VIDAS TROPONIN I (TNI), MODEL 30 445

K030950 · bioMerieux, Inc. · MMI · Jun 16, 2003 · Clinical Chemistry

Device Facts

Record IDK030950
Device NameVIDAS TROPONIN I (TNI), MODEL 30 445
ApplicantbioMerieux, Inc.
Product CodeMMI · Clinical Chemistry
Decision DateJun 16, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1215
Device ClassClass 2

Intended Use

VIDAS Troponin I Assay is an automated quantitative test for use on the VIDAS instruments for the determination of human cardiac troponin 1 in serum or plasma (lithium heparinate) using the ELFA (Enzyme-Linked Fluorescent Assay) technique

Device Story

VIDAS Troponin I (TNI) Assay is an automated enzyme-linked fluorescent immunoassay (ELFA) for quantitative detection of cardiac troponin I in human serum or plasma. The device utilizes a Solid Phase Receptacle (SPR) as both a solid phase for antigen capture and a pipettor. Sample and conjugate (alkaline phosphatase-labeled anti-cardiac troponin I antibodies) are cycled in and out of the SPR; troponin I binds to antibodies coating the SPR interior. After washing, a fluorescent substrate (4-methylumbelliferyl phosphate) is added; enzyme on the SPR wall catalyzes conversion to 4-methylumbelliferone. An optical scanner measures fluorescence intensity. The instrument computer automatically analyzes results, generates a test value, and prints a report. Used in clinical laboratory settings; operated by laboratory personnel. Provides quantitative troponin I levels to assist clinicians in cardiac assessment.

Clinical Evidence

Clinical testing compared 107 plasma samples and 32 serum samples between VIDAS TNI and the predicate (SCS Cardiac Troponin I). Results showed a correlation coefficient of 0.92 for both sample types. Nonclinical testing included precision (1.3-2.5% CV), analytical sensitivity (<0.01 µg/l), and interference studies showing no significant interference from bilirubin, hemoglobin, or triglycerides at specified concentrations.

Technological Characteristics

Automated ELFA system. Uses Solid Phase Receptacle (SPR) for antigen capture. Reagents contained in sealed strips. Detection via optical scanner measuring fluorescence of 4-methylumbelliferone produced by alkaline phosphatase-labeled antibodies. Quantitative assay for cardiac troponin I in serum/plasma.

Indications for Use

Indicated for the quantitative determination of human cardiac troponin I in serum or plasma (lithium heparinate) using the ELFA technique on VIDAS instruments.

