INFINITY ACUTE CARE SYSTEM, INFINITY MCABLE- MASIMO RAINBOW SET
K113798 · Draeger Medical Systems, Inc. · MHX · Mar 22, 2012 · Cardiovascular
Device Facts
Record ID
K113798
Device Name
INFINITY ACUTE CARE SYSTEM, INFINITY MCABLE- MASIMO RAINBOW SET
Applicant
Draeger Medical Systems, Inc.
Product Code
MHX · Cardiovascular
Decision Date
Mar 22, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The IACS is intended for multi-parameter, physiologic patient monitoring of adult, pediatric and neonatal patients in environments where patient care is provided by trained healthcare professionals. The IACS obtains the physiologic, multi-parameter data from the connection to the M540 monitor and optional medical devices and displays. The transfer of this data is accomplished by the Infinity network. The Infinity M540 is intended for the monitoring of multi-parameter, physiological patient information obtained from connected hardware in environments where patient care is provided by trained healthcare professional. The M540 is intended to monitor one patient at a time.
Device Story
IACS is a multi-parameter physiological monitoring system; comprises M540 patient monitor, M500 docking station, and C500/C700 Medical Cockpits. M540 acquires real-time vital signs via connected sensors; displays data locally; transmits data via Infinity network to Medical Cockpits. Used in clinical environments by trained healthcare professionals. Provides continuous trending, visual/audible alarms for out-of-limit parameters. Masimo Rainbow SET MCable adds non-invasive monitoring of SpO2, PI, pulse rate, SpCO, SpMet, SpHb, PVI, and SpOC. System supports bedside monitoring; M540 functions as portable, handheld monitor. Output allows clinicians to assess patient status, track trends, and respond to physiological changes, facilitating timely clinical intervention.
Clinical Evidence
Bench testing only. No clinical data provided. System tested in accordance with applicable standards (e.g., IEC 60601-1) and internal design control procedures to verify safety and performance.
Technological Characteristics
Multi-parameter physiological monitor; includes M540 handheld unit, docking station, and Medical Cockpit displays. Connectivity via Infinity network. Electrical safety per IEC 60601-1. Software-based monitoring; non-sterile; non-patient contacting (transient contact only).
Indications for Use
Indicated for adult, pediatric, and neonatal patients requiring multi-parameter physiological monitoring in clinical settings. Parameters include heart rate, arrhythmia (adult/pediatric), 12-lead analysis, ST segment/TruST (adult/pediatric), apnea, respiration, invasive/non-invasive pressure, temperature, cardiac output, SpO2, pulse rate, PI, SpHb (adult/pediatric), SpCO (adult/pediatric), SpMet, PVI, and etCO2.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
Infinity® Acute Care System™ (IACS) Monitoring Solution (K093788)
Masimo Radical 7 Pulse CO-Oximeter and Accessories (K110028)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The letters are black and the background is white. The two dots above the "a" are also black.
# 510(k) SUMMARY of SAFETY AND EFFECTIVENESS
MAR 2 2 2012
Draeger Medical Systems, Inc. Submitter : 6 Tech Drive Andover, MA 01810 Tel : (978) 379-8265 Fax: (978) 379-8335 Contact : Beth A. Zis, RAC Director of Requlatory Affairs
February 29, 2012 Date Prepared :
Device Names/Common Names/Classification Names:
Trade Name: Infinity® Acute Care System™ (IACS) Monitoring Solution Physiological Patient Monitor Common Name : Classification Name : Monitor, Physiological, Patient (with arrhythmia detection or Alarms) Product Code : MHX
| Trade Name : | Infinity MCable - Masimo Rainbow SET ® |
|-----------------------|----------------------------------------|
| Common Name : | Pulse CO-Oximeter |
| Classification Name : | Oximeter |
| Product Code : | DQA |
Class : ll
Regulation Numbers : 21 CFR 870.1025, 21 CFR 870.2700
#### Identification of Predicate or Legally Marketed Devices :
The modified devices are substanitally equivalent to the previously cleared :
Infinity® Acute Care System™ (IACS) Monitoring Solution and the Infinity® MCable - Masimo SET® MCable (smart pod) manufactured by Draeger Medical Systems, Inc. cleared under 510(k) #K093788 and #K061329.
Masimo Radical 7 Pulse CO-Oximeter and Accessories manufactured by Masimo Corporation cleared under 510(k) #K110028.
#### Device Description :
The Infinity Acute Care System (IACS) Monitoring Solution is a multi-parameter physiological patient monitoring system that acquires and displays patient data at the bedside. The IACS Monitoring Solution is a combination of two devices ; Infinity M540 patient monitor with Infinity M500 docking station integrated with the Infinity C500 Medical Cockpit or optional C700 (larger screen size) Medical Cockpit and respective software.
The physiological patient data acquired from the interconnected components is displayed on the Infinity M540 patient, and is transmitted via network to the patient bedside Infinity Medical
Page 1 of 3
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font, with two dots above the "a". Below the word "Dräger" is the number "K113798". The number is written in a handwritten style, with some of the digits overlapping.
Cockpit. The Infinity M540 patient monitor is a light weight hand held portable patient mor that displays real-time vital signs, provides continuous trending and produces visual and audible alarms if any of the physiological parameters monitored vary beyond present limits and transmits the data over the network. The Infinity M540 patient monitor can be used as a stand alone medical device without any integration with the Infinity Medical Cockpits similar to other Draeger Medical Pick-And-Go patient monitors.
The Infinity Mcable Masimo Rainbow SET is a new Pulse CO-Oximeter model for use with the IACS Infinity M540 Patient Monitor to provide continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), perfusion index (PI), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin (SpMet), total hemoglobin (SpHb), pleth variability index (PVI) and total oxygen content (SpOC).
