Wearable Device ; Cradle ; Small Strap, Medium Strap, Large Strap ; Multiple Device Charger ; Hospital Starter Package
Applicant
Snap40, Ltd.
Product Code
MSX · Cardiovascular
Decision Date
Nov 14, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Indications for Use
snap40 is intended for reusable bedside, mobile and central multi-parameter, continuous and intermittent physiologic patient monitoring of adult patients in environments where patient care is provided by trained healthcare professionals. snap40 is intended to provide visual and audible physiologic multi-parameter alarms. snap40 is intended for temperature monitoring temperature at the upper arm is clinically indicated. snap40 continuously monitors the following parameters in adults: - Pulse rate - Oxygen saturation - Temperature - Movement snap40 is intended for intermittent or spot-check monitoring of respiration rate. snap40 is not intended for use in high-acuity environments, such as ICU or operating rooms or for use in the home snap40 is not intended for use on acutely ill cardiac patients with the potential to develop life threatening arrhythmias e.g. very fast atrial fibrillation. For these patients, they should be monitored using a device with continuous ECG. snap40 is not a substitute for an ECG monitor. snap40 is not intended for SpO2 monitoring in conditions of high motion or low perfusion.
Device Story
Wearable device worn on upper arm; monitors pulse rate, SpO2, temperature, and movement continuously; respiration rate intermittently. Inputs: photoplethysmography (SpO2/pulse), thermistor (temperature), accelerometer/gyroscope (movement/respiration). Data transmitted via WiFi to software platform; provides visual/audible alarms on mobile devices and central station. Used in general medical/surgical hospital units by trained healthcare professionals. Output: vital sign parameters and alarms. Benefits: enables remote/centralized monitoring of patient status in non-critical areas; facilitates timely clinical intervention. Not for high-acuity settings or home use; not a substitute for ECG.
Clinical Evidence
Clinical studies conducted on human subjects. SpO2 accuracy validated in controlled desaturation study (n=12) per IEC 80601-2-61, demonstrating +/- 2 digits accuracy (70-100% range). Respiration rate accuracy validated against end-tidal CO2 (n=37) across various postures, demonstrating +/- 1 breath per minute accuracy (6-60 rpm range).
Indicated for adult patients in hospital environments requiring continuous monitoring of pulse rate, SpO2, temperature, and movement, or intermittent/spot-check monitoring of respiration rate. Contraindicated for high-acuity environments (ICU, OR), home use, acutely ill cardiac patients prone to life-threatening arrhythmias (e.g., fast atrial fibrillation), and SpO2 monitoring during high motion or low perfusion.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 14, 2018
snap40 Ltd Edwin Lindsay QA/RA Manager 24 Forth Street Edinburgh, EH1 3LH Gb
Re: K182543
Trade/Device Name: Wearable Device; Cradle; Small Strap, Medium Strap, Large Strap; Multiple Device Charger; Hospital Starter Package Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MSX, DQA, BZQ, FLL, DRG Dated: September 17, 2018 Received: September 20, 2018
Dear Edwin Lindsay:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Arielle Drummond -S
For
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K182543
Device Name snap40 System
Indications for Use (Describe)
snap40 is intended for reusable bedside, mobile and central multi-parameter, continuous and intermittent physiologic patient monitoring of adult patients in environments where patient care is provided by trained healthcare professionals.
snap40 is intended to provide visual and audible physiologic multi-parameter alarms.
snap40 is intended for temperature monitoring temperature at the upper arm is clinically indicated.
snap40 continuously monitors the following parameters in adults:
- · Pulse rate
- · Oxygen saturation
- · Temperature
- · Movement
snap40 is intended for intermittent or spot-check monitoring of respiration rate.
snap40 is not intended for use in high-acuity environments, such as ICU or operating rooms or for use in the home
snap40 is not intended for use on acutely ill cardiac patients with the potential to develop life threatening arrhythmias e.g. very fast atrial fibrillation. For these patients, they should be monitored using a device with continuous ECG. snap40 is not a substitute for an ECG monitor.
snap40 is not intended for SpO2 monitoring in conditions of high motion or low perfusion.
