AcQGuide® VUE Steerable Sheath

K221044 · Acutus Medical, Inc. · DRA · May 5, 2022 · Cardiovascular

Device Facts

Record IDK221044
Device NameAcQGuide® VUE Steerable Sheath
ApplicantAcutus Medical, Inc.
Product CodeDRA · Cardiovascular
Decision DateMay 5, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1280
Device ClassClass 2
Attributes3rd-Party Reviewed

Intended Use

The AcQGuide® VUE Steerable Sheath with electrodes is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart. The AcQGuide® VUE deflection facilitates catheter positioning. The electrodes help facilitate visualization of the sheath when used with a compatible localization system, such as the AcQMap System.

Device Story

AcQGuide® VUE Steerable Sheath is a single-use, percutaneous catheter introducer; provides maneuverability for interventional catheters in heart chambers. Features composite single-lumen shaft, ergonomic handle for torque/deflection, hemostasis valve, and side port for aspiration/infusion. Includes distal platinum/iridium electrodes connecting to AcQMap High Resolution Imaging and Mapping System via cable. Used in clinical settings by physicians; facilitates sheath visualization on mapping system during procedures. Enhances catheter positioning accuracy; aids navigation via localization system integration. Optional valve bypass tool included for circular catheter insertion.

Clinical Evidence

No clinical testing submitted. Substantial equivalence supported by bench testing, including dimensional inspection, leak testing, functional/compatibility testing, mechanical testing, handle torque testing, and physician simulated use in an animal model.

Technological Characteristics

Single-lumen shaft (PTFE liner, Pebax polymers), ergonomic handle, hemostasis valve (silicone). Distal platinum/iridium electrodes. 12.4 Fr ID, 15.2 Fr OD, 70 cm usable length. 180-degree deflection. EtO sterilized. CF defibrillation-proof applied part. Connectivity via 2mm pins to AcQMap System.

Indications for Use

Indicated for percutaneous catheter introduction into the vasculature and heart chambers in patients requiring catheter-based cardiac procedures.

