← Product Code [DTN](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTN) · K070839 # MODIFICATION TO CAPIOX FLEXIBLE VENOUS RESERVOIR (K070839) _Terumo Cardiovascular Systems Corp. · DTN · Apr 24, 2007 · Cardiovascular · SESE_ **Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTN/K070839 ## Device Facts - **Applicant:** Terumo Cardiovascular Systems Corp. - **Product Code:** [DTN](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTN.md) - **Decision Date:** Apr 24, 2007 - **Decision:** SESE - **Submission Type:** Special - **Regulation:** 21 CFR 870.4400 - **Device Class:** Class 2 - **Review Panel:** Cardiovascular - **Attributes:** Therapeutic ## Intended Use The Capiox Flexible Venous Reservoir is intended for use as a blood collection and storage device during cardiopulmonary bypass procedures. The device is intended for use in conjunction with blood-gas oxygenators in the extra-corporeal circuit. The device may be used for procedures lasting up to 6 hours. ## Device Story Capiox Flexible Venous Reservoir functions as blood collection/storage device during cardiopulmonary bypass (CPB). Device positioned below patient level; blood enters via gravity through inlet ports at base of reservoir bag. Internal mesh filter facilitates air removal by disintegrating air boluses; upper air purge feature allows collection/aspiration of air. Blood exits via gravity through outlet port for pumping through remainder of CPB circuit. Used in clinical settings by perfusionists/surgical staff. Benefits patient by managing venous return and removing air from extracorporeal circuit during procedures up to 6 hours. ## Clinical Evidence No clinical data provided; bench testing only. ## Technological Characteristics Flexible venous reservoir bag with integrated mesh filter for air removal and air purge port. Gravity-fed operation. Materials and design identical to predicate. Class II device. ## Regulatory Identification A cardiopulmonary bypass blood reservoir is a device used in conjunction with short-term extracorporeal circulation devices to hold a reserve supply of blood in the bypass circulation. ## Special Controls *Classification.* Class II (special controls), except that a reservoir that contains a defoamer or filter is classified into the same class as the defoamer or filter. The device, when it is a cardiopulmonary bypass blood reservoir that does not contain defoamers or blood filters, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9. ## Predicate Devices - Capiox Flexible Venous Reservoir ([K040023](/device/K040023.md)) ## Related Devices - [K040023](/device/K040023.md) — CAPIOX FLEXIBLE VENOUS RESERVOIR · Terumo Cardiovascular Systems Corp. · Mar 1, 2004 - [K024065](/device/K024065.md) — SOFT VENOUS RESERVOIR (SVRXX) WITH GBS COATING · Gish Biomedical, Inc. · Mar 5, 2003 - [K223361](/device/K223361.md) — BMR 1900 PHISIO · Sorin Group Italia S.R.L. · Jun 10, 2023 - [K050195](/device/K050195.md) — V-BAG WITH GBS COATING · Gish Biomedical, Inc. · Mar 17, 2005 - [K090690](/device/K090690.md) — VENOUS SOFTBAG RESERVOIRS WITH SOFTLINE COATING · Maquet Cardiopulmonary, AG · Dec 23, 2009 ## Submission Summary (Full Text) {0}------------------------------------------------ K070839 # APR 2 4 2007 ### CAPIOX® Flexible Venous Reservoir Submitter Information: This premarket notification is submitted by: > Garry A. Courtney, MBA, RAC Terumo Cardiovascular Systems Regulatory Affairs Specialist Telephone: 1-800-283-7866, Ext. 7420 Date of Preparation: March 23, 2007 ### Device Names: | Proprietary Name: | Capiox® Flexible Venous Reservoir | |-------------------|-----------------------------------------------------------------| | Common Name: | Blood Reservoir | | Classification: | CPB Reservoirs (w/ filters) are classified as Class II devices. | #### Predicate Device: The Capiox Flexible Venous Reservoir is substantially equivalent in intended use, materials, design, performance, technology and principles of operation to the (unmodified/predicate) Capiox Flexible Venous Reservoir (K040023). #### Intended Use: The Capiox Flexible Venous Reservoir is intended for use as a blood collection and storage device during cardiopulmonary bypass procedures. The device is intended for use in conjunction with blood-gas oxygenators in the extra-corporeal circuit. The device may be used for procedures lasting up to 6 hours. #### Principles of Operation and Technology: The Capiox Flexible Venous Reservoir is used as a blood storage device during cardiopulmonary bypass procedures. The reservoir bag is to be positioned at a level that is below patient (blood supply) level. When the bag is properly positioned, the blood that is drawn from the patient enters the reservoir via gravity into the blood inlet ports that are positioned at the base edge of the reservoir bag. As the patient's blood enters the device, it passes through a mesh filter that facilitates the removal of air from the blood. This removal of air via the mesh filter is accomplished when an air bolus makes contact with the mesh and is subsequently disintegrated. Additionally, there is an air purge feature at the upper region of the bag that allows for the collection and subsequent aspiration of air from the bag. Blood exits the device via gravity through the blood outlet port and is subsequently pumped through the remainder of the cardiopulmonary bypass circuit. {1}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 APR 2 4 2007 Terumo Cardiovascular Systems Garry A. Courtney, Regulatory Manager 125 Blue Ball Rd. Elkton, MD 21921 Re: K070839 Trade/Device Name: Capiox Flexible Venous Reservoir Regulation Number: 21 CFR 870.4400 Regulation Name: Cardiopulmonary bypass blood reservoir Regulatory Class: Class II Product Code: DTN Dated: March 23, 2007 Received: March 27, 2007 Dear Mr. Courtney: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Page 2 - Mr. Courtney Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, P. Lachner Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ K070839 Image /page/3/Picture/1 description: The image shows the logo for Terumo. The logo consists of a stylized letter "T" inside of a circle, followed by the word "TERUMO" in bold, sans-serif font. A small registration mark is present to the right of the "O" in "TERUMO". Terumo Cardiovascular Systems Garry A. Courtney, MBA, RAC Office of Regulatory Management 125 Blue Ball Road Elkton, MD 21921 Telephone: (800) 283-7866 (x7420) ### SECTION 4 Indications for Use 510(k) Number (if known): Unknown at time of Submission 4670839 CAPIOX® Flexible Venous Reservoirs Device Name: ## Indications For Use: The CAPIOX® Flexible Venous Reservoir is intended for use as a blood collection and storage device during cardiopulmonary bypass procedures. The device is intended for use in conjunction with blood-gas oxygenators in the extra-corporeal circuit. The device may be used in procedures lasting up to 6 hours. Parry A. Courtney, MBA, RAC 3/23/2007 Cardiovascular Systems Prescription Use -X (Part 21 CFR 801 Subpart D) OR Over-The-Counter Use -(Part 21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Duna R. lo. Lunes (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number_K090839 --- **Source:** [https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTN/K070839](https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTN/K070839) **Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact). **Cite as:** FDA Device Explorer (Innolitics), https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/DTN/K070839
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...