DORNIER MEDILA H
K983963 · Dornier Medtech America, Inc. · GEX · Nov 25, 1998 · General, Plastic Surgery
Device Facts
| Record ID | K983963 |
| Device Name | DORNIER MEDILA H |
| Applicant | Dornier Medtech America, Inc. |
| Product Code | GEX · General, Plastic Surgery |
| Decision Date | Nov 25, 1998 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 878.4810 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Dornier Medilas H Laser is intended to be used in cutting, vaporization, ablation, and coagulation of soft tissue in coniunction with endoscopic equipment (including laparoscopes, hysteroscopes, bronchoscopes, gastroscopes, cystoscopes, and colonoscopes), or in incision/excision, vaporization, ablation and coagulation of soft tissue in contact or non-contact open surgery (with or without hand piece). The Dornier Medilas H Laser is indicated for use in medicine and surgery, in the following specialties: Urology, Pulmonology, Arthroscopy, Lithotripsy, Gastroenterology, Gynecology, ENT, and General Surgery.
Device Story
Pulsed Holmium:YAG laser (2080 nm) for soft tissue surgery; utilizes bare quartz glass fibers (400/600 µm) for contact cutting and non-contact coagulation/vaporization. Operated by physicians in clinical/surgical settings. System features microprocessor-controlled operation, photoelectric power meters, and automatic calibration. Safety mechanisms include dual parallel microprocessors monitoring a rotating magnetic shutter; system halts laser production upon fault detection. Output parameters, time functions, and status displayed on graphic panel. Benefits include precise, temperature-controlled tissue interaction. Modified version incorporates new housing and updated 635 nm aiming beam (vs 630 nm predicate).
Clinical Evidence
No clinical data. Compliance demonstrated via bench testing and adherence to performance standards for Class IV light-emitting products (21 C.F.R. § 1040.10, §1040.11) and voluntary standards (IEC-601, IEC 825/VDE 0837/2.86).
Technological Characteristics
Pulsed Ho:YAG laser, 2080 nm wavelength. Quartz glass fibers (400/600 µm). Air-cooled, internal closed-circuit water cooling. Microprocessor-controlled with dual-processor safety monitoring. 635 nm aiming beam. Class IV laser. Complies with IEC-601 and IEC 825.
Indications for Use
Indicated for soft tissue cutting, vaporization, ablation, and coagulation in patients requiring endoscopic or open surgery across urology, pulmonology, arthroscopy, lithotripsy, gastroenterology, gynecology, ENT, and general surgery specialties.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
- Unmodified Dornier Medilas H Laser (K981718)
Related Devices
- K981718 — DORNIER MEDILAS H · Dornier Surgical Products, Inc. · Jul 17, 1998
- K132607 — NS 2500 HOLMIUM LASER SYSTEM · New Star Lasers, Inc. · Mar 21, 2014
- K253951 — Holmium Laser Therapeutic Apparatus (HZ-A); Holmium Laser Therapeutic Apparatus (HZ-B); Holmium Laser Therapeutic Apparatus (HZ-E) · Guangzhou Potent Medical Equipment Joint-Stock Co., Ltd. · Mar 6, 2026
- K061455 — DORNIER MEDILAS H 20 LASER · Dornier Medtech America, Inc. · Aug 1, 2006
- K183563 — Holinwon 30 · Wontech Co., Ltd. · Sep 5, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
**NOV 25 1998**
# 983962
## 510(k) SUMMARY Dornier Surgical Products, Inc's Medilas H Pulsed Laser
Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Walter Payerl President 1155 Roberts Boulevard Kennesaw, GA 30144
Phone: (770) 426-1315 Facsimile: (770) 514-6288
Contact Person: Carol Wernecke Director of Regulatory and Clinical Affairs 1155 Roberts Boulevard Kennesaw, GA 30144
Date Prepared: November 5, 1998
### Name of Device and Name/Address of Sponsor
#### Medilas H Pulsed Holmium YAG Laser
Dornier Surgical Products, Inc. 1155 Roberts Boulevard Kennesaw, GA 30144
### Classification Name
HO: FDA has not specifically classified YAG lasers.