Type of Use (Select one or both, as applicable)
|X Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
For snap40 system
### 510(k) Summary
This 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92.
### Submitter's Name:
snap40 Ltd
### Submitter's Address:
24 Forth Street, Edinburgh EH1 3LH United Kingdom
Telephone: +44 (0) 131 560 1137
### Establishment Registration Number:
Still to be established
### Contact Person:
Edwin Lindsay
Telephone +44 (0) 7917134922
### Date Prepared:
4th September 2018
Below summaries the Device Classification Information regarding the snap40 System:
### Primary Product Code:
| Regulation<br>Number | Device | Device<br>Class | Product<br>Code | Classification<br>Panel |
|----------------------|-----------------------------------------------------------------|-----------------|-----------------|-------------------------|
| 870.2300 | System, Network and<br>Communication,<br>Physiological Monitors | Class 2 | MSX | Cardiovascular |
### Secondary Product Codes:
| Regulation<br>Number | Device | Device<br>Class | Product<br>Code | Classification<br>Panel |
|----------------------|------------------------------------------------------------------------|-----------------|-----------------|-------------------------|
| 880.2910 | Thermometer, Electronic,<br>Clinical | Class 2 | FLL | General<br>Hospital |
| 870.2700 | Oximeter | Class 2 | DQA | Cardiovascular |
| 868.2375 | Monitor, Breathing<br>Frequency | Class 2 | BZQ | Anesthesiology |
| 870.2910 | Transmitters and<br>Receivers, Physiological<br>Signal, Radiofrequency | Class 2 | DRG | Cardiovascular |
{4}------------------------------------------------
For snap40 system
### Device Trade Name:
snap40 system
### Device Common Name:
snap40 system
#### Intended/ Indications Use:
snap40 is intended for reusable bedside, mobile and central multi-parameter, physiologic patient monitoring of adult patients in environments where patient care is provided by trained healthcare professionals.
snap40 is intended to provide visual and audible physiologic multi-parameter alarms.
snap40 is intended for temperature monitoring where monitoring temperature at the upper arm is clinically indicated.
snap40 is intended for continuous monitoring of the following parameters in adults:
- Pulse rate
- . Oxygen saturation
- Temperature ●
- Movement ●
snap40 is intended for intermittent or spot-check monitoring of respiration rate in adults.
snap40 is not intended for use in high-acuity environments, such as ICU or operating rooms or for use in the home
snap40 is not intended for use on acutely ill cardiac patients with the potential to develop life threatening arrhythmias e.g. very fast atrial fibrillation. For these patients, they should be monitored using a device with continuous ECG. snap40 is not a substitute for an ECG monitor.
snap40 is not intended for SpO2 monitoring in conditions of high motion or low perfusion.
