K122373 · Neomed, Inc. · FMF · Sep 20, 2012 · General Hospital
Device Facts
Record ID
K122373
Device Name
NEOMED ORAL/ENTERAL SYRINGE
Applicant
Neomed, Inc.
Product Code
FMF · General Hospital
Decision Date
Sep 20, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The device is indicated for use as a dispenser, a measuring device and an oral fluid transfer device. It is used to inject fluids into the body via extension sets and feeding tubes in neonatal and small pediatric patients.
Device Story
NeoMed Oral/Enteral Syringe is a piston-style syringe designed for administration of enteral liquids. Device features non-luer-compatible tip to prevent accidental connection to intravenous lines; ensures dedicated enteral-only system connectivity with NeoMed extension sets and feeding tubes. Syringe includes translucent barrel for volume visualization and orange markings for identification. Operated by clinicians in clinical settings to measure and transfer fluids into neonatal/pediatric patients via feeding tubes. Prevents wrong-route administration by physical design incompatibility with IV systems.
Clinical Evidence
Bench testing only. Design verification testing confirmed performance equivalence to the predicate device. Biocompatibility verified per ISO 10993.
Technological Characteristics
Piston syringe; 0.5mL to 100mL capacity. Materials comply with ISO 10993-1. Non-luer-lock tip design; compatible with NeoMed enteral-only connectors. Complies with ISO 7886-1, ISO 7886-2, and ISO 80369-1. Sterilized to SAL 10^-6.
Indications for Use
Indicated for neonatal and small pediatric patients requiring administration of enteral liquids via extension sets and feeding tubes; used as a dispenser, measuring device, and oral fluid transfer device.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
K092908 — NEOMED STERILE SYRINGE, MODELS SD-S1EO, SD-S3EO, SD-S630, SD-S12EO, SD-S20EO, SD-S35EO, SD-S60EO · Neomed, Inc. · Mar 10, 2010
K143344 — NeoMed NeoConnect Enteral Syringes with ENFit Connector · Neomed, Inc. · Apr 9, 2015
K152857 — NeoMed NeoConnect Enteral Syringes with ENFit Connector and compatible NeoSecure Tip Caps · Neomed, Inc. · Dec 17, 2015
K110853 — VR MEDICAL ENTERAL ORAL SYRINGE · Vr Medical · May 13, 2011
K100099 — ORAL/ENTERAL SYRINGES · Philips Children'S Medical Ventures · Jun 14, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
# K122373
# 510K Summary of Safety & Effectiveness
July 30, 2012
NEOMED
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| APPLICANT | NeoMed, Inc. |
|---------------------------|------------------------------------------------|
| | FDA Owner/Operator #10022926 |
| | Establishment Registration #3006520777 |
| | 100 Londonderry Court |
| | Suite 112 |
| | Woodstock, GA 30188 |
| | Tony Lair, President |
| | Tel: 770-516-2225 |
| | Fax: 770-516-2448 |
| | Email: lair1@concentric.net |
| OFFICIAL<br>CORRESPONDENT | Melinda Harrison, RAC |
| | NeoMed, Inc. |
| | Director of Quality and Regulatory Affairs |
| | 100 Londonderry Court |
| | Suite 112 |
| | Woodstock, GA 30188 |
| | Tel: 770-516-2225 |
| | Fax: 770-516-2448 |
| | Email: mharrison@neomedinc.com |
| TRADE NAME: | NeoMed Oral / Enteral Syringe (0.5ml to 100ml) |
| CLASSIFICATION<br>NAME: | Piston Syringe |
| DEVICE<br>CLASSIFICATION | Class II per 21 CFR §880.5560 |
| PRODUCT CODE | FMF |
| PREDICATE<br>DEVICE NAME | NeoMed Sterile Syringe (K092908) |
### SUBSTANTIAL EQUIVALENCE:
.
The principle device of this 510(k) premarket notification is the result of device modification (increase of product family size range) to the predicate device K092908 which were conducted in accordance with Quality System Regulations, 21 CFR 820.
