K210133 · Current Health · MSX · Sep 3, 2021 · Cardiovascular
Device Facts
Record ID
K210133
Device Name
Current Health Monitoring System Gen 2 (G2)
Applicant
Current Health
Product Code
MSX · Cardiovascular
Decision Date
Sep 3, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Indications for Use
The Current Wearable Health Monitoring System is intended for reusable bedside, mobile and central multi-parameter, physiologic patient monitoring of adult patients in professional healthcare facilities, such as hospitals or skilled nursing facilities, or their own home. It is intended for monitoring of patients by trained healthcare professionals. The Current Wearable Health Monitoring System is intended to provide visual and audible physiologic multi-parameter alarms. The Current Wearable Health Monitoring System is intended for temperature monitoring temperature at the upper arm is clinically indicated. The Current Wearable Health Monitoring System is intended for continuous monitoring of the following parameters in adults: Pulse rate, Oxygen saturation, Temperature, Movement. The Current Wearable Health Monitoring System is intermittent or spot-check monitoring, in adults, of: Respiration rate, Non-invasive blood pressure, Lung function & spirometry, Weight. The Current Wearable Health Monitoring System is not intended for use in high-acuity environments, such as ICU or operating rooms. The Current Wearable Health Monitoring System is not intended for use on acutely ill cardiac patients with the potential to develop life threatening arrhythmias e.g. very fast atrial fibrillation. For these patients, they should be monitored using a device with continuous ECG. The Current Wearable Health Monitoring System is not a substitute for an ECG monitor. The Current Wearable Health Monitoring System is not intended for SpO2 monitoring in conditions of high motion or low perfusion.
Device Story
Wearable device worn on upper arm; monitors physiological data (pulse rate, SpO2, temperature, movement, respiration rate) and integrates with peripheral devices (blood pressure, spirometry, weight). Data transmitted via secure server to mobile app/web-dashboard for healthcare professional review. Provides visual/audible alarms. Used in hospitals, skilled nursing facilities, or home by trained professionals. Enables remote patient monitoring; facilitates clinical decision-making through continuous/intermittent vital sign tracking. Benefits include remote oversight of patient status without requiring physical presence.
Clinical Evidence
Bench testing only. No clinical or animal data included. Verification and validation included electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), pulse rate/SpO2 accuracy (ISO 80601-2-61), respiration rate accuracy, temperature accuracy (ISO 80601-2-56), and biocompatibility (ISO 10993-1).
Indicated for adult patients (18+) in professional healthcare facilities or home settings for continuous monitoring of pulse rate, SpO2, temperature, and movement, and intermittent/spot-check monitoring of respiration rate, non-invasive blood pressure, lung function/spirometry, and weight. Not for high-acuity environments (ICU/OR) or acutely ill cardiac patients prone to life-threatening arrhythmias. Not for SpO2 monitoring during high motion or low perfusion.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
Predicate Devices
Current Health Wearable Monitoring System (K191272)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 3, 2021
Current Health Phil Bromley VP/OA/RA Playfair House, 6 Broughton Street Lane Edinburgh, EH 1 3LY United Kingdom
Re: K210133
Trade/Device Name: Current Health Monitoring System Gen 2 (G2) Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MSX, FLL, DQA, BZQ, DRG, BZG Dated: May 18, 2021 Received: June 4, 2021
Dear Phil Bromley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
Device Name
Current Health Monitoring System Gen 2
#### Indications for Use (Describe)
The Current Wearable Health Monitoring System is intended for reusable bedside, mobile and central multi-parameter, physiologic patient monitoring of adult patients in professional healthcare facilities, such as hospitals or skilled nursing facilities, or their own home. It is intended for monitoring of patients by trained healthcare professionals.
The Current Wearable Health Monitoring System is intended to provide visual and audible physiologic multi-parameter alarms. The Current Wearable Health Monitoring System is intended for temperature monitoring temperature at the upper arm is clinically indicated.
The Current Wearable Health Monitoring System is intended for continuous monitoring of the following parameters in adults:
- · Pulse rate
- · Oxygen saturation
- · Temperature
- · Movement
The Current Wearable Health Monitoring System is intermittent or spot-check monitoring, in adults, of:
- · Respiration rate
- · Non-invasive blood pressure
- · Lung function & spirometry
- · Weight
The Current Wearable Health Monitoring System is not intended for use in high-acuity environments, such as ICU or operating rooms.
