MEDOS HILITE INFANT RESERVOIR, MODEL MVC 0730

K091062 · Gish Biomedical, Inc. · DTN · Aug 4, 2009 · Cardiovascular

Device Facts

Record IDK091062
Device NameMEDOS HILITE INFANT RESERVOIR, MODEL MVC 0730
ApplicantGish Biomedical, Inc.
Product CodeDTN · Cardiovascular
Decision DateAug 4, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4400
Device ClassClass 2
AttributesTherapeutic, Pediatric

Intended Use

The MEDOS HILITE MVC 0730 Hardshell Venous Reservoir is indicated for use during cardiopulmonary bypass surgery as a storage reservoir for gravity and augmented venous return blood and to filter and defoam intrathoracic suctioned blood prior to its return to the extracorporeal circuit. The MVC 0730 is a neonate, infant reservoir intended for use in procedures at a maximum flow rate of 1.2 liters per minute for periods up to six hours (6.0) hours.

Device Story

Sterile, single-use, disposable hardshell venous reservoir; used in cardiopulmonary bypass (CPB) circuits. Collects, stores, and filters blood. Input: venous drainage (via 1/4" inlet) and suctioned blood. Internal components: defoamer sponge and 30-micron filter. Operation: venous blood enters bottom; suctioned blood passes through defoamer/filter cartridge. Output: filtered blood returned to extracorporeal circuit. Used in OR by perfusionists/surgeons. Benefits: provides venous blood storage and blood processing during neonatal/infant cardiac procedures.

Clinical Evidence

No clinical data. Bench testing only; performance and safety testing conducted to verify functional requirements and compliance with industry standards.

Technological Characteristics

Polycarbonate hardshell reservoir; internal defoamer sponge and 30-micron filter. Max venous flow 1.2 L/min; max cardiotomy flow 0.9 L/min. Sterile, non-pyrogenic, single-use. Mechanical filtration/defoaming principle.

Indications for Use

Indicated for neonate and infant patients undergoing cardiopulmonary bypass surgery. Used for storage of gravity/augmented venous return and filtration/defoaming of intrathoracic suctioned blood. Max flow rate 1.2 L/min; max duration 6 hours.

Regulatory Classification

Identification

A cardiopulmonary bypass blood reservoir is a device used in conjunction with short-term extracorporeal circulation devices to hold a reserve supply of blood in the bypass circulation.