The NeoMed Oral / Enteral Syringes ranging from 0.5ml to 100ml are substantially equivalent to the predicate device, given that both the principle and predicate devices:
- have the same indications for use ●
- have the same method of operation to dispense, measure and transfer fluids
| Special 510(K)<br>NeoMed, Inc. | NeoMed Oral / Enteral Syringe<br>Confidential |
|--------------------------------|-----------------------------------------------|
|--------------------------------|-----------------------------------------------|
{1}------------------------------------------------
- meet the applicable requirements of ISO 7886-1 and ISO 7886-2 .
- have the same materials which comply with ISO10993-1 as applicable to the . intended use of the device
- are sterilized to Sterility Assurance Level (SAL) of 10-6 .
- demonstrate equivalent performance during design verification testing .
- meet the applicable requirements of ISO 80369-1 .
Design Verification testing as a result of the risk analysis has demonstrated that the NeoMed Oral / Enteral Syringes ranging from 0.5mL to 100mL are functionally equivalent to predicate NeoMed Sterile Syringes and the new range of sizes do not affect safety or effectiveness when compared to the predicate device.
### DESCRIPTION OF THE DEVICE:
The NeoMed Oral / Enteral Syringe is specifically designed for administration of enteral liquids to neonatal and small pediatric patients. The NeoMed Oral / Enteral Syringe is a standard piston syringe which is incompatible with luer-lock and intravenous devices . The NeoMed Oral / Enteral Syringes are compatible with NeoMed enteral-only connectors, the NeoMed extension sets and NeoMed feeding tubes to form a dedicated system that prevents wrong-route administration of non-IV fluids and other competitive enteral/feeding tube, extension set type products that do not utilize a luer-lock system. They possess translucent barrels to provide visualization of fluid contents and volume, and patented orange lettering/gradient markings which coordinate with NeoMed orange extension sets and feeding tubes. NeoMed Oral / Enteral Syringes are manufactured as a single piece molded barrel that does not rely on adapters to create an oral tip. The NeoMed Oral / Enteral Syringes possess design features that prevent connectivity to luer type devices and hence prevents the chance that a f eeding tube could be inf used into the IV line.
The NeoMed Oral / Enteral Syringe device consists of the following components:
- > Syringe Barrel
- > Syringe Plunger
- > Syringe Gasket
- > Syringe Tip
- > Syringe Cap
#### INDICATIONS FOR USE:
The device is indicated for use as a dispenser, a measuring device and an oral fluid transfer device. It is used to inject fluids into the body via extension sets and feeding tubes in neonatal and small pediatric patients.
#### PERFORMANCE DATA: .
The NeoMed Oral / Enteral Syringe materials that come in direct contact with the patient have a long history of use in syringe manufacturing and are biocompatible according to ISO 10993 test results. Design verification test results demonstrate that the NeoMed Oral / Enteral Syringe performs as intended and the extended siz e range from the predicate does not affect safety or effectiveness.
### CONCLUSION:
Based on the method of operation, indications for use, materials of construction and performance testing, it can be concluded that the NeoM ed Oral / Enteral Syringes ranging i from 0.5ml to 100ml are equivalent to the predicate NeoMed Sterile Syringe with respect to intended use and technological characteristics.
Special 510(K) NeoMed, Inc.
NeoMed Oral / Enteral Syringe Confidential
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, with its wings forming a protective shape over a group of people.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
SEP
2 0
2012
NeoMed, Incorporated Ms. Melinda Harrison Director of Quality and Regulatory Affairs 100 Londonderry Court, Suite 112 Woodstock, Georgia 30188
Re: K122373
Trade/Device Name: NeoMed Oral / Enteral Syringe (0.5ml to 100ml) Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF Dated: August 30, 2012 Received: September 4, 2012
Dear Ms. Harrison:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
# Page 2- Ms. Harrison
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
For Dt. Steele
Smether.
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
### Enclosure
{4}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known):
Device Name: NEOMED ORAL / ENTERAL SYRINGE
Indications For Use:
The device is indicated for use as a dispenser, a measuring device and an oral fluid transfer device. It is used to inject fluids into the body via extension sets and feeding tubes in neonatal and small pediatric patients.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mld Chape 9/12/12
(Division Sign-Off) (Division Sign-On)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K12237
Page 1 of
Special 510(K) NeoMed, Inc.
NeoMed Oral / Enteral Syringe Confidential
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.