PATHFAST® cTnl-II test is an in vitro diagnostic test for the quantitative measurement of cardiac Troponin I (cTn!) in heparinized or EDTA whole blood and plasma. Measurements of cardiac Troponin I are used as an aid in the diagnosis of acute myocardial infarction. This method is for use in clinical laboratory or point of care (POC) settings. The PATHFAST® cTnl Calibrators are for calibration of the PATHFAST® system when used for the quantitative determination of cardiac Troponin I in human heparinized or EDTA whole blood and plasma.
Device Story
The PATHFAST cTnI-II is a quantitative chemiluminescent enzyme immunoassay used to measure cardiac Troponin I (cTnI) in whole blood or plasma. The device is used in clinical laboratories or point-of-care (POC) settings and is operated by healthcare personnel. The operator dispenses the patient sample into a reagent cartridge, which is then processed by the PATHFAST Analyzer. The system uses antibody-coated magnetic particles and an alkaline phosphatase conjugate to bind cTnI; Magtration technology performs bound/free separation. A chemiluminescent substrate is added, and the resulting photon emission is detected by a photomultiplier tube. The analyzer converts the light signal into a concentration value using a master calibration curve. The output provides clinicians with quantitative cTnI levels to aid in the diagnosis of acute myocardial infarction. The device requires serial testing for clinical interpretation.
Clinical Evidence
Prospective study of 333 patients (72 MI, 261 non-MI) presenting with suspected MI. Serial samples collected at 0-2h, 2-6h, and 6-12h. Clinical diagnosis by adjudication panel per 2007 ESC/ACCF/AHA/WHF guidelines. At 0.029 ng/mL cutoff, sensitivity ranged 73.6%-93.1% and specificity 91.6%-92.7% across time points. At 0.264 ng/mL cutoff, sensitivity ranged 23.6%-80.6% and specificity 97.7%-99.2%.
Technological Characteristics
Chemiluminescent enzyme immunoassay. Uses antibody-coated magnetic particles and alkaline phosphatase conjugate. Magtration technology for B/F separation. Photomultiplier tube detection. Reportable range 0.019–50 ng/mL. Storage 2-8°C. Calibration via master curve and user-performed calibration.
Indications for Use
Indicated for quantitative measurement of cardiac Troponin I in heparinized or EDTA whole blood and plasma to aid in the diagnosis of acute myocardial infarction. For use in clinical laboratory or point-of-care settings. Not for risk stratification. Requires serial sampling for interpretation.
Regulatory Classification
Identification
A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
Predicate Devices
Stratus CS Acute care Troponin I Testpak (k033487)
Related Devices
K231974 — PATHFAST®hs-cTnI-II · Phc Corporation · Mar 20, 2024
K242870 — Access hsTnI · Beckman Coulter, Inc. · Jun 16, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with a double helix intertwined around a staff. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular pattern around the caduceus.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Mitsubishi Chemical Medience Co. c/o Ms. Judi Smith, LLC Principal PO Box 103 Baldwin. MD 21013
Food & Drug Administration 10903 New Hampshire Avenue Building 66 Silver Spring, MD 20993
MAY 1 6 2011
k100130 Re:
Trade Name: PATHFAST® cTnI-II test; PATHFAST cTnI Calibrators Regulation Number: 21 CFR §862.1215 Regulation Name: Creatine Phosphokinase/creatine kinase or isoenzymes test svstem Regulatory Class: Class II Product Codes: MMI, JIT Dated: May 5, 2011 Received: May 11, 2011
Dear Ms. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{1}------------------------------------------------
Page 2 -
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
CJC.
Courtney Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indication for Use
510(k) Number (if known): K100130
Device Name: PATHFAST® cTnl-II test
Indication For Use:
PATHFAST® cTnl-II test is an in vitro diagnostic test for the quantitative measurement of cardiac Troponin I (cTn!) in heparinized or EDTA whole blood and plasma. Measurements of cardiac Troponin I are used as an aid in the diagnosis of acute myocardial infarction. This method is for use in clinical laboratory or point of care (POC) settings.
Prescription Use X (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Carol C. Benson
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K100130
{3}------------------------------------------------
# Indication for Use
510(k) Number (if known): K100130
Device Name: PATHFAST® cTnl Calibrators ·
Indication For Use:
The PATHFAST® cTnl Calibrators are for calibration of the PATHFAST® system when used for the quantitative determination of cardiac Troponin I in human heparinized or EDTA whole blood and plasma.
Prescription Use _ × (21 CFR Part 801 Subpart D)
And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Carl C. Benson
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K.100/3D
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for Precision for Medicine. The logo has a blue square, a lighter blue square, and a gold square on the left side. To the right of the squares is the text "PRECISION FOR MEDICINE" in blue. Below the text is the phrase "ACCELERATING VALUE. IMPROVING OUTCOMES."
