K033461 · Candela Corp. · GEX · Jan 16, 2004 · General, Plastic Surgery
Device Facts
Record ID
K033461
Device Name
CANDELA VBEAM PULSE DYE LASER SYSTEM
Applicant
Candela Corp.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jan 16, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Candela Vbeam Pulse Dye Laser is indicated for the following uses in: General Surgery: Photocoagulation of benign cutaneous vascular lesion and benign cutaneous lesions. Dermatology/Plastic Surgery: For treatment of benign cutaneous vascular lesion, such as facial and leg telangiectasia, rosacea, port wine stains, hemangiomas, angioma, spider angioma, Poikiloderma of Civatte, and benign cutaneous lesions, such as warts, scars, striae and psoriasis. and the treatment of wrinkles. New indication : Treatment of Inflammatory Acne Vulgaris Gynecology: Photocoagulation of benign cutaneous lesion and benign vascular lesion in gynecology. Podiatry: Treatment of benign cutaneous lesion, such as warts.
Device Story
595 nm flash-lamp excited pulse dye laser; controlled by embedded microprocessor. Laser energy delivered via optical fiber to handpiece; produces 5, 7, or 10 mm circular beams. Includes Dynamic Cooling Device (DCD) providing cryogen spray burst prior to laser pulse. Used in clinical settings by physicians for photocoagulation of vascular/cutaneous lesions, wrinkle treatment, and inflammatory acne vulgaris. Safety interlock systems protect patients/operators; user-selected parameters via onboard control panel. Device provides therapeutic benefit through targeted photocoagulation of lesions.
Clinical Evidence
No clinical data provided; substantial equivalence based on bench testing and comparison to predicate devices.
Technological Characteristics
595 nm flash-lamp excited pulse dye laser; embedded microprocessor control; optical fiber delivery; 5, 7, 10 mm spot sizes; integrated cryogen-based Dynamic Cooling Device. Conforms to 21 CFR 1040 (Laser Performance Standard), UL 2601 (Electrical Safety), and EN 60601-1-2.
Indications for Use
Indicated for patients requiring treatment of benign cutaneous vascular lesions (telangiectasia, rosacea, port wine stains, hemangiomas, angioma, spider angioma, Poikiloderma of Civatte), benign cutaneous lesions (warts, scars, striae, psoriasis), wrinkles, and inflammatory acne vulgaris.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the Candela logo with the text "Cbeam 510k Summary" at the top right. The date "JAN 1 6 2004" is printed above the Candela logo. The logo consists of a stylized flame-like shape on the left and the word "CANDELA" in bold, sans-serif font on the right.
KD33461 1 of 2
510(k) Summary
### General Information:
This 510(k) is to provide notification of substantial equivalence for the Candela Vbcam Pulse This JTV(x) is to provision is substantially equivalent to previously marketed device intended for photocoagulation of benign cutaneous vascular lesions, such as facial and leg telangiectasia, rosacea, port wine stains, hemangiomas, angioma, spider angioma, Poikiloderma of Civates and benign cutaneous lesions, such as warts, scars, striea and psoriasis. Photocoagulation of benign cutaneous lesion and benign vascular lesion in Gynecology, treatment of wrinkles and the treatment of inflammatory acne vulgaris. There have been no modifications to design to the Vbeam Pulse Dye Laser as cleared under K 021180.
| Submitted by: | Candela Corporation<br>530 Boston Post Road<br>Wayland, MA 01778-1886 |
|--------------------|---------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Lorraine Calzetta |
| Date prepared: | October 10, 2003 |
| Classification: | Class II (21 CFR § 878.4810 Laser Surgical Instrument for use<br>in General and Plastic Surgery and in Dermatology) |
| Common Name: | Dermatology Laser, Vbeam Pulse Dye Laser System |
| Predicate Devices: | Candela Vbeam Pulse Dye Laser (K021180), ICN Photonics<br>NLite (K024189) |
### Description:
The Vbeam is a 595 nm flash-lamp excited pulse dye medical laser, controlled by an embedded microprocessor., to be used for the treatment of benign cutaneous vascular lesions, such as facial and leg telangiectasia, rosacea, port wine stains, hemangiomas, angioma, spider angioma, Poikiloderma of Civatee and benign cutaneous lesions, such as warts, scars, striea and psoriasis. Photocoagulation of benign cutaneous lesion and benign vascular lesion in Gynecology, for the treatment of wrinkles and the treatment of inflammatory acne vulgaris.. The Vbeam laser may be used with the Candela Dynamic Cooling Device, which provides a short burst of cryogen spray prior to firing the laser pulse. The laser output energy is delivered via an optical fiber to a handpiece, which produces circular beams with diameters of 5, 7 or 10 millimeters on the skin. The cryogen, which is housed within the laser enclosure, is delivered via a hose to a nozzle located in the handpiece.
