VITROS TROPONIN I ES ASSAY, INCLUDING REAGENT PACK, CALIBRATORS AND RANGE VERIFIERS, MODELS 680 2301, 2302 AND 2303
Applicant
Ortho-Clinical Diagnostics
Product Code
MMI · Clinical Chemistry
Decision Date
Dec 19, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1215
Device Class
Class 2
Indications for Use
VITROS® Troponin I ES Reagent Pack: For in vitro diagnostic use only. For the quantitative measurement of cardiac Troponin I (cTnI) in human serum and plasma (heparin and EDTA) using the VITROS Immunodiagnostic System, to aid in the assessment of myocardial damage and risk stratification. Cardiac Troponin I measurement aids in the diagnosis of acute myocardial infarction and in the risk stratification of patients with non-ST-segment elevation acute coronary syndromes with respect to relative risk of mortality, myocardial infarction (MI) or increased probability of ischemic events requiring urgent revascularization procedures. VITROS® Troponin I ES Calibrators: For in vitro diagnostic use only. For use in the calibration of the VITROS Immunodiagnostic System for the quantitative measurement of cardiac Troponin I (cTnI) in human serum and plasma (heparin and EDTA). VITROS® Troponin I ES Range Verifiers: For in vitro diagnostic use only. For in vitro use in verifying the calibration range of the VITROS Immunodiagnostic System when used for the quantitative measurement of cardiac Troponin I (cTnI).
Device Story
VITROS Troponin I ES assay is an in vitro diagnostic test for quantitative measurement of cardiac Troponin I (cTnI) in human serum and plasma. The device utilizes the VITROS Immunodiagnostic System, an automated instrumentation platform. The assay employs a chemiluminescence immunoassay principle with enzyme-labeled mouse monoclonal antibodies. The system processes patient samples (80µL) to provide cTnI concentration values. Healthcare providers use these results to assess myocardial damage, diagnose acute myocardial infarction, and perform risk stratification for patients with non-ST-segment elevation acute coronary syndromes. The device is intended for clinical laboratory use.
Clinical Evidence
Clinical study of 506 patients (289 male, 217 female, age 23-95) presenting with ACS symptoms. 73 diagnosed with AMI, 433 non-AMI. Using a 0.120 ng/mL cutoff, sensitivity was 94.5% and specificity was 93.1% (AUC 0.963). Risk stratification study (n=395) showed patients with baseline cTnI > 99th percentile (0.034 ng/mL) had significantly higher 30-day risk of death or recurrent ischemia (33.0% event rate vs 6.3% for ≤ 0.012 ng/mL; p < 0.0001).
Technological Characteristics
Chemiluminescence immunoassay; uses streptavidin-coated wells, biotinylated mouse monoclonal anti-cTnI, and HRP-labeled mouse monoclonal anti-cTnI conjugate. Employs Intellicheck Technology for system monitoring. Measuring range 0.012–80.0 ng/mL. Compatible with VITROS Immunodiagnostic System (ECi/ECiQ). Performance validated per CLSI guidelines (EP5-A, EP6-A, EP7-A, EP9-A, EP17-A, GP10-A, C28-A2).
Indications for Use
Indicated for quantitative measurement of cardiac Troponin I in human serum and plasma to aid in assessment of myocardial damage, diagnosis of acute myocardial infarction, and risk stratification of patients with non-ST-segment elevation acute coronary syndromes.
Regulatory Classification
Identification
A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
Predicate Devices
Beckman Access® AccuTnI Troponin I (k974075)
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k062838
B. Purpose for Submission:
New device
C. Measurand:
Troponin I
D. Type of Test:
Quantitative
E. Applicant:
Ortho-Clinical Diagnostics, Inc.
F. Proprietary and Established Names:
VITROS Immunodiagnostic Products Troponin I ES Reagent Pack
VITROS Immunodiagnostic Products Troponin I ES Calibrators
VITROS Immunodiagnostic Products Troponin I ES Range Verifiers
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| MMI | Class II | 21 CFR 862.1215, Immunoassay method, troponin subunit | 75, Clinical Chemistry (CH) |
| JIT | Class II | 21 CFR 862.1150, Calibrator, Secondary | 75, Clinical Chemistry (CH) |
| JJX | Class I | 21 CFR 862.1660, Single (specified) analyte controls (assayed and unassayed) | 75, Clinical Chemistry (CH) |
{1}
H. Intended Use:
1. Intended use(s):
VITROS Immunodiagnostic Products Troponin I ES Reagent Pack
For in vitro diagnostic use only.
