K031230 · GE Medical Systems · LNH · May 2, 2003 · Radiology
Device Facts
Record ID
K031230
Device Name
GE PROPELLER IMAGING OPTION FOR MRI
Applicant
GE Medical Systems
Product Code
LNH · Radiology
Decision Date
May 2, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
When used with diffusion sensitizing gradients on a minimum of 3 axes, it will augment the standard Echo Planar Imaging (EPI) based diffusion imaging by providing improved image quality in areas of high susceptibility. This may aid the trained physician in the visualization of pathology in areas traditionally obscured by susceptibility artifacts. It is intended for use in anatomical regions where significant susceptibility differences exist between adjacent structures (e.g. tissue/air and tissue/bone). This includes, but is not limited to, inferior brain areas such as cerebellum, internal auditory canal, vertebrae and orthopedic areas. Other high susceptibility regions include tissue in the presence of MRI compatible metallic implants, artificial joint, etc. When used in the place of high resolution FSE, it will yield improved contrast-to-noise and signal-to-noise with comparable resolution and overall scan time, or improved contrast-to-noise and signal-to-noise with minimal in-plane motion-induced artifact. This may aid the trained physician in visualization of areas with reduced signal and image contrast or areas obscured by motion artifacts. It is intended for use in high-resolution anatomical regions where FSE is traditionally used, particularly in neurological imaging, where improved contrast-to-noise and signal-to-noise is needed.
Device Story
Propeller Imaging Option is a Fast Spin Echo (FSE)-based MRI pulse sequence; utilizes alternate k-space trajectory to improve signal-to-noise (SNR) and contrast-to-noise (CNR) ratios; reduces motion-induced artifacts. When configured with diffusion sensitizing gradients (min 3 axes), it augments standard EPI-based diffusion imaging; improves image quality in high-susceptibility regions (tissue/air, tissue/bone interfaces, metallic implants). Operated by trained MRI technicians/physicians in clinical settings; output is diagnostic MRI images. Physician uses images to visualize pathology in areas traditionally obscured by artifacts; benefits patient through clearer diagnostic visualization in challenging anatomical regions.
Clinical Evidence
Bench testing only. Evaluated against IEC 60601-1-4 (Programmable Electrical Medical Systems) and IEC 60601-2-33 (Magnetic Resonance Systems safety). No clinical data provided.
Technological Characteristics
Software-based MRI pulse sequence; Fast Spin Echo (FSE) architecture; utilizes alternate k-space trajectory; compatible with diffusion sensitizing gradients (min 3 axes). Operates on GE MRI hardware platforms. Complies with IEC 60601-1-4 and IEC 60601-2-33 standards.
Indications for Use
Indicated for patients requiring MRI imaging in anatomical regions with high susceptibility (e.g., inferior brain, cerebellum, internal auditory canal, vertebrae, orthopedic areas, or near metallic implants) or regions where high-resolution FSE is used. Aids physicians in visualizing pathology obscured by susceptibility or motion artifacts.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
GE Medical System Diffusion Weighted EPI Imaging Option (K972990)
Echo Planar Imaging Option including FSE (K944979)
Submission Summary (Full Text)
{0}------------------------------------------------
## SUMMARY OF SAFETY AND EFFECTIVENESS
- . This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h).
- . ldentification of Submitter Larry A. Kroger, Ph.D.,
MAY - 2 2003
- . Phone:
262-5443464
- . Fax:
262-548-4768
- Date Prepared:
March 21, 2003
- . Identification of the Product Propeller Imaging Option for MRI
Manufactured by: GE Medical Systems 3200 N Grandview Blvd. Waukesha, WI 53188
- . Common Name Propeller Imaging Option for MRI
- . Classification Name Magnetic Resonance Diagnostic System, 21 CFR 892.1000, 90-LNH
- . Device Description
The Propeller Imaging Option is a Fast Spin Echo (FSE) based pulse sequence that provides improved signal to noise and contrast to noise compared to traditional FSE with comparable resolution and scan time, and may be used to reduce motioninduced artifacts. With the addition of Diffusion gradients, Propeller may be used to differentiate tissues with restricted diffusion from tissues with normal diffusion, similar to Diffusion Weighted EPI.
