VRF40 VENOUS RESERVOIR WITH INTEGRAL CARDIOTOMY FILTER
Applicant
Sorin Biomedical, Inc.
Product Code
DTN · Cardiovascular
Decision Date
Sep 3, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4400
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The VRF40 is intended to allow for the collection and autotransfusion of shed mediastinal blood following cardiopulmonary bypass procedures or other procedures where shed blood is expected. The VRF40 Venous Reservoir with Integral Cardiotomy Filter is intended for use in CPB circuits and postoperative blood salvage procedures.
Device Story
Sterile, disposable, hardshell venous reservoir with integral cardiotomy filter; contains polyurethane defoamer sponge and outer gross filter screen. Used in clinical settings for collection and autotransfusion of shed mediastinal blood following cardiopulmonary bypass or other procedures where shed blood is expected. Device functions as a passive reservoir and filter system; collects blood, removes air via defoamer, and filters debris. Operated by perfusionists or clinical staff during or after surgery. Output is processed blood ready for autotransfusion; facilitates blood salvage, reducing need for donor blood products.
Clinical Evidence
Bench testing only. In vitro physical and functional tests included vacuum integrity, hemolysis, residual volume, and filter pressurization. Biocompatibility testing performed per Tripartite Guidelines for short-term blood contacting materials.
Technological Characteristics
Hardshell venous reservoir; polyurethane defoamer sponge; outer gross filter screen. Sterile, nonpyrogenic, disposable. Passive filtration and defoaming mechanism. No electronic components or software.
Indications for Use
Indicated for patients undergoing cardiopulmonary bypass (CPB) or procedures requiring postoperative blood salvage and autotransfusion of shed mediastinal blood.
Regulatory Classification
Identification
A cardiopulmonary bypass blood reservoir is a device used in conjunction with short-term extracorporeal circulation devices to hold a reserve supply of blood in the bypass circulation.
Special Controls
*Classification.* Class II (special controls), except that a reservoir that contains a defoamer or filter is classified into the same class as the defoamer or filter. The device, when it is a cardiopulmonary bypass blood reservoir that does not contain defoamers or blood filters, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
K964907 — VENOMIDICARD PEDIATRIC VENOUS RESERVOIR WITH INTEGRALCARDIOTOMY FILTER/MIDICARD PEDIATRIC CARDIOTOMY RESERVOIR WITH FILT · Sorin Biomedical, Inc. · Aug 21, 1997
K081947 — GISH CAPVRF44 HARDSHELL VENOUS RESERVOIR WITH HA COATING · Gish Biomedical, Inc. · Feb 4, 2009
K990239 — COBE VVR 4000 FILTERED HARDSHELL VENOUS RESERVOIR · Cobe Cardiovascular, Inc. · May 25, 1999
K030726 — GISH CAPVRF44 HARDSHELL VENOUS RESERVOIR WITH GBS COATING · Gish Biomedical, Inc. · May 30, 2003
K961836 — MAXIMA FORTE HARDSHELL VENOUS RESERVOIR · Medtronic Vascular · Nov 1, 1996
Submission Summary (Full Text)
{0}
Sorin Biomedical Inc.
VRF40 Venous Reservoir with Integral Cardiotomy Filter
510(k) Notification
K964908
Sept. 3, 1997
# 510(k) SUMMARY
## A. Name and Address of Submitter
Sorin Biomedical Inc.
17600 Gillette Avenue
Irvine, California 92614-5751
## B. Telephone and Fax Numbers of Submitter
Telephone: (714) 250-8320
Fax: (714) 250-1524
## C. Name of Contact Person
Richard J. DeRisio
Vice President
Quality & Regulatory Affairs
## D. Date Summary was Prepared
SORIN BIOMEDICAL INC. CONFIDENTIAL
10
{1}
Sorin Biomedical Inc.