For the quantitative measurement of cardiac Troponin I (cTnI) in human serum and plasma (heparin and EDTA) using the VITROS Immunodiagnostic System, to aid in the assessment of myocardial damage and risk stratification.
Cardiac Troponin I measurement aids in the diagnosis of acute myocardial infarction and in the risk stratification of patients with non-ST-segment elevation acute coronary syndromes with respect to relative risk of mortality, myocardial infarction (MI) or increased probability of ischemic events requiring urgent revascularization procedures.
VITROS Immunodiagnostic Products Troponin I ES Calibrators
For in vitro diagnostic use only.
For use in the calibration of the VITROS Immunodiagnostic System for the quantitative measurement of cardiac Troponin I (cTnI) in human serum and plasma (heparin and EDTA).
VITROS Immunodiagnostic Products Troponin I ES Range Verifiers
For in vitro diagnostic use only.
For in vitro use in verifying the calibration range of the VITROS Immunodiagnostic System when used for the quantitative measurement of cardiac Troponin I (cTnI).
2. Indication(s) for use:
See Intended Use section above
3. Special conditions for use statement(s):
Prescription use only
4. Special instrument requirements:
This device is intended for use with the VITROS Immunodiagnostic System which includes the ECi and ECiQ model analyzers.
I. Device Description:
One VITROS Troponin I ES Reagent Pack contains:
{2}
- 100 coated wells (streptavidin, binds ≥ 2 ng biotin/well)
- 7.0 mL conjugate reagent (HRP-mouse monoclonal anti-cTnI, binds ≥ 205 ng cTnI/mL) in buffer with bovine serum albumin and antimicrobial agent.
- 7.0 mL biotinylated antibody reagent (biotin-mouse monoclonal anti-cTnI, binds ≥ 205 ng cTnI/mL) in buffer with horse serum, bovine serum albumin, bovine gamma globulin and antimicrobial agent.
VITROS Troponin I ES Calibrators contain: 2 sets of Troponin I ES calibrators 1, 2 and 3 (2 mL liquid, human cTnI in serum with antimicrobial agent); nominal values 0, 0.08, and 15 ng/mL (exact values encoded on the lot calibration card), lot calibration card, protocol card and 24 calibrator bar code labels (8 for each calibrator).
Human blood products provided as components of the VITROS Troponin I ES Calibrators have been obtained from donors who were tested individually and who were found to be negative for hepatitis B surface antigen, and for antibodies to human immunodeficiency virus (HIV 1+2) and hepatitis C virus (HCV), using approved methods (enzyme immunoassays).
VITROS Troponin I ES Range Verifiers contain: 2 sets of VITROS Troponin I ES Range Verifiers, low and high (2 mL liquid, human cTnI in serum with antimicrobial agent); nominal values 0, 76 ng/mL.
Human blood products provided as components of the VITROS Troponin I ES Range Verifiers have been obtained from donors who were tested individually and who were found to be negative for hepatitis B surface antigen and for antibodies to human immunodeficiency virus (HIV 1+2) and hepatitis C virus (HCV), using approved methods (enzyme immunoassays, EIA).
