Angiography Injector

K242143 · Shandong Int Medical Instruments Co., Ltd. · DXT · Dec 23, 2024 · Cardiovascular

Device Facts

Record IDK242143
Device NameAngiography Injector
ApplicantShandong Int Medical Instruments Co., Ltd.
Product CodeDXT · Cardiovascular
Decision DateDec 23, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1650
Device ClassClass 2

Indications for Use

The Angiography Injector is intended to aspirate and inject contrast media in the interventional procedure. It can also be used to inject fluids into, or withdraw fluids from the body.

Device Story

Manual angiography injector; used in surgical operating rooms by clinicians. Device consists of piston, plunger cap, push-button, barrel, plunger, and adapter. Clinician manually operates device to aspirate or inject contrast media or other fluids. Provides visual feedback via graduated barrel. Single-use, sterile, disposable device. Benefits include controlled fluid delivery/withdrawal during interventional procedures.

Clinical Evidence

No clinical data. Bench testing only. Performance testing included appearance, lubricant, graduated scale, Luer connector (ISO 80369-7), nozzle lumen, force to operate, leakage, capacity tolerance, dead space, and particulate evaluation. Biocompatibility testing included cytotoxicity, hemolysis, intracutaneous reactivity, skin sensitization, acute systemic toxicity, and pyrogenicity.

Technological Characteristics

Manual syringe; volumes 6-12mL. Materials: PC barrel, ABS plunger/push-button, polyisoprene piston, HDPE plunger cap, PC/silicone adapter. Luer connector per ISO 80369-7; injector per ISO 7886-1. Sterilization: Ethylene Oxide (ISO 11135-1). Shelf life: 3 years (ASTM 1980).

Indications for Use

Indicated for aspiration and injection of contrast media during interventional procedures, or for injection/withdrawal of fluids from the body in patients requiring such procedures.