Regulatory Classification

Identification

A steerable catheter is a catheter used for diagnostic and monitoring purposes whose movements are directed by a steering control unit.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. May 5, 2022 Acutus Medical, Inc. % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114 Re: K221044 Trade/Device Name: AcQGuide® VUE Steerable Sheath Regulation Number: 21 CFR 870.1280 Regulation Name: Steerable catheter Regulatory Class: Class II Product Code: DRA Dated: April 7, 2022 Received: April 8, 2022 Dear Prithul Bom: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K221044 Device Name AcQGuide® VUE Steerable Sheath Indications for Use (Describe) The AcQGuide® VUE Steerable Sheath with electrodes is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart. The AcQGuide® VUE deflection facilitates catheter positioning. The electrodes help facilitate visualization of the sheath when used with a compatible localization system, such as the AcQMap System. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Acutus Medical. The logo is composed of the word "ACUTUS" in a sans-serif font, with the letters in gray. Below the word "ACUTUS" is the word "MEDICAL" in a smaller, sans-serif font, with the letters in orange. The logo is simple and modern, and the use of gray and orange gives it a professional look. # 510(k) SUMMARY 510(k) Number: K221044 Date Prepared: May 4, 2022 # SUBMITTER INFORMATION [807.92(a)(1)] Manufacturer: Acutus Medical, Inc. 2210 Faraday Ave., Suite 100 Carlsbad, CA 92008, U.S.A Phone: +1-442-232-6080 Fax: +1-442-232-6081 # Contact Person: Sarah Clay Regulatory Affairs Specialist Acutus Medical, Inc. 2210 Faraday Ave., Suite 100 Carlsbad, CA 92008, U.S.A Fax: +1-442-232-6081 E-mail: Sarah.Clay@acutus.com # DEVICE INFORMATION [807.92(a)(2)] | Trade Name | AcQGuide® VUE Steerable Sheath | |----------------------------|--------------------------------| | Common Name | Steerable Sheath | | Classification Name | Catheter, Steerable | | Regulation | 21 CFR § 870.1280 | | Product Code | DRA | | Regulatory Classification: | Class II | | Device Panel: | Cardiovascular | # PREDICATE DEVICE [807.92(a)(3)] The predicate device to support substantial equivalence of the AcQGuide® VUE Steerable Sheath, Model 900201 is the Acutus Medical, Inc. AcQGuide® MAX Steerable Sheath, Model 900200, cleared under 510(k) K211100. This 510(k) is also citing the CARTO VIZIGO Bi-Directional Guiding Sheath cleared under K170997 as a reference device to support substantial equivalence for device containing electrodes. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Acutus Medical. The word "ACUTUS" is in large, gray, sans-serif font, with a small orange dot above the left side of the "A". Below "ACUTUS", the word "MEDICAL" is written in a smaller, orange, sans-serif font. The logo is simple and modern. | Predicate Device | Manufacturer | FDA 510(k) | Product Code | |--------------------------------|----------------------|------------|--------------| | AcQGuide® MAX Steerable Sheath | Acutus Medical, Inc. | K211100 | DRA | | Reference Device | Manufacturer | FDA 510(k) | Product Code | |-----------------------------------------------|------------------|------------|--------------| | CARTO VIZIGO Bi-Directional<br>Guiding Sheath | Biosense Webster | K170997 | DYB | # DEVICE DESCRIPTION [807.92(a)(4)] The AcQGuide® VUE Steerable Sheath, Model 900201 is a single use, percutaneous catheter introducer designed to provide additional maneuverability to interventional catheters that are advanced through the sheath and into the right or left chambers of the heart. The distal portion of the sheath is comprised of a composite structured single lumen shaft. At the proximal end, an ergonomic handle provides torque and active deflection, a hemostasis valve allows safe introduction of an interventional catheter, and a side port provides access for aspiration, fluid flushes and fluid/medication infusions. There is a cable exiting from the handle of the sheath to provide connection for the distal electrodes. This cable terminates with four standard 2mm pins that are connected to the Auxiliary Interface Box or Auxiliary Catheter Cable of the AcQMap High Resolution Imaging and Mapping System, Model 900100 or 900000. The pin connections are assigned by the user for electrode identification (D1, 2, 3, 4) to the corresponding channel on the AcQMap System 900100. No designated order of pin connection is required. The AcQGuide® VUE Steerable Sheath includes a valve bypass tool as an optional accessory to facilitate the insertion of a loop or circular catheter. The dilator is designed to introduce the Steerable Sheath into the vasculature and into chambers of the heart. The dilator has a smooth tapered tip and provides a smooth transition to the round edge of the non-traumatic tip of the sheath. The dilator is able to track over a 0.035″ guidewire. The hub section is attached to the shaft. It is a standard female luer made of high-density polyethylene (HDPE) material. Some of the key design attributes of the Steerable Sheath include: - Deflectability - Hemostasis - Kink-resistance - Visibility under fluoroscopy - Easily flushed during the procedure - Single-handed operation - Biocompatible materials - Sterile, single use - Combined three-way stopcock {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Acutus Medical. The logo is composed of two lines of text. The top line says "ACUTUS" in a sans-serif font, with a small orange dot above the left side of the "A". The bottom line says "MEDICAL" in a smaller, orange sans-serif font. The logo is simple and modern. The AcQGuide® VUE Steerable Sheath contents include the following: - One (1) Acutus Medical AcQGuide® VUE Steerable Sheath - One (1) Acutus Medical AcQGuide® Dilator - One (1) Acutus Medical Valve Bypass Tool - Product Information/Documentation (Instructions for Use) #### INDICATIONS FOR USE [807.92(a)(5)] The AcQGuide® VUE Steerable Sheath with electrodes is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart. The AcQGuide® VUE deflection facilitates catheter positioning. The electrodes help facilitate visualization of the sheath when used with a compatible localization system, such as the AcQMap System. # COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The subject device AcQGuide® VUE Steerable Sheath, Model 900201 has the same intended use and fundamental scientific technology as the predicate device. The predicate device, AcQGuide MAX Steerable Sheath, Model 900200(K211100), and the AcQGuide® VUE Steerable Sheath are identical in most technological characteristics, except for the addition of electrodes and related components used for sheath localization when used with a compatible mapping system. In addition, the AcQGuide® VUE when compared to the reference device provides electrical reference for visualization on the AcQMap System. The purpose of this submission is to introduce the AcQGuide® VUE Steerable Sheath. Table 1 below provides a comparison of the predicate AcQGuide® MAX Steerable Sheath and the reference device CARTO VIZIGO Bi-Directional Guiding Sheath, with regards to classification, comparison of the indications for use, and a comparison of the technological characteristics against the subject device. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for Acutus Medical. The logo is composed of the word "ACUTUS" in a sans-serif font, with the letters in gray. Below the word "ACUTUS" is the word "MEDICAL" in a smaller, orange sans-serif font. The dot above the letter "i" in Acutus is also orange. # TABLE 1: COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE | Table 1: COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE | | | | | |--------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Characteristics | Subject Device<br>AcQGuide® VUE<br>Steerable Sheath<br>(K221044) | Predicate Device<br>AcQGuide MAX<br>Steerable Sheath<br>K211100 | Reference Device<br>CARTO VIZIGO Bi-<br>Directional Guiding<br>Sheath K170997 | Analysis of<br>Differences | | Classification | 21 CFR § 870.1280 | 21 CFR § 870.1280 | 21 CFR § 870.1340 | The subject device and<br>predicate device have the<br>same classification number.<br>The different classification<br>number for the reference<br>device does not raise a new<br>or different question of<br>safety and effectiveness. | | Product Code | DRA | DRA | DYB | The subject device and<br>predicate device have the<br>same product code. The<br>different product code for<br>the reference device does<br>not raise a new or different<br>question of safety and<br>effectiveness. | | Indications<br>for Use | The AcQGuide® VUE<br>Steerable Sheath with<br>electrodes is intended for<br>percutaneous catheter<br>introduction into the<br>vasculature and into the<br>chambers of the heart.<br>The AcQGuide® VUE<br>deflection facilitates<br>catheter positioning. The<br>electrodes help facilitate<br>visualization of the sheath<br>when used with a<br>compatible localization<br>system, such as the<br>AcQMap System. | The AcQGuide MAX is<br>intended for<br>percutaneous<br>catheter introduction<br>into the vasculature<br>and into the<br>chambers of the<br>heart. The AcQGuide<br>MAX deflection<br>facilitates catheter<br>positioning. | The Biosense Webster<br>CARTO VIZIGO™ 8.5F<br>Bi-Directional Guiding<br>Sheath is indicated for<br>introducing various<br>cardiovascular<br>catheters into the<br>heart, including the left<br>side of the heart<br>through the interatrial<br>septum. The sheath<br>curve can be visualized<br>when used with<br>compatible CARTO 3 EP<br>Navigation Systems. | Similar indication for use<br>language compared to<br>the predicate. Proposed<br>indications for use fall<br>within the intended use<br>for the predicate device.