{5}------------------------------------------------
### snap40 Ltd. Traditional 510(k) For snap40 system
### Summary of Substantial Equivalence:
The following predicate devices have been that the snap40 system can claim equivalence with and these are detailed below
### General Comparison
| General Information | | | | | |
|---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Property | Proposed<br>Device snap40 | Primary<br>Predicate<br>IACS | Infinity Central<br>Station | IntelliVue CL<br>Respiration Pod | IntelliVue Guardian<br>Software |
| Common Name | System,<br>Network and<br>Communication,<br>Physiological<br>Monitors | Monitor,<br>Physiological,<br>Patient (With<br>Arrhythmia<br>Detection or<br>Alarms) | Monitor,<br>Physiological,<br>Patient (With<br>Arrhythmia<br>Detection or<br>Alarms) | Monitor, Breathing<br>Frequency | Display, Cathode-<br>Ray Tube |
| Device Manufacturer | snap40 Ltd | Draeger | Draeger | Philips | Philips |
| Device Classification | II | II | II | II | II |
| Primary Product<br>Code | MSX | MHX | MHX | BZQ | DXJ |
| Secondary Product<br>Code | FLL DQA BZQ<br>DRG | MSX, DRT, DQA,<br>BZQ, FLL, DSK,<br>FLS, MLD, DXN,<br>CCK | - | DRG MSX | DQK NSX OUG |
| 510k Numbers | N/A | K113798 | K151860 | K132320 | K161767 |
| Target Population | Adult | Adult, Paediatric<br>and Neonatal | Adult, Paediatric<br>and Neonatal | Adult | N/A |
| Environment | Hospital | Hospital | Hospital | Hospital | Hospital |
| Intended Use/<br>Indication for Use | snap40 is<br>intended for<br>reusable<br>bedside, mobile<br>and central<br>multi-parameter,<br>physiologic<br>patient<br>monitoring of<br>adult patients in<br>environments<br>where patient<br>care is provided<br>by trained | The IACS is<br>intended for<br>multi-<br>parameter,<br>physiologic<br>patient<br>monitoring of<br>adult, pediatric<br>and neonatal<br>patients in<br>environments<br>where patient<br>care is<br>provided by | The Infinity<br>CentralStation<br>(ICS) is<br>intended for use<br>by trained<br>healthcare<br>professionals<br>for the purpose<br>of centralized<br>monitoring of<br>adult, pediatric<br>and neonatal<br>patient data<br>within the | The IntelliVue<br>CL<br>Respiration<br>Pod is<br>indicated for<br>use by<br>healthcare<br>professionals<br>whenever<br>there is a<br>need for<br>intermittent or<br>spot-check<br>acquisition | The IntelliVue<br>GuardianSoftw<br>are is<br>indicated for<br>use by<br>healthcare<br>providers<br>whenever<br>there is a need<br>for the<br>generation of a<br>patient record.<br>The IntelliVue |
| General Information | | | | | |
| Property | Proposed Device snap40 | Primary Predicate IACS | Infinity Central Station | IntelliVue CL Respiration Pod | IntelliVue Guardian Software |
| | healthcare professionals. | trained healthcare professionals. | hospital or clinical environment. Centralized monitoring involves the display and management of data from networked patient monitors including the annunciation of visual and audible physiologic parameter alarms at a central monitoring workstation. Infinity CentralStation with Rest ECG is intended for the production and interpretation of diagnostic electrocardiograms for adult and pediatric patients when connected to a monitor with diagnostic 12-Lead ECG monitoring enabled. | and monitoring of physiological patient parameters respiration rate and pulse rate wirelessly in specific hospital areas. The IntelliVue CL Respiration Pod is mainly indicated for use in general medical and surgery wards and in waiting areas of emergency rooms. It is not indicated for use in hospital areas in which continuous patient monitoring is needed, such as intensive care units or operating rooms. The intended use of the IntelliVue Cl | GuardianSoftw are is intended for use in the collection, storage and management of data from Philips specified Measurements and Philips Patient Monitors that are connected through networks. |
| | snap40 is intended to provide visual and audible physiologic multi-parameter alarms. | The IACS obtains the physiologic, multi- parameter data from the connection to the M540 monitor and optional medical devices and displays. The transfer of this data is accomplished by the Infinity network. | | | |
