEM posterior array coil | IMRIS Pelvic coil<br>(PCC150) | Comparison |
| MRI system Compatibility | GE 1.5T MRI Scanners | Siemens 1.5T MRI<br>Scanners<br>IMRIS 1.5T MRI system | Same field<br>strength -<br>1.5T MRI<br>Scanners |
| Coil Type | Receive-only up to 36<br>channel phased array coil | Receive-only eleven<br>channel phased array coil | Similar |
| System connection | The coil plugs into the MRI<br>System | The coil plugs into the MRI<br>System | Same |
| RF Cable Interface | Interface Cable with<br>Insulated Cable Traps | Interface Cable with<br>Insulated Cable Traps | Same |
| Tune and Match | No tune, no match | No tune, no match | Same |
| Safety features | Active and Passive<br>Decoupling | Active and Passive<br>Decoupling | Same |
| | | RF Fuse | New |
| Coil Enclosure<br>Material and design | Plastic, Vinyl coated closed<br>cell foam<br>Flexible | Plastic, Vinyl coated<br>closed cell foam<br>Flexible | Same |
| IEC 60601-1<br>Compliance | Yes | Yes | Same |
.
:
:
·
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows a sequence of handwritten digits and a letter. The sequence starts with the letter 'K', followed by the numbers '121997'. The numbers are written in a simple, slightly slanted style, and the overall impression is that of a quickly jotted down code or identifier.
# IMRIS
510(k) Summary for Oncology Package
### 7. SUMMARY OF NON-CLINICAL DATA
### Design Verification and Validation Test (Bench Testing)
The IMRIS Oncology Package passed the following tests and meets product specifications. IMRIS has performed a number of V&V tests. The main tests include
- IEC 60601-1 compliance .
- . IEC 60601-2-33 compliance
- Clinical image comparison .
- MRI compatibility test (MR image artifacts test, MR heating test), .
- Surface heating (normal and single fault conditions) .
- Single fault condition unplugged (passive detuning test) .
- Workflow .
- Loading test .
- The Head and Neck coil (HNC150) and Pelvic coil (PCC 150) Image Non-Uniformity and . SNR was measured and is reported in accordance with NEMA MS 9-2008 Characterization of Special Purpose Coils for Diagnostic Resonance Images. The NEMA MS 9-2008 references the ALTERNATE MEASUREMENT PROCEDURE as described in NEMA MS 6-2008.
The oncology package is tested for electrical, mechanical, and flammability hazards. The IMRIS Oncology package complies with voluntary standards (IEC 60601-2-33, and UL 94). The Oncology Package provided clinical images which demonstrate the clinical effectiveness of the Oncology package. The HNC 150 and PCC 150 coils are tested for MR image artifacts and surface heating test. The HNC 150 and PCC 150 coils SNR and Image non-uniformity are tested according to NEMA standards. The tests outlined above have been executed with acceptable results. Performance data demonstrate safety and effectiveness of the IMRIS Oncology Package.
### 8. CONCLUSION
The IMRIS Oncology Package has the same intended use and indications for use as the predicate devices. Performance data demonstrate safety and effectiveness of the IMRIS Oncology Package with the new characteristics.
The IMRIS Oncology Package verification results and performance/safety standard results show that the device is safe and effective and substantially equivalent to the currently available predicate device, GE MR Radiation Oncology options.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or other bird-like figure, with three curved lines representing its wings or body.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
—AUG—1-0—2012—
### Mr. Sanjay Shah QA and Regulatory Engineer IMRIS Inc. 100-1370 Sony Place WINNIPEG MANITOBA R3T 1N5 CANADA .
Re: K121997
Trade/Device Name: IMRIS Oncology Package Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH and MOS Dated: July 4, 2012 Received: July 9, 2012
### Dear Mr. Shah:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such 11 your dowled to the birting major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease of actived and I Drimination that your device complies with other requirements of the Act that 1 Dr Has Intact a and regulations administered by other Federal agencies. You must or any I vatual the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{7}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice mourements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket will anow you to oogin finding of substantial equivalence of your device to a legally marketed nothleation. THE FDTS imanly of tication for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21-GFR-Parts 801-and If you desire specific advice of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket 5-150. Thise, productions are veganding the reporting the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Tou may other bates generational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
### Indications for Use
| 510(k) Number (if known): | <121997 |
|---------------------------|---------|
|---------------------------|---------|
. Device. Name: IMRIS Oncology Package
-Indications for Use: -
The IMRIS Oncology package uses the IMRISneuro family of Intra-operative MRI systems/Siemens MAGNETOM MR systems indicated for use as a diagnostic imaging device that produce transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and / or spectra, and that display the internal structure and / or function of the entire body, including, but not limited to head, neck, and pelvis regions. Depending on the region of interest, contrast agents may be used. These images and / or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
The IMRIS Oncology Package provides an additional patient tabletop for IMRIS MR systems /Siemens MAGNETOM that allows patients to be imaged on a flat surface. The flat patient surface enables IMRIS MR system/Siemens MAGNETOM to acquire images in patient positions similar to other modalities that also utilize a flat patient surface such as X-ray, CT, PET and radiation therapy.
The IMRIS MR systems/Siemens MAGNETOM MR Systems with IMRIS Oncology Package may also be used with MR safe/MR conditional patient positioning and immobilization accessories to assist in obtaining consistent patient positions throughout multiple imaging sessions.
Prescription Use ---------X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) Number K121997
IMRIS Oncology package 510(k) Premarket Notification
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.