K163597 · Merit Medical Systems, Inc. · PUR · Jul 27, 2017 · General Hospital
Device Facts
Record ID
K163597
Device Name
VacLok AT Vacuum Syringe
Applicant
Merit Medical Systems, Inc.
Product Code
PUR · General Hospital
Decision Date
Jul 27, 2017
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.5860
Device Class
Class 2
Indications for Use
VacLok™ AT Vacuum Syringe is used to inject fluids into, or withdraw fluids from the body. It can also be used in cases where a vacuum syringe is preferred (e.g. thrombus, abscess fluid, bile, urine etc.).
Device Story
VacLok™ AT Vacuum Syringe is a manual piston syringe used for fluid injection or aspiration. Device consists of a clear polycarbonate barrel, ABS plunger, and silicone piston seal. Features a locking mechanism allowing the plunger to be locked at specific positions along the barrel to maintain a vacuum. Operated manually by a clinician to advance or withdraw the plunger. Used in clinical settings for procedures requiring fluid management or vacuum-assisted aspiration. Output is the physical movement of fluids; healthcare providers use the device to perform aspiration or injection tasks, potentially improving procedural efficiency in fluid removal or delivery.
Clinical Evidence
Bench testing only. No clinical data. Performance testing included ISO 7886-1 (cleanliness, graduation, leakage, capacity) and ISO 594-2 (luer lock fittings, torque, leakage). Biocompatibility testing conducted per ISO 10993-1, including cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogenicity, and hemolysis.
Technological Characteristics
Piston syringe with male luer lock connector. Materials: polycarbonate barrel, ABS plunger, silicone seal. Manual operation. Vacuum capability via cam locking mechanism. Sterilization: Ethylene oxide (ISO 11135). Compliant with ISO 7886-1 (manual syringes) and ISO 594-2 (luer fittings).
Indications for Use
Indicated for injection or withdrawal of fluids from the body, including cases where vacuum is preferred (e.g., thrombus, abscess fluid, bile, urine).
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized design featuring three overlapping, curved shapes that resemble an abstract representation of a human figure or a bird in flight.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 27, 2017
Merit Medical Systems, Inc. Ms. Angela Brady Senior Regulatory Affairs Specialist 1600 W Merit Parkway South Jordan, Utah 84095
Re: K163597
Trade/Device Name: Merit VacLok™ AT Vacuum Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston syringe Regulatory Class: Class II Product Code: PUR Dated: July 25, 2017 Received: July 26, 2017
Dear Ms. Angela Brady:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Mark S. Fellman -S
for
Lori A. Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163597
Device Name VacLok™ AT Vacuum Syringe
Indications for Use (Describe)
VacLok™ AT Vacuum Syringe is used to inject fluids into, or withdraw fluids from the body. It can also be used in cases where a vacuum syringe is preferred (e.g. thrombus, abscess fluid, bile, urine etc.).
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| General<br>Provisions | Submitter Name:<br>Address:<br>Telephone Number:<br>Fax Number:<br>Contact Person:<br>Date Prepared:<br>Registration Number: | Merit Medical Systems, Inc.<br>1600 West Merit Parkway<br>South Jordan, UT 84095<br>(801) 316-4818<br>(801) 316-4878<br>Ms. Angela Brady<br>December 20, 2016<br>1721504 |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device | Trade Name:<br>Common/Usual Name:<br>Classification Name:<br>Regulatory Class:<br>Product Code:<br>21 CFR §:<br>Review Panel: | VacLok™ AT Vacuum Syringe<br>Piston Syringe<br>Syringe, Piston<br>2<br>PUR<br>880.5860<br>General Hospital |
| Predicate<br>Device | Trade Name:<br>Classification Name:<br>Regulatory Class:<br>Product Code:<br>21 CFR §:<br>Premarket Notification:<br>Manufacturer: | Welmed Hypodermic Syringe<br>Syringe, Piston<br>2<br>FMF<br>880.5860<br>K070936<br>Welmed, Inc. |
| Reference<br>Device | Trade Name:<br>Classification Name:<br>Regulatory Class:<br>Product Code:<br>21 CFR §:<br>Premarket Notification:<br>Manufacturer: | VacLok Syringe<br>Syringe, Piston<br>2<br>FMF<br>880.5860<br>K994253<br>Merit Medical Systems, Inc. |
| Device<br>Description | The VacLok™ AT Vacuum Syringe is a general piston syringe<br>constructed using a barrel, plunger, piston seal, and locking<br>mechanism. It is designed to lock in any position along the length of<br>the barrel with the capability of holding a vacuum when the cam locking<br>mechanism is engaged. | |
| Indications for<br>Use | | The VacLok™ AT Vacuum Syringe is used to inject fluids into, or withdraw<br>fluids from the body. It can also be used in cases where a vacuum syringe<br>is preferred (e.g. thrombus, abscess fluid, bile, urine etc.).<br>The Indications for Use statement for the VacLok™ AT Vacuum<br>Syringe is not identical to the predicate device; however, the |
# 510 (K) Summary
{4}------------------------------------------------
differences do not alter the intended therapeutic use of the device nor do they affect the performance of the device. Both the subject and predicate devices have the same intended use to inject fluids into, or withdraw fluids from, the body, as stated in 21 CFR 880.5860.