J. Substantial Equivalence Information:
1. Predicate device name(s): Beckman Access® AccuTnI Troponin I
2. Predicate K number(s): k974075
3. Comparison with predicate
{3}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Principle | Chemiluminescence | Chemiluminescence |
| Antibody | Mouse monoclonal | Mouse monoclonal |
| Sample Type | serum and plasma | serum and plasma |
| Intended Use | For the quantitative measurement of cardiac Troponin I (cTnI) in human serum and plasma to aid in the assessment of myocardial damage and risk stratification. Cardiac Troponin I measurement aids in the diagnosis of acute myocardial infarction and in the risk stratification of patients with non-ST-segment elevation acute coronary syndromes with respect to relative risk of mortality, myocardial infarction (MI) or increased probability of ischemic events requiring urgent revascularization procedures. | For the quantitative determination of cardiac Troponin I (cTnI) levels in human serum and plasma to aid in the diagnosis and treatment of myocardial infarction and cardiac muscle damage. Cardiac troponin I determination aids in the risk stratification of patients with unstable angina or non-ST segment elevation acute coronary syndromes with respect to relative risk of mortality, myocardial infarction or increased probability of ischemic events requiring urgent revascularization procedures. |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Measuring range | 0.012 - 80.0 ng/mL | 0.01 – 100 ng/mL |
| Hook Effect | No high dose hook up to 14,000 ng/mL | No high dose hook up to 1,920 ng/mL |
K. Standard/Guidance Document Referenced (if applicable):
| STANDARDS |
| --- |
| Title and Reference Number |
| CLSI - Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline (EP9-A 1995) |
| CLSI - Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guideline (EP5-A) |
| CLSI - Interference Testing in Clinical Chemistry; Approved Guideline (EP7-A) |
| CLSI - Evaluation of Linearity; Approved Guideline (EP6-A) |
| CLSI - Limits of Detection and Quantitation; Approved Guideline (EP17-A) |
| CLSI - Assessment of the Clinical Accuracy of Laboratory Tests Using Receiver |
{4}
5
Operating Characteristic (ROC) Plots; Approved Guideline (GP10-A)
CLSI – How to Define and Determine Reference Intervals in the Clinical Laboratory; Approved Guideline (C28-A2)
L. Test Principle:
The VITROS Troponin I ES assay is performed using the VITROS Troponin I ES Reagent Pack and the VITROS Troponin I ES Calibrators on the VITROS Immunodiagnostic System using Intellicheck® Technology. An immunometric immunoassay technique is used, which involves the simultaneous reaction of cardiac Troponin I present in the sample with a biotinylated antibody (mouse monoclonal anti-cTnI) and a horseradish peroxidase (HRP)-labeled antibody conjugate (mouse monoclonal anti-cTnI). The antigen-antibody complex is captured by streptavidin on the wells. Unbound materials are removed by washing. The bound HRP conjugate is measured by a luminescent reaction. A reagent containing luminogenic substrates (a luminol derivative and a peracid salt) and an electron transfer agent, is added to the wells. The HRP in the bound conjugate catalyzes the oxidation of the luminol derivative, producing light. The electron transfer agent (a substituted acetanilide) increases the level of light produced and prolongs its emission. The light signals are read by the VITROS Immunodiagnostic System. The amount of HRP conjugate bound is directly proportional to the concentration of cTnI present.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Precision was evaluated consistent with CLSI document EP5. Two replicates of each of 2 frozen control sera and 4 frozen human sample pools were assayed on 2 separate occasions per day on at least 20 different days. The experiment was performed using 3 reagent lots on 3 different VITROS Immunodiagnostic Systems.
{5}
| | Units = ng/mL (μg/mL) | | | | | | | No. of Observations | No. of Days |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Mean Troponin I Conc. | Within-run* | | Within-calibration** | | Within-lab*** | | | |
| | | SD | CV (%) | SD | CV (%) | SD | CV (%) | | |
| System 1 | 0.024 | 0.001 | 4.2 | 0.003 | 12.5 | 0.002 | 8.3 | 84 | 29 |
| | 0.063 | 0.001 | 1.6 | 0.003 | 4.8 | 0.003 | 4.7 | 84 | 29 |
| | 0.091 | 0.001 | 1.1 | 0.003 | 3.3 | 0.003 | 3.3 | 84 | 29 |
| | 0.413 | 0.016 | 4.0 | 0.017 | 4.2 | 0.018 | 4.3 | 84 | 29 |
| | 5.23 | 0.050 | 1.0 | 0.112 | 2.2 | 0.106 | 2.0 | 84 | 29 |
| | 56.3 | 0.810 | 1.5 | 1.75 | 3.2 | 1.49 | 2.6 | 84 | 29 |
| System 2 | 0.028 | 0.001 | 3.7 | 0.003 | 11.1 | 0.003 | 10.7 | 84 | 29 |
| | 0.064 | 0.001 | 1.6 | 0.004 | 6.3 | 0.004 | 6.2 | 84 | 29 |
| | 0.092 | 0.001 | 1.1 | 0.004 | 4.4 | 0.004 | 4.3 | 84 | 29 |
| | 0.433 | 0.005 | 1.2 | 0.011 | 2.6 | 0.015 | 3.4 | 84 | 29 |
| | 5.58 | 0.047 | 0.9 | 0.117 | 2.1 | 0.139 | 2.5 | 84 | 29 |
| | 61.4 | 0.618 | 1.0 | 1.82 | 3.0 | 1.76 | 2.8 | 84 | 29 |
| System 3 | 0.027 | 0.001 | 3.7 | 0.002 | 7.4 | 0.003 | 11.1 | 84 | 29 |
| | 0.059 | 0.002 | 3.4 | 0.004 | 6.8 | 0.005 | 8.5 | 84 | 29 |
| | 0.086 | 0.001 | 1.2 | 0.005 | 5.9 | 0.006 | 7.0 | 84 | 29 |
| | 0.393 | 0.005 | 1.3 | 0.011 | 2.8 | 0.016 | 4.0 | 84 | 29 |
| | 5.05 | 0.065 | 1.3 | 0.101 | 2.0 | 0.183 | 3.6 | 84 | 29 |
| | 62.8 | 0.823 | 1.3 | 1.75 | 2.8 | 2.96 | 4.7 | 84 | 29 |
* Within-run (Repeatability): Between-duplicate precision was determined using duplicate determinations.
