Intersept Filtered Cardiotomy Reservoir

K151110 · Medtronic, Inc. · DTN · May 27, 2015 · Cardiovascular

Device Facts

Record IDK151110
Device NameIntersept Filtered Cardiotomy Reservoir
ApplicantMedtronic, Inc.
Product CodeDTN · Cardiovascular
Decision DateMay 27, 2015
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.4400
Device ClassClass 2

Intended Use

The Intersept filtered cardiotomy reservoir is indicated for use in the cardiopulmonary bypass circuit during surgery for: · an air/fluid separation chamber; - · a temporary storage reservoir for priming solution and blood; - · the filtration of particulates from bank blood and filtration of blood recovered from the field by suction: - · the addition of medications or other fluids. The Intersept filtered cardiotomy reservoir is also indicated for use after open heart surgery for: • the post-operative collection of autologous blood from the chest and the aseptic return of blood to the patient for volume replacement.

Device Story

Single-use cardiotomy reservoir; 2,600 ml capacity; 2 LPM max flow. Functions as air/fluid separation chamber, storage for priming solution/blood, and filtration unit for bank/field-recovered blood. Features eight access ports (suction, pre/post-defoamer, vent). Contains open-cell polyurethane defoamer and 20-micron microaggregate filter with polyester sleeve. Used in OR during cardiopulmonary bypass or post-operatively for autotransfusion. Operated by perfusionists or clinical staff. Filters particulates; allows medication/fluid addition; enables aseptic return of autologous blood to patient for volume replacement. Benefits patient by reducing particulate load and facilitating blood salvage.

Clinical Evidence

Bench testing only. Testing included filtration efficiency, blood trauma, and biocompatibility assessments. All tests passed.

Technological Characteristics

Single-use reservoir; 2,600 ml capacity; 2 LPM max flow. Materials: Polypropylene frame, open-cell polyurethane defoamer, 20-micron microaggregate filter, polyester sleeve. Eight luer/0.6 cm ID access ports. Sterile, non-pyrogenic fluid path. Mechanical filtration and air/fluid separation principle.

Indications for Use

Indicated for patients undergoing cardiopulmonary bypass surgery or post-operative autologous blood collection. Contraindicated in cases of lung air leaks, chest wall perforations, infection, malignancy, gross contamination (bone cement, etc.), aspiration of non-parenteral agents, open chest with vacuum, post-protamine administration, return to surgery, or use of unregulated vented chest tubes.

Regulatory Classification

Identification

A cardiopulmonary bypass blood reservoir is a device used in conjunction with short-term extracorporeal circulation devices to hold a reserve supply of blood in the bypass circulation.

