K231506 · Current Health , Ltd. · MSX · Aug 24, 2023 · Cardiovascular
Device Facts
Record ID
K231506
Device Name
Current Health System
Applicant
Current Health , Ltd.
Product Code
MSX · Cardiovascular
Decision Date
Aug 24, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2300
Device Class
Class 2
Indications for Use
The Current Wearable Health Monitoring System is intended for reusable bedside, mobile and central multi-parameter, physiologic monitoring of patients aged 14 years old and above in professional healthcare facilities, such as hospitals or skilled nursing facilities, or their own home. It is intended for monitoring of patients by trained healthcare professionals. The Current Wearable Health Monitoring System is intended to provide visual and audible physiologic multi-parameter alarms. The Current Wearable Health Monitoring System is intended for temperature monitoring temperature at the upper arm is clinically indicated. The Current Wearable Health Monitoring System is intended for continuous monitoring of the following parameters: · Pulse rate · Oxygen saturation · Temperature · Movement The Current Wearable Health Monitoring System is intended for intermittent or spot-check monitoring of: · Respiration rate · Non-invasive blood pressure · Lung function & spirometry · Weight The Current Wearable Health Monitoring System is not intended for use in high-acuity environments, such as ICU or operating rooms. The Current Wearable Health Monitoring System is not intended for use on acutely ill cardiac patients with the potential to develop life threatening arrhythmias e.g. very fast atrial fibrillation. For these patients, they should be monitored using a device with continuous ECG. The Current Wearable Health Monitoring System is not a substitute for an ECG monitor. The Current Wearable Health Monitoring System is not intended for SpO2 monitoring in conditions of high motion or low perfusion.
Device Story
Remote patient monitoring system; wearable device worn on upper arm; software platform with web/mobile interface. Inputs: physiologic vital sign data (pulse rate, SpO2, temperature, movement, respiration rate, NIBP, lung function, weight). Operation: wearable continuously monitors/transmits encrypted data to secure server; software platform processes data to generate visual/audible multi-parameter alarms and notifications for clinical care teams. Used in hospitals, skilled nursing facilities, or home by trained healthcare professionals. Output: vital signs data and alarms presented on mobile/web dashboard. Benefits: enables remote physiologic monitoring of patients aged 14+; facilitates timely clinical intervention via alarm notifications.
Clinical Evidence
Bench testing only; no clinical or animal data included.
Technological Characteristics
Wearable monitor worn on upper arm with strap. Reusable. Non-sterile. Connectivity: secure server transmission to mobile/web interface. Software-based alarm system. Identical to predicate K222550.
Indications for Use
Indicated for patients aged 14 years and older in professional healthcare facilities or home settings. Not for high-acuity environments (ICU/OR) or acutely ill cardiac patients prone to life-threatening arrhythmias. Not for SpO2 monitoring during high motion or low perfusion.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
Predicate Devices
Current Wearable Health Monitoring System (K222550)
Reference Devices
Infinity Acute Care System (IACS) Monitoring Solution (K113798)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 24, 2023
Current Health Ltd Giovanni Maggi Regulatory Affairs Manager The Stamp Office, Level 3, 10 Waterloo Place Edinburgh, EH1 3EG United Kingdom
Re: K231506
Trade/Device Name: Current Health System Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac monitor (including cardiotachometer and rate alarm) Regulatory Class: Class II Product Code: MSX, FLL, DQA, BZQ, DRG, BZG Dated: July 28, 2023 Received: July 28, 2023
Dear Giovanni Maggi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Shruti N. Mistry -S
for
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231506
Device Name Current Health System
#### Indications for Use (Describe)
The Current Wearable Health Monitoring System is intended for reusable bedside, mobile and central multi-parameter, physiologic monitoring of patients aged 14 years old and above in professional healthcare facilities, such as hospitals or skilled nursing facilities, or their own home. It is intended for monitoring of patients by trained healthcare professionals.
The Current Wearable Health Monitoring System is intended to provide visual and audible physiologic multi-parameter alarms. The Current Wearable Health Monitoring System is intended for temperature monitoring temperature at the upper arm is clinically indicated.
