K023271 · Terumo Medical Corp. · FMF · Dec 18, 2002 · General Hospital
Device Facts
Record ID
K023271
Device Name
TERUMO SYRINGE WITH/WITHOUT NEEDLE
Applicant
Terumo Medical Corp.
Product Code
FMF · General Hospital
Decision Date
Dec 18, 2002
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Terumo Syringe with/without needle is intended to be used for medical purposes to inject fluids into or withdraw fluids from the body.
Device Story
Manual piston syringe; used for fluid injection or withdrawal. Components: polypropylene barrel, plunger, thermoplastic elastomer gasket, stainless steel needle (optional). Operated manually by clinician or user. Barrel features clear graduations for dosage control; gasket facilitates smooth plunger movement. Needle design minimizes skin surface resistance. Individually peel-blister packed; sterile. Used in clinical or home settings for general medical fluid management.
Clinical Evidence
Bench testing only. Performance evaluations included cleanliness, leakage, sliding resistance, plunger/gasket fit force, nozzle-needle fit force, initial movement force, stopper force, dead space, graduation tolerance, primary container seal strength, silicone amount, elution, heavy metals, and pH. Sterilization validated per ISO 11137.
Indicated for medical purposes to inject fluids into or withdraw fluids from the body. No specific patient population, age, or gender restrictions are defined.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
Terumo Medical Corporation Hypodermic Syringe (K980181)
Submission Summary (Full Text)
{0}------------------------------------------------
Section II 510(k) Summary Terumo Syringe with/without Needle Terumo (Philippines) Corporation
#### 510(k) Summary
#### Submitter Information:
Name and Address: Terumo (Philippines) Corporation #124 East Main Ave. Laguna Technopark, Biñan, Laguna, Philippines 4026
Contact Person:
Ms. Sandi Hartka Manager Regulatory Affairs Terumo Medical Corporation 950 Elkton Blvd., 410-392-7243 Phone Number: Fax Number: 410-398-6079
#### Device Name:
| Propriety Name: | Terumo Syringe with/without Needle |
|----------------------|----------------------------------------------------------------|
| Common Name: | Piston Syringe, Disposable Syringe |
| Classification Name: | Syringe, Piston with or without Hypodermic Single Lumen Needle |
#### Predicate Device:
The Terumo Syringe with/without needle that is the subject of this premarket notification is substantially equivalent to the predicate device, Terumo Medical Corporation Hypodermic Syringe, which is legally marketed and was cleared by FDA (K980181).
#### Intended Use:
The Terumo Syringe with/without needle is intended to be used for medical purposes to inject fluids into or withdraw fluids from the body.
#### Principle of Operation and Technology:
Each Terumo Syringe with/without needle is designed for manual use therefore it is operated manually.
{1}------------------------------------------------
Section II 510(k) Summary Terumo Syringe with/without Needle Terumo (Philippines) Corporation
#### Design and Materials
The Terumo Syringe with/without needle consists mainly of 3 parts or 4 parts: a barrel, a plunger, a gasket and a needle (for the needle type). Barrel is made from polypropylene and is designed with clear graduations and figures for easy use. The gasket is made from a highly inert thermoplastic elastomer material facilitates smooth movement of the plunger and aids the user in controlling dosage. The needle is made from stainless steel and the needlepoint is carefully designed to minimize resistance at the skin surface and aids in smooth, easy insertion. The Terumo Syringe with/without Needle is individually packed in peel blister that ensures the sterility of the device until the package is opened.
#### Performance Evaluations:
The Terumo Syringe with/without needle submitted in this premarket notification was subjected to the following tests to demonstrate the safety and efficacy of the device:
- . Cleanliness
- Leakage Test .
- Sliding Resistance ●
- Plunger and Gasket Fit Force ●
- Nozzle-needle Fit Force .
- Initial Movement Force .
- Stopper Force .
- Dead Space .
- Equal to Nominal Capacity Graduation Tolerance .
- Between Main Capacity Graduation Tolerance .
- Primary Container Seal Strength .
- Silicone Amount Determination .
- Elution Test Method .
- Analysis of Heavy Metals .
- Determination of pH .
