K200629 · Control Medical Technology · FMF · May 29, 2020 · General Hospital
Device Facts
Record ID
K200629
Device Name
Aspire Mechanical Aspirator G
Applicant
Control Medical Technology
Product Code
FMF · General Hospital
Decision Date
May 29, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Aspire Aspirator G is a piston syringe to inject fluids into, or aspirate fluids from, the body.
Device Story
Hand-held, single-use mechanical aspirator; consists of barrel, plunger, and handles. Squeezing handles creates mechanical advantage to move plunger within barrel; handle design prevents plunger removal. Used for fluid injection or aspiration. Operated manually by clinician. Includes integrated pressure gauge for monitoring. Output is physical aspiration or injection of fluids; pressure gauge provides visual feedback to user. Benefits include controlled manual suction/injection capability. Device is identical to predicate in materials, design, and function, with addition of pressure gauge.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via operational/actuation tests, vacuum integrity, leakage tests (ISO 7886-1:2017, ISO 80369-20:2015), button force tests (MIL-STD-1472F), biocompatibility (ISO 10993-1), sterilization (ISO 11135-1:2014), and packaging/shipping integrity (ISO 1167-1:2006, ASTM F1608, ASTM D4169).
Technological Characteristics
Manual piston syringe; polycarbonate barrel and plunger; torsion spring; ISO Luer lock tip. Integrated pressure gauge. Single-use. Sterilized via ETO. Biocompatibility per ISO 10993. Mechanical actuation via handle-driven plunger.
Indications for Use
Indicated for use as a piston syringe to inject fluids into, or aspirate fluids from, the body.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
K072299 — POWRSYRINGE ASPIRATOR · Pinyons Medical Technology, Inc. · Oct 4, 2007
K072345 — POWRSYRINGE MONITOR · Pinyons Medical Technology, Inc. · Sep 21, 2007
K072318 — POWRSYRINGE INJECTOR · Pinyons Medical Technology, Inc. · Sep 21, 2007
K242143 — Angiography Injector · Shandong Int Medical Instruments Co., Ltd. · Dec 23, 2024
K163597 — VacLok AT Vacuum Syringe · Merit Medical Systems, Inc. · Jul 27, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
May 29, 2020
Control Medical Technology Shawn Fojtik President 2757 South 300 West Suite F (ZIEN) Salt Lake City, Utah 84115
Re: K200629
Trade/Device Name: Aspire Mechanical Aspirator G Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: DXE Dated: April 30, 2020 Received: May 01, 2020
Dear Shawn Fojtik:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Alan Stevens Assistant Director THT3C1: Injection Devices DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name
Indications for Use (Describe)
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5. 510(k) Summary K200629
In accordance with the Safe Medical Devices Act (SMDA) of 1990 and Title of the Code of Federal Regulations Part 807 (21 CFR §807), and in particular §807.92, the following summary of safety and effectiveness information is provided.
### 1. Submitter
Control Medical Technology 2757 South 300 West Salt Lake City, Utah 84115 USA Phone (383) 444-2666
### FDA Establishment Registration Number: #3007282893
### Contact
Control Medical Technology 2757 South 300 West Salt Lake City, Utah 84115 USA Phone (383) 444-2666
Date of Preparation: May 28, 2020
# 2. Subject Device
Trade Name:
Common Name: Classification Name: Product Code: Regulation:
Aspire Mechanical Aspirator G ("Aspire Aspirator G" and/or "Subject Device") Syringe and Aspirator Syringe, Piston FMF Class II, 21 CFR 880.5860
### 3. Predicate Device
Trade Names:
Common Name: Classification Name: Product Code: Regulation:
Aspire Mechanical Aspirator K072299 ("Aspire Aspirator" and/or "Predicate") Syringe and Aspirator Syringe, Piston FMF Class II, 21 CFR 880.5860
Image /page/3/Picture/18 description: The image shows a medical device labeled as a "PREDICATE DEVICE". The device appears to be a type of aspirator, specifically an "Aspire Aspirator". The text below the device also includes the numbers "K072299. Also K113757 and K131998".
Image /page/3/Picture/19 description: The image shows a close-up of a medical device labeled as a "Subject Device". The device appears to be a hand-operated aspirator, possibly used for suctioning fluids. It features a clear cylindrical chamber, a pressure gauge on top, and a black handle with a lever mechanism for creating suction. The text "Aspire Aspirator G" is visible at the bottom of the image.