The Current Wearable Health Monitoring System is intended for continuous monitoring of the following parameters:
- · Pulse rate
- · Oxygen saturation
- · Temperature
- Movement
The Current Wearable Health Monitoring System is intended for intermittent or spot-check monitoring of:
- · Respiration rate
- · Non-invasive blood pressure
- · Lung function & spirometry
- · Weight
The Current Wearable Health Monitoring System is not intended for use in high-acuity environments, such as ICU or operating rooms.
The Current Wearable Health Monitoring System is not intended for use on acutely ill cardiac patients with the potential to develop life threatening arrhythmias e.g. very fast atrial fibrillation. For these patients, they should be monitored using a device with continuous ECG. The Current Wearable Health Monitoring System is not a substitute for an ECG monitor.
The Current Wearable Health Monitoring System is not intended for SpO2 monitoring in conditions of high motion or low perfusion.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Special 510(k) Ref: K231506
Special 510(k) Summary
Version: 1.0
# Special 510(k) Summary
This summary of 510(k) substantial equivalence information is being submitted in accordance with the requirement of 21 CFR 807.92
| Submitter Information: | |
|------------------------------------|------------------------------------------------------------------------------------------|
| Name: | Current Health Ltd. |
| Address: | The Stamp Office<br>Level 3, 10 Waterloo Place<br>Edinburgh<br>EH1 3EG<br>United Kingdom |
| Establishment Registration Number: | 3015134004 |
| Owner/Operator Number: | 10059040 |
| Phone: | +44 (0) 131 285 8101 |
| Contact: | Giovanni Maggi<br>Regulatory Affairs Manager |
| E-mail: | giovanni.maggi@currenthealth.com |
| Date of Summary: | 27-Jul-2023 |
# Device Information:
Below summarises the Device Classification information regarding the Current Health System.
| Device Proprietary Name | Current Health System |
|-------------------------|-------------------------------------------|
| Common Name: | Remote Patient Monitor |
| Trade Name: | Current Wearable Health Monitoring System |
| Product Code(s): | MSX; FLL; DQA; BZQ; DRG; BZG (see below) |
Primary Product Code
| Regulation<br>Number (21 CFR) | Device | Product Class | Product Code | Classification Panel |
|-------------------------------|-----------------------------------------------------------------|---------------|--------------|----------------------|
| 870.2300 | System, Network and<br>Communication,<br>Physiological Monitors | Class II | MSX | Cardiovascular |
Secondary Product Codes
| Regulation<br>Number (21 CFR) | Device | Device Class | Product Code | Classification Panel |
|-------------------------------|--------------------------------------|--------------|--------------|----------------------|
| 880.2910 | Thermometer,<br>Electronic, Clinical | Class II | FLL | General Hospital |
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a blue abstract shape. The shape is a closed loop that resembles a combination of a figure eight and a circle. The loop is not perfectly symmetrical, and the lines are smooth and rounded.
Special 510(k) Ref: K231506
Section 05 – Special 510(k) Summary
Version: 1.0
| 870.2700 | Oximeter | Class II | DQA | Cardiovascular |
|----------|------------------------------------------------------------------------|----------|-----|-----------------|
| 868.2375 | Monitor,<br>Breathing<br>Frequency | Class II | BZQ | Anaesthesiology |
| 870.2910 | Transmitters and<br>Receivers, Physiological<br>Signal, Radiofrequency | Class II | DRG | Cardiovascular |
| 686.1840 | Spirometer, Diagnostic | Class II | BZG | Anaesthesiology |
Substantial Equivalence
| Manufacturer | Trade Name | Regulation & Product<br>Code | 510(k) Number |
|--------------------|----------------------------------------------|---------------------------------|---------------|
| Current Health Ltd | Current Wearable Health<br>Monitoring System | MSX; FLL; DQA; BZQ;<br>DRG; BZG | K222550 |
Referenced Device
| Manufacturer | Trade Name | Regulation & Product<br>Code | 510(k) Number |
|--------------|----------------------------------------------------------|-------------------------------------------------------------|---------------|
| Draeger | Infinity Acute Care System<br>Monitoring Solution (IACS) | MHX; BZQ; CCK; DQA,<br>DRT; DSK; DXN; FLL; FLS;<br>MLD; MSX | K113798 |
# Submission Description
This Special 510(k) covers an extension of the age range to 14 years old and above patients on the software platform and G2 wearable of the Current Health System. This was also presented in the Q-Submission, Q212340.
