FLEXCATH ADVANCE STEERABLE SHEATH (12 FRENCH)

K123591 · Medtronic Cryocath, LP · DRA · Dec 27, 2012 · Cardiovascular

Device Facts

Record IDK123591
Device NameFLEXCATH ADVANCE STEERABLE SHEATH (12 FRENCH)
ApplicantMedtronic Cryocath, LP
Product CodeDRA · Cardiovascular
Decision DateDec 27, 2012
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1280
Device ClassClass 2

Intended Use

The FlexCath Advance Steerable Sheath is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart. The sheath deflection facilitates catheter positioning.

Device Story

Percutaneous introducer sheath and dilator; used for catheter access to heart chambers. Features handle with rotating actuator for unidirectional deflection; side-port with stopcock for infusion, aspiration, and pressure monitoring. Operated by physicians in clinical settings. Modifications from predicate include updated stopcock, shaft pullwire assembly, deflection stopper, and barrel/slider assembly. Device facilitates navigation of catheters through vasculature; provides stable access for cardiac procedures. Benefits include improved maneuverability and positioning of diagnostic or therapeutic catheters.

Clinical Evidence

Bench testing only. Verification and validation included mechanical testing (leak/flush, deflection, tensile strength, kink resistance), shelf-life (12-month accelerated aging), and biocompatibility (ISO 10993-1/USP Class VI). No clinical data presented.

Technological Characteristics

12 French steerable sheath and dilator. Features: rotating actuator for unidirectional deflection, hemostasis valve, side-port with stopcock. Materials: biocompatible per ISO 10993. Sterilization: standard process. Connectivity: none (mechanical device).

Indications for Use

Indicated for percutaneous catheter introduction into the vasculature and heart chambers to facilitate catheter positioning.