The Current Wearable Health Monitoring System is not intended for use on acutely ill cardiac patients with the potential to develop life threatening arrhythmias e.g. very fast atrial fibrillation. For these patients, they should be monitored using a device with continuous ECG. The Current Wearable Health Monitoring System is not a substitute for an ECG monitor.
The Current Wearable Health Monitoring System is not intended for SpO2 monitoring in conditions of high motion or low perfusion.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Version: 2
510(k) Summary
# K210133 510(k) Summary
### Summary of Safety and Effectiveness
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92
| Submitters Information: | |
|------------------------------------|-------------------------------------------------------------------------------------|
| Name: | Current Health Ltd. |
| Address: | Playfair House<br>6A Boughton Street Lane<br>Edinburgh<br>EH1 3LY<br>United Kingdom |
| Establishment Registration Number: | 3015134004 |
| Owner/Operator Number: | 10059040 |
| Phone: | +44 (0) 131 285 8101 |
| Contact: | Phil Bromley<br>VP of QA/RA |
| E-mail: | phil.bromley@currenthealth.com |
| Date of Summary: | 26-Apr-21 |
### Device Information:
Below summarises the Device Classification information regarding the Current Wearable Health Monitoring System, Gen 2, Device.
| Device Proprietary Name | Second Generation (G2) Wearable |
|-------------------------|-------------------------------------------|
| Common Name: | Remote Patient Monitor |
| Trade Name: | Current Wearable Health Monitoring System |
| Product Code(s): | MSX; FLL; DQA; BZQ; DRG; BZG (see below) |
Primary Product Code
| Regulation<br>Number (21 CFR) | Device | Product Class | Product Code | Classification Panel |
|-------------------------------|-----------------------------------------------------------------|---------------|--------------|----------------------|
| 870.2300 | System, Network and<br>Communication,<br>Physiological Monitors | Class II | MSX | Cardiovascular |
Secondary Product Codes
| Regulation<br>Number (21 CFR) | Device | Device Class | Product Code | Classification Panel |
|-------------------------------|--------|--------------|--------------|----------------------|
|-------------------------------|--------|--------------|--------------|----------------------|
{5}------------------------------------------------
Version: 2
K210133
#### 510(k) Summary
| 880.2910 | Thermometer,<br>Electronic, Clinical | Class II | FLL | General Hospital |
|----------|------------------------------------------------------------------------|----------|-----|------------------|
| 870.2700 | Oximeter | Class II | DQA | Cardiovascular |
| 868.2375 | Monitor, Breathing<br>Frequency | Class II | BZQ | Anaesthesiology |
| 870.2910 | Transmitters and<br>Receivers, Physiological<br>Signal, Radiofrequency | Class II | DRG | Cardiovascular |
| 686.1840 | Spirometer, Diagnostic | Class II | BZG | Anaesthesiology |
#### Substantial Equivalence
| Manufacturer | Trade Name | Regulation & Product Code | 510(k) Number |
|--------------------|----------------------------------------------|---------------------------------|---------------|
| Current Health Ltd | Current Wearable Health<br>Monitoring System | MSX; FLL; DQA; BZQ;<br>DRG; BZG | K191272 |
### Device Description
This 510(k) for the introduction of the Second Generation (Gen 2) wearable is component of the Current Wearable Health Monitoring System. There is no substantial change to the healthcare professional monitoring applications (mobile apps and web-dashboard) and data infrastructure (secure server) with this 510(k) submission.
#### General Description
The Current Wearable Health Monitoring System is a remote patient monitoring system that consists of a single monitoring device (the wearable) worn on the upper arm by adult patients (aged 18 years old and over), a software platform (containing an alarming system) and a user interface to allow presentation of vital signs data both on mobile devices and a central station. The Current Wearable Health Monitoring System is also integrated with specific devices for monitoring of blood pressure, spirometry, lung function, and weight.
