OR TABLE AND HEADHOLDER/MAGNETOM HARMONY, SYMPHONY, SONATA SYSTEMS

K012495 · Siemens Medical Solutions USA, Inc. · LNH · Oct 31, 2001 · Radiology

Device Facts

Record IDK012495
Device NameOR TABLE AND HEADHOLDER/MAGNETOM HARMONY, SYMPHONY, SONATA SYSTEMS
ApplicantSiemens Medical Solutions USA, Inc.
Product CodeLNH · Radiology
Decision DateOct 31, 2001
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1000
Device ClassClass 2

Indications for Use

The MAGNETOM Harmony, Symphony, and Sonata System with the OR Table and Headholder is a whole body scanner. The MAGNETOM Systems with the OR Table and Headholder is indicated for use as diagnostic imaging device to produce transversal, sagittal, coronal and oblique images of the internal structures of the head or body. The images produced by MAGNETOM Systems with the OR Table and Headholder reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis. The OR Table with headholder will be an option for intra-operative and interventional procedures.

Device Story

Device consists of optional OR table and headholder for MAGNETOM Harmony, Symphony, and Sonata MRI systems. System functions as whole-body scanner; inputs are magnetic resonance signals from patient protons; outputs are transversal, sagittal, coronal, and oblique diagnostic images based on proton density, T1, T2, and flow. Used in clinical settings by physicians for diagnostic, interventional, and intra-operative procedures. Table/headholder integration allows patient positioning during surgery or intervention. Physician interprets images to assist in diagnosis. Benefit includes expanded utility of existing MRI systems for surgical/interventional environments without compromising safety or image quality.

Clinical Evidence

Bench testing only. No clinical data provided. Evaluation focused on interference testing of the OR table to ensure no impact on safety or effectiveness of the existing MAGNETOM system.

Technological Characteristics

Optional OR table and headholder accessory for MRI systems. System utilizes magnetic resonance imaging (proton density, T1, T2, flow). No changes to magnet, gradients, RF, or operating software of the base MAGNETOM systems. Standard MR safety parameters maintained.

Indications for Use

Indicated for patients requiring diagnostic MR imaging of head or body structures; enables intra-operative and interventional procedures.

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ # 10 Appendix: 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92. #### General Information. I. #### Establishment: - Siemens Medical Systems, Inc. . Address: 186 Wood Avenue South Iselin, N.J. 08830 2240869 Registration Number: - Contact Person: Mr. Jamie Yieh Technical Specialist, Regulatory Affairs Phone: (732) 321-4625 Email: jamie.yieh@sms.siemens.com Fax: (732) 321-4841 #### Date of Summary Preparation: 8/2/01 #### Device Name: - •Trade Name: OR Table and Headholder with the MAGNETOM Harmony, Symphony, Sonata - ·Classification Name: Magnetic Resonance Diagnostic Device, CFR § 892.1000 Class II •Classification: ·Performance Standards: None established under Section 514 the Food, Drug, and Cosmetic Act. {2}------------------------------------------------ # II. Safety and Effectiveness Information Supporting Substantial Equivalence. ## • Device Description: ## ·Intended Use The MAGNETOM Harmony, Symphony, and Sonata System with the OR Table and Headholder is a whole body scanner. The MAGNETOM Systems with the OR Table and Headholder is indicated for use as diagnostic imaging device to produce transversal, sagittal, coronal and oblique images of the internal structures of the head or body. The images produced by MAGNETOM Systems with the OR Table and Headholder reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis. The OR Table with headholder will be an option for intra-operative and interventional procedures. #### · Technological Characteristics The MAGNETOM Harmony, Symphony, Sonata with the new OR Table and Headholder is substantially equivalent to the Neuro II system. ### · General Safety and Effectiveness Concerns: The introduction of the new OR Table and headholder has no significant effect on the following MR safety and performance parameters: # [Safety] - . Maximum Static Field - . Rate of Change of Magnetic Field - ◆ RF Power Deposition - . Acoustic Noise Level # [Performance] - . Specification Volume - . Signal to Noise - Image Uniformity - . Geometric Distortion - . Slice Profile, Thickness and Gap - High Contrast Spatial Resolution Since we are simply adding an optional OR Table and headholder to the already existing MAGNETOM Harmony, Symphony, Sonata system, which will have exactly the same components (i.e. magnet, gradients, RF, and operating software). The optional OR Table and headholder will not have any significant effect on the currently available systems. The OR Table and headholder will be used in various applications ranging from regular diagnostic scans to interventional/intra-operative procedures. To ensure that the OR Table or headholder does not influence the MAGNETOM {3}------------------------------------------------ Harmony, Symphony, Sonata, Siemens did an evaluation on interference from the OR table which could affect the safety or effectiveness with the currently available systems. ### • Substantial Equivalence: Laboratory testing were performed to support this claim of substantial equivalence and to show that the technological differences do not raise any new questions pertaining to safety and effectiveness. {4}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT 31 2001 Mr. Jamie Yieh Technical Specialist, Regulatory Affairs Siemens Medical Systems, Inc. 186 Wood Avenue South ISELIN NJ 08830 Re: K012495 Trade/Device Name: OR Table and Headholder for MAGNETOM Harmony, Symphony Sonata Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: 90 LNH Dated: August 2, 2001 Received: August 3, 2001 Dear Mr. Yieh: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {5}------------------------------------------------ This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4610 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on by received to premainted the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely vours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ 510(k) Number (if known) K) צעצ OR Table and Headholder for MAGNETOM Harmony, Device Name: Symphony, and Sonata Systems #### Indications for Use: The MAGNETOM Harmony, Symphony, and Sonata System with the OR Table and Headholder is a whole body scanner. The MAGNETOM Systems with the OR Table and Headholder is indicated for use as diagnostic imaging device to produce transversal, sagittal, coronal and oblique images of the internal structures of the head or body. The images produced by MAGNETOM Systems with the OR Table and Headholder reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis. The OR Table with headholder will be an option for intra-operative and interventional procedures. (please do not write below this line- continue on another page if needed) Concurrence of CDRH, Office of Device Evaluation | Prescription Use <span style="text-decoration:underline;"></span> | |-------------------------------------------------------------------| |-------------------------------------------------------------------| OR | Over-The-Counter Use <span style="text-decoration:underline;"></span> | |-----------------------------------------------------------------------| |-----------------------------------------------------------------------| David A. Segerson (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number K012495
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...