Wearable Device, Home Package in Carry Case with Homehub, Home Package in Carry Case without Homehub
K190073 · Current Health , Ltd. · MSX · Mar 27, 2019 · Cardiovascular
Device Facts
Record ID
K190073
Device Name
Wearable Device, Home Package in Carry Case with Homehub, Home Package in Carry Case without Homehub
Applicant
Current Health , Ltd.
Product Code
MSX · Cardiovascular
Decision Date
Mar 27, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Indications for Use
The Current Wearable Health Monitoring System is intended for reusable bedside, mobile and central multi-parameter, physiologic patient monitoring of adult patients in professional healthcare facilities, such as hospitals or skilled nursing facilities, or their own home. It is intended for monitoring of patients by trained healthcare professionals. The Current Wearable Health Monitoring System is intended to provide visual and audible physiologic multi-parameter alarms. The Current Wearable Health Monitoring System is intended for temperature monitoring where monitoring temperature at the upper arm is clinically indicated. The Current Wearable Health Monitoring System is intended for continuous monitoring of the following parameters in adults: Pulse rate, Oxygen saturation, Temperature, Movement. The Current Wearable Health Monitoring System is intermittent or spot-check monitoring of respiration rate, non-invasive blood pressure and weight in adults. The Current Wearable Health Monitoring System is not intended for use in high-acuity environments, such as ICU or operating rooms. The Current Wearable Health Monitoring System is not intended for use on acutely ill cardiac patients with the potential to develop life threatening arrhythmias e.g. very fast atrial fibrillation. For these patients, they should be monitored using a device with continuous ECG. The Current Wearable Health Monitoring System is not a substitute for an ECG monitor. The Current Wearable Health Monitoring System is not intended for SpO2 monitoring in conditions of high motion or low perfusion.
Device Story
Wearable upper-arm device; monitors pulse rate, SpO2, temperature, movement; integrates with external blood pressure and weight devices. Data transmitted to software platform/central station for professional healthcare review. Used in hospitals, skilled nursing, or home; patient applies device/charges; healthcare professionals monitor remotely or on-site. Provides visual/audible alarms. Aids clinical decision-making by providing physiologic data; benefits include continuous monitoring outside high-acuity settings.
Clinical Evidence
No clinical studies conducted. Evidence based on bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), alarm system verification (IEC 60601-1-8), pulse rate and SpO2 accuracy (ISO 80601-2-61), respiratory rate accuracy, temperature accuracy (ISO 80601-2-56), wireless coexistence, and usability (IEC 62366).
Technological Characteristics
Upper-arm wearable device; wireless radio communication; QI charging; battery-powered. Biocompatible materials (ISO 10993-1). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, ISO 80601-2-61, ISO 80601-2-56, IEC 62366, ASTM D4169. Software-based alarm system and data platform.
Indications for Use
Indicated for adult patients in professional healthcare facilities (hospitals, skilled nursing) or home settings for continuous monitoring of pulse rate, SpO2, temperature, and movement, and intermittent monitoring of respiration rate, non-invasive blood pressure, and weight. Contraindicated for high-acuity environments (ICU, OR) and acutely ill cardiac patients prone to life-threatening arrhythmias (e.g., fast atrial fibrillation). Not for SpO2 monitoring during high motion or low perfusion.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
Predicate Devices
snap40 Wearable Vital Signs Monitoring System (K182543)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 27, 2019
Current Health Ltd. Edwin Lindsay OA/RA Manager Playfair House. 12A Broughton Street Lane Edinburgh, EH1 3LY Gb
Re: K190073
Trade/Device Name: Current Wearable Health Monitoring System Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MSX, DQA, FLL, BZQ, DRG Dated: January 11, 2019 Received: January 16, 2019
Dear Edwin Lindsay:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Jessica E. Paulsen -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K190073
#### Device Name
Current Wearable Health Monitoring System
#### Indications for Use (Describe)
The Current Wearable Health Monitoring System is intended for reusable bedside, mobile and central multi-parameter, physiologic patient monitoring of adult patients in professional healthcare facilities, such as hospitals or skilled nursing facilities, or their own home. It is intended for monitoring of patients by trained healthcare professionals.