There are no significant changes presented to the other software components previously cleared in K191272 – specifically, there is no change to the display of data from the wearable or how alarms are presented/notified to the healthcare professional. There are no changes to the wearable hardware component, as cleared in 510(k) K210133. There are no changes to the alarms system, as cleared in 510(k) K222550. Well-established methods have been used to evaluate the change and the data to be reviewed is provided in a summary in this submission.
Well-established methods have been used to evaluate the change and the date to be reviewed in a summary in this submission.
# Device Description
#### General Description
The Current Wearable System is a remote patient monitoring system that consists of a monitoring device (the wearable) worn on the upper arm by 14 years old and above patients, a software platform (containing the alarming system) and a user interface to allow presentation of vital signs data both on mobile devices and web-based dashboard. The Current Wearable System is also integrated with third-party devices for displaying and monitoring physiological signs.
The Wearable is intended to continuously monitors physiological vital sign data from the person being monitored and securely transmit the encrypted data via the secure server. The wearable is
{7}------------------------------------------------
urrent health
Special 510(k) Ref: K231506
Section 05 – Special 510(k) Summary
Version: 1.0
intended for use in professional healthcare facilities, such as hospitals or skilled nursing facilities, or the home by trained healthcare professionals.
The healthcare professional can securely access the patient physiological signs remotely via a mobile application or a web-interface which is also intended to provide visual and audible physiologic multi-parameter alarms.
# Intended / Indications for Use
The Current Wearable Health Monitoring System is intended for reusable bedside, mobile and central multiparameter, physiologic monitoring of patients aged 14 years old and above in professional healthcare facilities, such as hospitals or skilled nursing facilities, such as hospitals or their own home. It is intended for monitoring of patients by trained healthcare professionals.
The Current Wearable Health Monitoring System is intended to provide visual and audible physiologic multiparameter alarms. The Current Wearable Health Monitoring System is intended for temperature monitoring where monitoring temperature at the upper arm is clinically indicated.
The Current Wearable Health Monitoring System is intended for continuous monitoring of the following parameters:
- . Pulse rate
- Oxygen saturation
- Temperature
- Movement
The Current Wearable Health Monitoring System is intermittent or spot-check monitoring of the following parameters of:
- . Respiration rate
- Non-invasive blood pressure
- Lung function & spirometry .
- Weight
The Current Wearable Health Monitoring System is not in high-acuity environments, such as ICU or operating rooms.
The Current Wearable Health Monitoring System is not intended for use on acutely ill cardiac patients with the potential to develop life-threatening arrhythmias e.g., very fast atrial fibrillation. For these patients, they should be monitored using a device with continuous ECG. The Current Wearable Health Monitoring System is not a substitute for an ECG monitor.
The Current Wearable Health Monitoring System is not intended for SpO2 monitoring in conditions of high motion or low perfusion.
### Comparison with the Predicate and Previously Cleared Device
The candidate device is substantially equivalent to the predicate, K222550, the Current Health Wearable Monitoring System and a comparison of the key characteristics is summarised in Table 1.
