EXCELART™ with Pianissimo XG/AG SPIN Edition is a 1.5T MRI system. Inputs include RF signals from patient anatomy via various RF coils (e.g., QD Torso SPEEDER, 8ch phased array, 70mm circular coil). System transforms signals into diagnostic images using high-speed reconstruction computer. Features include True Steady State Free Precession (SSFP) sequence, parallel imaging, and redesigned gating unit for improved sequence triggering. Operated by trained clinicians in clinical settings. Output is diagnostic MR imagery used for clinical decision-making across whole-body applications. Benefits include reduced scan times via parallel imaging and improved image processing speeds. Safety features include dB/dt limitation control (1st operating mode) and emergency ramp-down unit.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and performance parameters (static field strength, SAR, acoustic noise, gradient dimensions) compared against predicate to demonstrate equivalence.
Technological Characteristics
1.5T MRI system. Materials: standard medical-grade components. Energy: RF and magnetic gradient fields. Connectivity: Standalone system. Software: Improved UI, RF coil control, dB/dt limitation control, True SSFP sequence, parallel imaging. Gradient coil dimensions: 1050 x 1175 x 50 mm. Max dB/dt: 46 T/sec.
Indications for Use
Indicated for whole-body MRI imaging, including head, abdomen, pelvis, limbs, spine, neck, TMJ, heart, and vasculature. Supports specialized applications like MRCP, MR fluoroscopy, BOLD, perfusion/diffusion, and proton spectroscopy. No specific age or gender restrictions noted.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
ਤੇ ਨੇ ।।
5 2002 NOV
Appendix 1
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
| 1. DEVICE NAME: | Magnetic Resonance Diagnostic Device Accessory |
|---------------------------------|--------------------------------------------------------------------------------------------------------------|
| Model Name: | MRT-1501/P3,P2 |
| Trade/Proprietary Name: | EXCELART™ with Pianissimo XG/AG SPIN Edition |
| 2. ESTABLISHMENT REGISTRATION: | 2020563 |
| 3. U.S. Agent Name and Address: | TOSHIBA AMERICA MEDICAL SYSTEMS, INC.<br>2441 MICHELLE DRIVE<br>TUSTIN, CA 92780 |
| Contact Person: | Michaela Mahl<br>(714) 730 - 5000 |
| 4. Manufacturing Site: | TOSHIBA CORPORATION<br>MEDICAL SYSTEMS COMPANY<br>1385 Shimoishigami<br>Otawara-shi, Tochigi 324-8550, Japan |
| 5. DATE OF SUBMISSION: | September 27, 2002 |
#### DEVICE DESCRIPTION 6.
The EXCELART™ with Pianissimo XG/AG SPIN Edition system has the following features compared to the current EXCELART™ with Pianissimo XG/AG system.
- Extension of a receiving system is enabled a maximum of 8 ch. .
- Use of QD Torso SPEEDER corresponding to Parallel imaging which can shorten scan time is . enabled.
- The computer only for reconstruction was carried for high-speed image processing. .
- . dB/dt enabled use by 1st operating mode specified in IEC 60601-2-33 by the TrueSSFP sequence.
- Max. field strength is changed 25mT/m to 30mT/m. .
- . The marginal performance (Min.TR/Min.TE / Min.Slice thickness / Imaging area) of a sequence has been improved.
- The user interface of software was changed for the improvement of operativity. '
- . Gating unit was redesigned to improve the making triggers for sequence.
| Model Number with suffix | Trade/Proprietary Name |
|--------------------------|--------------------------------------------------------|
| MRT-1501/P3 | EXCELART <sup>TM</sup> with Pianissimo XG SPIN Edition |
| MRT-1501/P2 | EXCELART <sup>TM</sup> with Pianissimo AG SPIN Edition |
{2}------------------------------------------------
#### SUMMARY OF MAJOR HARDWARE CHANGES 6.1.
A. The computer only for reconstruction was carried for high-speed image processing.
- B. Gating unit was redesigned to improve the making triggers for sequence.
C: 8ch phased array kit is added in the optional items.