VRF40 Venous Reservoir with Integral Cardiotomy Filter
510(k) Notification
# 510(k) SUMMARY (cont.)
## E. Device Name
*Trade or Proprietary Name:*
VRF 40 Venous Reservoir with Integral Cardiotomy Filter
*Common Name:*
Cardiotomy Reservoir, Venous Reservoir with Integral Cardiotomy Filter
*Classification Name:*
Cardiopulmonary Bypass Blood Reservoir
Cardiopulmonary Bypass Cardiotomy Suction Line Blood Filter
## F. Summary of Substantial Equivalence
The VRF40 Venous Reservoir with Integral Cardiotomy Filter is judged to be substantially equivalent in intended use, materials, design, and performance characteristics to the Sorin Hardshell Venous Reservoir (HSVRF). The HSVRF was cleared for chest drainage on September 30, 1987 (K872719). The VRF40 was designed for use in chest drainage and is marketed by DIDECO for that intended use outside the United States.
## G. Device Description
The VRF40 is a sterile, nonpyrogenic, disposable, hardshell venous reservoir with an integral cardiotomy filter. It contains a polyurethane defoamer sponge and outer gross filter screen.
## H. Intended Use
The VRF40 is intended to allow for the collection and autotransfusion of shed mediastinal blood following cardiopulmonary bypass procedures or other procedures where shed blood is expected.
## I. Summary of Comparison of Technological Characteristics
The technological characteristics of the VRF40 is similar to the Sorin Hardshell Venous Reservoir with respect to chest drainage autotransfusion. These devices have a similar number of cardiotomy inlet ports, have similar filter elements, and are similar in setup and use for chest drainage.
SORIN BIOMEDICAL INC. CONFIDENTIAL
108
26
{2}
Sorin Biomedical Inc.
VRF40 Venous Reservoir with Integral Cardiotomy Filter
510(k) Notification
# 510(k) SUMMARY (cont.)
## J. Summary of Nonclinical Tests
Substantial equivalence was based on a comparison of test results from the following *in vitro* physical and functional tests:
**Physical Tests**
Vacuum Integrity
**Functional Tests**
Hemolysis
Residual Volume
Filter Pressurization
All testing was conducted on nonaged, sterile devices. Test results show similar performance characteristics between the VRF40 and the Sorin Hardshell Venous Reservoir with respect to the expanded use of chest drainage autotransfusion.
**Biocompatibility**
Biocompatibility testing was previously performed on a finished, sterilized device in accordance with the Tripartite Guidelines for Biocompatibility testing for short term blood contacting materials. All test results were comparable to the negative control groups. Therefore, the materials used in the VRF40 are considered acceptable for their intended use.
## K. Conclusions
Based upon the above information, Sorin Biomedical Inc. concludes that the VRF40 Venous Reservoir with Integral Cardiotomy Filter is substantially equivalent to the Sorin Hardshell Venous Reservoir for use in chest drainage autotransfusion or any procedure where shed blood is expected.
SORIN BIOMEDICAL INC. CONFIDENTIAL
104
27
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20856
Mr. Paul Butorac
Manager, Quality Engineering Department
Sorin Biomedical, Inc.
17600 Gillette Avenue
P.O. Box 19503
Irvine, California 92713-9503
SEP - 3 1997
Re: K964908
VRF40 Venous Reservoir with Integral Cardiotomy Filter
Regulatory Class: II (Two)
Product Code: 74 DTN
Dated: June 10, 1997
Received: June 12, 1997
Dear Mr. Butorac:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}
Page 2 - Mr. Paul Butorac
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html."
Sincerely yours,

Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular, Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{5}
270.4408
Page 1 of 1
510(k) Number (if known): K964908
Device Name: VRF40 Venous Reservoir with Integral Cardiotomy Filter
Indications for Use: The VRF40 Venous Reservoir with Integral Cardiotomy Filter is intended for use in CPB circuits and postoperative blood salvage procedures.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number K 964908
Prescription Use ☑
(Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐
(Optional Format 1-2-96)
3
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.