<br>Similar indication for use<br>compared to the<br>reference device, related<br>to utilizing electrodes to<br>help visualization of the<br>sheath. The differences in<br>Indications for Use do not<br>raise different questions<br>of safety or<br>effectiveness. | | Table 1: COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE | | | | | | Characteristics | Subject Device<br>AcQGuide® VUE<br>Steerable Sheath<br>(K221044) | Predicate Device<br>AcQGuide MAX<br>Steerable Sheath<br>K211100 | Reference Device<br>CARTO VIZIGO Bi-<br>Directional Guiding<br>Sheath K170997 | Analysis of<br>Differences | | Intended Use | Catheter Delivery<br>System with sheath<br>visualization when used<br>with compatible<br>mapping system | Catheter Delivery<br>System | Catheter Delivery<br>System with shaft<br>visualization when used<br>with a compatible<br>CARTO 3EP navigation<br>system. | Same intended use as the<br>predicate device with<br>regard to catheter<br>delivery. Similar intended<br>use as the reference<br>device with respect to<br>use of electrodes with<br>catheter delivery system<br>to assist with visualization<br>when used with<br>compatible mapping<br>system. The differences<br>in Intended Use do not<br>raise different questions<br>of safety or<br>effectiveness. | | Key Components | Employs single<br>deflection curve with<br>atraumatic tip. Sheath<br>deflection facilitates<br>accurate catheter<br>positioning. Able to<br>facilitate use with<br>catheters up to 12 Fr in<br>diameter. Hemostasis<br>valve to prevent air<br>incursion and minimizes<br>blood loss. Flush port<br>provides ability to<br>administer saline flush<br>throughout procedure.<br>Usable length of up to 70<br>cm. Visibility of distal<br>electrode under<br>fluoroscopy. Electrodes<br>that aide in visualization<br>of sheath when used<br>with the AcQMap<br>System. | Employs single<br>deflection curve with<br>atraumatic tip. Sheath<br>deflection facilitates<br>accurate catheter<br>positioning. Facilitates<br>use with catheters up<br>to 12 Fr in diameter.<br>Hemostasis valve to<br>prevent air incursion<br>and minimizes blood<br>loss. Flush port<br>provides ability to<br>administer saline flush<br>throughout<br>procedure.<br>Usable length of up to<br>70 cm. | Bi-directional steerable<br>sheath fitted with a<br>hemostasis valve to<br>minimize blood loss.<br>Radiopaque tip marker<br>to allow fluoroscopic<br>visualization. 4<br>electrodes spaced<br>along shaft to enable<br>shaft visualization. | Identical key components<br>as the predicate device with<br>regards to usable length,<br>French size compatibility,<br>sheath deflection,<br>hemostasis valve, and side<br>flush port.<br>Identical key components<br>as the reference device<br>with regards to sheath<br>visibility of distal<br>electrode under<br>fluoroscopy, and<br>electrodes that aide in<br>visualization of sheath<br>when used with a<br>compatible mapping<br>system. | | Deflection<br>and Reach | Maximum deflection:<br>180 degrees<br>Reach: 5.0 cm at 90<br>degrees | Maximum deflection:<br>180 degrees<br>Reach: 5.0 cm at 90<br>degrees | Maximum deflection:<br>180 degrees clockwise<br>and counter clockwise<br>Reach: Not available | Identical deflection<br>mechanism and reach as<br>the predicate device. | | Characteristics | Subject Device<br>AcQGuide® VUE<br>Steerable Sheath<br>(K221044) | Predicate Device<br>AcQGuide MAX<br>Steerable Sheath<br>K211100 | Reference Device<br>CARTO VIZIGO Bi-<br>Directional Guiding<br>Sheath K170997 | Analysis of<br>Differences | | Side Port for<br>Flushing<br>(Y/N) | Yes | Yes | Yes | Identical | | Device Inner<br>Diameter (ID)<br>(French) | 12.4 Fr (4.1 mm) | 12.4 Fr (4.1 mm) | 8.5Fr | Identical to predicate<br>device. | | Device Outer<br>Diameter<br>(OD) | 15.2 Fr (5.1)mm | 15.2 Fr (5.1)mm | 11.5Fr | Identical to the predicate<br>device. | | Device Usable<br>Length | 70 cm | 70 cm | 71 cm | Identical to predicate<br>device. | | Device Overall<br>Length | 85cm | 85cm | Not available | Identical to the predicate<br>device | | Device Materials | Sheath: PTFE Liner for<br>main lumen Pebax 4033<br>with 20-25% BaSO4,<br>Pantone 422C, Pebax 5533<br>Pantone 419C, Pebax 3533<br>Pantone 422C, Pebax 6333<br>Pantone 422C, Pebax 7233<br>Pantone 422C<br>Hemostasis Valve: Silicone<br>.80 NVR 50D (100578) | Sheath: PTFE Liner for<br>main lumen Pebax 4033<br>with 20-25% BaSO4,<br>Pantone 422C, Pebax 5533<br>Pantone 419C, Pebax 3533<br>Pantone 422C, Pebax 6333<br>Pantone 422C, Pebax 7233<br>Pantone 422C<br>Hemostasis Valve:<br>Silicone .80 NVR 50D<br>(100578) | Not available | Identical to the predicate<br>device. | | Device Color<br>Additives | Dupont 2020T 9002-88-4,<br>Pebax | Dupont 2020T 9002-88-<br>4, Pebax | Not available | Identical to the predicate<br>device. | | Radiopaque<br>Marker | Total Length: .039" (0.99<br>mm). Distal sections are<br>visible when using<br>standard or low-level<br>fluoroscopy | Total Length: .039"<br>(0.99 mm). Distal<br>sections are visible<br>when using standard<br>or low-level<br>fluoroscopy | Radiopaque tip marker<br>to allow fluoroscopic<br>visualization. | Identical to predicate<br>device. | | Characteristics | Subject Device<br>AcQGuide® VUE<br>Steerable Sheath<br>(K221044) | Predicate Device<br>AcQGuide MAX<br>Steerable Sheath<br>K211100 | Reference Device<br>CARTO VIZIGO Bi-<br>Directional Guiding<br>Sheath K170997 | Analysis of<br>Differences | | Guide wire<br>compatibility | .035" | .035" | Not available | Identical to predicate<br>device. | | Dilator | Yes<br>Overall Length: 90.02cm<br>Usable Length: 87.4 cm<br>Outer Diameter: 4 mm<br>Inner Diameter: .039"<br>(0.99 mm) | Yes<br>Overall Length:<br>90.02cm<br>Usable Length: 87.4 cm<br>Outer Diameter: 4 mm<br>Inner Diameter: .039"<br>(0.99 mm) | Not available | Identical to predicate<br>device. | | Electrodes | Yes<br>Platinum / Iridium | No | Yes<br>Not available | Similar electrode material<br>as the reference device.