## VIA FEDERAL EXPRESS AND ELECTRONIC TRANSMISSION
January 29, 2015
U.S. Food and Drug Administration Center for Devices and Radiological Health Document Mail Center - WO66-G609 10903 New Hampshire Avenue Silver Spring, MD 20993-0002
- ATTN: Courtney H. Lias, PhD, Director, Division of Chemistry and Toxicology Devices
- 510(k) Add-To-File for Revised 510(k) Summary and Product Insert Re: PATHFAST CTNI-II TEST, PATHFAST CTNI CALIBRATORS, PATHFAST SAMPLE DILUENT 2 (K100130) LSI Medience Corporation
Dear Dr. Lias,
This submission is in reqards to the above referenced 510(k) Premarket Notification, PATHFAST CTNI-II TEST, PATHFAST CTNI CALIBRATORS, PATHFAST SAMPLE DILUENT 2 (K100130), cleared May 16, 2011.
We are requesting FDA replace the 510(k) Summary on file for K100130 with the attached 510(k) Summary due to two typographical errors found on page 4 of the 510(k) Summary originally provided to FDA. Specifically, in the section entitled, "Diagnosis of AMI-Serial Samples," in the first table describing the results for the 0.029 ng/mL Cutoff, the 95% confidence intervals for Clinical Sensitivity for the 2 to 6 hour and the 6 to 12 hour collection times were incorrectly reported. These errors have been corrected in the attached revised 510(k) Summary. The following table shows the corrections highlighted in yellow.
| | | | # / TOTAL | 95% CI |
|-----------------------|-----------|------|-----------------|-----------------|
| Cutoff<br>0.029 ng/mL | 0 to 2 h | Sens | 73.6% (53/72) | (61.9% - 83.3%) |
| | | Spec | 92.7% (242/261) | (88.9% - 95.6%) |
| | 2 to 6 h | Sens | 93.1% (67/72) | (84.5% - 97.7%) |
| | | Spec | 93.1% (243/261) | (89.3% - 95.9%) |
| | 6 to 12 h | Sens | 91.7% (66/72) | (82.7% - 96.9%) |
| | | Spec | 91.6% (239/261) | (87.5% - 94.6%) |
# Revised 510(k) Summary Table for Diagnosis of AMI - Serial Samples
{5}------------------------------------------------
Image /page/5/Figure/0 description: The image is an abstract design composed of three rectangles. The largest rectangle is on the left and is a dark blue color. To the right of the large rectangle are two smaller rectangles stacked vertically. The top rectangle is a light blue color, and the bottom rectangle is a gold color.
Please note that all information included within the original 510(k) submission that was reviewed and cleared by FDA was accurate and correctly reported. Thank you for your assistance in this matter.
This 510(k) Add-to-File is being provided in duplicate, via one hard copy and one eCopy. The eCopy is an exact duplicate of the paper copy.
Confidentiality
We request that this 510(k) Add-to-File, including commercial information, be maintained by FDA in confidence pursuant to 21 CFR 807.95 for the maximum period allowed by 21 CFR 807.95(b) and (c), and including the maximum postdetermination period specified in 21 CFR 807.95(e). Please notify me directly of any request for release of information pertaining to this 510(k) prior to public disclosure of such information.
Please contact the undersigned at (240) 316-3377 or Judi.Smith@precisionformedicine.com if any additional information is required.
Respectfully Submitted,
Judi Sans
Judi Smith, MS, MT (ASCP) Vice President, In Vitro Diagnostics and Quality Precision for Medicine
For: LSI Medience Corporation
{6}------------------------------------------------
#### 510(k) SUMMARY
## CONTACT:
Judi Smith Precision for Medicine 2 Bethesda Metro Center, Suite 850 Bethesda, MD 20814
### NAME OF DEVICE:
| Trade Name: | PATHFAST® cTnI-II Test, PATHFAST® Calibrators, PATHFAST® Sample Diluent 2 cTnl |
|----------------------------|--------------------------------------------------------------------------------|
| Common Names/Descriptions: | Cardiac troponin immunoassay |
| Classification Name: | Immunoassay method, troponin subunit |
## PREDICATE DEVICE:
Siemens Stratus CS Acute Care Troponin I Testpak
### DEVICE DESCRIPTION:
#### INTENDED USE:
PATHFAST cTnl-II test is an in vitro diagnostic test for the quantitative measurement of cardiac Troponin I (cTnl) in heparinized or EDTA whole blood and plasma. Measurements of cardiac Troponin I are used as an aid in the diagnosis of acute myocardial infarction. This method is for use in clinical laboratory or point of care (POC) settings.
