COMFORT EZ

K122787 · Simple Diagnostics, Inc. · FMF · Dec 6, 2012 · General Hospital

Device Facts

Record IDK122787
Device NameCOMFORT EZ
ApplicantSimple Diagnostics, Inc.
Product CodeFMF · General Hospital
Decision DateDec 6, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5860
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Comfort EZ™ disposable sterile insulin syringes are intended for injection of U100 insulin only.

Device Story

Comfort EZ™ is a disposable, sterile, single-use piston syringe with a fixed needle. It consists of a calibrated hollow barrel, plunger, gasket, and needle cap. Available in 28-31 gauge needle sizes and 0.3 ml/cc or 1 ml/cc capacities. Operated manually by patients or healthcare providers to measure and inject U100 insulin. The device functions via standard piston syringe mechanics; the needle is fixed in the barrel nozzle lumen and cannot be exchanged. The device is intended for over-the-counter use. By providing a sterile, calibrated delivery mechanism, it enables accurate insulin administration, supporting patient disease management.

Clinical Evidence

No clinical data. Bench testing only. Testing included biocompatibility (cytotoxicity, sensitization, intracutaneous toxicity, hemolysis, systemic toxicity, pyrogenicity), mechanical hub/needle bond strength, and sterility/EO residue testing. Compliance with ISO 8537:1991(E) for sterile, single-use insulin syringes was demonstrated.

Technological Characteristics

Materials: Polypropylene (barrel), stainless steel (needle), natural rubber (gasket), polyethylene (packaging film). Lubricant: Polydimethysiloxane (0.25 mg/cm2 max). Sterilization: Ethylene oxide (SAL 10^-6). Design: Fixed needle piston syringe, ISO 8537 compliant. Dimensions: 0.3 ml/cc and 1 ml/cc capacities; 28-31 gauge needles.

Indications for Use

Indicated for the injection of U100 insulin in patients requiring insulin therapy.

