BARD QUANTUM SVR MODEL NUMBER H-6440VR

K981628 · C.R. Bard, Inc. · DTN · Aug 5, 1998 · Cardiovascular

Device Facts

Record IDK981628
Device NameBARD QUANTUM SVR MODEL NUMBER H-6440VR
ApplicantC.R. Bard, Inc.
Product CodeDTN · Cardiovascular
Decision DateAug 5, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4400
Device ClassClass 2

Intended Use

The Bard® Quantum™ SVR is designed to function in an extracorporeal circuit. The flexible venous reservoir receives dynamic blood from: (1) venous return (2) suctioned blood from the cardiotomy reservoir. The venous reservoir stores this blood prior to returning it to the circuit. The Bard® Quantum™ SVR is designed to function for up to 6 hours in an extracorporeal circuit.

Device Story

Flexible venous reservoir for extracorporeal circuits; receives venous return and suctioned blood from cardiotomy reservoir; stores blood before returning to circuit; used during cardiopulmonary bypass; operated by perfusionists/clinicians in surgical settings; provides temporary blood storage to maintain circuit flow; benefits patient by facilitating extracorporeal circulation.

Clinical Evidence

Bench testing only. Performance and integrity testing compared subject device to predicates; results demonstrated equivalence in minimum operating volume and capacity. Biocompatibility testing performed and passed.

Technological Characteristics

Flexible venous reservoir; materials identical to predicate devices; designed for extracorporeal circuit use; 6-hour operational limit.

Indications for Use

Indicated for use in extracorporeal systems during cardiopulmonary bypass to receive and store venous return and suctioned blood from the cardiotomy reservoir for up to 6 hours.

Regulatory Classification

Identification

A cardiopulmonary bypass blood reservoir is a device used in conjunction with short-term extracorporeal circulation devices to hold a reserve supply of blood in the bypass circulation.