The PATHFAST cTnl Calibrators are for calibration of the PATHFAST system when used for the quantitative determination of cardiac Troponin I in human heparinized or EDTA whole blood and plasma.
PATHFAST SAMPLE DILUENT 2 is an in vitro diagnostic product used to dilute samples when PATHFAST assay values exceed the reportable assay range.
### DESCRIPTION:
The PATHFAST cTnl-II assay is for the quantitative measurement of human cardiac troponin I in heparinized or EDTA whole blood or plasma. The assay is designed for use on the PATHFAST instrument.
The PATHFAST cTnl-II test is a chemiluminescent enzyme immunoassay performed on the PATHFAST instrument. Patient samples, whole blood or plasma, are dispensed by the operator into the designated area on the reagent cartridge. The instrument combines the patient sample, the antibody coated magnetic particles, and the alkaline phosphatase conjugate and incubates the mixture for 5 minutes at 37°C. During this incubation, the analyte in the patient sample binds to the antibody on the
{7}------------------------------------------------
coated particles, and the alkaline phosphatase conjugate binds to the analyteantibody-coated particle.
After the incubation, the instrument performs Bound/Free (B/F) separation using Magtration® technology to remove any excess unbound reagents. The chemiluminescent substrate is then added. The substrate is catalyzed by the bound alkaline phosphatase, which results in emission of photons.
The photo-multiplier tube in the PATHFAST instrument detects the photons that are emitted during the reaction. The chemiluminescent count is converted to analyte concentration values by the instrument based on the master calibration curve for the reagent lot.
# SUBSTANTIAL EQUIVALENCE:
The PATHFAST cTnl-II assay is substantially equivalent to the Stratus CS cTnl TestPak (predicate device - K033487). The following table summarizes the similarities and differences.
| Comparison between PATHFAST cTnI-II and predicate device |
|----------------------------------------------------------|
| PATHFAST cTnI-II |
| | PATHFAST cTnI-II | Stratus CS cTnI TestPak |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | An in vitro diagnostic test for the quantitative measurement of cardiac Troponin I (cTnI) in heparinized or EDTA whole blood and plasma.<br><br>The PATHFAST cTnl Calibrators are for calibration for the PATHFAST system when used for the quantitative determination of cardiac Troponin I in human heparinized or EDTA whole blood and plasma. | An in vitro diagnostic test for the measurement cardiac Troponin I in heparinized plasma. |
| Indications for Use | PATHFAST cTnI-II is an in vitro diagnostic test for the quantitative measurement of cardiac Troponin I (cTnI) in heparinized or EDTA whole blood and plasma.<br>Measurements of cardiac Troponin I are used as an aid in the diagnosis of acute myocardial infarction. This method is for use in clinical laboratory or point of care (POC) settings. | The Stratus® CS Acute Care™ Troponin I method is an in vitro diagnostic test for the measurement of cardiac Troponin I in heparinized plasma. Cardiac Troponin I measurements can be used as an aid in the diagnosis of myocardial infarction. Cardiac Troponin I can also be used as an aid in the risk stratification of patients with acute coronary syndromes with respect to their relative risk of mortality. |
| Sample type | Plasma, Whole blood | Plasma |
| Anticoagulant | Heparin, EDTA | Heparin |
| Calibration | Reagent lot: Initially by master calibration code, then by user with enclosed calibrators. Recalibration required every four weeks | Reagent lot: Initially by master calibration code, then by user with recommended calibrators.<br>Recalibration recommended every 60 days |
| Controls | Recommended | Recommended |
| Storage | 2 - 8°C | 2 - 8°C |
| Test Methodology | Chemiluminescent enzyme immunoassay | Sandwich-type immunofluorometric assay |
{8}------------------------------------------------
| | PATHFAST cTnI-II | Stratus CS cTnI TestPak |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Reportable range | 0.019 – 50 ng/mL | 0 – 50 ng/mL |