Page 1 of 5 Section 5 510k Summary
{4}------------------------------------------------
### 4. Reference Device
| Trade Name: | PowrSyringe K072345 |
|----------------------|---------------------------|
| Common Name: | Syringe and Aspirator |
| Classification Name: | Syringe, Piston |
| Product Code: | FMF |
| Regulation: | Class II, 21 CFR 880.5860 |
Reference device utilized in submission because the pressure gauge is the same material, assembly, and processes as Subject Device.
### 5. Device Description
The Subject Device is a single-use, hand-held, and general-purpose aspirator configured with a barrel, plunger, and handles. The front handle is pivotally connected the back handle and distally connected to the barrel. The back handle is pivotally connected to the front handle and distally connected to the plunger. Squeezing the handles creates a mechanical advantage to move the plunger inside the barrel. The handle design also prevents the plunger from being pulled completely out of the barrel during use.
Image /page/4/Picture/7 description: The image shows two medical devices, labeled as "PREDICATE DEVICE" and "SUBJECT DEVICE". The "PREDICATE DEVICE" is identified as "Aspire Aspirator K072299. Also K113757 and K131998". The "SUBJECT DEVICE" is identified as "Aspire Aspirator G". Both devices appear to be hand-operated suction devices with a similar design, featuring a handle and a clear cylinder.
The Subject Device's design, intended use, fundamental scientific principle, operation, technology, actuation, use, barrel, plunger, front handle, button, dimensions, luers, connectors, polymers, molding, assembly, packaging, and sterilization is the same as the Predicate. Users may connect pressure gauges to Predicate during use. The main difference between the Subject Device and the Predicate is the addition of a basic off-the-shelf pressure gauge and change from a one-piece to two-piece button between the handles.
### 6. Indications for Use
The Subject Device is the same Indication as the Predicate: The Subject Device is a piston syringe to inject fluids into, or aspirate fluids from, the body.
### 7. Comparison to Predicate Devices
{5}------------------------------------------------
The Subject Device is substantially equivalent to the Predicate based on equivalence in:
Same Science Same Fundamental Scientific Principle
## Same Device Construction
- Design Mechanical Function Valves Barrels Handles
Mechanism of Action Manufacturing Dimensions Plungers
# Same Labeling
Same Indication for Use Warnings
Instructions for Use
### Same Manufacturing
Molding Packaging Assembly Sterilization
| | Predicate<br>Control Medical Technology<br>Aspire Aspirator K072299 | Subject Device<br>Control Medical Technology<br>Aspire Aspirator G | Comments |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| Common Name | Syringe and Aspirator | Syringe and Aspirator | Same as Predicate. |
| Classification Name | Syringe, Piston | Syringe, Piston | Same as Predicate. |
| Committee | General Hospital | General Hospital | Same as Predicate. |
| FDA Product Code | FMF | FMF | Same as Predicate. |
| Regulation | Class II, 21 CFR 880.5860 | Class II, 21 CFR 880.5860 | Same as Predicate. |
| Intended Use | The Aspire Aspirator is a piston<br>syringe to inject fluids into, or aspirate<br>fluids from, the body. | The Aspire Aspirator G is a piston<br>syringe to inject fluids into, or aspirate<br>fluids from, the body. | Same as Predicate. |
| Fundamental<br>Scientific Principle | The front handle is pivotally connected<br>to the back handle and distally<br>connected to the barrel. The back<br>handle is pivotally connected to the<br>front handle and distally connected to<br>the plunger. Squeezing the handles<br>creates a mechanical advantage to<br>move the plunger inside the barrel.<br>The handle design also prevents the<br>plunger from being pulled completely<br>out of the barrel during use. | The front handle is pivotally connected<br>to the back handle and distally<br>connected to the barrel. The back<br>handle is pivotally connected to the<br>front handle and distally connected to<br>the plunger. Squeezing the handles<br>creates a mechanical advantage to<br>move the plunger inside the barrel.<br>The handle design also prevents the<br>plunger from being pulled completely<br>out of the barrel during use. | Same as Predicate. |