** Within-calibration: Total within calibration precision was determined using a single lot of reagent over a single calibration interval.
***Within-lab: Total within-lab precision was estimated using a single reagent lot calibrated weekly.
To assess the low-end precision, eight low troponin I serum sample pools were tested in singleton, twice a day on 20 occasions over 28 days using 2 lots of reagent. From the resulting imprecision profiles, the coefficient of variation (CV) at the $99^{\text{th}}$ percentile URL of $0.034 \, \text{ng/mL}$ was $10\%$ . This $\%$ CV was determined using a combined mean value using both lots tested. This demonstrates that the VITROS Troponin I ES assay meets the recommendations as described by the ESC/ACC that the assay should have the imprecision (total CV) at the $99^{\text{th}}$ percentile cutoff of $10\%$ or less.
## b. Linearity/assay reportable range:
Two buffer based pools with mean analyte concentrations of $0.011\mathrm{ng / mL}$ and $83.9\mathrm{ng / mL}$ respectively, were mixed in varying proportions to give a total of 11 pools. A total of 30 singleton determinations of pools 1 and 11, and 9 singleton determinations of pools 2 to 10 were made using 3 Master Lots. Calculated concentrations of pools 2 to 10 were determined from the mean measured concentrations of pools 1 and 11. These calculated concentrations were then compared with the mean measured concentrations of pools 2 to 10. The mean measured concentrations of pools 2 to 10, when expressed as a percentage of the calculated concentrations, gave a mean result of $97.5\%$ , with a range of $96.4\%$ to $99.6\%$ . Analysis by linear regression indicated that the
{6}
assay is linear across the range tested (0.011 to 83.9 ng/mL) with an R² of 0.9996.
To assess low end linearity, two buffer based pools with mean analyte concentrations of 0.002 ng/mL and 0.155 ng/mL respectively, were mixed in varying proportions to give a total of 11 pools. A total of 30 singleton determinations of pools 1 and 11, and 9 singleton determinations of pools 2 to 10 were made using 3 Master Lots. Calculated concentrations of pools 2 to 10 were determined from the mean measured concentrations of pools 1 and 11. These calculated concentrations were then compared with the mean measured concentrations of pools 2 to 10. The mean measured concentrations of pools 2 to 10, when expressed as a percentage of the calculated concentrations, gave a mean result of 103%. The mean measured concentrations of pools 3 to 10, when expressed as a percentage of the calculated concentrations, gave a mean result of 106%. Analysis by linear regression also indicated that the assay is linear across the range of 0.015 to 0.155 ng/mL with an R² of 0.9955.
Hook Effect:
A series of samples having Troponin I concentrations ranging from 1.44 to 14,442 ng/mL were assayed in the VITROS Troponin I ES assay. Singleton determinations of each sample were made with each of 3 Master Lots. The high dose hook of the VITROS Troponin I ES assay was shown to be greater than 14,442 ng/mL.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The calibration for the VITROS Troponin I ES assay is defined by a set of Reference Calibrators. The Reference Calibrators are calibrated against a commercially available Troponin I ES assay. The Reference Calibrators are used to construct a number of Reference Calibration Curves (RCCs) from which a Master Curve may be established. This was used to determine the Troponin I ES concentration in the Calibrators for each Master Lot, as well as providing the Master Curve data for the Lot Calibration card.