| Characteristic | Current Health Wearable<br>Monitoring System with<br>Modification | Current Health<br>Wearable<br>Monitoring System<br>K222550<br>(Primary Predicate) | Draeger IACS<br>K113798<br>(Referenced<br>Device) | Equivalence<br>with Primary<br>Predicate |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|
| Device Name | Current Health Monitoring<br>System | Current Health<br>Monitoring System | Infinity Acute Care<br>System Monitoring<br>Solution (IACS) | Equivalent |
| Characteristic | Current Health Wearable<br>Monitoring System with<br>Modification | Current Health<br>Wearable<br>Monitoring System<br>K222550<br>(Primary Predicate) | Draeger IACS<br>K113798<br>(Referenced<br>Device) | Equivalence<br>with Primary<br>Predicate |
| Manufacturer | Current Health Ltd | Current Health Ltd | Draeger | Equivalent |
| Device<br>Classification | II | II | II | Equivalent |
| Primary<br>Product Code | MSX | MSX | MHX | Equivalent |
| Secondary<br>Product Code | FLL; DQA; BZQ; DRG | FLL; DQA; BZQ; DRG | BZQ; CCK; DQA,<br>DRT; DSK; DXN;<br>FLL; FLS; MLD; MSX | Equivalent |
| Target<br>Population | 14 years and above | Adult | Adult, Paediatric<br>and Neonatal | Substantially<br>equivalent |
| Indications for<br>Use | The Current Wearable<br>Health Monitoring System is<br>intended for reusable<br>bedside, mobile and central<br>multi-parameter,<br>physiologic monitoring of<br>patients aged 14 years old<br>and above in professional<br>healthcare facilities, such as<br>hospitals or skilled nursing<br>facilities, such as hospitals<br>or skilled nursing facilities,<br>or their own home. It is<br>intended for monitoring of<br>patients by trained<br>healthcare professionals.<br>The Current Wearable<br>Health Monitoring System is<br>intended to provide visual<br>and audible physiologic<br>multi-parameter alarms.<br>The Current Wearable<br>Health Monitoring System is<br>intended for temperature<br>monitoring where<br>monitoring temperature at<br>the upper arm is clinically<br>indicated.<br>The Current Wearable<br>Health Monitoring System is<br>intended for continuous<br>monitoring of the following<br>parameters:<br>● Pulse rate<br> | The Current Wearable<br>Health Monitoring<br>System is intended for<br>reusable bedside,<br>mobile and central<br>multi-parameter,<br>physiologic patient<br>monitoring of adult<br>patients in<br>professional<br>healthcare facilities,<br>such as hospitals or<br>skilled nursing<br>facilities, or their own<br>home. It is intended<br>for monitoring of<br>patients by trained<br>healthcare<br>professionals.<br>The Current Wearable<br>Health Monitoring<br>System is intended to<br>provide visual and<br>audible physiologic<br>multi-parameter<br>alarms. The Current<br>Wearable Health<br>Monitoring System is<br>intended for<br>temperature<br>monitoring where<br>monitoring<br>temperature at the<br>upper arm is clinically<br>indicated. | The IACS<br>Monitoring Solution<br>is a combination of<br>two devices; Infinity<br>M540 patient<br>monitor with<br>Infinity M500<br>docking station<br>integrated with the<br>Infinity C500<br>Medical Cockpit or<br>optional C700<br>(larger screen size)<br>Medical Cockpit and<br>respective software.<br>The IACS is<br>intended for multi-<br>parameter,<br>physiologic patient<br>monitoring of adult,<br>pediatric and<br>neonatal patients in<br>environments<br>where patient care<br>is provided by<br>trained healthcare<br>profess sionals. The<br>IACS obtains the<br>physiologic, multi-<br>parameter data<br>from the<br>connection to the<br>M540 monitor and<br>optional medical<br>devices and | Substantially<br>Equivalent |
{8}------------------------------------------------
current health
Special 510(k) Ref: K231506
# Section 05 – Special 510(k) Summary
{9}------------------------------------------------
| ) { current health |
|--------------------|
| |
Special 510(k) Ref: K231506
# Section 05 – Special 510(k) Summary
| Characteristic | Current Health Wearable<br>Monitoring System with<br>Modification | Current Health<br>Wearable<br>Monitoring System<br>K222550<br>(Primary Predicate) | Draeger IACS<br>K113798<br>(Referenced<br>Device) | Equivalence<br>with Primary<br>Predicate |
