The PoleStar N-30 is a Magnetic Resonance Diagnostic Device intended to produce transverse, sagittal, coronal, and oblique 2D and 3D images of sections of the head selected by the physician. The images produced by the PoleStar N-30 reflect the spatial distribution of protons (Hydrogen Nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and T2*. Anatomical regions: sections of the head selected by the physician. Nuclei excited: H-1 Diagnostic uses: T1, T2, T2* and density weighted imaging. The PoleStar N-30 is intended to be used intraoperatively in a standard operating room. When interpreted by trained physicians, the MR images provide information that can be useful in determining a diagnosis.
Device Story
PoleStar N30 is a compact, open-design intraoperative MRI system utilizing a permanent magnet (0.13T) to acquire 2D and 3D images of the head. Device operates in standard operating rooms; movable between operating table and storage cabinet. Input: H-1 proton magnetic resonance signals. Processing: pulse sequences including spin echo, gradient echo, fast spin echo, and steady state free precession. Output: transverse, sagittal, coronal, and oblique MR images displayed for physician interpretation. System aids clinical decision-making by providing real-time diagnostic information during surgery. Benefits include intraoperative visualization without moving patient to radiology suite.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing, technical specifications, and comparison of imaging performance parameters (reconstruction times, matrix sizes, pulse sequence capabilities) against the predicate device.
Technological Characteristics
Permanent magnet MRI system; 0.13T field strength; open-bore design (24.9 x 42 cm). Imaging modes: 2D/3D, spin echo, gradient echo, fast spin echo. Connectivity: DICOM 3.0 interface. Power: 3x208V. Passive and active shimming. Gradient strength: 23.5 mT/m. Computer: P4 2.8GHz, 160GB storage. No cryogen use.
Indications for Use
Indicated for diagnostic imaging of head sections in patients requiring intraoperative MRI. Intended for use by trained physicians in a standard operating room environment.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the Medtronic logo. The logo consists of a circular symbol above the word "Medtronic". The symbol depicts three human figures in different poses, arranged in a circular pattern. The text "Medtronic" is written in a bold, sans-serif font.
MAR 1 6 2010
# Traditional 510(k) Summary of Safety and Effectiveness
K092308
The Following Traditional 510(k) Summary of Safety and Effectiveness has been prepared pursuant to requirements for 510(k) summaries specified in 21 CFR § 807.92(a).
## 807.92(a)(1) - Submitter Details:
| Submitter name: | Yair Penias -Quality and Regulatory Manager |
|-----------------|---------------------------------------------|
| Address: | P.O. Box 548, Yokneam Elit 20692, ISRAEL |
| Phone: | +972-4-909-2306 |
| Fax: | +972-4-959-1011 |
| E-mail: | yair.penias@medtronic.com |
| Contact Person: | Yair Penias -Quality and Regulatory Manager |
| Date: | March 11, 2009 |
### 807.92(a)(2) - Device Details:
| Trade Name and Common Name: | PoleStar N30 - Magnetic Resonance<br>Diagnostic Device, Also known as<br>"PoleStar® N30 Surgical MRI System". |
|-----------------------------|---------------------------------------------------------------------------------------------------------------|
| Classification: | 21 CRF 892.1000 Magnetic Resonance<br>Diagnostic Device. |
| Class: | II<br>MRDD were reclassified by FDA from<br>Class III to Class II effective July 28,<br>1998. |
| Product Code: | LNH – Magnetic Resonance Imaging<br>System |
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the Medtronic logo. The logo consists of a circular emblem above the word "Medtronic". The emblem depicts a stylized human figure in motion, repeated in a circular pattern to suggest continuous movement or progress. The word "Medtronic" is written in a bold, sans-serif font.
### 807.92(a)(3) - Predicate Devices:
The PoleStar N30 is comparable to:
| Medical<br>Device Name | Applicant Name | 510(k)<br>Number | Classification |
|------------------------|------------------------|------------------|-----------------|
| PoleStar N-20 | ODIN Technologies Ltd. | K032541 | Class II device |
Additional Substantial Equivalence Information is provided in the attached Substantial Equivalence Comparison Table.
