This in vitro diagnostic method is intended to quantitatively measure the concentration of cardiac Troponin I (TnI) in human serum and plasma (lithium heparin) using the Bayer Immuno 1™ system. When used in conjunction with other clinical data such as presenting symptoms and diagnostic procedures, measurement of cardiac Troponin I aids in the diagnosis of acute myocardial infarction (AMI). This diagnostic method is not intended for use on any other system.
Device Story
Bayer Immuno 1 Troponin I assay is an in vitro diagnostic enzyme-linked sandwich immunoassay. Input: human serum or plasma samples. Operation: sample mixed with fluorescein-labeled monoclonal antibody, alkaline phosphatase (ALP)-labeled goat affinity-purified antibody, and magnetizable particles coated with anti-fluorescein antibody. Complex formation (fluorescein-conjugate:Troponin I:ALP-conjugate) captured on magnetic particles; particles washed; para-nitrophenyl phosphate substrate added. ALP converts substrate to para-nitrophenoxide; absorbance measured at 405 nm and 450 nm. Output: Troponin I concentration (0-200 ng/mL). Used in clinical laboratories; operated by trained personnel. Results assist clinicians in diagnosing acute myocardial infarction when combined with patient symptoms and other diagnostic data.
Clinical Evidence
Clinical correlation study conducted at two sites with 279 patient samples comparing Bayer Immuno 1 Troponin I to Dade Stratus Troponin I. Results: r = 0.98, slope = 0.96, intercept = -0.2. Linearity recoveries: 88-110%. Total imprecision: 2.1% to 4.3% CV. Minimum detectable concentration: 0.1 ng/mL. Healthy donor samples (n=197) were <0.1 ng/mL. AMI patient samples (n=41) were ≥0.9 ng/mL.
Technological Characteristics
Solid-phase enzyme immunoassay. Reagents: monoclonal/polyclonal antibodies, magnetizable particles, para-nitrophenyl phosphate substrate. Detection: spectrophotometric absorbance at 405/450 nm. Range: 0-200 ng/mL. System: Bayer Immuno 1. Calibration: linear point fit using liquid calibrators.
Indications for Use
Indicated for the quantitative measurement of cardiac Troponin I in human serum and plasma to aid in the diagnosis of acute myocardial infarction (AMI) in patients presenting with symptoms suggestive of cardiac events.
Regulatory Classification
Identification
A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
Predicate Devices
Dade Stratus® Troponin I kit
Related Devices
K993353 — TROPONIN I ASSAY FOR THE BAYER IMMUNO 1 SYSTEM ( IN VITRO DIAGNOSTIC SYSTEM) · Bayer Corp. · Dec 6, 1999
K991177 — OPUS TROPONIN I (CTN) TEST MODULES · Dade Behring, Inc. · Jul 1, 1999
K973650 — TROPONIN (TROP) METHOD · Dade Chemistry Systems, Inc. · Oct 20, 1997
K242870 — Access hsTnI · Beckman Coulter, Inc. · Jun 16, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
K973616
DEC - 5 1997
# 510K SUMMARY OF SAFETY AND EFFECTIVENESS
# BAYER IMMUNO 1™ SYSTEM TROPONIN I METHOD, SETPOINT CALIBRATORS AND TESTPOINT CONTROLS FOR THE IMMUNO 1 SYSTEM. Bayer Corporation, 511Benedict Avenue, Tarrytown, New York 10591
Listed below is a comparison of the performance between the Immuno 1 Troponin I method ( T01-3887-51) and a similar device that was granted FDA determination of substantial equivalence: the Dade Stratus® Troponin I kit. The information used in this summary of Safety and Effectiveness was extracted from the Troponin I Method Sheet and from data on file at Bayer Corp.
#### INTENDED USE
This in vitro diagnostic procedure is a solid- phase enzyme immunoassay intended for the quantitative determination of Troponin I in human serum and plasma on the Bayer Immuno 1 system. When used in combination with other clinical data such as presenting symptoms and diagnostic procedures, measurement of Troponin I aids in the diagnosis of acute myocardial infarction.
### ASSAY DESCRIPTION
The Baver Immuno 1 Troponin I™ assay is an enzyme label sandwich assay using a monoclonal and a polvclonal antibody. A Troponin I specific monoclonal antibody is labeled with fluorescein and a Troponin I specific goat affinity purified antibody is labeled with alkaline phosphatase (ALP). The solid phase consists of a suspension of magnetizable particles coated with antibody to fluorescein (mIMP reagent). Sample or cailbrator, R1 reagent containing fluorescein - antibody conjugate. R2 reagent containing ALP-antibody conjugate and mIMP reagent are mixed and incubated at 37°C. In the presence of Troponin I a (fluorescein-conjugate: Troponin I:ALPconjugate) complex is formed and captured by the anti fluorescein antibodies on the magnetic particles. The particles are washed and para-nitrophenvl phosphate solution is added. The ALP in the antibody conjugate reacts with the substrate to form para-nitrophenoxide and phosphate. Increasing absorbance due to the formation of para-nitrophenoxide is monitored at 405 nm and 450 nm. The dose response curve is directly proportional to the concentration of Troponin I in the sample. A linear point fit is used to construct the dose response curve. The Baver Immuno 1 Troponin ITM assay has a range of 0 to 200 ng/ml and liquid calibrators are provided with values of 0, 5, 10, 20, 60 and 200 ng/ml Troponin I. Bayer Immuno 1 Troponin I TEST point controls are provided with values of 3, 7, 50 ng/mL Troponin I.