The Wearable is intended to continuously monitors physiological vital sign data from the person being monitored and securely transmit the encrypted data via the secure server. The wearable is intended for use in professional healthcare facilities, such as hospitals or skilled nursing facilities, or the home by trained healthcare professionals.
The healthcare professional can securely access the patient physiological signs remotely via a mobile application or a web-interface which is also intended to provide visual and audible physiologic multi-parameter alarms.
It is intended to continuously monitor pulse rate (PR), oxygen saturation (SpO2), temperature (TEMP) and movement (MOVEMENT). Current is intended for intermittent or spot-checking monitoring of respiration rate (RESP), blood pressure (BP), spirometry and lung function, and weight (WEIGHT).
In the home environment, the patient will have responsibility for applying the device to their arm, charging the device, and plugging in the Homehub to mains power. The data will still be made directly available to healthcare professionals. These healthcare professionals will be at a remote location e.g., an office or within the hospital or could be with the patient in their own home.
### Intended/ Indications for Use
The Current Wearable Health Monitoring System is intended for reusable mobile and central multiparameter, physiologic patient monitoring of adult patients in professional healthcare facilities, such as
{6}------------------------------------------------
hospitals or skilled nursing facilities, or their own home. It is intended for monitoring of patients by trained healthcare professionals.
The Current Wearable Health Monitoring System is intended to provide visual and audible physiologic multiparameter alarms. The Current Wearable Health Monitoring System is intended for temperature monitoring where monitoring temperature at the upper arm is clinically indicated.
The Current Wearable Health Monitoring System is intended for continuous monitoring of the following parameters in adults:
- Pulse rate
- Oxygen saturation
- Temperature
- Movement
The Current Wearable Health Monitoring System is intermittent or spot-check monitoring of the following parameters in adults, of:
- . Respiration rate
- . Non-invasive blood pressure
- Lung function & spirometry ●
- Weight in adults
The Current Wearable Health Monitoring System is not in high-acuity environments, such as ICU or operating rooms.
The Current Wearable Health Monitoring System is not intended for use on acutely ill cardiac patients with the potential to develop life-threatening arrhythmias e.g., very fast atrial fibrillation. For these patients, they should be monitored using a device with continuous ECG. The Current Wearable Health Monitoring System is not a substitute for an ECG monitor.
The Current Wearable Health Monitoring System is not intended for SpO2 monitoring in conditions of high motion or low perfusion.
### Substantial Equivalence
The candidate device is substantially equivalent to the predicate, K191272, the Current Health Wearable Monitoring System (Gen 1) and a comparison of the key characteristics is summarised in Table 1.
| Characteristic | Current Health Gen 2 Wearable<br>(Candidate) | Current Health Wearable<br>Monitoring System K191272<br>(Predicate( | Equivalence |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Device Name | Current Health Gen 2 | Current Health Monitoring System | n/a |
| Manufacturer | Current Health Ltd | Current Health Ltd | n/a |
| Device Classification | II | II | Equivalent |
| Primary Product Code | MSX | MSX | Equivalent |
| Secondary Product<br>Code | FLL; DQA; BZQ; DRG | FLL; DQA; BZQ; DRG | Equivalent |
| Indications for Use | The Current Wearable Health<br>Monitoring System is intended for<br>reusable bedside, mobile and central<br>multi-parameter, physiologic patient<br>monitoring of adult patients in<br>professional healthcare facilities, such<br>as hospitals or skilled nursing facilities, | The Current Wearable Health<br>Monitoring System is intended for<br>reusable bedside, mobile and central<br>multi-parameter, physiologic patient<br>monitoring of adult patients in<br>professional healthcare facilities, such<br>as hospitals or skilled nursing facilities, | Equivalent |
| Characteristic | Current Health Gen 2 Wearable<br>(Candidate) | Current Health Wearable<br>Monitoring System K191272<br>(Predicate) | Equivalence |
| | or their own home. It is intended for<br>monitoring of patients by trained<br>healthcare professionals. | or their own home. It is intended for<br>monitoring of patients by trained<br>healthcare professionals. | |