The Current Wearable Health Monitoring System is intended to provide visual and audible physiologic multi-parameter alarms. The Current Wearable Health Monitoring System is intended for temperature monitoring temperature at the upper arm is clinically indicated.
The Current Wearable Health Monitoring System is intended for continuous monitoring of the following parameters in adults:
- · Pulse rate
- · Oxygen saturation
- · Temperature
- Movement
The Current Wearable Health Monitoring System is intermittent or spot-check monitoring of respiration rate, non-invasive blood pressure and weight in adults.
The Current Wearable Health Monitoring System is not intended for use in high-acuity environments, such as ICU or operating rooms.
The Current Wearable Health Monitoring System is not intended for use on acutely ill cardiac patients with the potential to develop life threatening arrhythmas e.g. very fast atrial fibrillation. For these patients, they should be monitored using a device with continuous ECG. The Current Wearable Health Monitoring System is not a substitute for an ECG monitor.
The Current Wearable Health Monitoring System is not intended for SpO2 monitoring in conditions of high motion or low perfusion.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
For Current Wearable Health Monitoring System
### 510(k) Summary
This 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92.
### Submitter's Name:
Current Health Ltd
### Submitter's Address:
Playfair House, 12A Broughton Street Lane Edinburgh EH1 3LY
Telephone: +44 (0) 131 560 1137
### Establishment Registration Number:
Still to be established
### Contact Person:
Edwin Lindsay
Telephone +44 (0) 7917134922
### Date Prepared:
11th January 2019
Note: snap40 Ltd has recently changed their name to Current Health Ltd.
{5}------------------------------------------------
For Current Wearable Health Monitoring System
### Below summaries the Device Classification Information regarding the Current Wearable Health Monitoring System:
### Primary Product Code:
| Regulation<br>Number | Device | Device<br>Class | Product<br>Code | Classification<br>Panel |
|----------------------|-----------------------------------------------------------------|-----------------|-----------------|-------------------------|
| 870.2300 | System, Network and<br>Communication,<br>Physiological Monitors | Class 2 | MSX | Cardiovascular |
### Secondary Product Codes:
| Regulation<br>Number | Device | Device<br>Class | Product<br>Code | Classification<br>Panel |
|----------------------|------------------------------------------------------------------------|-----------------|-----------------|-------------------------|
| 880.2910 | Thermometer, Electronic,<br>Clinical | Class 2 | FLL | General<br>Hospital |
| 870.2700 | Oximeter | Class 2 | DQA | Cardiovascular |
| 868.2375 | Monitor, Breathing<br>Frequency | Class 2 | BZQ | Anesthesiology |
| 870.2910 | Transmitters and<br>Receivers, Physiological<br>Signal, Radiofrequency | Class 2 | DRG | Cardiovascular |
#### Device Trade Name:
Current Wearable Health Monitoring System
### Device Common Name:
Current Wearable Health Monitoring System
### Intended/ Indications Use:
The Current Wearable Health Monitoring System is intended for reusable bedside, mobile and central multi-parameter, physiologic patient monitoring of adult patients in professional healthcare facilities, such as hospitals or skilled nursing facilities, or their own home. It is intended for monitoring of patients by trained healthcare professionals.
The Current Wearable Health Monitoring System is intended to provide visual and audible physiologic multi-parameter alarms. The Current Wearable Health Monitoring System is intended for temperature monitoring where monitoring temperature at the upper arm is clinically indicated.
The Current Wearable Health Monitoring System is intended for continuous monitoring of the following parameters in adults:
- Pulse rate
- . Oxygen saturation
- Temperature
- Movement ●
{6}------------------------------------------------
For Current Wearable Health Monitoring System
The Current Wearable Health Monitoring System is intermittent or spot-check monitoring of respiration rate, non-invasive blood pressure and weight in adults.
The Current Wearable Health Monitoring System is not intended for use in high-acuity environments, such as ICU or operating rooms.
The Current Wearable Health Monitoring System is not intended for use on acutely ill cardiac patients with the potential to develop life threatening arrhythmias e.q. very fast atrial fibrillation. For these patients, they should be monitored using a device with continuous ECG. The Current Wearable Health Monitoring System is not a substitute for an ECG monitor.
The Current Wearable Health Monitoring System is not intended for SpO2 monitoring in conditions of high motion or low perfusion.