| | snap40 is intended for temperature monitoring where monitoring temperature at the upper arm is clinically indicated. | | | | |
| | snap40 is intended for continuous monitoring of the following parameters in adults:<br>Pulse rate<br>Oxygen saturation<br>Temperature<br>Movement | The Infinity M540 is intended for the monitoring of multi- parameter, physiological patient information obtained from connected hardware in environments where patient care is provided by trained healthcare | | | |
| | snap40 is intended for intermittent or spot-check monitoring of | | | | |
| General Information | | | | | |
| Property | Proposed<br>Device snap40 | Primary<br>Predicate<br>IACS | Infinity Central<br>Station | IntelliVue CL<br>Respiration Pod | IntelliVue Guardian<br>Software |
| | respiration rate<br>in adults. | professional.<br>The M540 is<br>intended to<br>monitor one<br>patient at a<br>time. | | Respiration<br>Pod when<br>used together<br>with a patient<br>monitor is for<br>intermittent or<br>spot-check<br>monitoring<br>and recording<br>of, and to<br>generate<br>alarms for,<br>respiration<br>rate and pulse<br>rate of adult<br>patients. | |
| | snap40 is not<br>intended for use<br>in high-acuity<br>environments,<br>such as ICU or<br>operating<br>rooms or for<br>use in the home | The M540<br>monitors the<br>following<br>parameters:<br>• Heart rate<br>• Arrhythmia<br>(adult and<br>pediatric only)<br>• 12-lead<br>analysis<br>• ST segment<br>analysis<br>including<br>TruST® (adult<br>and pediatric<br>only)<br>• 12-lead ST<br>segment<br>analysis (adult<br>and pediatric<br>only)<br>• Apnea<br>• Respiration<br>rate<br>• Invasive<br>pressure<br>• Non-invasive<br>pressure<br>• Temperature<br>• Cardiac<br>output (only<br>available when<br>the M540 is<br>docked in an | | The IntelliVue<br>CL<br>Respiration<br>Pod is also<br>intended for<br>acquisition of<br>respiration<br>rate and pulse<br>rate data of<br>adult patients<br>for a clinical<br>information<br>management<br>system. | |
| | snap40 is not<br>intended for use<br>on acutely ill<br>cardiac patients<br>with the<br>potential to<br>develop life<br>threatening<br>arrhythmias e.g.<br>very fast atrial<br>fibrillation. For<br>these patients,<br>they should be<br>monitored using<br>a device with<br>continuous<br>ECG. snap40 is<br>not a substitute<br>for an ECG<br>monitor. | | | The Intellivue<br>CL<br>Respiration<br>Pod is<br>intended for<br>use by<br>healthcare<br>professionals.<br>It is not | |
| | snap40 is not<br>intended for<br>SpO2<br>monitoring in<br>conditions of<br>high motion or<br>low perfusion. | | | | |
| General Information | | | | | |
| Property | Proposed<br>Device snap40 | Primary<br>Predicate | Infinity Central<br>Station | IntelliVue CL<br>Respiration Pod | IntelliVue Guardian<br>Software |
| | | IACS<br>IACS<br>configuration)<br>· Arterial<br>oxygen<br>saturation<br>(SpO 2)<br>· Pulse rate<br>· Perfusion<br>Index (P1)<br>· Total<br>hemoglobin<br>(SpHb) - adult<br>and pediatric<br>only<br>· Total oxygen<br>content (SpCO)<br>- adult and<br>pediatric only<br>· Methemoglobin<br>saturation<br>(SpMet)<br>· Pleth<br>variability<br>index (PVI)<br>· Mainstream<br>etCO2 | | intended for<br>home use. It<br>is not a<br>therapeutic<br>device.<br><br>The IntelliVue<br>CL<br>Respiration<br>Pod is not<br>intended for<br>use on<br>patients with<br>extremely<br>high values<br>for respiration<br>rate (above 60<br>rpm).<br><br>The IntelliVue<br>CL<br>Respiration<br>Pod is not<br>intended for<br>use on<br>acutely ill<br>cardiac<br>patients with<br>the potential<br>to develop life<br>threatening<br>arrhythmias,<br>e.g. very fast<br>atrial<br>fibrillation or<br>ventricular<br>tachycardia<br>(rapid<br>irregular<br>pulse rate).<br>For | |
| General Information | | | | | |
| Property | Proposed<br>Device snap40 | Primary<br>Predicate<br>IACS | Infinity Central<br>Station | IntelliVue CL<br>Respiration Pod | IntelliVue Guardian<br>Software |
| | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.