To establish the open-stability of the VITROS Troponin I ES Range Verifiers, the verifiers were pooled and stored in sample cups at 2-8°C. The pooled Range Verifiers were tested at the initial time point (day 0) and days 4 and 7 of the trial using three Master Kit Lots and three lots of VITROS Troponin I ES Range Verifiers. The data supports the storage of the VITROS Troponin I ES Range Verifiers at 2-8°C for up to 7 days.
To establish the open-stability of the VITROS Troponin I ES Calibrators, the calibrators were opened, pooled and stored in sample cups at 2-8°C. The pooled calibrators were tested at the initial time point. The calibrators stored at 2-8°C were tested at weeks 2, 4, 8 and 10. The data supports the storage of the calibrators at 2-8°C after opening for up to 10 weeks.
7
{7}
8
d. Detection limit:
The limit of detection (LoD) for the VITROS Troponin I ES assay using human serum pools is 0.012 ng/mL, determined consistent with CLSI document EP17 and with proportions of false positives less than 1% and false negatives less than 1%; based on 600 determinations, with 150 blank and 450 low-level samples. The limit of quantitation (LoQ) is 0.012 ng/mL as determined by the lowest concentration at which precision and accuracy design requirements are still met and are within the linear range of the assay.
e. Analytical specificity:
The VITROS Troponin I ES assay was evaluated for interference consistent with CLSI document EP7. Commonly encountered substances were tested on 3 lots of reagents. Of the compounds tested, hemoglobin may interfere with the VITROS Troponin I ES assay resulting in a positive bias as shown in the table below.
| Interferent | Interferent Concentration | | Units = ng/mL (μg/L)
Analyte Conc*
Bias** | |
| --- | --- | --- | --- | --- |
| Hemoglobin | 0.062 mmol/L | 100 mg/dL | 0.006 | 0.003 |
| Hemoglobin | 0.124 mmol/L | 200 mg/dL | 0.006 | 0.030 |
| Hemoglobin | 0.186 mmol/L | 300 mg/dL | 0.006 | 0.031 |
| Hemoglobin | 0.248 mmol/L | 400 mg/dL | 0.006 | 0.033 |
| Hemoglobin | 0.310 mmol/L | 500 mg/dL | 0.006 | 0.034 |
* Average test concentration of replicate determinations using three different lots of reagent, concentration below the measuring range.
** Estimate of the average difference observed.
The VITROS Troponin I ES assay was evaluated for interference consistent with CLSI document EP7. Of the compounds tested, none was found to cause a bias of >10% with the assay at the concentrations indicated at clinically relevant cTnI concentrations of 0.400 ng/mL
{8}
| Compound | Concentration | Tested |
| --- | --- | --- |
| Acetaminophen | 1324 μmol/L | 200 μg/mL |
| Albumin | na | 8 g/dL |
| Allopurinol | 2740 μmol/L | 400 μg/mL |
| Ambroxol | 1050 μmol/L | 400 μg/mL |
| Ampicillin | 144 μmol/L | 50 μg/mL |
| Ascorbic acid | 342 μmol/L | 60 μg/mL |
| Aspirin | 3.62 mmol/L | 650 μg/mL |
| Atenolol | 38 μmol/L | 10 μg/mL |
| Bilirubin | 342 μmol/L | 20 mg/dL |
| Biotin | 10.2 nmol/L | 0.25 μg/dL |
| Caffeine | 515 μmol/L | 100 μg/mL |
| Captopril | 230 μmol/L | 50 μg/mL |
| Cinnarizine | 1090 μmol/L | 400 μg/mL |
| Cocaine | 33 μmol/L | 10 μg/mL |
| Diclofenac | 169 μmol/L | 50 μg/mL |
| Digoxin | 7.8 nmol/L | 0.61 μg/mL |
| Dipyrone | 30 mmol/L | 1000 mg/dL |
| Dopamine | 4250 μmol/L | 650 μg/mL |
| Erythromycin | 273 μmol/L | 200 μg/mL |
| Fibrinogen | na | 1000 mg/dL |
| na = not applicable | | |
| Compound | Concentration | Tested |
| --- | --- | --- |
| Furosemide | 1210 μmol/L | 400 μg/mL |
| Ibuprofen | 2425 μmol/L | 500 μg/mL |
| Low MW Heparin | na | 5 U/mL |
| Methyldopa | 105 μmol/L | 25 μg/mL |
| Nifedipine | 173 μmol/L | 60 μg/mL |
| Nitrofurantoin | 269 μmol/L | 64 μg/mL |
| Nystatin | na | 7.0 μg/mL |
| Oxytetracycline | 10.9 μmol/L | 5.0 μg/mL |
| Phenytoin | 365 μmol/L | 100 μg/mL |
| Propranolol | 24 μmol/L | 5.0 μg/mL |
| Quinidine | 62 μmol/L | 20 μg/mL |
| Sodium Azide | 3.08 mmol/L | 200 μg/mL |
| Sodium Heparin | na | 8 U/mL |
| Streptokinase | na | 1.96U/mL |
| Theophylline | 1380 μmol/L | 250 μg/mL |
| t-PA | na | 2.3 μg/mL |
| Trimethoprim | 258 μmol/L | 75 μg/mL |
| Triolein | 33.9 mmol/L | 3000 mg/dL |
| Verapamil | 353 μmol/L | 160 μg/mL |
| Warfarin | 32.4 μmol/L | 10 μg/mL |
The cross-reactivity of the VITROS Troponin I ES assay was evaluated by adding the following substances to samples containing no analyte.