Special Controls

*Classification.* Class II (special controls), except that a reservoir that contains a defoamer or filter is classified into the same class as the defoamer or filter. The device, when it is a cardiopulmonary bypass blood reservoir that does not contain defoamers or blood filters, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### AUG 6 1998 # 510(k) SUMMARY FOR THE BARD® QUANTUM™ CVR This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of SMDA 1990. - A. Submitter's Information | Name: | C.R. Bard, Inc., Bard Cardiopulmonary Division | |----------------------|------------------------------------------------| | Address: | 25 Computer Drive, Haverhill, MA 01832 | | Phone: | (978) 373-1000 extension 3371 | | Fax: | (978) 374-6200 | | Contact Person: | Sandra Perreand, Regulatory Affairs Manager | | Date of Preparation: | April 23, 1998 | - B. Device Name: | Trade Name: | Bard® Quantum™ SVR | |----------------------|----------------------------------------| | Common/Usual Name: | Cardiotomy and Venous Reservoir | | Classification Name: | Cardiopulmonary Bypass Blood Reservoir | ### C. Predicate Device Name(s): - 1) Avecor Affinity 321 Softshell Venous Reservoir Bag, K935717 - 2) Medtronic MVR 1600 Softshell Venous Reservoir Bag, K920774 ### D. Device Description/Indications for Use: The Bard® Quantum™ SVR is designed to function in an extracorporeal circuit. The flexible venous reservoir receives dynamic blood from: - (1) venous return - (2) suctioned blood from the cardiotomy reservoir The venous reservoir stores this blood prior to returning it to the circuit. #### E. Technological Characteristics Summary The "510(k) Substantial Equivalence Decision-Making Process (Detailed)" decision tree (FDA 92-415, Premarket Notification 510(k): Regulatory Requirements for Medical Devices, Page 51) was utilized to make a determination of substantial equivalence as follows: {1}------------------------------------------------ ### Does New Device Have Same Indication Statements? 1. The Bard® Quantum™ SVR (Bard H6440VR) has the same Yes. indications as the AVECOR Affinity Venous Reservoir Bag and the Medtronic MVR 1600 Softshell Venous Reservoir. Although the wording in the Indications for Use for all of these devices is not identical, all of these devices are indicated for use in extracorporeal systems during cardiopulmonary bypass. ### 2. Does New Device Have Same Technological Characteristics, e.g., Design, Materials, etc.? Yes. The Bard H6440VR has the same technological characteristics and the same materials as at least one of the predicate devices in each category as shown in Table IV-1 or the difference is insignificant as follows. As explained above, the Bard H6440VR has the same indications as the Avecor Affinity and the Medtronic MVR 1600 although the Bard H6440VR includes additional detail. The contraindications for the proposed device are standard phraseology and do not reflect a restriction in the use of the device. The minimum operating volume for the proposed device was statistically equivalent to both the Medtronic MVR 1600 Softshell Venous Reservoir and the Avecor Affinity VR 321 Softshell Reservoir for minimum volume to re-open the reservoir. The maximum volume of the device is simply a measure of capacity and is not a safety or efficacy concern for the device and is similar to those of the predicate devices. The materials of the device are the same as those materials used elsewhere in the device, and each material is used in one of the predicate devices in a similar location. The device was tested for biocompatibility as described in Section III and Appendix 1 and passed all testing. ### 3. Are the Descriptive Characteristics Precise Enough to Ensure Equivalence? No. Although the materials and technological characteristics are the same, differences between vendors and manufacturing techniques and the fact that neither of the predicate devices individually include all the same materials and technological characteristics as those of the proposed device necessitate bench testing to ensure equivalence. {2}------------------------------------------------ ### Are Performance Data Available to Assess Equivalence? 4. Yes. The Bard H6440VR, the Avecor Affinity, and the Medtronic MVR Series were compared for performance and integrity characteristics as described in Section IV.5. Results for this testing may be found in Appendix 2. In addition, as previously mentioned, the Bard H6440VR was subjected to biocompatibility testing (see summary in Section III and results in Appendix 1). ### 5. Performance Data Demonstrate Equivalence? Yes. The performance of the Bard H6440VR was equivalent to or better than those of the predicate devices or the differences were clinically insignificant as further described in Section IV.5. A summary of the performance and integrity testing may be found in Section IV.5. More in depth protocols and results may be found in Appendix 2. Following the summary of the performance and integrity testing is a summary of the biocompatibility testing. ## SUBSTANTIALLY EQUIVALENT DETERMINATION: The Bard H6440VR is substantially equivalent to the predicate devices, the Avecor Affinity and the Medtronic MVR 1600 Softshell Venous Reservoirs. {3}------------------------------------------------ Image /page/3/Picture/2 description: The image shows the seal of the Department of Health & Human Services USA. The seal features an eagle with its wings spread, symbolizing protection and strength. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. 5 1038 તાલુ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Sandra Perreand Bard Cardiopulmonary Division C.R. Bard, Inc. 25 Computer Drive Haverhill, MA 01832 Re : K981628 Bard® Quantum™ SVR (Softshell Venous Reservoir) Requlatory Class: II (Two) Product Code: DTN May 6, 1998 Dated: Received: May 7, 1998 Dear Ms. Perreand: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations. {4}------------------------------------------------ ### Page 2 - Ms. Sandra Perreand This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, # # Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radioloqical Health Enclosure {5}------------------------------------------------ #### 1-981628 INDICATIONS FOR USE D. Device Name: Bard® Quantum™ SVR (Softshell Venous Reservoir) ## Indications for Use: The Bard® Quantum™ SVR is designed to function in an extracorporeal circuit. The flexible venous reservoir receives dynamic blood from: - venous return (1) - suctioned blood from the cardiotomy reservoir (2) The venous reservoir stores this blood prior to returning it to the circuit. The Bard® Quantum™ SVR is designed to function for up to 6 hours in an extracorporeal circuit. Concurrence of CDRH, Office of Device Evaluation (ODE) | (Division Sign-Off) | | |-------------------------------------------------------------------|----------| | Division of Cardiovascular, Respiratory, and Neurological Devices | | | 510(k) Number | K9816281 | | Prescription Use (Per 21 CFR 801.109) | <b>X</b> | OR | Over-the-Counter Use ______ | |---------------------------------------|----------|----|-----------------------------| |---------------------------------------|----------|----|-----------------------------| Bard <sup>®</sup> Quantum <sup>TM</sup> SVR 510(k) Submission 8
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...