PATHFAST®hs-cTnI-II is an in vitro diagnostic test for the quantitative measurement of cardiac Troponin I (cTnI) in heparinized or EDTA whole blood and plasma. Measurements of cardiac Troponin I are used as an aid in the diagnosis of acute myocardial infarction (AMI). PATHFAST® hs-cTnI-II is for use in clinical laboratory or point of care (POC) settings.
Device Story
PATHFAST® hs-cTnI-II is a chemiluminescent enzyme immunoassay for quantitative measurement of cardiac Troponin I (cTnI). Input: heparinized or EDTA whole blood or plasma samples. Operation: operator dispenses sample into reagent cartridge; instrument incubates sample with antibody-coated magnetic particles and alkaline phosphatase conjugate; Magtration® technology performs bound/free separation; chemiluminescent substrate (CDP-Star) added; photomultiplier tube detects emitted photons. Output: analyte concentration values calculated via master calibration curve. Used in clinical laboratories or point-of-care settings by healthcare professionals. Output assists clinicians in diagnosing acute myocardial infarction (AMI). Benefits: provides rapid, quantitative cardiac biomarker assessment to support timely clinical decision-making.
Clinical Evidence
No new clinical trials were conducted. The submission relies on analytical performance data (linearity, LoB, LoD, LoQ) and re-evaluation of clinical data from the predicate device (K100130). Linearity was demonstrated from 4.1 to 50,000 ng/L with deviations <10%. LoQ was established at 4.1 ng/L (CV <20%). The 99th percentile cutoff remains 29 ng/L. Clinical sensitivity and specificity are unchanged from the predicate.
Technological Characteristics
Chemiluminescent enzyme immunoassay; quantitative measurement of cTnI. Reportable range: 4.1 to 50,000 ng/L. Analyzed on PATHFAST Analyzer. Standards: CLSI EP6 (linearity), CLSI EP17-A2 (detection capability).
Indications for Use
Indicated for the quantitative measurement of cardiac Troponin I in heparinized or EDTA whole blood and plasma to aid in the diagnosis of acute myocardial infarction (AMI) in clinical laboratory or point-of-care settings.
Regulatory Classification
Identification
A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
K100130 — PATHFAST CTNI-II TEST, PATHFAST CTNI CALIBRATORS, PATHFAST SAMPLE DILUENT 2 · Mitsubishi Chemical Medience CO · May 16, 2011
K242870 — Access hsTnI · Beckman Coulter, Inc. · Jun 16, 2025
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K231974
B Applicant
PHC Corporation
C Proprietary and Established Names
PATHFAST@hs-cTnI-II
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| MMI | Class II | 21 CFR 862.1215 - Creatine Phosphokinase/Creatine Kinase Or Isoenzymes Test System | CH - Clinical Chemistry |
## II Submission/Device Overview:
A Purpose for Submission:
Modification of an existing device
B Measurand:
Troponin I
C Type of Test:
Quantitative, chemiluminescent enzyme immunoassay
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
K231974 - Page 2 of 5
# III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
PATHFAST®hs-cTnI-II is an in vitro diagnostic test for the quantitative measurement of cardiac Troponin I (cTnI) in heparinized or EDTA whole blood and plasma. Measurements of cardiac Troponin I are used as an aid in the diagnosis of acute myocardial infarction (AMI).
PATHFAST® hs-cTnI-II is for use in clinical laboratory or point of care (POC) settings.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
D Special Instrument Requirements:
PATHFAST Analyzer
# IV Device/System Characteristics:
A Device Description:
Same as K100130
B Principle of Operation:
Same as K100130
V Substantial Equivalence Information:
A Predicate Device Name(s):
PATHFAST cTnI-II test
B Predicate 510(k) Number(s):
K100130
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K231974 | K100130 |
| --- | --- | --- |
| Device Trade Name | PATHFAST®hs-cTnI-II | PATHFAST cTnI-II |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | For the quantitative measurement of cardiac Troponin I (cTnI) in heparinized or EDTA | Same |
{2}
| Device & Predicate Device(s): | K231974 | K100130 |
| --- | --- | --- |
| | whole blood and plasma. Measurements of cardiac Troponin I are used as an aid in the diagnosis of acute myocardial infarction (AMI). For use in clinical laboratory or point of care (POC) settings. | |
| Methodology | Chemiluminescent enzyme immunoassay | Same |
| General Device Characteristic Differences | | |
| Reportable range | 4.1 to 50,000 ng/L | 0.019 to 50 ng/mL |
VI Standards/Guidance Documents Referenced:
CLSI EP6 2nd Edition: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline
CLSI EP17-A2: Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
Previously established in K100130.
2. Linearity:
Linearity studies using EDTA and lithium heparin whole blood and plasma samples were designed and analyzed following the recommendations in the Guideline CLSI EP6 2nd Edition: Evaluation of the Linearity of Quantitative Measurement Procedures.
EDTA whole blood and plasma samples, and lithium heparin whole blood and plasma samples with low and high cTnI values were mixed to produce up to 12 levels with concentrations ranging from 1 to 64,500 ng/L. Three dilution series of each sample matrix were prepared and tested with three different lots of reagents (one lot for each dilution series). Each level was tested in replicates of three. The maximum deviation from linearity observed in the study across all sample matrices was 9.7%.
K231974 - Page 3 of 5
{3}
The studies support that the device is linear from 4.1 to 50,000 ng/L.
3. Analytical Specificity/Interference:
Previously established in K100130.
4. Assay Reportable Range:
The measuring range of the PATHFAST hs-cTnI-II is from 4.1 ng/L to 50,000 ng/L.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Unchanged from K100130.
6. Detection Limit:
The limit of blank (LoB) was determined by testing four lots of reagents and four blank samples. The limit of detection (LoD) for EDTA and lithium heparin whole blood and plasma was determined by testing 4 low level samples and four lots of reagents for each sample matrix.
The limit of quantitation (LoQ) study was conducted using EDTA and lithium heparin whole blood and plasma samples at concentrations expected to be around the proposed LoQ. The LoQ was determined by testing six samples per sample type and two reagent lots.
The LoB was calculated using the classical approach using nonparametric analysis as described in CLSI EP17-A2 Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures. The LoD was calculated per the classical approach using parametric analysis as described in the CLSI EP17-A2 Guideline. The LoQ was determined based on a precision performance goal of less than 20 % CV.
The results of the study support the assay LoQ of 4.1 ng/L.
| | EDTA (ng/L) | | LiHep (ng/L) | |
| --- | --- | --- | --- | --- |
| | WB | Plasma | WB | Plasma |
| LoB | 1.466 | 1.466 | 1.466 | 1.466 |
| LoD | 2.991 | 2.958 | 2.942 | 3.002 |
| LoQ | 4.1 | 4.1 | 4.1 | 4.1 |
7. Assay Cut-Off:
See K100130
K231974 - Page 4 of 5
{4}
K231974 - Page 5 of 5
B Comparison Studies:
1. Method Comparison with Predicate Device:
See K100130
2. Matrix Comparison:
See K100130
C Clinical Studies:
1. Clinical Sensitivity:
See K100130
2. Clinical Specificity:
See K100130
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
D Clinical Cut-Off:
See K100130
E Expected Values/Reference Range:
See K100130
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.