| | The Current Wearable Health<br>Monitoring System is intended to<br>provide visual and audible physiologic<br>multi-parameter alarms. The Current<br>Wearable Health Monitoring System is<br>intended for temperature monitoring<br>where monitoring temperature at the<br>upper arm is clinically indicated. | The Current Wearable Health<br>Monitoring System is intended to<br>provide visual and audible physiologic<br>multi-parameter alarms. The Current<br>Wearable Health Monitoring System is<br>intended for temperature monitoring<br>where monitoring temperature at the<br>upper arm is clinically indicated. | |
| | The Current Wearable Health<br>Monitoring System is intended for<br>continuous monitoring of the following<br>parameters in adults: | The Current Wearable Health<br>Monitoring System is intended for<br>continuous monitoring of the following<br>parameters in adults: | |
| | • Pulse rate<br>• Oxygen saturation<br>• Temperature<br>• Movement | • Pulse rate<br>• Oxygen saturation<br>• Temperature<br>• Movement | |
| | The Current Wearable Health<br>Monitoring System is intended for<br>intermittent or spot-check monitoring<br>of the following parameters in adults,<br>of: | The Current Wearable Health<br>Monitoring System is intended for<br>intermittent or spot-check monitoring<br>of the following parameters in adults,<br>of: | |
| | • Respiration rate<br>• Non-invasive blood pressure<br>• Lung function & spirometry<br>• Weight in adults | • Respiration rate<br>• Non-invasive blood pressure<br>• Lung function & spirometry<br>• Weight in adults | |
| | The Current Wearable Health<br>Monitoring System is not intended for<br>use in high-acuity environments, such<br>as ICU or operating rooms. | The Current Wearable Health<br>Monitoring System is not intended for<br>use in high-acuity environments, such<br>as ICU or operating rooms. | |
| | The Current Wearable Health<br>Monitoring System is not intended for<br>use on acutely ill cardiac patients with<br>the potential to develop life-<br>threatening arrhythmias e.g., very fast<br>atrial fibrillation. For these patients,<br>they should be monitored using a<br>device with continuous ECG. The<br>Current Wearable Health Monitoring<br>System is not a substitute for an ECG<br>monitor. | The Current Wearable Health<br>Monitoring System is not intended for<br>use on acutely ill cardiac patients with<br>the potential to develop life-<br>threatening arrhythmias e.g., very fast<br>atrial fibrillation. For these patients,<br>they should be monitored using a<br>device with continuous ECG. The<br>Current Wearable Health Monitoring<br>System is not a substitute for an ECG<br>monitor. | |
| | The Current Wearable Health<br>Monitoring System is not intended for<br>SpO2 monitoring in conditions of high<br>motion or low perfusion. | The Current Wearable Health<br>Monitoring System is not intended for<br>SpO2 monitoring in conditions of high<br>motion or low perfusion. | |
| Characteristic | Current Health Gen 2 Wearable<br>(Candidate) | Current Health Wearable<br>Monitoring System K191272<br>(Predicate( | Equivalence |
| Intended user/Location | Professional healthcare facilities &<br>home environments | Professional healthcare facilities &<br>home environments | Equivalent |
| Site of application | Upper arm with a strap | Upper arm with a strap | Equivalent |
| Wearable physiological<br>monitoring | Pulse rate; Oxygen saturation;<br>Temperature; Respiration Rate;<br>Movement | Pulse rate; Oxygen saturation;<br>Temperature; Respiration Rate;<br>Movement | Equivalent |
| Instructions of use | Instructions for use included that<br>includes graphical instructions, text<br>and relevant warnings and cautions | Instructions for use included that<br>includes graphical instructions, text<br>and relevant warnings and cautions | Equivalent |
| Sterile | No | No | Equivalent |
| Re-usable | Yes | Yes | Equivalent |
| Materials | Plastic case and ISO 10993-1:2018<br>compliant | Plastic case and ISO 10993-1:2018<br>compliant | Equivalent |
| Standards Applied | IEC 60601-1; IEC 60601-1-2; IEC<br>62304; IEC 80601-2-61; IEC 80601-<br>2-56 | IEC 60601-1; IEC 60601-1-2; IEC<br>62304; IEC 80601-2-61; IEC 80601-<br>2-56 | Equivalent |
{7}------------------------------------------------
current health
510(k) Submissions - Gen 2
Version: 2
K210133
510(k) Summary
{8}------------------------------------------------
rrent health
510(k) Submissions - Gen 2
Version: 2
K210133
### 510(k) Summary
Table 1: Comparison of characteristics between Gen 2 and Gen 1 Wearable Devices
# Technological Characteristics
The Gen 2 wearable device has identical indications for use, operating principles, performance, and technical specification as the predicate device, the Current Health Wearable Monitoring system (Gen 1) wearable.
