K990239 · Cobe Cardiovascular, Inc. · DTN · May 25, 1999 · Cardiovascular
Device Facts
Record ID
K990239
Device Name
COBE VVR 4000 FILTERED HARDSHELL VENOUS RESERVOIR
Applicant
Cobe Cardiovascular, Inc.
Product Code
DTN · Cardiovascular
Decision Date
May 25, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4400
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The COBE® VVR 4000" Filtered Hardshell Venous Reservoir is intended to be used in adult surgical procedures requiring cardiopulmonary bypass for periods of up to six hours, and for postoperative chest drainage collection and autotransfusion.
Device Story
Sterile, single-use hardshell venous reservoir; capacity 4000 ml. Inputs: venous blood (vacuum or non-vacuum assisted) and cardiotomy suction blood. Operation: collects blood during cardiopulmonary bypass; defoamer removes gross air; integral cardiotomy filter removes particles. Post-bypass, device collects shed blood for autotransfusion (storage <6 hours). Used in OR by perfusionists/surgeons. Output: filtered blood returned to patient circulation. Benefits: blood volume management, air removal, and autotransfusion capability.
Clinical Evidence
Bench testing only. In-vitro testing performed to verify filtration efficiency, minimum operating volume, pressure relief valve operation, reservoir implosion resistance, and blood cell damage.
Technological Characteristics
Hardshell reservoir with bucket, defoamer, and cardiotomy filter. Materials: sterile, non-pyrogenic. Flow rates: up to 8 L/min (venous), 4 L/min (cardiotomy), 8 L/min (combined). Features: gasketed lid joint, vent port, pressure relief valve. Single-use, non-resterilizable.
Indications for Use
Indicated for adult patients undergoing surgical procedures requiring cardiopulmonary bypass (up to 6 hours) and for postoperative chest drainage collection and autotransfusion.
Regulatory Classification
Identification
A cardiopulmonary bypass blood reservoir is a device used in conjunction with short-term extracorporeal circulation devices to hold a reserve supply of blood in the bypass circulation.
Special Controls
*Classification.* Class II (special controls), except that a reservoir that contains a defoamer or filter is classified into the same class as the defoamer or filter. The device, when it is a cardiopulmonary bypass blood reservoir that does not contain defoamers or blood filters, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
K961836 — MAXIMA FORTE HARDSHELL VENOUS RESERVOIR · Medtronic Vascular · Nov 1, 1996
K081947 — GISH CAPVRF44 HARDSHELL VENOUS RESERVOIR WITH HA COATING · Gish Biomedical, Inc. · Feb 4, 2009
K964908 — VRF40 VENOUS RESERVOIR WITH INTEGRAL CARDIOTOMY FILTER · Sorin Biomedical, Inc. · Sep 3, 1997
K022281 — SAFE MAXI VENOUS/CARDIOTOMY RESERVIOR, MODEL 016045 · Polystan A/S · Aug 6, 2002
K151110 — Intersept Filtered Cardiotomy Reservoir · Medtronic, Inc. · May 27, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
MAY 25 1999
## 510(k) Summary
| SUBMITTER: | COBE Cardiovascular, Inc.<br>14401 W. 65th Way<br>Arvada, CO 80004 |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT PERSON: | Lynne Leonard<br>Phone: (303) 467-6586<br>Fax:<br>(303) 467-6429 |
| DATE PREPARED: | January 21, 1999 |
| DEVICE TRADE NAME: | COBE® VVR 4000™ Filtered Hardshell Venous Reservoir |
| COMMON/USUAL NAME: | Filtered Venous Reservoir with Integral Cardiotomy Filter |
| CLASSIFICATION NAMES: | Cardiopulmonary Bypass Blood Reservoir with Defoamer<br>and Cardiotomy Suction Line Blood Filter;<br>Autotransfusion Apparatus |
| PREDICATE DEVICES: | COBE® HVR™ 4000 Filtered Hardshell Venous Reservoir<br>Baxter/Bentley BMR-4500G Filtered Venous Reservoir<br>Baxter/Bentley HSR-4000 Filtered Venous Reservoir |
### DEVICE DESCRIPTION:
The COBE® VVR 4000" Filtered Hardshell Venous Reservoir, a sealed hardshell venous reservoir with defoamer and integral cardiotomy filter, is a sterile device with non-pyrogenic fluid pathways, for single use only, and is not to be resterilized by the user. It is designed for collection and return of postoperative chest drainage, and for storage and filtration of blood during adult surgical procedures requiring cardiopulmonary bypass for periods of up to six hours.