### Intended Use: Infinity® Acute Care System™ (IACS) Monitoring Solution:
The IACS is intended for multi-parameter, physiologic patient monitoring of adult, pediatric and neonatal patients in environments where patient care is provided by trained healthcare profess sionals. The IACS obtains the physiologic, multi-parameter data from the connection to the M540 monitor and optional medical devices and displays. The transfer of this data is accomplished by the Infinity network.
### Intended Use: Infinity M540 Patient Monitor:
The Infinity M540 is intended for the monitoring of multi-parameter, physiological patient information obtained from connected hardware in environments where patient care is provided by trained healthcare professional. The M540 is intended to monitor one patient at a time.
Indications for Use: The M540 monitors the following parameters:
- Heart rate
- Arrhythmia (adult and pediatric only)
- 12-lead analvsis
- ST segment analysis including TruST® (adult and pediatric only)
- 12-lead ST seament analysis (adult and pediatric only)
- Apnea
- Respiration rate
- Invasive pressure
- Non-invasive pressure
- Temperature
- Cardiac output (only available when the M540 is docked in an IACS configuration)
- Arterial oxygen saturation (SpO2)
- Pulse rate
- Perfusion Index (PI)
- Total hemoglobin (SpHb) adult and pediatric only
- Total oxygen content (SpCO) adult and pediatric only
- Methemoglobin saturation (SpMet)
- Pleth variability index (PVI)
- Mainstream etCO2
Page 2 of 3
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. Below the word "Dräger" is the number "K113798" in a handwritten style. The word "Dräger" is the name of a German company that makes medical and safety technology products. The number "K113798" may be a serial number or product code.
## MRI Compatibility Statement:
The IACS, Infinity M540 and any connected optional hardware are not intended for use in hyperbaric chambers and environments containing MRI equipment.
## Substantial Equivalence:
The modified Infinity Acute Care System (IACS) Monitoring Solution VG2 software and Infinity MCable - Masimo Rainbow SET have been tested in accordance with applicable standards and internal design con trol procedures and were determined to be as safe and effective for its intended use as the predicate devices.
#### Biocompatibility:
Not applicable – The Infinity Acute Care System Monitoring Solution and its components are not intended to contact the patient. If patient contact is made, it is transient with intact skin.
#### Sterilization:
Not applicable – The Infinity Acute Care System (IACS) Monitoring Solution and its components are not supplied sterile.
## Standards:
Electrical Safety: EC60601-1: Medical electrical equirements for safety and essential performance; and applicable collateral standards.
Page 3 of 3
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" around the perimeter. In the center of the logo is an abstract image of an eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MAR 2 2 2012
Draeger Medical Systems, Inc. c/o Ms. Beth A. Zis Director of Regulatory Affairs 6 Tech Drive Andover, MA 01810
Re: K113798
Trade/Device Name: Infinity acute care system, Infinity Mcable-Masimo rainbow SET Regulation Number: 21 CFR 878.1025
Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm)
Regulatory Class: Class II
Product Code: MHX, MSX, DRT, DQA, BZQ, FLL, DSK, FLS, MLD, DXN, CCK Dated: March 7, 2012 Received: March 8, 2012
Dear Ms. Zis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
# Page 2 - Ms. Beth A. Zis
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bran D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Additional Information for K113798 Infinity® Acute Care System™ Monitoring Solution
Image /page/5/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. Above the "a" are two dots, which is a common feature in the German spelling of the word. Below the word "Dräger" is the number "K113798" written in a handwritten style.
# Indications for Use
510(k) Number (if known): K113798
#### Infinity® Acute Care System™ (IACS) Monitoring Solution Device Name:
Intended Use: Infinity® Acute Care System™ (IACS) Monitoring Solution:
The IACS is intended for multi-parameter, physiologic patient monitoring of adult, pediatric and neonatal patients in environments where patient care is provided by trained healthcare professionals. The IACS obtains the physiologic, multi-parameter data from the connection to the M540 monitor and optional medical devices and displays. The transfer of this data is accomplished by the Infinity network.
Intended Use: Infinity M540 Patient Monitor:
The Infinity M540 is intended for the monitoring of multi-parameter, physiological patient information obtained from connected hardware in environments where patient care is provided by trained healthcare professional. The M540 is intended to monitor one patient at a time.
## Indications for Use
The M540 monitors the following parameters:
- Heart rate
- Arrhythmia (adult and pediatric only)
- 12-lead analysis
- ST segment analysis including TruST® (adult and pediatric only)
- 12-lead ST segment analysis (adult and pediatric only)
- Apnea
- Respiration rate
- Invasive pressure
- Non-invasive pressure
- Temperature
- Cardiac output (only available when the M540 is docked in an IACS configuration)
- Arterial oxygen saturation (SpO2)
- Pulse rate
- Perfusion Index (PI)
- Total hemoglobin (SpHb) (adult and pediatric only)
- Total oxygen content (SpCO) (adult and pediatric only)
- Methemoglobin saturation (SpMet)
- Pleth variability index (PVI)
- Mainstream etCO2
Page 1 of 2
{6}------------------------------------------------
Additional Information for K113798 Infinity® Acute Care System™ Monitoring Solution
Image /page/6/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font, with two dots above the "a". Below the word, there is a handwritten number "K113798". The number appears to be a serial number or identification code.
# MRI Compatibility Statement:
The IACS, Infinity M540 and any connected optional hardware are not intended for use in hyperbaric chambers and environments containing MRI equipment.
#### AND/OR Prescription Use × (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular Devices
Division of Cardiovascular Devices
510(k) Number
Page 2 of 2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.