### 807.92(a)(4) - Device Description:
The PoleStar N30 utilizes a permanent magnet to acquire 2D single-slice, multi slice, and 3d volume images. A wide variety of pulse sequences are provided to the operator, including spin echo, gradient echo, fast spin echo, and steady state free precession acquisitions. The PoleStar N30 is a widely open and compact Intraoperative MRI unit intended to be used in a typical pre-existing operating room. The PoleStar N30 can be moved within the room between procedures, from the operating table to its Magnet Storage Cabinet, thus allowing the operating room to be used for any type of surgery.
> Page 2 of 6 K092308
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the Medtronic logo. The logo consists of a circular emblem above the word "Medtronic". The emblem depicts a figure in motion, with three overlapping silhouettes suggesting movement. The word "Medtronic" is written in a bold, sans-serif font.
#### 807.92(a)(5) - Device Indication For Use:
#### The general purpose of the device as defined in 21 CFR 892.1000:
The PoleStar N30 is a Magnetic Resonance Diagnostic Device intended to produce transverse, sagittal, coronal, and oblique 2D and 3D images of sections of the head selected by the physician. The images produced by the PoleStar N30 reflect the spatial distribution of protons (Hydrogen Nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and T2*.
- Anatomical regions: sections of the head selected by the physician. .
- . Nuclei excited: H-1
- T1, T2, T2* and density weighted imaging. Diagnostic uses: .
The PoleStar N30 is intended to be used intraoperatively in a standard operating room. When interpreted by trained physicians, the MR images provide information that can be useful in determining a diagnosis
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Medtronic logo. The logo consists of a circular symbol above the word "Medtronic". The symbol depicts a person in three different positions within the circle, suggesting movement or transformation. The word "Medtronic" is written in a bold, sans-serif font.
.
# 807.92(a){6) – Substantial Equivalence Comparison Table:
| Model | Odin | Medtronic Navigation |
|-------------------------------------------|----------------------------------------------------|----------------------------------------------------|
| parameter | PoleStar N-20 (K032541) | PoleStar (Model: N30) |
| Clinical application | Extremities and selected<br>sections of the head | sections of the head<br>selected by the physician |
| Magnet type | Permanent | Permanent |
| Field strength | 0.13T | 0.13T |
| 5 gauss fringe field<br>(radial/axial, m) | 2.2 | 2.2 |
| Shimming | Passive, active | Passive, active |
| Gradient subsystem | | |
| Strength mT/m | 22 | 23.5 |
| Rise time to 10mT/m | <1 | <0.15 |
| msec | | |
| Computer system | | |
| - CPU: | Pentium 586 | P4 2.8GHZ |
| - Memory Cache size [MB] | 1 | 1 |
| array processor | 4xDSP C44 TI | 4xDSP C44 TI |
| - Memory size [GB] | 40 | 160 |
| storage media | magnetic disk, floppy disk | magnetic disk, floppy disk |
| number of images stored | 1,310,720 | 5,242,880 |
| Imaging modes: | | |
| - single | Yes | Yes |
| - multislice | Yes | Yes |
| - volume study | Yes | Yes |
| - other | No | No |
| Reconstruction time: | | |
| - single slice, sec | <3/slice | <2/slice |
| - multislice, sec | <3/slice | <1/slice |
| - volume sec | <20/volume | <16/volume |
| Cardiac gating | No | No |
| Model | Odin | Medtronic Navigation |
| parameter | PoleStar N-20 (K032541) | PoleStar (Model: N30) |
| (ECG/peripheral) | No | No |
| Respiratory gating | No | No |
| Angiography | Optional | Optional |
| Spectroscopy | No | No |
| Imaging; | | |
| - pulse sequence | Spin Echo, Fast Spin Echo,<br>Gradient Echo, 2D 3D | Spin Echo, Fast Spin Echo,<br>Gradient Echo, 2D 3D |
| - repetition time, msec | 10-5000 increments of 1 | 10-5000 increments of 1 |
| - echo time, msec | 3-150 | 2.5-150 |
| - inversion time, msec | N/A | N/A |
| - slice thickness, mm | 2-10 | 2-10 |