<br>Minor differences in<br>material do not raise<br>different questions of<br>safety or effectiveness. | | Able to visualize<br>sheath on<br>Mapping System | Yes<br>Acutus AcQMap System<br>Model 900100 or 900000 | No | Yes | Subject device is able to be<br>visualized on the AcQMap<br>System which is identical<br>to the reference device<br>used with compatible<br>mapping system. | | Sterilization<br>Method | Ethylene Oxide (EtO) | Ethylene Oxide (EtO) | Ethylene Oxide (EtO) | Identical | | Shelf Life | 24 months | 18 months | Not available | Minor differences in shelf<br>life do not raise different<br>questions of safety or<br>effectiveness. | | Biocompatible<br>Blood, Body and<br>Fluid Contacting<br>Materials | Yes | Yes | Yes | Identical | {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for Acutus Medical. The word "ACUTUS" is written in a sans-serif font in gray color. Below the word "ACUTUS", the word "MEDICAL" is written in a smaller sans-serif font in orange color. {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for Acutus Medical. The logo is composed of the word "ACUTUS" in a stylized, sans-serif font, with a small orange dot above the left side of the letter A. Below the word "ACUTUS" is the word "MEDICAL" in a smaller, sans-serif font, with each letter spaced out. The word "MEDICAL" is also in orange. {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the logo for Acutus Medical. The logo features the company name in a stylized font, with the word "ACUTUS" in gray and "MEDICAL" in orange. The letters are spaced out, and the overall design is modern and clean. {10}------------------------------------------------ Image /page/10/Picture/0 description: The image shows the logo for Acutus Medical. The logo is composed of two lines of text. The top line says "ACUTUS" in gray, sans-serif font. The bottom line says "MEDICAL" in a smaller, orange, sans-serif font. # PERFORMANCE DATA [807.92(b)] All necessary testing was conducted on the AcQGuide® VUE Steerable Sheath, Model 900201 to support a determination of substantial equivalence to the predicate device. #### NONCLINICAL TESTING SUMMARY - BENCH [807.92(b)(1)] The necessary bench testing was performed on the AcQGuide® VUE Steerable Sheath, Model 900201 in order to ensure that it conformed to the design specifications and to support a determination of substantial equivalence to the predicate device. The bench testing performed on the AcQGuide® VUE Steerable Sheath, Model 900201 consisted of the following: - Packaging and shelf-life - Sterilization - Biocompatibility - Design Verification and Validation studies, including: - O Dimensional inspection of Sheath and Dilator - Visual inspection for any defects O - Leak testing O - Functional and compatibility testing O - Mechanical testing O - Handle torque testing O - . Physician simulated use in an animal model The collective results of the nonclinical testing demonstrate that the materials chosen, the manufacturing processes, and design of the AcQGuide® VUE Steerable Sheath, Model 900201 meets the specifications necessary for consistent performance during its intended use. In addition, the collective bench testing demonstrates that the AcQGuide® VUE Steerable Sheath, Model 900201 does not have any potential impact on the safety or effectiveness for percutaneous catheter introduction into the vasculature and into chambers of the heart, when compared to the predicate device. The AcQGuide® VUE Steerable Sheath, Model 900201 was tested to verify that the device met its established performance specifications. This was done by utilizing dimensional testing to ensure that all components met the required dimensions per specifications, visual inspection to ensure that all components were free of defects, leak testing to ensure the packaging meets its specifications, functional and compatibility testing to ensure the functional and compatibility requirements are met, mechanical testing to ensure the components are able to withstand the expected forces, and handle torque testing to ensure it can withstand the expected torques. {11}------------------------------------------------ Image /page/11/Picture/0 description: The image shows the logo for Acutus Medical. The word "ACUTUS" is written in a sans-serif font in gray color. Below the word "ACUTUS", the word "MEDICAL" is written in a smaller sans-serif font in orange color. #### CLINICAL