{7}------------------------------------------------
For Current Wearable Health Monitoring System
### Summary of Substantial Equivalence:
The following predicate devices have been that the Current Wearable Heath Monitoring System can claim equivalence with and these are detailed below
### General Comparison
| Property | Proposed Device<br>Current Wearable Health<br>Monitoring System | Primary Predicate<br>snap40 Wearable Vital<br>Signs Monitoring System | Secondary Predicate:<br>VitalConnect Platform | Secondary Predicate:<br>IACS | Comment |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common<br>Name | System, Network and<br>Communication,<br>Physiological Monitors | System, Network and<br>Communication,<br>Physiological Monitors | Transmitters and<br>Receivers, Physiological<br>Signal, Radiofrequency | Monitor, Physiological,<br>Patient (With Arrhytmia<br>Detection or Alarms) | N/A |
| Device<br>Manufacturer | Current Health Ltd | Current Health Ltd | Philips | Draeger | N/A |
| Device<br>Classification | II | II | II | II | N/A |
| 510(k)<br>Number | N/A | K182543 | K141167 | K113798 | N/A |
| Primary<br>Product Code | MSX | MSX | DRG | MHX | N/A |
| Secondary<br>Product Code | FLL DQA BZQ DRG | FLL DQA BZQ DRG | DSI MHX | MSX, DRT, DQA, BZQ,<br>FLL, DSK, FLS, MLD,<br>DXN, CCK | N/A |
| Target<br>Population | Adult | Adult | Adult | Adult, Paediatric and<br>Neonatal | The proposed device and<br>predicates are identical |
| Environment | Professional Healthcare<br>Facilities & home | Hospital | Professional Healthcare<br>Facilities & Home | Hospital | The proposed device and<br>secondary predicate are identical |
| Property | Proposed Device<br>Current Wearable Health<br>Monitoring System | Primary Predicate<br>snap40 Wearable Vital<br>Signs Monitoring System | Secondary Predicate:<br>VitalConnect Platform | Secondary Predicate:<br>IACS | Comment |
| Intended Use/<br>Indication for<br>Use | The Current Wearable<br>Health Monitoring<br>System is intended for<br>reusable bedside, mobile<br>and central multi-<br>parameter, physiologic<br>patient monitoring of<br>adult patients in<br>professional healthcare<br>facilities, such as<br>hospitals or skilled<br>nursing facilities, or their<br>own home. It is intended<br>for monitoring of patients<br>by trained healthcare<br>professionals. | snap40 is intended for<br>reusable bedside,<br>mobile and central<br>multi-parameter,<br>physiologic patient<br>monitoring of adult<br>patients in<br>environments where<br>patient care is<br>provided by trained<br>healthcare<br>professionals.<br><br>snap40 is intended to<br>provide visual and<br>audible physiologic<br>multi-parameter<br>alarms.<br><br>Snap40 is intended for<br>temperature<br>monitoring where<br>monitoring<br>temperature at the<br>upper arm is clinically<br>indicated.<br><br>Snap40 is intended for<br>continuous<br>monitoring of the<br>following parameters<br>in adults:<br><br>Pulse rate<br>Oxygen saturation<br>Temperature<br>Movement<br><br>snap40 is intended for<br>intermittent or spot-<br>check monitoring of<br>respiration rate in<br>adults. | The VitalConnect<br>Platform is a wireless<br>remote monitoring<br>system intended for use<br>by healthcare<br>professionals for<br>continuous collection of<br>physiological data in<br>home and healthcare<br>settings. This includes<br>heart rate,<br>electrocardiography<br>(EGG), heart rate<br>variability (R-R interval),<br>respiratory rate, skin<br>temperature, activity<br>(including step count),<br>and posture (body<br>position relative to<br>gravity including fall).<br>Data is transmitted<br>wirelessly to a central<br>location where it is<br>stored for analysis. The<br>Vital Connect Platform<br>can be configured by<br>Authorized Persons to<br>notify healthcare<br>professionals when<br>physiological data falls<br>outside selected<br>parameters.<br><br>The device is intended<br>for use on general care<br>patients who are 18 years<br>of age or older as a<br>general patient monitor,<br>to provide physiological<br>information. The data<br>from the VitalConnect<br>Platform is intended for<br>use by healthcare | The IACS is intended<br>for multi-parameter,<br>physiologic patient<br>monitoring of adult,<br>pediatric and neonatal<br>patients in<br>environments where<br>patient care is<br>provided by trained<br>healthcare<br>professionals.