The proposed VacLok™ AT Vacuum Syringe incorporates the same intended use, as well as similar materials, design and principle of operation as the predicate Welmed Hypodermic Syringe. The subject device is substantially equivalent to the predicate device.
| | Attribute | Subject Device –<br>VacLok™ AT Vacuum<br>Syringe | Predicate Device -<br>Welmed Hypodermic<br>Syringe |
|--------------------------------------|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Design | Standard piston syringe<br>constructed with a clear<br>hollow barrel into which is<br>inserted a closely fitting<br>movable plunger and<br>tip/seal. Fitting offered<br>with male luer lock<br>connector. | Standard piston syringe<br>constructed with a clear<br>hollow barrel into which is<br>inserted a closely fitting<br>movable plunger and<br>tip/seal. Fitting offered<br>with male luer lock<br>connector. |
| Comparison to<br>Predicate<br>Device | Material | The barrel is constructed<br>from clear polycarbonate;<br>the plunger from ABS<br>material; the seal is made<br>of silicone material | The barrel is constructed<br>from clear polycarbonate.<br>The plunger/piston seal<br>unknown |
| | Principle of<br>Operation | Manually operated by<br>advancing and<br>withdrawing the plunger<br>within the barrel. | Manually operated by<br>advancing and<br>withdrawing the plunger<br>within the barrel. |
| | Operational<br>Volume | Operational volume of 20<br>and 30 mL. | Operational volume of 3,<br>5, 10, 20, 30 and 60 mL. |
| | Graduation | Printed with accurate<br>graduation lines that are<br>compliant with ISO 7886-<br>1. | Printed with accurate<br>graduation lines that are<br>compliant with ISO 7886-<br>1. |
| | Intended<br>Use | General purpose fluid<br>injection and aspiration<br>from the body, as outlined<br>in 21 CFR 880.5860 | General purpose fluid<br>injection and aspiration<br>from the body, as outlined<br>in 21 CFR 880.5860 |
{5}------------------------------------------------
The reference device supports the vacuum technology for the subject device.
| Reference<br>Device<br>Comparison | Attribute | Subject Device –<br>VacLok™ AT Vacuum<br>Syringe | Reference Device –<br>VacLok Syringe |
|-----------------------------------|--------------------------------|--------------------------------------------------|--------------------------------------|
| Vacuum<br>Technology | Variable camlock<br>technology | Fixed stop position<br>technology | |
FDA quidance and recognized consensus standards have been established for Piston Svringes under FDA Product Code FMF and 21 CFR 880.5860. A battery of tests was performed based on the requirements of the below recognized consensus standards and guidance, as well as biocompatibility, sterilization, and packaging standards and guidance. Conformity to these standards demonstrates that the proposed VacLok™ AT Vacuum Syringe met the standards' established acceptance criteria for the device.
- . ISO 7886-1:1993, Sterile hypodermic syringes for single use - Part 1: Syringes for manual use [Including Technical Corrigendum 1 (1995)]
- ISO 594-2:1998, Conical fittings with 6% (Luer) taper for syringes, ● needles and certain other medical equipment - Part 2: Lock fittings .