| Cross-reactant Tested | Concentration Tested | Mean cTnI Result of Cross-reactant Pool Units=ng/mL | % Cross-reactivity |
| --- | --- | --- | --- |
| Skeletal Troponin I | 1000 ng/mL | 0.082 | 0.008 |
| Cardiac Troponin C (Recombinant) | 1000 ng/mL | 0.017 | 0.002 |
| Cardiac Troponin T (Recombinant) | 1000 ng/mL | ND* | ND* |
| Actin (from Rabbit Muscle) | 1000 ng/mL | ND* | ND* |
| Myosin (Recombinant) | 1000 ng/mL | ND* | ND* |
| Tropomyosin (from porcine muscle) | 1000 ng/mL | ND* | ND* |
| CK-MB (Recombinant) | 1000 ng/mL | ND* | ND* |
| Myoglobin (Recombinant) | 1000 ng/mL | ND* | ND* |
*ND = Not Detectable. Concentration was below the measuring range of the assay.
Cross-reactivity was expressed as the mean result obtained for the cross-reactant pool divided by the cross-reactant concentration in percentage term.
f. Assay cut-off:
The 99th Percentile Upper reference Limit (URL) is 0.034 ng/mL.
The VITROS Troponin I ES assay AMI diagnostic cutoff is 0.120 ng/mL. See Clinical Cut-off section below for more details.
{9}
10
2. Comparison studies:
a. Method comparison with predicate device:
A method comparison was performed based on CLSI document EP9 using samples analyzed on the VITROS Immunodiagnostic System with those analyzed using the Beckman Unicel™ DxI 800 Access Immunoassay System. The relationship between the two methods was determined by Deming regression. VITROS Troponin I ES = 0.9790 x Beckman Access AccuTnI + 0.2191 ng/mL with samples ranging from 0.015 to 73.5 ng/mL (n = 260).
b. Matrix comparison:
Individual results from 20 donors using three determinations of each sample type (serum, EDTA plasma, and heparin plasma) and after storage at 2-8° C for 5 and 7 days and after 4 weeks at -20 ° C were tabulated. The mean value for each specimen type within each storage condition was calculated from the three determinations. The overall mean, SD and CV (%) for each sample type across storage conditions were also calculated. For each condition (sample type or storage), % differences were calculated from the baseline (serum for specimen type and fresh for storage condition). The mean and range of the % differences, across all samples, for each condition were calculated. The results indicated that all sample types are suitable (< 10% mean percent difference) for use in the VITROS Troponin I ES assay and that their suitability is not affected by storage.