<br><br>The IACS obtains the<br>physiologic, multi-<br>parameter data from<br>the connection to the<br>M540 monitor and<br>optional medical<br>devices and displays.<br>The transfer of this<br>data is accomplished<br>by the Infinity<br>network.<br><br>The Infinity M540 is<br>intended for the<br>monitoring of multi-<br>parameter,<br>physiological patient<br>information obtained<br>from connected<br>hardware in<br>environments where<br>patient care is<br>provided by trained<br>healthcare<br>professional. The<br>M540 is intended to<br>monitor one patient at<br>a time.<br><br>The M540 monitors<br>the following<br>parameters. | The intended use and indications for use of<br>the proposed device and the predicates are<br>the same except that the primary predicate<br>is only intended for use in the hospital.<br><br>The intended use and indications for use of<br>the proposed device and the secondary<br>predicate, the Vital Connect, are the same in<br>that they both offer multi-parameter<br>physiologic monitoring in healthcare setting<br>and in the home.<br><br>The intended use and indications for use<br>of the proposed device and the<br>secondary predicate, the Draeger IACS,<br>are the same in that they both monitor<br>blood pressure non-invasively. |
| Property | Proposed Device<br>Current Wearable Health<br>Monitoring System | Primary Predicate<br>snap40 Wearable Vital<br>Signs Monitoring System | Secondary Predicate:<br>VitalConnect Platform | Secondary Predicate:<br>IACS | Comment |
| | The Current Wearable<br>Health Monitoring<br>System is intended for<br>intermittent or spot-<br>check monitoring of<br>respiration rate, non-<br>invasive blood pressure<br>and weight in adults. | Snap40 is not<br>intended for use in<br>high-acuity<br>environments, such<br>as ICU or operating<br>rooms or for use in<br>the home | professionals as an aid<br>to diagnosis and<br>treatment. It is not<br>intended for use on<br>critical care patients. | · Heart rate<br>· Arrhythmia (adult<br>and pediatric only)<br>· 12-lead analysis<br>· ST segment<br>analysis including<br>TruST® (adult and<br>pediatric only)<br>· 12-lead ST segment<br>analysis (adult and<br>pediatric only)<br>· Apnea<br>· Respiration rate<br>· Invasive pressure<br>· Non-invasive<br>pressure<br>· Temperature<br>· Cardiac output<br>(only available when<br>the M540 is docked in<br>an IACS<br>configuration)<br>· Arterial oxygen<br>saturation (SpO 2)<br>· Pulse rate<br>· Perfusion Index<br>(P1)<br>· Total hemoglobin<br>(SpHb) - adult and<br>pediatric only<br>· Total oxygen<br>content (SpCO) - adult<br>and pediatric only<br>· Methemoglobin<br>saturation (SpMet)<br>· Pleth variability<br>index (PVI)<br>· Mainstream etCO2 | |
| | The Current Wearable<br>Health Monitoring<br>System is not intended<br>for use in high-acuity<br>environments, such as<br>ICU or operating rooms. | snap40 is not<br>intended for use on<br>acutely ill cardiac<br>patients with the<br>potential to develop<br>life threatening<br>arrhythmias e.g. very<br>fast atrial fibrillation.<br>For these patients,<br>they should be<br>monitored using a<br>device with<br>continuous ECG.<br>Snap40 is not a<br>substitute for an ECG<br>monitor. | | | |
| | The Current Wearable<br>Health Monitoring<br>System is not intended<br>for use on acutely ill<br>cardiac patients with the<br>potential to develop life<br>threatening arrhythmias<br>e.g. very fast atrial<br>fibrillation. For these<br>patients, they should be<br>monitored using a device<br>with continuous ECG.<br>The Current Wearable<br>Health Monitoring<br>System is not a<br>substitute for an ECG<br>monitor. | Snap40 is not<br>intended for SpO2<br>monitoring in<br>conditions of high<br>motion or low<br>perfusion | | | |
{8}------------------------------------------------
For Current Wearable Health Monitoring System
Page 5 of 10
K190073
{9}------------------------------------------------
For Current Wearable Health Monitoring System
Any technical differences have been justifically and using performance testing. These do not affect the safety or effectiveness of the proposed device.