## Performance Data
- ISO 11135:2014. Sterilization of health care products Ethylene oxide - Requirements for the development, validation, and routine control of a sterilization process for medical devices
- ISO 10993-7:2008. Biological evaluation of medical devices Part ● 7: Ethylene oxide sterilization residuals
- AAMI TIR 28:2009, Product adoption and process equivalency for ● ethylene oxide sterilization
- ISO 11607-1:2006, Packaging for Terminally Sterilized Medical Devices - Part 1: Requirements for Materials, Sterile Barrier Systems and Packaging Systems [Including: Amendment 1 (2014)]
- ASTM D4169-14: 2014, Standard Practice for Performance Testing of Shipping Containers and Systems
- ASTM F1980-07:2007, Standard Guide for Accelerated Aging of ● Sterile Barrier Systems for Medical Devices (Reapproved 2011)
- . ISO 2233:2000. Packaging - Complete, filled transport packages and unit loads – Conditioning for testing
- ISO 10993-1:2009, Biological evaluation of medical devices Part . 1: Evaluation and Testing within a risk management process [Including: Technical Corrigendum 1 (2010)]
{6}------------------------------------------------
| | • ISO 10993-4:2002 (Amd.1:2006), Biological evaluation of medical devices - Part 4: Selection of tests for interactions with blood [Including: Amendment 1(2006)] |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | • ISO 10993-5:2009, Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity |
| | • ISO 10993-10:2010, Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization |
| | • ISO 10993-11:2006, Biological evaluation of medical devices – Part 11: Tests for systemic toxicity |
| | • ASTM F756-08:2008, Standard Practice for Assessment of Hemolytic Properties of Materials |
| | • AAMI/ANSI ST72:2011, Bacterial Endotoxins – Test methods, routine monitoring, and alternatives to batch testing |
| | • United States Pharmacopeia 37, National Formulary 32, 2014 <151> Pyrogen Test |
| | • FDA Guidance Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process, June 2016 |
| | • FDA Guidance Document, Guidance on the Content of Premarket Notification [510(k)] Submissions for Piston Syringes, April 1993 |
| | • FDA Guidance, The New 510(k) Paradigm – Alternate Approaches to Demonstrating Substantial Equivalence in Premarket Notifications, March 1998 |
| Performance Data cont'd | • FDA Guidance, Recognition and Use of Consensus Standards, September 2007 |
# Performance Bench Testing
The VacLok™ AT Vacuum Syringe complies with the FDA recognized consensus standards ISO 7886-1 and ISO 594-2, as outlined within this submission. Results of the testing demonstrate that the subject device met the acceptance criteria sufficient for its intended use. Testing included the following from these standards:
| ISO 7886-1 | |
|----------------------------------------|------------------------------------------------------------------|
| Cleanliness | Graduation Scale |
| Limits for acidity or alkalinity | Barrel |
| Limits for extractable metals | Piston/Plunger Assembly |
| Lubricant | Nozzle |
| Tolerance on Graduation | Performance - Freedom from air<br>and liquid leakage past piston |
| Capacity | |
| ISO 594-2 | |
| Gauging | Unscrewing torque |
| Liquid leakage | Ease of assembly |
| Air leakage | Resistance to overriding |
| Separation force | Stress cracking |
| Vacuum Performance Testing | |
| Air Leakage (Annex B of ISO<br>7886-1) | Vacuum Hold (Annex B of ISO<br>7886-1, 2 minutes) |
{7}------------------------------------------------
| <b>Biocompatibility testing</b><br>Biocompatibility evaluation for the VacLok™ AT Vacuum Syringe was<br>conducted in accordance with ISO 10993-1 "Biological evaluation of<br>medical devices - Part 1: Evaluation and testing within a risk<br>management process", and the FDA guidance "Use of International<br>Standard ISO 10993-1, “Biological evaluation of medical devices - Part<br>1: Evaluation and testing within a risk management process,” June<br>2016. Testing included the following: | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <b>Performance<br/>Data cont'd</b> | Cytotoxicity Sensitization Irritation Acute Systemic Toxicity Pyrogenicity Hemolysis USP Physicochemical The VacLok™ AT Vacuum Syringe is considered an externally<br>communicating device with indirect blood contact for a duration of less<br>than 24 hours. |
| <b>Summary of<br/>Substantial<br/>Equivalence</b> | Based on the intended use, materials, design, and performance testing,<br>the VacLok™ AT Vacuum Syringe meets the requirements that are<br>considered essential for its intended use and is considered substantially<br>equivalent to the predicate device, the Welmed Hypodermic Syringe,<br>K070936. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.