|----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|
| | Oxygen saturation Temperature Movement The Current Wearable<br>Health Monitoring System is<br>intended for intermittent or<br>spot-check monitoring of<br>the following parameters of: Respiration rate Non-invasive blood<br>pressure Lung function &<br>spirometry Weight The Current Wearable<br>Health Monitoring System is<br>not intended for use in high-<br>acuity environments, such<br>as ICU or operating rooms.<br>The Current Wearable<br>Health Monitoring System is<br>not intended for use on<br>acutely ill cardiac patients<br>with the potential to<br>develop life-threatening<br>arrhythmias e.g., very fast<br>atrial fibrillation. For these<br>patients, they should be<br>monitored using a device<br>with continuous ECG. The<br>Current Wearable Health<br>Monitoring System is not a<br>substitute for an ECG<br>monitor.<br>The Current Wearable<br>Health Monitoring System is<br>not intended for SpO2<br>monitoring in conditions of<br>high motion or low<br>perfusion. | The Current Wearable<br>Health Monitoring<br>System is intended for<br>continuous<br>monitoring of the<br>following parameters<br>in adults:<br>Pulse rate Oxygen<br>saturation Temperature Movement The Current Wearable<br>Health Monitoring<br>System is intended for<br>intermittent or spot-<br>check monitoring of<br>the following<br>parameters in adults,<br>of: Respiration<br>rate Non-invasive<br>blood<br>pressure Lung<br>function &<br>spirometry Weight in<br>adults The Current Wearable<br>Health Monitoring<br>System is not<br>intended for use in<br>high-acuity<br>environments, such as<br>ICU or operating<br>rooms.<br>The Current Wearable<br>Health Monitoring<br>System is not<br>intended for use on<br>acutely ill cardiac<br>patients with the<br>potential to develop<br>life-threatening<br>arrhythmias e.g., very | displays. The<br>transfer of this data<br>is accomplished by<br>the Infinity<br>network.<br>The Infinity M540 is<br>intended for the<br>monitoring of multi-<br>parameter,<br>physiological<br>patient information<br>obtained from<br>connected<br>hardware in<br>environments<br>where patient care<br>is provided by<br>trained healthcare<br>professional. The<br>M540 is intended to<br>monitor one patient<br>at a time. | |
{10}------------------------------------------------
Special 510(k) Ref: K231506
Section 05 – Special 510(k) Summary
| Characteristic | Current Health Wearable<br>Monitoring System with<br>Modification | Current Health<br>Wearable<br>Monitoring System<br>K222550<br>(Primary Predicate) | Draeger IACS<br>K113798<br>(Referenced<br>Device) | Equivalence<br>with Primary<br>Predicate |
|-----------------------------------------|---------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|
| | | fast atrial fibrillation.<br>For these patients,<br>they should be<br>monitored using a<br>device with<br>continuous ECG. The<br>Current Wearable<br>Health Monitoring<br>System is not a<br>substitute for an ECG<br>monitor.<br>The Current Wearable<br>Health Monitoring<br>System is not<br>intended for SpO2<br>monitoring in<br>conditions of high<br>motion or low<br>perfusion. | | |
| Intended<br>user/Location | Professional healthcare<br>facilities & home<br>environments | Professional<br>healthcare facilities<br>& home<br>environments | Professional<br>healthcare<br>facilities | Equivalent |
| Site of<br>application | Wearable monitor upper<br>arm with a strap | Wearable monitor<br>upper arm with a<br>strap | Wearable monitor | Equivalent |
| Wearable<br>physiological<br>monitoring | Pulse rate; Oxygen<br>saturation; Temperature;<br>Respiration Rate;<br>Movement | Pulse rate; Oxygen<br>saturation;<br>Temperature;<br>Respiration Rate;<br>Movement | Heart rate;<br>Arrhythmia<br>(adult and<br>pediatric only);<br>12-lead analysis;<br>ST segment<br>analysis<br>including TruST®<br>(adult and<br>pediatric only);<br>12-lead ST<br>segment analysis<br>(adult and<br>pediatric only);<br>Apnea;<br>Respiration rate;<br>Invasive<br>pressure; Non-<br>invasive | Equivalent |
{11}------------------------------------------------
| ) current health |
|------------------|
| |
Special 510(k) Ref: K231506
Section 05 – Special 510(k) Summary