Special Controls

*Classification.* Class II (special controls), except that a reservoir that contains a defoamer or filter is classified into the same class as the defoamer or filter. The device, when it is a cardiopulmonary bypass blood reservoir that does not contain defoamers or blood filters, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a circular logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a symbol. The symbol consists of three stylized human profiles facing right, stacked on top of each other, with a ribbon-like element flowing beneath them. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 27, 2015 Medtronic, Inc. Chelsea Pioske Regulatory Affairs Specialist 7611 Northland Drive Minneapolis, Minnesota 55428 Re: K151110 Trade/Device Name: Intersept Filtered Cardiotomy Reservoir Regulation Number: 21 CFR 870.4400 Regulation Name: Cardiopulmonary Bypass Blood Reservoir Regulatory Class: Class II Product Code: DTN. DTP Dated: April 24, 2015 Received: April 27, 2015 Dear Chelsea Pioske: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR {1}------------------------------------------------ Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, M.A. Hilleman for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K151110 Device Name Intersept™ Filtered Cardiotomy Reservoir Indications for Use (Describe) The Intersept filtered cardiotomy reservoir is indicated for use in the cardiopulmonary bypass circuit during surgery for: · an air/fluid separation chamber; - · a temporary storage reservoir for priming solution and blood; - · the filtration of particulates from bank blood and filtration of blood recovered from the field by suction: - · the addition of medications or other fluids. The Intersept filtered cardiotomy reservoir is also indicated for use after open heart surgery for: • the post-operative collection of autologous blood from the chest and the aseptic return of blood to the patient for volume replacement. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary | Date Prepared: | April 23, 2015 | |-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Applicant: | Medtronic, Inc.<br>Medtronic Perfusion Systems<br>7611 Northland Drive<br>Minneapolis, MN 55428<br>Establishment Registration Number: 2184009 | | Contact Person: | Chelsea Pioske<br>Regulatory Affairs Specialist<br>Phone: (763) 514-9838<br>Fax: (763) 367-8361<br>E-mail: chelsea.pioske@medtronic.com | | Alternate Contact: | Susan Fidler<br>Senior Regulatory Affairs Manager<br>Medtronic Perfusion Systems<br>Phone: 763.514.9839<br>Fax: 763.367.8360<br>Email: susan.c.fidler@medtronic.com | | Trade Name: | Intersept™ Filtered Cardiotomy Reservoir | | Classification Name: | Reservoir, Blood, Cardiopulmonary Bypass; Defoamer,<br>Cardiopulmonary Bypass | | Regulation Number: | 21 CFR 870.4400, 21 CFR 870.4230 | | Product Classification: | Class II | | Product Code: | DTN, DTP | Name of Predicate Device: Intersept® Cardiotomy Reservoir with Filter #### Device Description The Intersept™ Filtered Cardiotomy Reservoir is a single-use, device with a sterile, nonpyrogenic fluid path. The maximum capacity of each reservoir is 2,600 ml. The maximum recommended flow rate is 2 liters per minute (LPM). Each reservoir has eight luer or 1/4" (0.6 cm) ID access ports; - four suction access ports . - two pre-defoamer and two post-defoamer access ports. One of the post-defoamer access . ports is intended for use as a vent. {4}------------------------------------------------ Each Intersept™ Filtered Cardiotomy Reservoir contains an open cell polyurethane defoamer with a 20 micron microaggregate filter covered with a polyester sleeve. ## Indications for Use The Intersept filtered cardiotomy reservoir is indicated for use in the cardiopulmonary bypass circuit during surgery for: - an air/fluid separation chamber; . - a temporary storage reservoir for priming solution and blood; - the filtration of particulates from bank blood and the storage and filtration of blood ● recovered from the field by suction; - . the addition of medications or other fluids. The Intersept filtered cardiotomy reservoir is also indicated for use after open heart surgery for: - the post-operative collection of autologous blood from the chest and the aseptic return of ● blood to the patient for volume replacement. ## Contraindications This product used for any other purposes than for the indicated use is the responsibility of the user. The cardiotomy reservoir is contraindicated for use in post-operative chest drainage/autotransfusion procedures when: - . there is an air leak in the lung or gross perforations to the chest wall exist; - pericardial, mediastinal, pulmonary or systemic infection or malignancy is present; - gross contamination (e.g., bone cement, lymphatic failure, perforated intestine) is present ● or suspected: - aspirating blood from a site containing topical hemostatic agents, bactericidal wound irritants or antibiotics not intended for parenteral administration; - the chest is open and vacuum is applied; - protamine has been administered prior to the reservoir being removed from the bypass circuit; - the patient is returned to surgery for any reason; - vented chest tubes not incorporating vent flow regulation such as stopcock are used. # Comparison to Predicate Devices A comparison of the Intersept™ Filtered Cardiotomy Reservoir to the predicate device (the Intersept® Cardiotomy Reservoir with Filter) indicates the following similarities: - Intended Use: The intended use is the same as predicate device. ● - Design: The overall design is the same as the predicate. - Materials: The material types used are the same as the predicate. - Principles of Operation and Technology: The principles of operation are the same as the predicate device. - . Performance: The performance is substantially equivalent to the predicate and/or reference device. {5}------------------------------------------------ #### Summary of Testing Testing has demonstrated that the Intersept™ Filtered Cardiotomy Reservoir is substantially equivalent to the predicate. The following tests were conducted: | Component | Base Material Changes | Verification/Validation | Results | |-----------|-------------------------|-------------------------|---------| | Frame | Current: Polypropylene | Filtration Efficiency | Pass | | Frame | | Blood Trauma | Pass | | Frame | Proposed: Polypropylene | Biocompatibility | Pass | ## Conclusion The data included in this submission is sufficient to demonstrate substantial equivalence of the Intersept™ Filtered Cardiotomy in terms of safety and effectiveness as compared to the predicate device, the Intersept® Cardiotomy Reservoir with Filter.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...