| - scan orientation | Transverse, coronal,<br>sagittal, oblique | Transverse, coronal,<br>sagittal, oblique |
| - measuring matrix | 64x64 to 256x256 steps of<br>1 in phase encoding | 64x64 to 256x256 steps of<br>1 in phase encoding |
| - display matrix | 1024x768 | 1024x768 |
| - pixel intensity | 0-4095 | 0-4095 |
| Surface coils/Anatomical<br>regions: | | |
| - spine | No | No |
| - knee | Yes | No |
| - neck | No | No |
| - TMJ | No | No |
| - extremity | Yes | No |
| - head | Yes | Yes |
| - breast | No | No |
| - shoulder | No | No |
| - others | No | No |
| Bore diameter or WxH, cm | 25.2x42 | 24.9 x42 |
| Model parameter | Odin<br>PoleStar N-20 (K032541) | Medtronic Navigation<br>PoleStar (Model: N30) |
| Bore features | Open access to patient | Open access to patient |
| Cooling system type | Closed loop water cooling<br>(Gradients only). | No |
| Cryogen use | No | No |
| Magnet weight, kg | 400 | 330 |
| HxWxD, cm | 153x97x120 | 145x97x120 |
| Filed Of View (FOV),cm | 5-20 | 5-20 |
| Dicom 3.0 interface | Yes | Yes |
| Power requirements: | | |
| - line voltage, V | 3x208 (3 phase) | 3x208 (3 phase) |
| - Kva | 15 | 8 |
| - A/C, BTU/hr | <10000 | <10000 |
Page 4 of 6 K092308
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Medtronic logo. The logo consists of a circular symbol above the word "Medtronic". The symbol depicts a human figure in motion, repeated in a circular pattern. The word "Medtronic" is written in a bold, sans-serif font.
KO92308 Page 5 of 6
.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Medtronic logo. The logo consists of a circular graphic above the word "Medtronic". The graphic contains a stylized representation of a human figure in motion, repeated in a circular pattern. The word "Medtronic" is written in a bold, sans-serif font.
.
and the state of the same of the same
:
KC92308 Page 6 of 6
:
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings and a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The eagle is facing to the right, and the text is arranged around the circle.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Mr. Yair Penias Quality and Regulatory Manager Medtronic Navigation Israel, Ltd. Kochav Yokneam Bldg, P.O. Box 548 Yokneam Elit, 20692 ISRAEL
MAR 1 3 2010
Re: K092308
Trade/Device Name: PoleStar N-30 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: February 18, 2010 Received: February 22, 2010
Dear Mr. Penias:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21. Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{8}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510/8) rremarket notification. The FDA finding of substantial equivalence of your device to a legally ronaketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tollefree no (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Amiit J. Patel
Donald J. St. Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
510(k) Number (if known): K092308
Device Name: PoleStar N-30
Indication For Use:
The PoleStar N-30 is a Magnetic Resonance Diagnostic Device intended to produce transverse, sagittal, coronal, and oblique 2D and 3D images of sections of be bead selected by the physician. The images produced by the PoleStar N-30 reflect the spatial distribution of protons (Hydrogen Nuclei) exhibiting magnetic resonance.
The NMR properties that determine the image appearance are proton density, spinlattice relaxation time (T1), spin-spin relaxation time (T2) and T2*
Anatomical regions: sections of the head selected by the physician.
Nuclei excited: H-1
Diagnostic uses: T1, T2, T2* and density weighted imaging.
The PoleStar N-30 is intended to be used intraoperatively in a standard operating room. When interpreted by trained physicians, the MR images provide information that can be useful in determining a diagnosis.
(PLEASE DO NOT WRITE BELOW THE LINE – CONTINUE ON ANOTHER PAGE IF NECESSARY)
. Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use X (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Eva
510K K092308
Dogecoin2
Page 1 of 1
Attachment B
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.