| Characteristic | Current Health Wearable<br>Monitoring System with<br>Modification | Current Health<br>Wearable<br>Monitoring System<br>K222550<br>(Primary Predicate) | Draeger IACS<br>K113798<br>(Referenced<br>Device) | Equivalence<br>with Primary<br>Predicate |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|
| | | | Temperature;<br>Cardiac output;<br>Arterial oxygen<br>saturation (SpO<br>2); Pulse rate;<br>Perfusion Index<br>(PI); Total<br>hemoglobin<br>(SpHb) - adult<br>and pediatric<br>only; Total<br>oxygen content<br>(SpCO) - adult<br>and pediatric<br>only;<br>Methemoglobin<br>saturation<br>(SpMet); Pleth<br>variability index<br>(PVI);<br>Mainstream<br>etCO2 | |
| Instructions of<br>use | Instructions for use<br>included that includes<br>graphical instructions, text<br>and relevant warnings and<br>cautions | Instructions for use<br>included that<br>includes graphical<br>instructions, text and<br>relevant warnings<br>and cautions | Instructions for use<br>included that<br>includes graphical<br>instructions, text<br>and relevant<br>warnings and<br>cautions | Equivalent |
| Sterile | No | No | No | Equivalent |
| Re-usable | Yes | Yes | Yes | Equivalent |
| Generation of<br>Alarms | The alarm system inputs<br>data from the patient facts<br>database (data from the<br>wearable, existing<br>integrated devices and<br>from a broader range of<br>patient observations) to<br>generate appropriate<br>alarms, notifications and<br>quantified notes to the<br>clinical care team | The alarm system<br>inputs data from<br>the patient facts<br>database (data<br>from the wearable,<br>existing<br>integrated devices<br>and from a<br>broader range of<br>patient<br>observations) to<br>generate<br>appropriate alarms,<br>notifications | Arrhythmia<br>detector and alarm<br>(including ST-<br>segment<br>measurement and<br>alarm) | Equivalent |
{12}------------------------------------------------
| Image: Current Health logo | current health |
|----------------------------|----------------|
|----------------------------|----------------|
Special 510(k) Ref: K231506
Section 05 - Special 510(k) Summary
Version: 1.0
| Characteristic | Current Health Wearable<br>Monitoring System with<br>Modification | Current Health<br>Wearable<br>Monitoring System<br>K222550<br>(Primary Predicate) | Draeger IACS<br>K113798<br>(Referenced<br>Device) | Equivalence<br>with Primary<br>Predicate |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|------------------------------------------|
| | | and quantified notes<br>to the clinical<br>care team | | |
| Display of<br>Alarms on User<br>Interface | Alarms are presented<br>through visual and audible<br>notifications on the web<br>dashboard and mobile<br>apps | Alarms are<br>presented through<br>visual and audible<br>notifications on the<br>web dashboard and<br>mobile apps | Alarms are<br>presented through<br>user interface<br>displayable at<br>bedside | Equivalent |
Table 1: Comparison of characteristics between the Modified System and the Predicate Systems.
# Technological Characteristics
The proposed modification to extending the age range to 14 years old and above patients on the Current Health Wearable Monitoring System has identical indications for use, operating principles, performance, and technical specification as the predicate device, the Current Health Wearable Monitoring system.
# Summary of Non-Clinical Tests (Performance data)
The performance of the modification is identical to the predicate and previously cleared device in terms of technical specification and safety.
# Summary of Animal & Clinical Studies
Substantial equivalence is based on an assessment of non-clinical performance data and no animal or clinical performance data is included.
# Conclusion
Based on the information presented in this Special 510(k) submission, the Current Health Wearable Monitoring System with the modified intended patient population, is substantially equivalent to the predicate device (Current Health Wearable Monitoring System) in terms of safety, performance, functionality, and indications for use and is as safe and effective as the predicate device for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.