Regulatory Classification

Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(K) Summary, K122787 Simple Diagnostics® 6 2012 DEC 11555 Heron Bay Blvd., Suite 200 Coral Springs, Florida 33076 (877) 342-2385 な Contact person: Muhammad Arif Date prepared: October 17, 2012 - Trade Name: Comfort EZTM Insulin Syringe 1. Common Name: Insulin syringe Classification Name: Syringe, piston, product code FMF, Regulation: 880.5860 Class of device: Class II. - The legally marketed device to which we are claiming equivalence [807.92(a)(3)] : Feel-ject 2. Insulin Syringe made by Feel Tech (Korea) K070917 - Description of device: The Comfort EZ™ insulin syringe consists of a calibrated hollow 3. barrel which can contain the medication and the distal end of barrel is fixed with needle. The needle cannot be exchanged after assembling because needle is fixed in the barrel nozzle lumen. The plunger and gasket are the same shape as the conventional insulin syringes. The needle cap cover is intended to provide physical protection to the needle tube. The cap is color coded orange, same as equivalent insulin syringes. The syringes are available in 28 through 31 gauge needle sizes, in 0.3 ml/cc and 1 ml/cc capacities. They are supplied with a sterile fluid path, (EO), non-toxic, and non pyrogenic, for single use only, disposable. The devices operate on the principles of common piston syringes. - Intended use: For the injection of U100 insulin. 4. - Technological characteristics: The Comfort EZTM Insulin Syringes and the predicate 5. devices have identical technological characteristics and perform the same way as common piston svringes. The materials utilized remain the same as the predicate, mainly stainless steel for the needles and polypropylene for the barrel components. These syringes are EO sterilized and are for single use only. | Device Name | Feel-ject Insulin Syringe made by Feel Tech<br>(Korea) K070917 | Comfort EZTM insulin<br>syringe | |---------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------| | Intended Use | For the injection of U100 insulin. | Identical | | Method of Use | 1. To expose plunger, remove large white cap.<br>2. To expose needle, pull small orange cap straight<br>off, being careful not to bend needle.<br>3. To measure correct dosage, align top edge of<br>plunger tip with your dosage mark on the syringe<br>scale.<br>4. Use syringe only once. To prevent misuse after<br>injection, use sharps container | Identical | #### Comparison Table. 1 {1}------------------------------------------------ | Device Name | Feel-ject Insulin Syringe made by Feel Tech<br>(Korea) K070917 | Comfort EZTM insulin<br>syringe | |------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------| | Type | Type 8 of ISO8537:1991(E); syringe with fixed needle<br>tube and fitted with protective end cap. | Identical | | Barrel Marking | Scale: conforms to ISO 8537:1991(E) | Identical | | Lubricant | Polydimethysiloxane | Identical | | Lubricant amount<br>(mg/cm2) | 0.25 max | Identical | | Barrel transparency | Haze 25% max. (2mm sheet) | Identical | | Reuse | For only use | Identical | | Biocompatibility | Conforms to ISO 10993-1 | Identical | | Materials | 1) Plastic parts: polypropylene | Identical | | | 2) Gasket : natural rubber | Identical | | | 3) Packing film : Polyethylene film | Identical | | Sterility | Sterilized by ethylene oxide gas SAL = 10-6 | Identical | | Packaging | Each syringe is individually packaged in a Tyvek peel<br>pouch. There are 10 syringes packaged in a poly bag.<br>There are 10 poly bags packaged in a cardboard box. | Ten syringes are packaged in a<br>poly bag. There are 10 poly<br>bags packaged in a cardboard<br>box. (i.e. 100 syringes) | Performance: Bench and standards compliance testing was performed. 6. Bench testing included: Biocompatibility: (ref ISO 10993) Cytotoxicity, Maximization, Intracutaneous Toxicity, Hemolysis, Systemetic toxicity, and Pyrogen Testing, Mechanical testing: Hub/needle bond strength. ISO 7894:1993 Sterility testing (ISO 11607 and ISO 11135) including EO residues testing (ISO 10997-7.) Standards compliance: ISO 8537 Sterile, Single-Use Syringes, with or without Needle, for Insulin Scale: conforms to ISO8537:1991(E) The tests demonstrated that the device is as safe, as effective, and performs in a substantially equivalent manner to the predicate device. - 7. Clinical testing: Not required to establish equivalence. - Conclusion: The Comfort EZ™ insulin syringe complies with applicable standards for 8. insulin svringes, uses identical materials and manufacturing methods to the predicate, and has been tested for biocompatibility and performance characteristics. The syringe has the identical intended use to the predicate. We therefore conclude that the Comfort EZ™ syringe is substantially equivalent to the predicate device. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with its wings spread, and three wavy lines below it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular fashion around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 ## December 6, 2012 Simple Diagnostics, Incorporated C/O.Mr. Daniel Kamm, P.E. Kamm & Associates 8870 Ravello Court Naples, Florida 34114 Re: K122787 Trade/Device Name: Comfort EZTM Insulin Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF Dated: September 8, 2012 Received: September 11, 2012 Dear Mr. Kamm: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ #### Page 2 - Mr. Kamm Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Anthony Ls. m. ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ Anthony D. Watson, B.S., M.S., M.B.A. Director Director of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use K122787 510(k) Number (if known): Comfort EZ™ Insulin Syringe Device Name: Indications For Use: The Comfort EZ™ disposable sterile insulin syringes are intended for injection of U100 insulin only. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use X (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Digitally signed by Richard C. Chapman Date: 2012.12.12:06 08:30:06 -05'00' (Division Sign-Off) Division of Anestheslology, General Hospital Infection Control, Dental Devices K122787 510(k) Number: Page 1 of 1 {5}------------------------------------------------ # Indications for Use K122787 510(k) Number (if known): Comfort EZ™ Insulin Syringe Device Name: Indications For Use: The Comfort EZ™ disposable sterlle insulin syringes are intended for injection of U100 insulin only. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use X (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) .. Concurrence of CDRH, Office of Device Evaluation (ODE) | Gail G. Gantt | |--------------------------------------| | Digitally signed by Gail G. Gantt | | DN: c=US, o=U.S. Government, ou=HHS, | | ou=FDA, ou=People, cn=Gail G. Gantt, | | 0.9.2342.19200300.100.1.1=1300087388 | | Date: 2012.12.06 15:10:58 -05'00' | Page 1 of 1 (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices K122787 510(k) Number:
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...