Special Controls

*Classification.* Class II (special controls), except that a reservoir that contains a defoamer or filter is classified into the same class as the defoamer or filter. The device, when it is a cardiopulmonary bypass blood reservoir that does not contain defoamers or blood filters, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K09/062 ## MEDOS Medizintechnik AG Traditional 510(k) Section 5, 510(k) Summary # Summary This summary of 510(k)-safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR § 807.92. AUG 0 4 2009 ### Company making the submission: 1. | Name: | GISH Biomedical, Inc. | |------------|---------------------------------------| | | A member of the MEDOS group | | Address: | 22942 Arroyo Vista | | | Rancho Santa Margarita, CA 92688-2600 | | Telephone: | 949-635-6200 voice | | | 949-635-6299 fax | | Contact: | janetp@gishbiomedical.com | | | Janet Peets | | | Regulatory & Clinical Affairs | ### 2. Device: | Proprietary Name: | MEDOS HILITE MVC 0730 Hardshell Venous Reservoir | |----------------------|--------------------------------------------------| | Common Name: | Cardiopulmonary Blood Reservoir | | Classification Name: | Extracorpeal Circuit Blood Defoamer | | | Cardiopulmonary Bypass Blood Reservoir | #### 3. Predicate Devices: CAPIOX® RX05 Hollow Fiber Oxygenator with Hardshell Reservoir, K022115, manufactured by Terumo Cardiovascular Systems Corporation. #### Classifications Names & Citations: 4. 21 CFR 870.4230, 21 CFR 870.4400, Extracorpeal circuit blood defoamers, Cardiopulmonary bypass blood reservoir. Cardiopulmonary Bypass, Class II, DTN, Cardiovascular. #### 5. Description: The MEDOS HILITE MVC 0730 Hardshell Venous Reservoirs are sterile, non-pyrogenic, single use, disposable, device designed for collection, storage and filtration of blood during cardiopulmonary bypass. The MEDOS HILITE MVC 0730 has a clear polycarbonate shell and an internal defoamer filter cartridge. Venous drainage enters the 1/4" venous inlet at the bottom section of the reservoir. Venous suctioned blood enters the top section of the defoamer/filter cartridge and passed through a defoamer sponge and 30 micron filter. The maximum venous flow rate is 1.2 lpm. The maximum cardiotomy flow rate is 0.9 lpm. {1}------------------------------------------------ ### 6. Indications for use: The MEDOS HILITE MVC 0730 Hardshell Venous Reservoir is indicated for use during cardiopulmonary bypass surgery as a storage reservoir for gravity and augmented venous return blood and to filter and defoam intrathoracic suctioned blood prior to its return to the extracorporeal circuit. The MVC 0730 is a neonate, infant reservoir intended for use in procedures at a maximum flow rate of 1.2 liters per minute for periods up to six hours (6.0) hours. #### 7. Contra-indications: No contra-indications have been noted. # 8. Comparison: The MEDOS HILITE MVC 0730 Hardshell Venous Reservoir has the same device characteristics as the predicate device. # 9. Test Data: The MEDOS HILITE MVC 0730 Hardshell Venous Reservoir has been subjected to extensive safety, performance, and validations prior to release. Final testing for the system includes various performance tests designed to ensure that the device meets all of its functional requirements and performance specifications. Safety tests have further been performed to ensure the device complies with applicable industry and safety standards. # 10. Literature Review: A review of literature pertaining to the safety and effectiveness has been conducted. Appropriate safeguards have been incorporated in the design of MEDOS HILITE MVC 0730 Hardshell Venous Reservoir. ### 11. Conclusions: Based upon the testing and comparison to the predicate device, the MEDOS HILITE MVC 0730 Hardshell Venous Reservoir has the same intended use, with similar technological characteristics. MEDOS Medizintechnik AG therefore posits that its device is equivalent in safety and effectiveness to predicate devices. {2}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract image of an eagle. ### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-0609 Silver Spring, MD 20993-0002 # AUG 0 4 2009 GISH Biomedical, Inc. c/o Ms. Janet Peets Regulatory and Clinical Affairs 22942 Arroyo Vista Rancho Santa Margarita, CA 92688-2600 Lansing. MI 48917 K091062 Re: > Medos Hilite MVC 0730 Reservoir Regulation Number: 21 CFR 870.4400 Regulation Name: Cardiopulmonary Bypass Blood Reservoir Regulatory Class: Class II (two) Product Code: DTN Dated: June 24, 2009 Received: June 25, 2009 Dear Ms. Peets: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ # Page 2 - Ms. Janet Peets Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/odrh/industry/support/index.html. Sincerely yours, ina R. bulmer. Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## MEDOS Medizintechnik AG 510(k) Number KL)91062 ## Device Name: MEDOS HILITE MVC 0730 Reservoir # Indications for use: The MEDOS HILITE MVC 0730 Hardshell Venous Reservoir is indicated for use during cardiopulmonary bypass surgery as a storage reservoir for gravity and augmented venous return blood and to filter and defoam intrathoracic suctioned blood prior to its return to the extracorporeal circuit. The MVC 0730 is a neonate, infant reservoir intended for use in procedures at a maximum flow rate of 1.2 liters per minute for periods up to six hours (6.0) hours. ### Prescription Device: Federal Law (US) restricts this device to sale by or on the order of a physician. | Prescription Use : | Yes | |-----------------------|-----| | OR | | | Over-The-Counter Use: | No | # PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Dwna R. La Lune ivision Sian-Off Division of Cardiovascular Devices 510(k) Number Ko 9 10 6 2 # Section 4 510(k) Submission Page # \$\xi\$
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...