Regulatory Classification

Special Controls

*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. December 23, 2024 Shandong INT Medical Instruments Co., Ltd. Zhang Wei RA specialist No. 188. Fuzhou Road Rizhao. Shandong 276599 China Re: K242143 Trade/Device Name: Angiography Injector Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic Injector And Syringe Regulatory Class: Class II Product Code: DXT Dated: December 5, 2024 Received: December 6, 2024 Dear Zhang Wei: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {1}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2}------------------------------------------------ Sincerely, # Shruti N. Mistry -S Shruti Mistry Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ #### Indications for Use Submission Number (if known) #### K242143 Device Name Angiography Injector Indications for Use (Describe) The Angiography Injector is intended to aspirate and inject contrast media in the interventional procedure. It can also be used to inject fluids into, or withdraw fluids from the body. Type of Use (Select one or both, as applicable) < | Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ### K242143 - 510(K) summary #### I. Submitter | Submission Sponsor | Shandong INT Medical Instruments Co., Ltd.<br>No. 188, Fuzhou Road, Rizhao, Shandong 276599, China | |---------------------------|----------------------------------------------------------------------------------------------------| | Submission Correspondent: | Zhang Wei<br>RA specialist<br>Tel.: +8615601952252<br>E-mail: zhangwei@int-medical.com | | Submission Date: | December 5, 2024 | #### Proposed Device II. | Trade or Proprietary Name: | Angiography Injector | |----------------------------|-----------------------------------| | Common or Usual Name: | Angiographic Injector And Syringe | | Regulation Number: | 21 CFR 870.1650 | | Product code: | DXT | | Regulatory Class: | Class II | | Review Panel: | Cardiovascular | | Reason for Submission: | New device | #### III. Predicate Devices | Primary predicate device | | |--------------------------|--------------------------------| | 510(k) Number: | K163084 | | Trade name: | Merit Coronary Control Syringe | | Classification: | Class II | | Product Code: | DXT | | Manufacturer: | Merit Medical Systems, Inc. | #### IV. Indications for Use The Angiography Injector is intended to aspirate and inject contrast media in the interventional procedure. It can also be used to inject fluids into, or withdraw fluids from the body. #### V. Device Description {5}------------------------------------------------ The Angiography Injector is intended to provide the function of aspirating and injecting contrast media in the interventional procedure. It can also be used to inject fluids into, or withdraw fluids from the body. The proposed device is intended for single use and is provided sterile using EO sterilization. The Angiography Injector consists of six components: 1) Piston, 2) Plunger cap, 3) Push-button 4) Barrel, 5) Plunger and 6) Adapter (rotating or fixed). The proposed device is available in volume of 6mL, 8mL, 10mL and 12mL. The primary package is blister package and plastic paper package. The blister package and plastic paper package all consist of PE film and Tyvek®2FS paper. ### VI. Comparison to the Predicate Device The proposed Angiography Injector has the same indications for use, as well as similar materials, design and mode of action as the Merit Coronary Control Syringe predicate device. {6}------------------------------------------------ | Item | Proposed device | Predicate device<br>K163084 | Discussion | |----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------| | Product name | Angiography Injector | Merit Coronary Control Syringe | | | Product Code | DXT | DXT | Same | | Regulation No. | 21 CFR 870.1650 | 21 CFR 870.1650 | Same | | Class | Class II | Class II | Same | | Indications for<br>Use | The Angiography Injector is intended to<br>aspirate and inject contrast media in the<br>interventional procedure. It can also be<br>used to inject fluids into, or withdraw<br>fluids from the body. | The Merit Coronary Control Syringe is<br>intended to be used for the intraarterial<br>or intravenous administration of<br>radiographic contrast media. The<br>syringe can also be used to inject fluids<br>into, or withdraw fluids from the body. | Same | | Mode of Action | Intended for manual use | Intended for manual use | Same | | Environment of<br>use | Surgical operating room | Surgical operating room | Same | | Materials | Barrel - PC<br>Plunger - ABS<br>Piston - Polyisoprene rubber<br>Plunger Cap - HDPE<br>Push-button - ABS<br>Rotating adapter - PC + Silicone | Barrel - PC<br>Plunger - ABS<br>Seal - Silicone<br>Rotator O - Ring - Silicone | Similar1<br>(Discussion 1) | | Design Standards | Luer Connector to ISO 80369-7<br>Injector to ISO 7886-1 | Luer Connector to ISO 80369-7<br>Injector to ISO 7886-1 | Same | | Duration of Use | Less than 24h | Less than 24h | Same | | Volume | 6~12mL | 6~20 mL | Similar2<br>(Discussion 2) | | Graduation | Printed with accurate graduation lines | Printed with accurate graduation lines | Same | | | that are compliant with ISO 7886-1. | that are compliant with ISO 7886-1. | | | Sterile | Yes | Yes | Same | | Single for Use | Yes | Yes | Same | | Prescription<br>(Rx Only) | Yes | Yes | Same | | Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide | Same | | SAL | 10-6 | 10-6 | Same | | Shelf life | 3 years | 3 years | Same | | Barrel<br>Transparency | Clear | Clear | Same | | Physical and<br>Mechanical | ISO 7886-1 | ISO 7886-1 | Same | {7}------------------------------------------------ {8}------------------------------------------------ #### Comparison Discussion The following differences between the subject and predicate device do not raise new or different questions of safety and effectiveness. Discussion 11 The difference in the materials does not raise additional questions for safety and effectiveness of the device when compared to the predicate device. The biocompatibility evaluation test of the subject devices have been performed on the final finished device which includes all materials of construction. The test results show that the acceptance criteria have been met. Discussion 22 The volume of the proposed device is slightly different from the predicate device. This difference does not raise new or different questions of safety and effectiveness when compared to the predicate device. The performance tests have been carried out according to ISO 7886, and the test results show that the requirements of the standard have been met. #### VII. Bench Performance Testing The completion of the performance testing listed in the table below demonstrates that the proposed device meets the defined design specifications and is suitable for its intended use. ## Test Items: Appearance Lubricant Graduated scale Luer connector Nozzle lumen Force to operate the piston Freedom from air and liquid leakage past plunger stopper Tolerance on graduated capacity Fit of plunger stopper/plunger in barrel Dead space Particulate Evaluation Simulated use testing 5 / 7 Reducing (oxidizable) matter {9}------------------------------------------------ Limits for extractable metals Limits for acidity or alkalinity Residue of evaporation UV adsorption EO residual Sterility Bacterial endotoxin #### VIII. Biocompatibility Testing To demonstrate the biological safety of the body-contacting materials of the Angiography Injector, the following biocompatibility testing in accordance with "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"; Guidance for Industry and Food and Drug Administration Staff" were completed. #### Test Items: In Vitro Cytotoxicity Test In Vitro Hemolytic Intracutaneous Reactivity Skin Sensitization Acute Systemic Toxicity Pyrogen ### IX. Sterilization The subject device is sterilized by traditional Ethylene Oxide sterilization methods. The Ethylene Oxide sterilization process is validated as per ISO 11135-1 Sterilization of health care products -Ethylene Oxide - Part 1 (Requirements for development, validation and routine control of a sterilization process for medical devices), Method C (Overkill Method: This method is based on demonstrating that the sterilization of a microbial challenge (biological indicator) exceeds the challenge posed by the bioburden of the product. Confirmatory EO residual testing was {10}------------------------------------------------ conducted on the subject device to confirm that the design differences did not impact residual EO levels of the device. The testing confirmed that EO residuals were within the limits specified in ISO 10993-7. #### X. Packaging and Shelf Life The subject device is supplied sterile. The sterile barrier packaging is designed to maintain sterility until it open. The proposed shelf life of subject device is 3 years. The shelf life study for Angiography Injector was planned as per ASTM 1980 recommended conditions. The study was conducted with elevated temperature conditions and with real time conditions. As per ASTM 1980 recommended accelerated aging conditions were used to establish the shelf life of the Angiography Injector. The real time shelf time study is still on going. The results of the accelerated aging study has not shown any significant changes in product characteristics. ### XI. Clinical Testing No clinical study is included in this submission. ### XII. Conclusion The proposed device has the same indications and has similar design features and technological characteristic as the predicate device. Non-clinical testing data demonstrates that the proposed device is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...