K042227 · Alfa Scientific Designs, Inc. · MMI · Nov 19, 2004 · Clinical Chemistry
Device Facts
Record ID
K042227
Device Name
INSTANT-VIEW TROPONIN I TEST
Applicant
Alfa Scientific Designs, Inc.
Product Code
MMI · Clinical Chemistry
Decision Date
Nov 19, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1215
Device Class
Class 2
Intended Use
The Instant-View® Troponin I Test is an immunoassay for the rapid qualitative detection of cardiac troponin I (cTnI) in human whole blood or serum at a cutoff level of 1.5 ng/mL. It provides an aid in the diagnosis of myocardial infarction in emergency room, point-of-care, and hospital setting. The Instant-View® Troponin I Test provides a qualitative result rather than information about change in the level of cTnI with single testing. Serial testing should be performed to determine a temporal change in the level of cTnI. If desired, a quantitative method should be used to quantitate the concentration of cTnI. Clinical consideration and professional judgement should be applied when making a diagnosis decision based on this test result.
Device Story
Lateral flow immunoassay for qualitative detection of cardiac troponin I (cTnI) in human whole blood or serum; utilizes double antibody sandwich principle. Input: whole blood or serum sample applied to cassette or dip strip. Mechanism: colloidal gold-labeled mouse anti-cTnI conjugate binds cTnI; captured by immobilized mouse anti-cTnI on nitrocellulose membrane; burgundy-colored test line indicates presence of cTnI ≥ 1.5 ng/mL; control line confirms adequate sample volume and migration. Used in emergency rooms, point-of-care, and hospital settings by healthcare professionals. Output: visual qualitative result (presence/absence of test line). Results aid clinical diagnosis of myocardial infarction; serial testing required for temporal monitoring. Does not provide quantitative concentration; clinical judgment required for interpretation.
Clinical Evidence
Performance evaluated using 300 clinically confirmed serum specimens (150 positive, 150 negative). Sensitivity reported at 99.3% (149/150) and specificity at 96.0% (144/150), with 97.7% overall accuracy. Reproducibility tested at four sites (three physician offices, one medical lab) using 80 samples at four concentration levels (0, 0.1, 1.5, 10 ng/ml), showing >98.8% agreement. Interference and cross-reactivity studies confirmed no interference from common drugs/analytes (10 ug/ml) or cross-reactivity with other troponins, bilirubin, or cholesterol.
Technological Characteristics
One-step lateral flow chromatographic immunoassay. Components: nitrocellulose membrane, conjugate pad with colloidal gold-labeled anti-cTnI antibodies. Test line coated with anti-cTnI antibodies; control line coated with goat anti-mouse IgG. Formats: cassette (dip strip in plastic housing) or dip strip. Qualitative detection at 1.5 ng/ml cutoff. Standalone device; no energy source required. Manual operation.
Indications for Use
Indicated for rapid qualitative detection of cardiac troponin I (cTnI) in human whole blood or serum at a cutoff of 1.5 ng/mL to aid in myocardial infarction diagnosis in emergency, point-of-care, and hospital settings.
Regulatory Classification
Identification
A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
Predicate Devices
VBL Serum Troponin I Test by Vancouver Biotech LTD. (K023505)
Related Devices
K041619 — BIO CHECK WHOLE BLOOD/PLASMA/SERMON MODEL #801116 · Biocheck, Inc. · Dec 17, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
K042227
| NOV 19 2004 | | 510(k) Summary of<br>Safety and Effectiveness as Required by 21 CFR 807.92 |
|------------------------------|----------------------------|------------------------------------------------------------------------------------------------|
| Manufacture and<br>Submitter | Name: | Alfa Scientific Designs, Inc. |
| | Address: | 12330 Stowe Drive<br>Poway, CA 92064<br>Telephone: (858) 513-3888 x 308<br>Fax: (858) 513-8388 |
| | Contact Person:<br>E-mail: | Naishu Wang, MD, Ph.D.<br>wnss@alfascientific.com |
| | Device Name | Trade Name: |
| | | Common Name: |
| Classification: | | |
| | Creatine phosphokinase/creatine kinase or isoenzymes<br>test system |
|---------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date of Summary<br>Preparation | August 12, 2004 |
| Predicate Device | VBL Serum Troponin I Test by Vancouver Biotech LTD.<br>K023505 |
| Device Description | A one-step lateral flow chromatographic immunoassay. The test<br>strip in the device consists of 1) a burgundy-colored conjugate pad<br>containing colloidal gold coupled with anti- Troponin I antibodies,<br>and 2) nitrocellulose membrane containing a test line (T line) and a<br>control line (C line). The T line is coated with anti- Troponin I<br>antibodies, and the C line is coated with goat anti-mouse IgG<br>antibodies. |