The Candela Vbeam Pulse Dye Laser is equipped with safety interlock systems to protect patients and operators. Users of the device make selections from an onboard control panel to regulate operation during treatment.
8 02
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows a logo with a stylized flame-like shape on the left and the word "CANDELA" on the right. The flame shape is divided into two sections, with the left section being solid black and the right section consisting of horizontal lines. The word "CANDELA" is written in a simple, sans-serif font, with each letter clearly defined. The logo appears to be for a company or organization named Candela.
K033461 2 of 2
### Testing:
As a laser product, the Vbeam Pulsed Dye Laser is required to conform and does conform to the Laser Performance Standard (21 CFR 1040). In addition the Cbeam conforms to the UL 2601 Electrical Safety Standard and with the Harmonized Standard EN 60601-1-2, Part 2 established by and required by the European Community.
#### Summary of Substantial Equivalence:
The Candela Vbeam Laser has the same intended use, utilizes similar operating principles and matches key design aspects, including similar spot size, and maximum delivered power as the predicate device.
On the basis of similarities in methods of assembly, method of operation, and intended uses, Candela Corporation believes that its Candela Vbeam Laser System is substantially equivalent to the predicate device.
8 03
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or a stylized human figure, composed of three curved lines.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 1 6 2004
Ms. Lorraine Calzetta Patrovic Manger, Regulatory Affairs Candela Corporation 530 Boston Post Road Wayland, Massachusetts 01778
Re: K033461
Trade/Device Name: Candela Vbeam Pulsc Dye Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: October 27, 2003 Received: October 30, 2003
Dear Ms. Patrovic:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosule) (c regars) the enactment date of the Medical Device Amendments, or to conniered price to may 20, 2017 - 11:11 accordance with the provisions of the Federal Food, Drug, devices market Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmete Act (710) that do not required to the general controls provisions of the Act. The r ou may, dicrororo, manes are & Act include requirements for annual registration, listing of general controls provisions practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Spccial Controls) or class III (PMA), it If your device to such additional controls. Existing major regulations affecting your device can may be subject to subli additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination docs not mean Flease be advised that 1197 5 resuarce or our device complies with other requirements of the Act that I DT has made a contrimalations administered by other Federal agencies. You must of any I edical statutes and regaranchs, including, but not limited to: registration and listing (21 Comply with an the 7ter 51equirements, costs 801); good manufacturing practice requirements as set CITY art 8077, mooning (21 OS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (Scctions 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
# Page 2 - Ms. Lorraine Calzetta Patrovic
This letter will allow you to begin marketing your device as described in your Scction 510(k) This letter will anow you to obgin maing of substantial equivalence of your device to a legally premarket notification. The Pira mangssification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advise is. Jour 201) 594-4659. Also, please note the regulation entitled, Connect the Office of Comparket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
### 510(k) Number (if known): K033461
# Device Name: Candela Vbeam Pulse Dye Laser System
Indications For Use:
The Candela Vbeam Pulse Dye Laser is indicated for the following uses in:
General Surgery: Photocoagulation of benign cutaneous vascular lesion and benign cutaneous lesions.
Dermatology/Plastic Surgery: For treatment of benign cutaneous vascular lesion, such as facial and leg telangiectasia, rosacea, port wine stains, hemangiomas, angioma, spider angioma, Poikiloderma of Civatte, and benign cutaneous lesions, such as warts, scars, striae and psoriasis. and the treatment of wrinkles.
### New indication : Treatment of Inflammatory Acne Vulgaris
Gynecology: Photocoagulation of benign cutaneous lesion and benign vascular lesion in gynecology.
Podiatry: Treatment of benign cutaneous lesion, such as warts.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Miriam C Provost
(Division Sign-Off) Division of General, Restorative and Neurological Devices
Page 1 of /
K033461 010(k) Number ----------------------------------------------------------------------------------------------------------------------------------------------------------------