Infinity CentralStation Wide

K151860 · Draeger Medical Systems, Inc. · MHX · Sep 2, 2015 · Cardiovascular

Device Facts

Record IDK151860
Device NameInfinity CentralStation Wide
ApplicantDraeger Medical Systems, Inc.
Product CodeMHX · Cardiovascular
Decision DateSep 2, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1025
Device ClassClass 2
AttributesSoftware as a Medical Device, Pediatric

Intended Use

The Infinity CentralStation (ICS) is intended for use by trained healthcare professionals for the purpose of centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical environment. Centralized monitoring involves the display and management of data from networked patient monitors including the annunciation of visual and audible physiologic parameter alarms at a central monitoring workstation. Infinity CentralStation with Rest ECG is intended for the production and interpretation of diagnostic electrocardiograms for adult and pediatric patients when connected to a monitor with diagnostic 12-Lead ECG monitoring enabled.

Device Story

Infinity CentralStation (ICS) Wide is a centralized monitoring workstation for hospital/clinical use by trained healthcare professionals (physicians, nurses, technicians). It receives real-time multi-parameter physiological data and alarm signals from networked patient monitors, including wireless telemetry. The system displays waveforms, trends, and events; provides alarm annunciation; and supports diagnostic 12-lead ECG production. It functions as an IT platform running medical-grade software. Healthcare providers use the workstation to manage patient data, review trends, and respond to alarms, facilitating clinical decision-making through continuous patient surveillance. Optional features include web-based data review, HIS demographic integration, and ventilator data display via VentCentral.

Clinical Evidence

Validation testing performed in a hospital environment to assess user acceptance, satisfaction, and observations via surveys with pre-determined assessment questions and user ratings. Verification testing confirmed design outputs satisfy design inputs. No clinical outcome data or patient-contact testing required as the device is not used in the patient vicinity.

Technological Characteristics

Commercial-grade computer platform with widescreen/touch-screen displays. Connectivity via Infinity Network. Software-based central monitoring station. Complies with IEC 60601-1-2 (EMC) and IEC 60950-1 (IT equipment safety).

Indications for Use

Indicated for centralized monitoring of adult, pediatric, and neonatal patients in hospital or clinical settings. Includes production and interpretation of diagnostic 12-lead ECGs for adult and pediatric patients when connected to compatible monitors. No contraindications.