The Dade Stratus® Troponin I assay utilizes the binding of two Troponin I monoclonal antibodies in a two site "sandwich" immunoassay. Both the Dade Stratus® and Bayer Immuno 17M Troponin Tassays utilize an alkaline phospatase enzyme conjugated antibody. The Bayer Immuno 17M Troponin I assay uses a para-nitrophen; I phosphate substrate and measures
{1}------------------------------------------------
increasing absorbance due to the formation of para-nitrophenoxide at 405 nm and 450 nm, while the Dade Stratus® Troponin I assay uses 4-Methylumberlliferyl Phosphate as the substrate and measures front surface fluorescence. The Bayer Immuno 11M Troponin I test uses liquid calibrators prepared with a buffered bovine serum albumin and cardiac Troponin I at specific levels while the Dade Stratus® Troponin I uses frozen liquid calibrators prepared from buffered bovine matrix and human cardiac troponin I at specified levels.
## PERFORMANCE CHARACTERISTICS
Total imprecision of Bayer Immuno 1™ Troponin I method ranges from 4.3% with a low human control having a mean troponin I value of 2.9 ng/mL to 2.1% for a high control having a mean troponin I value of 47.8 ng/mL. These estimates of imprecision were made by analyzing human serum samples in duplicate on two Immuno 1 instruments on each of 18 days. The results are calculated from a calibration curve generated on each instrument on the first day.
Linearity recoveries range from 88% to 110% determined by diluting human AMI patient pools with troponin I free normal human serum. Recoveries determined by diluting AMI patient pools with Immuno 1 Sample diluent B range from 85% to 116% with an average recovery of 97.9%. A Correlation study at two clinical trial sites between the Bayer Immuno 1™ Troponin I method and the Dade Stratus® Troponin I method using a total of 279 patient samples yielded regression statistics of r = 0.98, Immuno 1 = 0.96 x Stratus - 0.2. Troponin I in the 279 patient samples ranged from 0 to 77.1 ng/mL.
Minimum Detectable Concentration of the Bayer Immuno 17M Troponin I method is 0.1 ng/mL. This is a multisystem estimate of two (2) times the with-in run standard deviation of the zero calibrator.
Cross Reactivity to human cardiac Troponin T, human cardiac Troponin C, human skeletal Troponin I, human skeletal Troponin T, bovine actin and bovine tropomyosin is minimal or undetectable.
#### EXPECTED VALUES
Specimens from 197 healthy Red Cross blood donors were all found to give results below the minimum detectable concentration of 0.1 ng/mL. Specimens from 64 patients hospitalized with a clinical diagnosis of angina gave results less than the minimum detectable concentration for 60 patients and in the range 0.2 to 0.6 ng/mL for the other 4. Specimens from 41 hospitalized patients with a clinical diagnosis of AMI had results greater than or equal to 0.9 ng/mL.
The Bayer Immuno 17ª Troponin I method is substantially equivalent to the Dade Stratus® Troponin I method currently approved for clinical use in the United States.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines forming its body and wings. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
DEC - 5 1997
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Gabriel Muraca, Jr. . Manager Regulatory Affairs Bayer Corporation 511 Benedict Avenue Tarrytown,, New York 10591-5097
Re: K973616 Troponin I Assay for Immuno 1™ System Requlatory Class: II Product Code: MMI November 18, 1997 Dated: November 19, 1997 Received:
Dear Mr. Muraca:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਮੁ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{3}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity (shirtor) in To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven Litman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radioloqical Health
Enclosure
{4}------------------------------------------------
Page 1 of 1
510(k) Number (if known): k973616
Device Name: Bayer Immuno 1TM System Troponin I (TnI)
Indications For Use:
This in vitro diagnostic method is intended to quantitatively measure the
reasure the manage in TV (D.) indexemen servim and plasma (II) This in vitro diagnostic method is intention is numan serum and plasma (lithium
concentration of cardiac Troponin 17% and we alling wed in conjunction with concentration of cardiac Tropolitir (111) in nation in conjunction with
heparin) using the Bayer Immuno 11M system. When used in conjunction with heparin) using the Bayer Inimalio I - System - and diagnostic procedures,
other clinical data such as presenting symptoms and diagnosis of acute myocardia other clinical data such as presenting symptoms and any
measurement of cardiac Troponin I aids in the diagnosis of acute myocardial infarction (AMI).
This diagnostic method is not intended for use on any other system.
(Division Sign-Off)
Division of Clir Laboratory zvices
510(k) Number. K973616
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use_ (Per 21 CFR 801.109) OR
Over-The-Counter Use_
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.