| Mechanism | Handle design creates mechanical<br>advantage to move the plunger into<br>and out of the barrel. | Handle design creates mechanical<br>advantage to move the plunger into<br>and out of the barrel. | Same as Predicate. |
| User Control | Manual | Manual | Same as Predicate. |
| Barrels | Polycarbonate | Polycarbonate | Same as Predicate. |
| Barrel Volume | 30ml | 30ml | Same as Predicate. |
| Barrel Length | 9.5cm | 9.5cm | Same as Predicate. |
| Barrel Outer<br>Diameter | 3.5cm | 3.5cm | Same as Predicate. |
| Barrel Tip | ISO Luer lock | ISO Luer lock | Same as Predicate. |
| Plunger | Polycarbonate | Polycarbonate | Same as Predicate. |
| Plunger Length | 12cm | 12cm | Same as Predicate. |
| O-ring | Yes | Yes | Same as Predicate. |
| O-ring & Barrel<br>Lubricant | Med 400 | Med 400 | Same as Predicate. |
| Plunger Retention? | Yes, Handle design prevents plunger<br>from being pulled out of barrel when<br>handles are fully squeezed. | Yes, Handle design prevents plunger<br>from being pulled out of barrel when<br>handles are fully squeezed. | Same as Predicate. |
| Front Handle | Two-piece polycarbonate | Two-piece polycarbonate | Same as Predicate. |
| Front Handle Length | 16.75cm | 16.75cm | Same as Predicate. |
| Front Handle Width | 1.75cm | 1.75cm | Same as Predicate. |
| Back Handle | Two-piece snap together<br>polycarbonate | Two-piece snap together<br>polycarbonate | Same as Predicate. |
| Back Handle Length | 16.50cm | 16.50cm | Same as Predicate. |
| Back Handle Width | 1.75cm | 1.75cm | Same as Predicate. |
| Spring | Torsion spring | Torsion spring | Same as Predicate. |
| Pressure Gauge | Users may connect a separate<br>pressure gauge to the Predicate. | Subject Device includes a pressure<br>gauge. | Users may connect a separate<br>pressure gauge to the Predicate.<br>Same gauge as Reference<br>Device. |
| Drain Bag and<br>connecting tubing? | Yes >250ml | Yes >250ml | Same as Predicate. |
| Valve at tip? | Yes | Yes | Same as Predicate. |
| Valve at plunger? | Yes | Yes | Same as Predicate. |
| Packaging | Tray, Tyvek Lid, and Box | Tray, Tyvek Lid, and Box | Same Tray, Tyvek lid, and Box as<br>Predicate. |
| Sterilization | ETO | ETO | Same sterilization as Predicate. |
| Same Instruction for<br>Use? | Yes | Yes | Same IFU for Predicate and<br>Subject Device. |
| Single Use Device? | Yes, Single Use | Yes, Single Use | Same as Predicate. |
| Bio-compatibility? | Yes. ISO 10993 compliant | Yes. ISO 10993 compliant | Same as Predicate. No new<br>materials. Barrels, plungers, and<br>tubing are the exact same as<br>Predicate. |
{6}------------------------------------------------
{7}------------------------------------------------
# 8. Non-clinical Tests.
Bench testing confirms the Subject Device performs same as the Predicate including:
- Operational & Actuation Tests
- Vacuum Integrity Tests ●
- ISO 7886-1:2017 Annex B Barrel Leakage Tests ●
- ISO 7886-1:2017 Annex B Plunger Leakage Tests
- ISO 7886-1:2017 Annex B Luer Leakage Tests ●
- ISO 80369-20:2015 Annex D Fluid Seal Integrity Tests
- MIL-STD-1472F Button Force Tests
- ISO 10993-1 Biocompatibility
- ISO 11135-1:2014 Sterilization to achieve SAL 106 and EO/ECH limits below ISO ● 10993-7 standards
- ISO 1167-1:2006 Packaging Integrity Tests ●
- ASTM F1608 Package Integrity Tests ●
- ASTM D4169 Shipping Integrity Tests ●
- 9. Clinical Tests. No clinical study is required or included.
- 10. Conclusion. The Subject Device is substantially equivalent to the Predicate based on comparison of the device classification, product code, basic operating principle, fundamental scientific technology, indication for use, technical characteristics, packaging, and sterilization methods. Testing confirms the suitability of Subject Device for its intended use.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.