Regulatory Classification

Identification

A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JUN 1 6 2003 Image /page/0/Picture/1 description: The image shows a handwritten string of characters, which appears to be an alphanumeric code. The code reads 'K030950'. The characters are written in a clear, legible style, with each digit and letter distinctly formed. The writing is dark and stands out against the background. # 510(k) SUMMARY ## VIDAS Troponin I (TNI) Assay #### A. Submitter Information: | Submitter's Name:<br>Address: | bioMérieux, Inc.<br>595 Anglum Road<br>Hazelwood, MO 63042 | |-------------------------------|------------------------------------------------------------| | Contact Person: | Nancy Weaver | | Phone Number: | 314-731-8695 | | Fax Number: | 314-731-8689 | | Date of Preparation: | January 17, 2003 | | B. Device Name: | | | Trade Name: | VIDAS Troponin I (TNI) Assay | | Common Name: | Enzyme-linked Fluorescent Immunoassay (ELFA) for | the quantitative detection of human cardiac troponin I. | Classification Name: | 21 CFR 862.1215,<br>Immunoassay method, Troponin Subunit | |----------------------|----------------------------------------------------------| |----------------------|----------------------------------------------------------| #### C. Predicate Device Name: Trade Name: Stratus CS Cardiac Troponin I Testpak #### D. Device Description: The VIDAS Troponin I (TNI) Assay is an enzyme-linked fluorescent immunoassay (ELFA) performed in an automated VIDAS® instrument. All assay steps and assay temperature are controlled by the instrument. A pipette tip-like disposable device, the Solid Phase Receptacle (SPR), serves as the solid phase as well as a pipettor for the assay. Reagents for the assay are in the sealed TNI Reagent Strips. The sample is transferred into the wells containing anti-cardiac troponin I antibodies labeled with alkaline phosphatase (conjugate). The sample/conjugate mixture is cycled in and out of the SPR for a specified length of time. Troponin I present in the specimen will #### bioMérieux, Inc. 595 Anglum Road, Hazelwood, Missouri 63042-2320, USA Phone: 314/731-8500 800/638-4835 Fax: 314/731-8700 http://www.biomerieux-usa.com {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the logo for bioMerieux. The logo consists of the company name in all caps, with a stylized sphere above it. A thin, curved line extends from the top and bottom of the sphere. bind to the anti-cardiac troponin I immunoglobulin coating the interior of the SPR. Unbound sample components are washed away. A fluorescent substrate, 4-methylumbelliferyl phosphate, is introduced into the SPR. Enzyme remaining on the SPR wall will catalyze the conversion of the substrate to the fluorescent product 4-methylumbelliferone. The optical scanner in the instrument measures the intensity of fluorescence. When the VIDAS TNI assay is completed, the results are analyzed automatically by the computer, a test value is generated, and a report is printed for each sample. #### E. Intended Use: VIDAS Troponin I Assay is an automated quantitative test for use on the VIDAS instruments for the determination of human cardiac troponin I in serum or plasma (lithium heparinate) using the ELFA (Enzyme-Linked Fluorescent Assay) technique. ### F. Technological Characteristics Summary: Major Similarities Include: - 1. Both tests are run on automated immunoassay systems. - 2. Both assays are quantitative tests for the detection of cardiac troponin I in human plasma. The VIDAS assay can also be used with serum. - 3. Both test use reagent test strips or packs. - 4. Both tests use mouse monoclonal antibodies for antigen capture and in the coniugate. - 5. Both assays use 4-methylumbelliferyl phosphate as a fluorescent substrate. #### Major Differences Include: - 1. The VIDAS TNI assay is a one-step immunoassay sandwich method. The SCS cTnl assay is two-site sandwich assay based on solid phase Radial Partition Immunoassay technology. - 2. The VIDAS TNI assay uses a Solid Phase Receptacle (SPR) to capture the antigen. The SCS cTnl assay uses glass fiber paper to capture the antigen. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for bioMerieux. The logo consists of the company name in all caps, with a stylized circle above it. A curved line bisects the circle and the company name. ### G. Performance Data: #### Nonclinical Testing: | | SCS cTnl | VIDAS TNI | |-------------------------------|-----------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------| | Within-run precision:<br>% CV | 2.7 - 4.3 | 1.3 - 2.5 | | Dilution % Recovery | 98.1 - 102.0 | 97 - 111 | | Analytical Sensitivity: | 0.03 µg/l | < 0.01 µg/l | | Expected Values | 95% of 101 apparently healthy<br>patients had values of 0.00 -<br>0.06 µg/l | 99.4% of 496 clinically healthy<br>patients had values below 0.1<br>µg/l | | Interference | No significant interference from:<br>Bilirubin 1026 µmol/L<br>Hemoglobin 1000 mg/dL<br>Triglycerides 3000 mg/dL | No significant interference from:<br>Bilirubin 491 µmol/L<br>Hemoglobin 300 µmol/L<br>Triglycerides 30 mg/ml | | Specificity: | Troponin-C (cardiac)<br>1000 µg/l:<br>0.01% | Troponin C<br>1000 µg/l:<br>< 0.001% | | | Troponin-T (cardiac)<br>1000 µg/l:<br>0.04% | Cardiac Troponin T<br>60 µg/l:<br>1.6% | | | Troponin-T (skeletal)<br>1000 µg/l:<br>0.04% | Skeletal Muscle Troponin T<br>1000 µg/l:<br>< 0.001% | | | Troponin-l (skeletal)<br>280 µg/l:<br>0.04% | Skeletal Muscle Troponin I<br>1000 µg/l:<br>< 0.001% | #### Clinical Testing: One hundred and seven plasma samples were tested with both the VIDAS TNI (Y) and the SCS Cardiac Troponin I (X). The results are as follows: Y = 1.016 X = 0.013 with a correlation coefficient of 0.92. Thirty-two serum samples were tested with both the VIDAS TNI and the SCS Cardiac Troponin I assay. The results are as follows: VIDAS TNI = 7.33 SCS Cardiac Troponin I Correlation coefficient = 0.92 #### H. Conclusion: The VIDAS Troponin I (TNI) Assay is substantially equivalent to the Stratus CS Cardiac Troponin I Testpak. #### bioMérieux, Inc. 595 Anglum Road, Hazelwood, Missouri 63042-2320, USA Phone: 314/731-8500 800/638-4835 Fax: 314/731-8700 http://www.biomerieux-usa.com {3}------------------------------------------------ Image /page/3/Picture/1 description: The image is a black and white seal for the U.S. Department of Health & Human Services. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of three human profiles facing right, with flowing lines above them that could represent hair or movement. JUN 1 6 2003 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Ms. Nancy Weaver Staff Regulatory Specialist BioMerieux, Inc. 595 Anglum Road Hazelwood, MO 63042-2320 Re: k030950 Trade/Device Name: VIDAS Troponin I Assay Regulation Number: 21 CFR 862.1215 Regulation Name: Creatine phosphokinase/creatine kinase or isoenzymes test system Regulatory Class: Class II Product Code: MMI Dated: March 12, 2003 Received: March 26, 2003 Dear Ms. Weaver: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {4}------------------------------------------------ #### Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Steven Putman Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: VIDAS Troponin I Assay Indications for Use: VIDAS Troponin I Assay is an automated quantitative test for use on the VIDAS instruments for the determination of human cardiac troponin 1 in serum or plasma (lithium heparinate) using the ELFA (Enzyme-Linked Fluorescent Assay) technique Sean Cooley Division Sign-Off Óffice of In Vitro Diagnostic Device Evaluation and Safety 510(k) K030950 ### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...