Regulatory Classification

Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/0/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three human profiles facing to the right, stacked on top of each other in a flowing, abstract design. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Draeger Medical Systems, Inc. Mr. Thomas Ostrowski Regulatory Affairs Specialist 6 Tech Drive Andover, Massachusetts 01810 Re: K151860 Trade/Device Name: Infinity CentralStation Wide Regulation Number: 21 CFR 870.1025 Regulation Name: Patient Physiological Monitor (with arrhythmia detection or alarms) Regulatory Class: Class II Product Code: MHX Dated: July 6, 2015 Received: July 8, 2015 Dear Mr. Thomas Ostrowski, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR {1}------------------------------------------------ Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Mitchell Stein for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ K151860 510(k) Number (if known): ## Device Name : Infinity CentralStation Wide ### Indications for Use: The Infinity CentralStation (ICS) is intended for use by trained healthcare professionals for the purpose of centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical environment. Centralized monitoring involves the display and management of data from networked patient monitors including the annunciation of visual and audible physiologic parameter alarms at a central monitoring workstation. Infinity CentralStation with Rest ECG is intended for the production and interpretation of diagnostic electrocardiograms for adult and pediatric patients when connected to a monitor with diagnostic 12-Lead ECG monitoring enabled. ### Contraindications : There are no known contraindications. | Prescription Use | <span> ✓ </span> | |----------------------|-----------------------------------------------| | (Per 21 CFR 801.109) | | OR | Over-The-Counter Use | | |----------------------|--| |----------------------|--| (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) {3}------------------------------------------------ ## K151860 ## Page 1 of 5 # 510(k) SUMMARY Draeger Medical Systems, Inc. I. Submitter: 6 Tech Drive Andover, MA 01810-2434 > Tel: (978) 379-8002 Fax: (978) 379-8335 Contact Person: Thomas Ostrowski Regulatory Affairs Specialist E-mail: tom.ostrowski@draeger.com Date Prepared: July 6, 2015 ## II. Device #### Names / Common Names / Classification Names: | Name of Device: | Infinity CentralStation Wide | |----------------------|---------------------------------------------------------------------| | Common Name: | Central Monitoring Station | | Classification Name: | Patient Physiological Monitor (with Arrhythmia detection or alarms) | | Product Code: | 74 MHX | | Regulatory Class: | II | | Regulation Number: | 21 CFR §870.1025 | ### III. Predicate Device: Infinity CentralStation VG1 MS26800 and components / accessories were cleared under K130711 on April 11, 2013. ### IV. Device Description: #### Device Identification: Infinity CentralStation (ICS) Wide (MS26800) consists of medical grade software installed and configured on an information technology platform available in one model with configurable software options and optional touch and/or non-touch widescreen displays. {4}------------------------------------------------ | K151860 | |-------------| | Page 2 of 5 | #### Device Description: The Infinity CentralStation (ICS) Wide with proposed VG2 modifications is a Central monitoring station capable of real-time display and storage of multi-parameter physiological patient data and alarm annunciation for patient monitors including but not limited to ambulatory and non-ambulatory wireless telemetry monitoring. The key device hardware components for ICS Wide consist of the following: - . Commercial grade computer with a central processing unit (CPU) (MS25707) - 22" Widescreen (MS26806) and/or optional Touch-screen (MS26807) displays ● - Keyboard ● - Mouse ● - External speakers ● - Optional hardware components include: ● - o Uninterruptible Power Supply (UPS) (MS23562) - o Laser printer - o CPU Mount (MS18840) - o R50-N strip recorder (5740068) Software for ICS Wide VG2 consists of a base system package which includes the following functionality: - Configurable Main Screen - Support for Wireless Telemetry monitoring - Central Patient Admission (Wireless Telemetry) - -Remote bed view - Alarm annunciation - Continuous 2 hour storage of up to 4 waveforms - 2-hour storage of up to 1000 events per patient - | 72 hour trend storage (graphical and tabular) - Support of 2-channel strip recordings (requires R50-N recorder) - -Support of printed reports (requires network laser printer) - -Support for up to 64 patient monitors (up to 32 main screen plus 32 surveillance) - Infinity® Network connectivity -- Optional features and functionality available separately for use with ICS Wide VG2 include: - ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Second screen - -RAID 1 Database hard drive {5}------------------------------------------------ K151860 Page 3 of 5 - RAID 1 Operating System hard drive | - -Expanded 28, 48, 72, 96 or 120 hour Full/Event Disclosure and Trends - Wireless Telemetry options (ST Analysis, Full arrhythmia, TruST derived 12-lead, — SpO2) - Diagnostic 12-L REST ECG Report (requires compatible 12-lead Draeger bedside) - - Web-based review of stored data (requires Infinity® Symphony) — - -Import of patient demographics from HIS (requires Infinity® Gateway) - Export of Full/Event disclosure data to Innovian (requires Innovian Critical or Perioperative Care) - VentCentral® application for displaying ventilator data received via patient monitors - l Uninterruptable Power Supply (UPS) ### Environment of Use: The device is intended to be used in an environment where patient care is provided by Healthcare Professionals, i.e. physicians, nurses, and technicians, trained on the use of the device, who will determine when use of the device is indicated, based upon their professional assessment of the patient's medical condition. ICS Wide is not intended to be used in the patient environment. ### V. Indications for Use / Intended Use: The Infinity CentralStation (ICS) is intended for use by trained healthcare professionals for the purpose of centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical environment. Centralized monitoring involves the display and management of data from networked patient monitors including the annunciation of visual and audible physiologic parameter alarms at a central monitoring workstation. Infinity CentralStation with Rest ECG is intended for the production and interpretation of diagnostic electrocardiograms for adult and pediatric patients when connected to a monitor with diagnostic 12-Lead ECG monitoring enabled. ### Contraindications: There are no known contraindications for ICS Wide. ### VI. Comparison of Technological Characteristics with the Predicate Device: The primary modification of the Infinity CentralStation Wide VG2 release is a change in the operating system of the device to support enhanced cybersecurity protection. Additional proposed modifications do not significantly alter technological characteristics and do not impact the safety or effectiveness of the device. {6}------------------------------------------------ K151860 Page 4 of 5 ## VII. Performance Data ### Verification Testing: ICS Wide VG2 has been verified through performance / functional testing to confirm that the proposed modifications have been effectively implemented and design outputs satisfy design inputs. The results of verification have been documented and included in the design history file of the device. Verification testing of the proposed modifications supports the claim of substantial equivalence with the predicate device and do not raise any new issues of safety and effectiveness. ### Validation Testing: Validation testing was performed in a hospital environment for the purpose of determining user acceptance, satisfaction and observations. User acceptance criteria was established and assessed based on user feedback obtained in the form of surveys with pre-determined assessment questions and user ratings for the various functionality tested. Results of Validation testing have been documented and support the claim of substantial equivalence and do not raise any new issues of safety and effectiveness. ### Biocompatibility: Not applicable - Infinity CentralStation Wide VG2 is not intended for use in the patient vicinity and does not come into contact with the patient. #### Sterilization: Not applicable - Infinity CentralStation Wide VG2 is not intended for use in the patient vicinity and does not require sterilization for use. #### Standards / Compliance testing: Infinity CentralStation Wide VG2 has been tested and complies with the following standards in support of EMC and electrical safety requirements: {7}------------------------------------------------ Page 5 of 5 | IEC60601-1-2 Edition 3: | | Medical Electrical Equipment - Part 1-2: General | | |---------------------------|-----------------|-----------------------------------------------------|--| | | 2007-03 | Requirements for Basic Safety and Essential | | | | | Performance - Collateral Standard: Electromagnetic | | | | | Compatibility - Requirements and Tests | | | IEC60950-1 | Edition 2: 2005 | Information technology equipment - Safety - Part 1: | | | | + Am 1:2009 + | General requirements | | | | Am 2:2013 | | | ### VIII. Conclusions #### Substantial Equivalence: The Infinity CentralStation Wide VG2 modifications have been tested in accordance with applicable standards and internal design control procedures. The results of verification and validation testing have demonstrated that the Infinity CentralStation Wide VG2 when compared to the predicate device Infinity CentralStation VG1 is as safe and effective as the predicate for its intended use.
Innolitics
510(k) Summary
Decision Summary
Classification Order