{10}------------------------------------------------
For Current Wearable Health Monitoring System
### Device Description:
The Current Wearable Health Monitoring System consists of a single monitoring device worn on the upper arm, a software platform (containing an alarming system) and a user interface to allow presentation of vital signs data both on mobile devices and a central station. The Current Wearable Health Monitoring System is also integrated with specific devices for monitoring of blood pressure and weight.
The system is intended to continuously monitor adult patient vital signs in professional healthcare facilities, such as hospitals or skilled nursing facilities, or their own home. It is intended for monitoring of patients by trained healthcare professionals. It is intended to continuously monitor pulse rate (PR), oxygen saturation (SpO2), temperature (TEMP) and movement (MOVEMENT). Current is intended for intermittent or spot-checking monitoring of respiration rate (RESP), blood pressure (BP) and weight (WEIGHT).
In the home environment, the patient will have responsibility for applying the device to their arm, charging the device and plugging in the Homehub to mains power. However, the data will still be made directly available to healthcare professionals. These healthcare professionals will be at a remote location e.q. an office or within the hospital or could be with the patient in their own home.
### Technological Characteristics:
A comparative review of the Current Wearable Health Monitoring System with the predicate device found that the technology, mode of operation, and general principles for treatment with this device were substantially equivalent.as the predicate device.
### Non-Clinical Tests (Performance/Physical Data):
The Current Wearable Health Monitoring System was evaluated for its safety and effectiveness based on the following testing:
| Test Name | Test Description | Results |
|---------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Electrical Safety | The Current Health System was tested to confirm<br>that it met the applicable standards for electrical<br>safety (IEC 60601-1) | Passed |
| EMC | The Current Health System was tested to confirm<br>that it met the applicable standards for<br>electromagnetic compatibility (EMC) (IEC 60601-1-<br>2) | Passed |
| QI Charger Testing | The Current Health Wearable was tested to confirm<br>the device could be charged using the QI Charger | Passed |
| Environment Testing | The Current Health System was tested to confirm<br>the storage and operating temperature ranges. | Passed |
| Test Name | Test Description | Results |
| Alarm Testing | The Current Health System was tested to confirm<br>that it met the applicable standards for basic safety<br>and essential performance for alarm systems (IEC<br>60601-1-8)<br><br>Alarm Function Verification Tests for each<br>parameter function were performed. | Passed |
| Pulse Rate Testing<br>Validation of the<br>accuracy of pulse rate<br>monitoring | The Current Health System was tested to confirm<br>the accuracy of pulse rate monitoring of the system<br>in accordance with ISO 80601-2-61 and the FDA<br>Pulse Oximeters - Premarket Notification<br>Submissions: Guidance for Industry and FDA Staff.<br>2007 | Passed |
| Battery Testing | The Current Health Wearable Device Battery Life<br>testing was performed to determine the battery life<br>of the Wearable Device while operating in various<br>modes as well as time to fully charge. | Passed |
| PDU Buffering Time<br>Testing | The Current Health system was tested to evaluate<br>the difference between the theoretical maximum<br>PDU buffering time and the observed maximum<br>PDU buffering time. | Passed |
| Usability Testing | The Current Health System was assessed with<br>regards to usability for compliance with IEC 62366 -<br>Medical devices - Application of usability<br>engineering to medical devices and IEC 60601-1-11 | Passed |
| Device Ship/Transport<br>Testing | Ensure device, enclosed in the selected shipping<br>container, meets ASTM D4169 specifications. | Passed |
| Biocompatibility<br>Testing | Testing and analysis of the Current Health System<br>has demonstrated compliance to ISO 10993-1:<br>Biological evaluation of medical devices – Guidance | Passed |
| SpO2 Testing<br>Validation of the<br>accuracy of SpO2<br>monitoring | Ensure the accuracy and communication of the<br>SpO2 functions within the Current Health system as<br>per ISO80601-2-61 and the FDA SpO2 guidance;<br>Pulse Oximeters-Premarket Notification<br>Submissions Guidance for Industry and Food and<br>Drug Staff, March 4, 2013 | Passed |
| Test Name | Test Description…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.