The PPG sensor in the Gen 2 wearable device is identical to the sensor used in the Gen 1 wearable for equivalent sensing of oxygen saturation and pulse rate and it is used in the same way. The Gen 2 wearable has different analogue front-end electronics and sampling rate to the Gen 1 wearable however this has been demonstrated to be equivalent through the performance testing presented in this 510(k). Temperature sensing in Gen 2 uses a different thermocouple which has been demonstrated to have equivalent performance to the Gen 1 temperature sensor in terms of accuracy and tolerance. Reparatory rate for Gen 2 is measured using the same technical method with different motion sensors and these have been demonstrated to have equivalent performance during performance testing.
### Summary of Non-Clinical Tests (Performance data)
The performance of the Gen 2 wearable is identical to the predicate device in terms of technical specification and safety. The primary differences are that the Gen 2 is smaller than the Gen 1 with the ability to charge the device on-arm while in use.
Verification and validation activities established the safety and performance of the Gen 2 based on the following tests, which were all passed:
Electrical Safety: The Current Health Gen 2 was tested to confirm that it met the applicable standards for electrical safety (IEC 60601-1)
Electromagnetic compatibility: The current Health Gen 2 was tested to confirm it meets the applicable standards for electromagnetic compatibility (EMC) (IEC 60601-1-2)
{9}------------------------------------------------
#### Version: 2
### 510(k) Summary
Pulse Rate Testing: Validation of the accuracy of pulse rate monitoring – the Current Health Gen 2 was tested to confirm the accuracy of pulse rate monitoring of the system in accordance with ISO 80601-2-61 and the FDA Pulse Oximeters – Premarket Notification Submissions: Guidance for Industry and FDA Staff. 2007
SpO2 Testing: Validation of the accuracy of SpO2 monitoring – Ensure the accuracy and communication of the SpO2 functions of the Current Health Gen 2 wearable as per ISO80601-2-61 and the FDA SpO2 guidance; Pulse Oximeters-Premarket Notification Submissions Guidance for Industry and Food and Drug Staff, March 4, 2013
Respiratory Rate Testing: Ensure accuracy of the Current Health Gen 2 measurement of respiration rate in comparison to respiration rate measured via end-tidal CO2 in a variety of postures
Temperature Measurement Accuracy: The Current Health Gen 2 was tested to confirm the Temperature Measurement Accuracy of the system in compliance with ISO 80601-2-56
Device Ship/Transport Testing: Ensure device, enclosed in the selected shipping container, meets ASTM D7386 specifications.
Biocompatibility Testing: Testing and analysis of the Current Health Gen 2 has demonstrated compliance to ISO 10993-1: Biological evaluation of medical devices – Guidance
System Verification and Validation Testing: The system verification testing was performed to verify the software and firmware of the Current Health Gen 2. This included testing of integration and interoperability of the peripheral devices for blood pressure and weight.
Software Verification and Validation Testing: Software verification testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, since a failure or latent design flaw could directly result in minor injury to the patient or a failure or latent flaw could indirectly result in minor injury to the patient or operator through incorrect or delayed information or through the action of a care provider.
### Summary of Animal & Clinical Studies
Substantial equivalence is based on an assessment of non-clinical performance data and no animal or clinical performance data is included.
### Conclusion
The Gen 2 wearable device has identical indications for use, operating principles, performance, and technical specification as the Current Health Wearable Monitoring system (Gen 1) wearable.
Based on the information presented in this 510(k) premarket notification, the Current Health Second Generation (Gen 2) wearable device is substantially equivalent to the predicate device (Current Health Wearable Monitoring System (Gen 1)) in terms of safety, performance, functionality and indications for use and is as safe and effective for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.