#### Specifications:
| PRODUCT | SYRINGE SIZES | NEEDLE<br>GAUGE | EXPOSED NEEDLE<br>LENGTH |
|---------------------------------------------------------------------------|-----------------------------------------------|----------------------------------|----------------------------------------------------------|
| Terumo Syringe<br>With/without Needle | 3 cc, 5 cc, 10 cc and<br>20 cc Terumo syringe | Gauges 20 - 27 | 16 mm (5/8") to 38 mm (1 ½) |
| Terumo Disposable<br>Hypodermic Syringe<br>(predicate device,<br>K980181) | 3 cc, 5cc, 10<br>20 cc hypodermic<br>syringe | Gauges 18 - 24<br>Gauges 25 - 27 | 19 mm (3/4) to 38 mm (1 ½)<br>13 mm (1/2") to 25 mm (1") |
- Page 7 -
{2}------------------------------------------------
510(k) Summary Section II Terumo Syringe with/without Needle Terumo (Philippines) Corporation
#### Substantial Equivalence Comparison:
The Terumo Syringe with/without needle is substantially equivalent to the precicate Terumo Medical Corporation Hypodermic Syringe as follows:
- Intended Use: Both Terumo Syringe with/without Needle and the predicate Terumo . Disposable Hypodermic Syringe, with or without hypodermic single lumen needle is intended to be used for injecting fluids or withdrawing fluids from the body.
- Principles of Operation: Terumo Syringe with/without Needle and the cleared Terumo . Disposable Hypodermic Syringe (K980181) are operated manually.
- Design and Materials: The design of the Terumo Syringe with/without Needle and the . predicate Terumo Disposable Hypodermic Syringe is basically the same. Both devices are comprised of a barrel, plunger and gasket (and needle) that are blister packed.
Both Terumo Syringe with/without Needle and the predicate Terumo Disposable Hypodermic Syringe is made of polypropylene and thermoplastic elastomer. The materials used in the molding of the devices are the same relative to device functioning. Any difference in these materials does raise any new issues of safety and/or effectiveness.
- Performance: The Terumo Syringe with/without Needle has been demonstrated to be safe . and effective during handling and for the duration if its expiration. The Terumo Syringe with/without Needle is substantially equivalent to the predicate Terumo Disposable Hypodermic Syringe.
#### Substantial Equivalence Summary:
In summary, the Terumo Syringe with/without Needle and the predicate Terumo Disposable Hypodermic Syringe are substantially equivalent in intended use, principles of operation, design and materials, and performance. Any noted differences between the two devices do not raise new issues of safety or effectiveness.
#### Additional Safety Information:
- Sterilization conditions have been validated in accordance with ISO 11137 to provide a . Sterility Assurance Level (SAL) of 10to.
- . Performance evaluations were conducted on aged and non-aged devices to ensure that the aging process does not adversely affect the performance of the device.
{3}------------------------------------------------
Section II 510(k) Summary Terumo Syringe with/without Needle Terumo (Philippines) Corporation
#### Conclusion:
In summary, the Terumo Syringe with/without Needle is substantially equivalent in intended use, operational principle, design and materials, and performance to the predicate Terumo Disposable Hypodermic Syringe (K980181).
Terumo (Phils.) Corporation's Statement that this device is substantially equivalent to any other device is done solely to comply with the requirements of the Federal Food, Drug and Cosmetic Act and is not intended to be the basis for patent infringement action.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# DEC 1 8 2002
Ms. Sandi Hartka Manager Regulatory Affairs Terumo Medical Corporation 950 Elkton Boulevard Elkton, Maryland 21921
Re: K023271
Trade/Device Name: Terumo Syringe with/without Needle Regulation Number: 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF Dated: September 9, 2002 Received: October 1, 2002
Dear Ms. Hartka:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/4/Picture/10 description: The image shows a partial view of a circular seal or emblem. The text "DEPARTMENT OF HEALTH" is visible along the left side of the circle. To the right, there is a stylized graphic of three human profiles facing the same direction, stacked one above the other. The overall impression is that of an official or organizational symbol related to health services.
{5}------------------------------------------------
#### Page 2 - Ms. Hartka
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrl/dsma/dsmamain.html
Sincerely yours,
Sina Punpe
Timothy A. Ulatowski Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
## 510(k) Number (if known): K02 327)
Device Name:
Terumo Syringe with/without Needle
#### Indications For Use:
The Terumo Syringe with/without needle is intended to be used for medical purposes to inject fluids into or withdraw fluids from the body. . .
G. Adolfo A. Pacheco
Geraldine A. Pasaba Regulatory Affairs Officer Terumo (Philippines) Corporation
### (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Patrone Cucurati
(Division Sign-Off) (Division Sign-On)
Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
510(k) Number: *K*
02327
Prescription Use
OR
Over-The-Counter Use
(Per 21 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.