Regulatory Classification

Identification

A steerable catheter is a catheter used for diagnostic and monitoring purposes whose movements are directed by a steering control unit.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### 5.0 510(k) Summary - K123591 ### 5.1 General Information: # DEC 2 7 2012 | Submitter's Name and Address: | Medtronic CryoCath LP<br>16771 Chemin Ste-Marie<br>Kirkland, Quebec<br>H9H 5H3, CANADA | |-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Official Correspondent: | Vanessa Ware<br>Principal Regulatory Affairs Specialist<br>Telephone: 763.526.2743<br>Fax: 763.367.9903<br>Medtronic CryoCath LP<br>Coral Sea Street<br>Mounds View, MN 55112 | | Date of Summary: | November 20, 2012-Revised December 7, 2012 | | Trade Name: | FlexCath Advance™ Steerable Sheath & Dilator<br>Steerable Catheter | | Common Name: | Steerable Sheath and Dilator | | Classification Name: | Steerable Catheter | | Classification: | Class II, 21 CFR § 870.1280 | | Product Code: | DRA | | Predicate Device(s): | FlexCath® Steerable Sheath & Dilator, Model 3FC12<br>(K102176) | #### 5.2 Device Description The FlexCath Advance Steerable Sheath is a sterile, single use percutaneous introducer fitted with a valve to allow for introduction, withdrawal and swapping of catheters and wires while preventing air ingress and minimizing blood loss. A side-port with stopcock is integrated to allow continuous drip infusion, injection through the center lumen, flushing, aspiration, blood sampling and pressure monitoring. The FlexCath can be deflected to provide additional maneuverability to catheters that are advance through the sheath and into the right or left chamber of the heart. The FlexCath Advance Steerable Sheath is comprised of two (2) main sections: the shaft and the handle. A dilator is included with each sheath. This application addresses modifications to the 12 French (12F) FlexCath Steerable Sheath which resulted in the creation of the 12F FlexCath Advance Steerable Sheath (4FC12). {1}------------------------------------------------ Modifications were made to the stopcock, shaft (pullwire and ring assembly), deflection stopper, and barrel and slider assembly. In addition, an updated Technical Manual (Instructions for Use) and updated packaging labeling was developed for the FlexCath Advance Steerable Sheath. ### Intended Use 5.3 . The FlexCath Advance Steerable Sheath is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart. The sheath deflection facilitates catheter positioning. ### Technological Characteristics Comparison of New Device to Predicate Device 5.3.1 The FlexCath Advance Steerable Sheath and Dilator, with the modified valve assembly (stopcock) and shaft pullwire assembly (both transparent to the user), modified deflection stopper (increased deflection with same reach) and modified barrel and threaded slider, has the following similarities to the predicate device previously cleared under KI02176: - Same Indication For Use . - Same fundamental scientific technology . - . Same unidirectional deflection - Same basic sheath design (dimensions, bullnose tip shape) . - Same user interface with handle (rotating actuator for deflection) . - Same materials . - Same sterilization process . #### 5.4 Summary of Performance Tests Verification and validation activities were performed to ensure that the FlexCath Advance Steerable Sheath and Dilator fulfilled system requirements and to ensure that the product design conforms to the user needs and intended uses. Performance testing included mechanical testing, shelf-life, and biocompatibility in accordance with applicable industry standards and FDA guidance. Testing included: - Sheath Valve Assembly (Stopcock) Leak and Flush Testing . - Sheath Shaft Pull Wire Assembly Deflection Testing, System Device Compatibility, . Interactions, Tensile Strength - Stopper Catheter Sheath Kink and Deflection . - Barrel and Threaded Slider Maintain Deflection, Sheath Shape . ## Test not repeated and leveraged from predicate because of the same design attributes include: - Bond pull forces . - Packaging integrity . - Dilator placement, compatibility, visibility under fluoroscopy . - Sheath deflection with a rotating actuator . - Hemostasis valve aspiration flush test and seal with fluid and under pressure . - Sheath shaft visibility under fluoroscopy . - Flushing and aspiration while loaded with catheter . {2}------------------------------------------------ ### Shelf Life Testing: Shelf life design verification testing was completed to ensure that device specifications of the product were met after a 12-month accelerated aging study after design modifications to the sheath valve assembly, pullwire assembly, deflection stopper, and barrel and threaded slider. All testing passed. Testing showed that the function and mechanical characteristics of the modified device are not compromised after a 12-month accelerated aging study. ### Biocompatibility Testing: The stopcock (Supplier: Elcam Medical) was assessed for biocompatibility with the guidance from ISO 10993-1 standard or USP Class VI by the supplier. All materials of the stopcock were tested and/or assessed to determine equivalency to a material that was already tested and determined to be biocompatible. The devices or materials were subjected to the following tests: - . Cytotoxicity - . Sensitization - Skin Irritation or Intracutaneous Reactivity . - Acute systemic toxicity . - Hemolysis . - Optional Tests: ● - Determination of physiochemical Attributes (in purified water) and/or o - Determination of physiochemical Attributes (in isopropyl alcohol) and/or o - Pyrogen Study o Leveraged biocompatibility of the predicate device included: - · Kligman Maximization Test per ISO 10993-10 (2002) - · Rabbit Pyrogen Test Material Meditated per ISO 10993-11 (2006) - · Intracutaneous Test per ISO 10993-10 (2002) - · Systemic Injection Test per ISO 10993-11 (2006) - · Complement Activation Assay per ISO 10993-4 (2002) - · UPTT per ISO 10993-4 (2002) - · Thrombogenicity In vivo canine per ISO 10993-4 (2002) All test successfully passed. The results if the bench testing demonstrate that the FlexCath Advance Steerable Sheath and Dilator meet the established specifications necessary for consistent performance for its intended use and is substantially equivalent to the predicate (FlexCath Steerable Sheath and Dilator). #### Discussion of Testing Supporting Substantial Equivalence ર્સ્ટ Non-clinical performance testing (bench testing) was performed to support substantial equivalence between the FlexCath Advance Steerable Sheath (4FC12) and the predicate FlexCath Steerable Sheath and Dilator (3FC12) previous cleared under K102176. All relevant key performance attributes were tested on the FlexCath Advance against performance {3}------------------------------------------------ requirements, and some test were leveraged from the predicate device and not repeated. The FlexCath Advance does not provide a new therapy and the intended use/indication for use remains equivalent to that of approved FlexCath, Model 3FC12. It was determined that modifications of features to the FlexCath Steerable Sheath had no impact on safety and performance and all testing demonstrated that acceptance criteria were met. In addition, the testing demonstrates that the device does not raise new questions of safety or effectiveness when compared to the predicate. ### રે. ર Conclusion Design verification testing was performed to verify that the performance of FlexCath Advance remains substantially equivalent to the predicate device. All test results demonstrate that the properties and performance of the device are suitable for its intended use. There are no differences between devices or in testing which would raise new issues of safety or effectiveness. {4}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the seal is a stylized symbol that resembles a human figure embracing or supporting another, with flowing lines suggesting movement or care. ### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 ## DEC 2 7 2012 Medtronic Cryocath Lp c/o Ms. Vanessa Ware 8200 Coral Sea Street NE MS MVS46 Mounds view, MN 55112 Re: K123591 Trade/Device Name: Flexcath advance steerable sheath (12 french) Regulation Number: 21 CFR 870.1280 Regulation Name: Steerable catheter Regulatory Class: Class II Product Code: DRA Dated: December 21, 2012 Received: December 26, 2012 Dear Ms. Ware: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {5}------------------------------------------------ Page 2 - Ms. Vanessa Ware comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Matthew G. Hillebrenner for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ### Statement of Indications for Use 1.0 510(k) Number (if known): K 123591 Device Name: FlexCath Advance ™ Steerable Sheath and Dilator ### Indications for Use: The FlexCath Steerable Sheath is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart. The Sheath deflection facilitates catheter positioning. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Cmgthlle Cardiovascular Devices 510(k) Numbe
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...