| Precision | Whole blood across instruments/within lot %CV from 0.5% to 6.0% with concentrations from 0.096 to 38.5 ng/mL and across lots/within instrument %CV from 1.9% to 9.1% with concentrations from 0.096 to 41.3 ng/mL<br>Plasma across instruments/within lot %CV from 4.3% to 8.0% with concentrations from 0.146 to 30.4 ng/mL and across lots/within instrument %CV from 1.6% to 10.8% with concentrations from 0.157 to 29.7 ng/mL<br>Plasma cutoff within-run %CV from 3.7% to 6.2% and total %CV from 3.9% to 7.1% with concentrations from 0.022 to 0.251 ng/mL | Within run %CV from 2.7% to 4.3% and total %CV from 3.4% to 5.1% with concentrations from 0.64 to 6.48 ng/mL |
| Limit of blank | 0.004 ng/mL | <0.03 ng/mL (analytical sensitivity) |
| Interfering substances | No interference observed with Bilirubin-conjugated and free (60 mg/dL); Hemoglobin (1000 mg/dL); Triglyceride (1000 mg/dL); Rheumatoid factor (500 IU/mL) | No interference observed with Bilirubin-icterus (60 mg/dl); Hemoglobin (1000 mg/dl); Triglyceride (3000 mg/dl) |
| Cross reactivity | No significant cross reactivity with cTnT (0.08%); cTnC (0.08%); skTnl (0.09%) | No significant cross reactivity with cTnT (1000 ng/mL); cTnC (1000 ng/mL); skTnl (280 ng/mL) |
| Reference interval | 490 heparinized plasma samples from apparently healthy individuals. | 101 heparinized plasma samples from apparently healthy individuals. |
| | 99th percentile = 0.029 ng/mL | 99th percentile = 0.00 – 0.07 ng/mL<br>(With Stratus CS STAT cardiac Troponin I Assay) 168 samples from 0.00 to 34.38 ng/mL. |
| Comparison with predicate | Comparison with predicate device<br>57 samples from 0.100 to 43.45 ng/mL.<br>$y = 0.947x - 0.005, r = 0.994$ | $y = 0.96x - 0.11, r = 0.99$ |
# Diagnosis of AMI – Serial Samples
An additional study was performed on prospectively collected banked serial plasma samples (heparin anticoagulant) from 333 consecutively eligible consented patients presenting to the ER with suspicion of myocardial infarction (MI). The study vielded 72 MI patients and 261 non-MI patients. The blood sample collection times were at 0-2 hours, at 2 to 6 hours, and at 6 to 12 hours of presentation to the ER. The clinical diagnosis of MI was performed by an adjudication panel according to ESC/ACCF/AHA/WHF 2007 guidelines. The adjudication panel used existing clinical data including ECG and other clinical information to make the diagnosis, but not the final discharge diagnosis or the PATHFAST cTnl-II Test results. Clinical sensitivity and clinical specificity were calculated by comparing the PATHFAST cTnl-II Test results to clinical diagnosis assigned by the adjudication panel. The calculations were performed with both the 99th percentile cutoff (≥0.029 ng/mL) and the ROC cutoff (≥0.264 ng/mL) of the assay.
{9}------------------------------------------------
| | | #/TOTAL | 95% CI | |
|-----------------------|-----------|---------|-----------------|-----------------|
| Cutoff<br>0.029 ng/mL | 0 to 2 h | Sens | 73.6% (53/72) | (61.9% - 83.3%) |
| | | Spec | 92.7% (242/261) | (88.9% - 95.6%) |
| | 2 to 6 h | Sens | 93.1% (67/72) | (84.5% - 97.7%) |
| | | Spec | 93.1% (243/261) | (89.3% - 95.9%) |
| | 6 to 12 h | Sens | 91.7% (66/72) | (82.7% - 96.9%) |
| | | Spec | 91.6% (239/261) | (87.5% - 94.6%) |
| | | | #/TOTAL | 95% Cl* |
| Cutoff<br>0.264 ng/mL | 0 to 2 h | Sens | 23.6% (17/72) | (14.4% - 35.1%) |
| | | Spec | 99.2% (259/261) | (97.3% - 99.9%) |
| | 2 to 6 h | Sens | 62.5% (45/72) | (50.3% - 73.6%) |
| | | Spec | 98.1% (256/261) | (95.6% - 99.4%) |
| | 6 to 12 h | Sens | 80.6% (58/72) | (69.5% - 88.9%) |
| | | Spec | 97.7% (255/261) | (95.1% - 99.2%) |
### Limitations of procedure
- 1. The instrument reporting system contains error codes to warn the operator of specific malfunctions. Any reports slip containing such error codes should be kept for follow-up. See the PATHFAST operator's manual.
- 2. When using the 99th percentile cutoff, the PATHFAST cTnl-II Test should be interpreted with at least 2 serial samples. When using the ROC cutoff, the test should be interpreted with 3 serial samples. When samples are collected in the early hours, it is not advisable to use the higher cutoff.
- 3. Patient samples may contain heterophilic antibodies that could react in immunoassays to give a falsely high or low result. This assay has been designed to minimize interference from heterophilic antibodies. Nevertheless, complete elimination of this interference from all patient specimens cannot be guaranteed. A test result that is inconsistent with the clinical picture and patient history should be interpreted with caution.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.