D. 70mm Circular coil is added in the optional items.
E. QD torso SPEEDER is added in the optional items.
F. QD Neurovascular array coil (K001870) is added in the optional items.
#### SUMMARY OF MAJOR SOFTWARE CHANGES 6.2.
A. Improved user interface.
- B. New RF coils control.
- C. dB/dt limitation control.
- D. True Steady State Free Precession (SSFP) sequence.
- E. Parallel imaging.
#### 7. SAFETY PARAMETERS
| | Current EXCELART™ with<br>Pianissimo XG/AG (No changes<br>from the previous submission,<br>K002531) | New EXCELART™<br>with Pianissimo<br>XG/AG SPIN Edition |
|-----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|--------------------------------------------------------|
| a. Static field strength: | 1.5 T | Same |
| b. Peak and A-weighted acoustic<br>noise: | 95.0 dB(A-weighted) | 95.4 dB(A-weighted) |
| c. Operational modes: | Normal operating mode | 1st operating mode for<br>dB/dt |
| i. Safety parameter display: | SAR | SAR, dB/dt |
| ii. Operating mode access<br>requirements: | Not applicable because used<br>only in normal operating mode | Same |
| d. Maximum SAR | < 1.5 W/kg | Same |
| e. Maximum dB/dt<br>and Gradient coil dimensions: | 19.3T/sec<br>1050 x 1175 x 51<br>(unit: mm) | 46 T/sec<br>1050 x 1175 x 50<br>(unit: mm) |
| f. Potential emergency conditions<br>and means provided for shutdown: | Shut down by Emergency Ramp<br>Down Unit for collision hazard<br>by ferromagnetic objects | Same |
| g. Biocompatibility of materials: | Not applicable | Same |
#### IMAGING PERFORMANCE PARAMETERS 8.
No changes from the previous submission, K002531.
#### 9. INTENDED USE
No changes from the previous submission, K002531.
{3}------------------------------------------------
## 10. EQUIVAALENCY INFORMATION
TOSHIBA Corporation Medical Systems Company believes that the new EXCELART™ with Pianissimo XG/AG SPIN Edition (model MRT-1501/P3, MRT-1501/P2) Magnetic Resonance Imaging (MRI) system is substantially equivalent to the current EXCELART™ with Pianissimo XG/AG (model MRT-1501/P3, MRT-1501/P2) (K002531) cleared on October 26, 2000 except for new RF coils and dB/dt limitation.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a black and white logo of the U.S. Department of Health & Human Services. The logo features a stylized eagle with three curved lines representing its body and wings. The words "DEPARTMENT of HEALTH & HUMAN SERVICES" are arranged vertically along the left side of the logo in a circular fashion.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
5 2002 NOV
Toshiba America Medical Systems, Inc. % Mr. Mark Job 510(k) Program Manager TÜV America, Inc. 1775 Old Highway 8 NEW BRIGHTON MN 55112 Re: K023511
Trade/Device Name: EXCELART™ with Pianissimo XG, AG SPIN Edition Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: 90 LNH
Dated: October 17, 2002 Received: October 21, 2002
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours.
Nancy C. Snigdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Page 1
KO2 35 11 510(k) Number (if known).
Device Name: EXCELART™ with Pianissimo XG, AG SPIN Edition
Indications for Use:
Imaging of:
- . The Whole Body (including head, abdomen, pelvis, limbs and extremities, spine, neck, TMJ, heart, blood vessels). [Application terms include MRCP (MR Cholangiopancreatography), MR Cisternography, MR Myelography, MR Myelography, MR Fluoroscopy, SAS (Surface Anatomy Scan), Dynamic Scan, Cine Imaging and Cardiac tagging.] { K002531 }
- Fluid Visualization { K002531 } ●
- . 2D / 3D Imaging { K002531}
- . MR Angiography / MR Vascular Imaging {K002531}
- Blood Oxygenation Level Dependent (BOLD) imaging { K002531} ●
- Perfusion / Diffusion Imaging {K993803, K002531}
- . Proton Spectroscopy { K010129}
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR -
】【
Over-The-Counter Use
(Optional Format 1-2-96)
.
(Division Sign Off)
Division: Digestive, Abdominal,
and GI Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.