3. Clinical studies:
a. Clinical Sensitivity:
Samples were collected and analyzed at two independent clinical sites for sensitivity and specificity. Samples were collected at the time intervals indicated in the chart below, determined from the time of admission. Sample results were generated using the VITROS Troponin I ES assay. Troponin I concentrations were evaluated in serial samples collected from a combined total of 506 prospectively enrolled individuals presenting with symptoms of Acute Coronary Syndrome. This patient population consisted of 289 males and 217 females ranging in age from 23 to 95. A total of 73 patients were diagnosed as AMI positive and 433 were diagnosed as non-AMI according to ACC/ESC/AHA criteria. A total of 1696 plasma and serum samples were collected from the 506 patients from the two clinical sites. Site 1 provided a total of 1393 samples. Site 2 provided a total of 303 samples. The data are presented in the table below:
{10}
11
| | Hours Post Admission | | | |
| --- | --- | --- | --- | --- |
| | | 0-6 hrs | 6-12 hrs | 12-24 hrs |
| VITROS Troponin I ES Assay (AMI cutoff = 0.120 ng/mL) | % Sensitivity | 70 (86/123) | 89 (78/88) | 90 (43/48) |
| | % Specificity | 96 (683/711) | 94 (420/447) | 94 (206/220) |
## Sensitivity/Specificity (Combined Data using Peak Sample from each Serial Draw) Sites 1 and 2: VITROS Troponin I ES
| N | 506 | | |
| --- | --- | --- | --- |
| Cut-Off value | >0.12 | | |
| | With AMI | Without AMI | Total |
| Positive Result | 69 | 31 | 100 |
| Negative Result | 4 | 402 | 406 |
| Total | 73 | 433 | 506 |
| Sensitivity | 94.52 % | | |
| Specificity | 93.08 % | | |
## b. Clinical specificity:
See Clinical Sensitivity section
## c. Other clinical supportive data (when a. and b. are not applicable):
### Risk Stratification
In a prospective study, 395 patients with myocardial ischemia symptoms suggestive of Acute Coronary Syndrome (ACS) were hospitalized in one medical center to rule in or rule out myocardial infarction (MI). Serial blood samples were obtained from each patient and tested using the VITROS Troponin I ES assay. These patients were followed up after baseline cTnI testing and monitored for short-term adverse cardiac events, which include re-hospitalization due to recurring ischemia, MI and all cause death.
The results were analyzed using the 99th percentile upper reference limit (0.034 ng/mL), as recommended by the ESC/ACC/AHA consensus document for the redefinition of MI.
The findings from this study verified previous clinical studies that showed patients with minor increases in cTnI values provide prognostic information about the long and short-term risk of death for patients with ACS. The data collected from this study indicate that patients with baseline cardiac troponin I values above the 99th percentile upper reference limit had a significantly higher short-term risk of death or recurrent ischemic events after presentation.
{11}
| 30 Day Cardiac Event including Cardiac Death (n = 395) | | | | |
| --- | --- | --- | --- | --- |
| | N | # Events | Cumulative Event Rate | P-Value |
| Troponin I ES | | | | p < 0.0001 |
| ≤ 0.012 | 161 | 10 | 6.3% | |
| 0.013 - 0.034 | 121 | 13 | 10.8% | |
| > 0.034 | 113 | 37 | 33.0% | |

30 Day Cardiac Event Rate by ES cTnI $(n = 395)$
# 4. Clinical cut-off:
The 99th Percentile Upper Reference Limit (URL) is $0.034\mathrm{ng / mL}$ . This value was based on 21 estimates of the URL using over 10,000 serum, heparin and EDTA plasma samples from individual donors and included variation from raw materials, analyzer, operator, manufacturing processes and reagent age. The URL of $0.034\mathrm{ng / mL}$ was selected as the highest value determined from the multiple determinations of the $99^{\text{th}}$ percentile. The $99^{\text{th}}$ percentile values from the multiple studies ranged from 0.012 to $0.034\mathrm{ng / mL}$ .
The VITROS Troponin I ES assay AMI diagnostic cutoff is $0.120\mathrm{ng / mL}$ . A clinical study was performed consistent with CLSI document GP10. The study was conducted at two external clinical sites using prospectively collected, serially drawn specimens. A total of 506 patients with symptoms of acute coronary syndrome (ACS) were diagnosed as AMI or non-AMI according to the ESC/ACC/AHA criteria. 73 patients were diagnosed as AMI and 433 patients
{12}
were diagnosed as non-AMI. The peak sample from each patient's serial draw was obtained and used in the analysis. The peak cTnI result is defined as the highest cTnI concentration observed in the serial draw obtained from each patient. The Receiver Operator Curve (ROC) compares clinical sensitivity and specificity at various decision thresholds. The optimum decision threshold maximizes the area under the curve (AUC) and represents the highest sensitivity and specificity for the assay. The AUC for the VITROS Troponin I ES assay was 0.963. Using the cutoff of 0.120 ng/mL cTnI, the sensitivity of the VITROS Troponin I ES assay was 95% (95% CI of 87%–99%) and the specificity was 93% (95% CI of 90%–95%).
5. Expected values/Reference range:
See Clinical cut-off section
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
13
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.