The maximum capacity of the COBE® VVR 4000™ Filtered Hardshell Venous Reservoir is approximately 4000 ml. The device operates at venous blood flow rates up to 8 liters/minute, cardiotomy flow rates up to 4 liters/minute, or at a combined maximum flow rate of 8 liters/minute.
During cardiopulmonary bypass, the device receives the patient's venous blood (under vacuum assisted or non-vacuum assisted venous return conditions) and also receives cardiotomy suction blood and filters it prior to returning it to the circulating blood volume. Entrained air is removed from both the venous blood and cardiotomy blood by the defoamer.
Following cardiopulmonary bypass, the device may be converted for use in postoperative chest drainage. Shed blood which has been stored in the reservoir less than six hours may be autotransfused for blood volume replacement.
The major components of the reservoir are the bucket, the defoamer, and the cardiotomy filter. The bucket serves both as a primary structural component of the device and as a transparent vessel to contain excess blood volume and to allow for the monitoring of changes in blood volume. The ports provide blood tubing connections between the reservoir and the patient, the defoamer serves to remove gross air from the incoming venous and cardiotomy suction blood, and the filter removes particles from cardiotomy suction blood.
{1}------------------------------------------------
#### INDICATIONS FOR USE
The COBE® VVR 4000" Filtered Hardshell Venous Reservoir is intended to be used in adult surgical procedures requiring cardiopulmonary bypass for periods of up to six hours, and for postoperative chest drainage collection and autotransfusion.
#### STATEMENT OF TECHNICAL CHARACTERISTICS COMPARISON
The COBE® VVR 4000" Filtered Hardshell Venous Reservoir is substantially equivalent to the currently marketed COBE® HVR™ 4000 Filtered Hardshell Venous Reservoir (K984456). The two devices are identical except that the COBE® VVR 4000™ Filtered Hardshell Venous Reservoir has a gasketed lid ioint, vent port, and pressure relief valve. Both the COBE® VVR 4000™ Filtered Hardshell Venous Reservoir and the COBE® HVR™ 4000 Filtered Hardshell Venous Reservoir are intended to be used in adult surgical procedures requiring cardiopulmonary bypass for periods of up to six hours. The COBE® VVR 4000™ Filtered Hardshell Venous Reservoir has the added indication of postoperative chest drainage collection and autotransfusion of the collected blood.
The COBE® VVR 4000™ Filtered Hardshell Venous Reservoir is also substantially equivalent to the Baxter/Bentley BMR-4500G Filtered Venous Reservoir. Both devices are sealed hardshell venous reservoirs with a defoamer and integral cardiotomy filter, designed for storage and filtration of blood during adult surgical procedures requiring cardiopulmonary bypass. The COBE® VVR 4000" Filtered Hardshell Venous Reservoir has the added indication of postoperative chest drainage collection and autotransfusion of the collected blood.
The COBE® VVR 4000™ Filtered Hardshell Venous Reservoir is also substantially equivalent to the Baxter/Bentley HSR-4000 Filtered Venous Reservoir. Both devices are sealed hardshell venous reservoirs with a defoamer and integral cardiotomy filter, designed for storage and filtration of blood during adult surgical procedures requiring cardiopulmonary bypass, and for postoperative chest drainage collection and autotransfusion of the collected blood.
Biocompatibility and in-vitro tests were performed to demonstrate that the COBE® VVR 4000" Filtered Hardshell Venous Reservoir is substantially equivalent to the COBE® HVR™ 4000 Filtered Hardshell Venous Reservoir, the Baxter/Bentley BMR-4500G Filtered Venous Reservoir, and the Baxter/Bentley HSR-4000 Filtered Venous Reservoir.
In-vitro testing consisted of:
Filtration Efficiency Minimum Operating Volume Pressure Relief Valve Operation Reservoir Implosion Resistance Blood Cell Damage
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 25 1999
Ms. Lynne Leonard Manager, Regulatory Submissions COBE Cardiovascular, Inc. 14401 W. 65th Way Arvada, CO 80004-3599
K990239 Re : COBE® VVR 4000™ Filtered Hardshell Venous Reservoir Regulatory Class: III (Three) Product Code: 74 DTP - DTN Dated: January 22, 1999 Received: January 25, 1999
Dear Ms. Leonard:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the marice is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Ms. Lynne Leonard
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callahon
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Premarket Notification
# Indications For Use
1
510(k) Number (If known): k910239
COBE® VVR 4000™ Filtered Hardshell Venous Reservoir Device Name:
Indications For Use:
The COBE® VVR 4000* Filtered Hardshell Venous Reservoir is intended to be used in adult surgical procedures requiring cardiopulmonary bypass for periods of up to six hours, and for postoperative chest drainage collection and autotransfusion.
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number
Prescription Use
(Per 21 CFR 801.109)
✓
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.