- . Indications for Use
When used with diffusion sensitizing gradients on a minimum of 3 axes, it will augment the standard EPI-based diffusion imaging by providing improved image quality in areas of high susceptibility. This may aid the trained physician in the visualization of pathology in areas traditionally obscured by susceptibility artifacts.
It is intended for use in anatomical regions where significant susceptibility differences exist between adjacent structures (e.g. tissue/air and tissue/bone). This includes, but is not limited to, inferior brain areas such as cerebellum, internal auditory canal,
{1}------------------------------------------------
vertebrae and orthopedic areas. Other high susceptibility regions include tissue in the presence of MRI compatible metallic implants, artificial joints, etc.
When used in the place of high resolution FSE, it will yield improved contrast-tonoise and signal-to-noise with comparable resolution and overall scan time, or improved contrast-to-noise and signal-to-noise with minimal in-plane motion-induced artifact. This may aid the trained physician in visualization of areas with reduced signal and image contrast or areas obscured by motion artifacts.
It is intended for use in high-resolution anatomical regions where FSE is traditionally used, particularly in neurological imaging, where improved contrast-to-noise and signal-to-noise is needed.
## . Comparison with Predicate
The Propeller Imaging Option is substantially equivalent to the features currently marketed GE Medical System Diffusion Weighted EPI Imaging Option (510k #K972990), and Echo Planar Imaging Option including FSE (K944979).
## Summary of Studies .
The Propeller Imaging Option was evaluated to thelEC60601-1-4, the Programmable Electrical Medical Systems standard, and IEC 60601-2-33 International medical equipment safety standard for Magnetic Resonance Systems.
## . Conclusions
It is the opinion of GE that the Propeller Imaging Option for MRI does not result in any new potential hazards.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with three curved lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the bird symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
GE Medical Systems % Mr. Heinz-Joerg Steneberg Division Manager, Medical Department TUV Rheinland of North America 12 Commerce Road NEWTON CT 06470
Re: K031230
MAY = 2 2003
Trade/Device Name: GE Propeller Imaging Option for MRI Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: 90 LNH Dated: April 17, 2003 Received: April 18, 2003
Dear Mr. Steneberg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): K� 3 |2 ]0
Device Name: GE Propeller Imaging Option for MRI
Indications For Use:
Propeller Imaging Option is a Fast Spin Echo (FSE) based sequence which employs a alternate k-space trajectory.
- 1. When used with diffusion sensitizing gradients on a minimum of 3 axes, it will augment the standard Echo Planar Imaging (EPI) based diffusion imaging by providing improved image quality in areas of high susceptibility. This may aid the trained physician in the visualization of pathology in areas traditionally obscured by susceptibility artifacts.
It is intended for use in anatomical regions where significant susceptibility differences exist between adjacent structures (e.g. tissue/air and tissue/bone). This includes, but is not limited to, inferior brain areas such as cerebellum, internal auditory canal, vertebrae and orthopedic areas. Other high susceptibility regions include tissue in the presence of MRI compatible metallic implants, artificial joint, etc.
- 2. When used in the place of high resolution FSE, it will yield improved contrast-tonoise and signal-to-noise with comparable resolution and overall scan time, or improved contrast-to-noise and signal-to-noise with minimal in-plane motion-induced artifact. This may aid the trained physician in visualization of areas with reduced signal and image contrast or areas obscured by motion artifacts.
It is intended for use in high-resolution anatomical regions where FSE is traditionally used, particularly in neurological imaging, where improved contrast-to-noise and signalto-noise is needed.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) for NB
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K031230
Prescription Use ***_***
(Per 21 CFR 801.109)
OR Over-The-Counter Use **_**
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.