| Summary of the<br>Similarity to the<br>Predicate Device | • Both are one-step lateral-flow chromatographic immunoassays.<br>• Both are qualitative tests.<br>• Both are in-vitro diagnostic devices.<br>• Both have a built-in quality control feature, C line, to indicate that<br>an adequate volume of sample is applied and the liquid flow<br>occurred properly |
{1}------------------------------------------------
| Intended Use | The INSTANT-VIEW Troponin I Test is an immunoassay for the rapid<br>qualitative detection of cardiac troponin I (cTnI) in human whole<br>blood or serum at a cutoff level of 1.5 ng/ml. It provides an aid in<br>the diagnosis of myocardial infarction in emergency room, point-of-<br>care, and hospital setting. |
|-----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sensitivity and<br>Specificity Study | The sensitivity and specificity of the device was evaluated against<br>300 clinically confirmed serum specimens, 150 positive and 150<br>negative. The results demonstrated that the INSTANT-VIEW® Troponin<br>I Test have a sensitivity of 99.3% (149/150) and a specificity of<br>96.0% (144/150). The overall accuracy of this device is 97.7%<br>(293/300). |
| Reproducibility<br>study | Reproducibility studies were performed on 80 samples with the<br>concentration of human cardiac troponin I evenly distributed at four<br>different levels: 0, 0.1, 1.5, 10 ng/ml. Each sample was tested at<br>three physician's office laboratories (POLs) and one medical<br>analysis laboratory. The agreements at the four sites were over<br>98.8%. |
| Interference and<br>Cross-reactivity<br>Study | No cross-reaction was observed with the closely related substances,<br>such as troponins other than human cardiac troponin bilirubin, and<br>cholesterols.<br>No interference was observed with endogenous substances including<br>commonly used drugs or analytes at a concentration of 10 ug/ml. |
| Formats of the<br>Device | The proposed device has two formats: Serum Test and Whole<br>Blood/Serum Test. The Serum Test has two sub-formats: Cassette<br>and Dip Strip. The Whole Blood/Serum Test only has the cassette<br>format. A cassette is a device that assembles a dip-strip in a plastic<br>housing. The studies demonstrate all the formats are equivalent. |
| Conclusion | The results of specificity, sensitivity, reproducibility, cross<br>reactivity, and interference studies demonstrate that the INSTANT-<br>VIEW Troponin I Test is substantially equivalent to the predicate<br>device. |
:
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services (HHS). The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The caduceus is depicted with three intertwined strands and a wing-like structure at the top. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
NOV 1 9 2004
Naishu Wang, MD., Ph.D. President Alfa Scientific Designs, Inc. 12330 Stowe Drive Poway, CA 92064
k042227 Re: Trade/Device Name: INSTANT-VIEW® Troponin I Serum Test (Cassette) INSTANT-VIEW® Troponin I Serum Test (Dip Strip) INSTANT-VIEW® Troponin I Whole Blood/Serum Test (Cassette) Regulation Number: 21 CFR 862.1215 Regulation Name: Creatine phosphokinase/creatine kinase or isoenzymes test system Regulatory Class: Class II Product Code: MMI Dated: October 28, 2004 Received: November 4, 2004
Dear Dr. Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21. Code of Federal Regulations (CFR). Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{3}------------------------------------------------
## Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Cornelius B. Loker
Cornelia B. Rooks, MA Acting Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): K04 2227
Device Name: INSTANT-VIEW® Troponin I Serum Test (Cassette) INSTANT-VIEW® Troponin I Serum Test (Dip Strip) INSTANT-VIEW® Troponin I Whole Blood/Serum Test (Cassette)
Indications For Use:
The INSTANT-VIEW Troponin I Test is an immunoassay for the rapid qualitative detection of cardiac troponin I (cTnI) in human whole blood or serum at a cutoff level of 1.5 ng/ml. It provides an aid in the diagnosis of myocardial infarction in emergency room, point-of-care, and hospital setting.
The INSTANT-VIEW Troponin I Test provides a qualitative result rather than information about change in the level of cTnI with single testing. Serial testing should be performed to determine a temporal change in the level of cTnI. If desired, a quantitative method should be used to quantitate the concentration of cTnI. Clinical consideration and professional judgment should be applied when making a diagnosis decision based on this test result.
Prescription Use_ ✓ (Pert 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Carol Beman
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K042227
Page 1 of &
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.