TESTING SUMMARY [807.92(b)(2)] This section is not applicable. No clinical testing is being submitted to support review of this 510(k) premarket notification. #### BIOCOMPATIBILITY The AcQGuide® VUE Steerable Sheath, Model 900201 is manufactured from materials similar to the predicate AcQGuide® MAX Steerable Sheath, Model 900200, and has passed all of the appropriate biocompatibility tests. Biocompatibility testing for the AcQGuide® VUE Steerable Sheath, Model 900201 was conducted in accordance with ISO 10993-1:2018 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process and the FDA Guidance Document titled, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" issued on September 4, 2020. The AcQGuide® VUE Steerable Sheath, Model 900201 has three components: a composite structured single lumen shaft, an ergonomic handle to provide torque and active deflection, and a hemostasis valve with side flush port. Each component was evaluated per the above standards and guidance documents to determine the appropriate biocompatibility testing. Per ISO 10993-1:2018. The AcQGuide® VUE Steerable Sheath is categorized as an external communicating device, which comes into contact with circulating blood for a limited (≤24 hours) contact duration. #### STERILIZATION AcQGuide® VUE is labeled and marketed as "Sterile" by Acutus Medical, Inc. AcQGuide® VUE is subjected to the identical ethylene oxide (EO) sterilization process as the predicate device, AcQGuide® MAX Steerable Sheath, Model 900200, to meet a sterility assurance level (SAL) of 10 ്. .Acutus Medical, Inc. conducted a technical review to establish that the subject device can be adopted into the existing AcQGuide EtO processing cycle number ACU-537 at Steris Corporation. The following aspects were used to perform the evaluation per AAMI TIR28:2009 and ISO 11135:2014: Packaging System Evaluation, Product Design/Materials, Barrier System, Load Configuration, and Bioburden Resistance. Shelf life study was performed to confirm acceptable product performance and ability to withstand sterilization. All materials identified were suitable for EtO sterilization and were not known to retain high EtO residual levels, or adversely impact the ability of EtO process parameters to be met. In addition, Acutus Medical, Inc. performed material mediated pyrogenicity testing, to evaluate its potential to produce a pyrogenic response when tested in New Zealand white rabbits, per 21 CFR Part 58 Compliance Good Laboratory Practice for Nonclinical Laboratory Studies. The subject AcQGuide® VUE Steerable Sheath, Model 900201 is considered non-pyrogenic per the requirements of ISO 10993-11 guidelines. {12}------------------------------------------------ Image /page/12/Picture/0 description: The image shows the logo for Acutus Medical. The logo is composed of the word "ACUTUS" in a gray, sans-serif font. Below the word "ACUTUS" is the word "MEDICAL" in a smaller, orange, sans-serif font. The logo is simple and modern. # ELECTRICAL SAFETY AND ELECTROMAGNETIC COMPATIBILITY (EMC) For the purpose of this submission, the subject device AcQGuide® VUE Steerable Sheath requires an electrical safety evaluation. The AcQGuide® VUE has been assessed for electrical safety in accordance with IEC 60601-1, Medical electrical equipment Part 1: General requirements for basic safety and essential performance, and IEC 60601-1-2, Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances – requirements and tests. The AcQGuide® VUE Steerable Sheath is a CF Defibrillation proof applied part, used with the AcQMap System or any other 3D cardiac mapping system. The verification testing of the AcQGuide® VUE was constrained to the subset of Clauses 8.1 - 8.11 of IEC 60601-1. The completion of the following tests demonstrated that the AcQGuide® VUE is safe for its intended use: - Defibrillation Proof Applied Parts (IEC 60601-1, Clause 8.5.5) - . Dielectric Voltage Withstand (IEC 60601-1, Clause 8.8.3) - Sheath Leakage Current (IEC 60601-1, Clause 8.7) The relevant test sections of IEC 60601-1-2 is as follows: - Qualification testing of the AcQMap System with applied parts containing cables and Platinum-Iridium Electrodes. - . Qualification testing of the AcQMap System using general-purpose electrophysiology catheters to connect to the Auxiliary inputs. The AcQGuide® VUE is built and constructed to be a general-purpose electrophysiology applied part. Therefore, using the AcQGuide® VUE as one of the applied parts does not invalidate or require retesting of the AcQMap Console. Additionally, retesting of the electromagnetic compatibility (EMC) is not required in order to release the AcQGuide® VUE catheter for use with the AcQMap System. #### CONCLUSIONS [807.92(b)(3)] Extensive nonclinical testing has been performed on the AcQGuide® VUE Steerable Sheath, Model 900201 to evaluate the overall performance of the device. The nonclinical tests demonstrated that the device is as safe and effective as the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...