### Predicate Devices
Unmodified Dornier Medilas H Laser
{1}------------------------------------------------
## Intended Use
The Dornier Medilas H Laser is intended to be used in cutting, vaporization, ablation, and coagulation of soft tissue in coniunction with endoscopic equipment (including laparoscopes, hysteroscopes, bronchoscopes, gastroscopes, cystoscopes, and colonoscopes), or in incision/excision, vaporization, ablation and coagulation of soft tissue in contact or non-contact open surgery (with or without hand piece).
The Dornier Medilas H Laser is indicated for use in medicine and surgery, in the following specialties: Urology, Pulmonology, Arthroscopy, Lithotripsy, Gastroenterology, Gynecology, ENT, and General Surgery.
Technological Characteristics and Substantial Equivalence
The Dornier Medilas H Pulsed Holmium YAG Laser is a compact pulsed HO: YAG laser emitting laser radiation in the invisible range of 2080 nm. The Medilas H provides a temperature-controlled method for contact cutting and non-contact coagulation and vaporization with a bare fiber.
The modified Medilas H has the same principles of operation and similar technological characteristics as the previously unmodified Medilas H (K981718).
The modified Medilas H and the unmodified Medilas H laser systems have photoelectric power meters and are automatically calibrated. They also have a 2-stage waterproof and explosion proof footswitch and both lasers incorporate a watchdog-monitored microprocessor.
Both Lasers depict a graphic display panel which show the laser operating parameters, application modes, time functions, system status and messages for the user.
The Medilas H lasers provide a temperature controlled method for contact cutting and non-contact coagulation and vaporization with fibers made of quartz glass with a bare fiber tip. These fibers are available in diameters of 400 and 600 µm.
The Medilas H lasers feature one mode of operation: Standard. The microprocessor controlled Dornier Medilas H lasers have been developed in accordance with the latest technical standards.
The lasers have a cooling system that includes an air-cooled, temperature controlled, internal closed circuit water system. The cooling systems are identical as concurred in 510(k) # K981718.
As a safety feature, both Medilas H lasers have a single rotating magnetic shutter which moves the filter out of the laser beam. One microprocessor controls the shutter. The Medilas H lasers contain a single shutter with two parallel running microprocessors. Whenever one microprocessor runs differently from the other, the Medilas H laser
{2}------------------------------------------------
hardware initiates a "system fault" routine. During any "system fault" laser production and release immediately halt.
There are minor differences between the Medilas H laser, and the unmodified cleared Medilas H laser (K981718). The modified laser system aiming beam is 635 nm, maximum power at aperture 1mW and the unmodified cleared Medilas H laser aiming beam is 630 nm, maximum power at aperture 1mW. The modified Medilas H is contained in new housing. The dimensional specification is visually different from the predicate (unmodified Medilas H) laser system. These modifications do not present any new issues of safety or effectiveness of the device.
## Performance Data
While no performance standard have been established for HO: YAG lasers under Section 514 of the Federal Food, Drug and Cosmetic Act, the Dornier Medilas H laser is in compliance with class IV performance standards for light emitting products promulgated under the Radiation Control Health and Safety Act of 1968. See 21 C.F.R. § 1040.10 and §1040.11. The laser also complies with the applicable requirements of the following voluntary standards: IEC-601, IEC 825/VDE 0837/2.86.
## Advisory:
This information was prepared for the sole purpose of compliance with the Safe Medical Devices Act of 1990. It does not imply that the procedures described herein can be performed with the equipment described without substantial risk of personal injury or death to patients due to operator error or in procedures requiring a high degree of skill.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the seal is an abstract image of an eagle.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 2 5 1998
Ms. Carol Wernecke Director Regulatory and Clinical Affairs Dornier Medical Systems, Inc 1155 Roberts Boulevard Kennesaw, Georgia 30144
Re: K983963
Trade Name: Dornier Medilas H Laser (Modified) Regulatory Class: II Product Code: GEX Dated: November 05, 1998 Received: November 06, 1998
Dear Ms. Wernecke:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
## Page 2 - Ms. Carol Wernecke
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
foeele
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# PREMARKET NOTIFICATION
# INDICATIONS FOR USE STATEMENT
| | K9P 3963 | |
|----------------|---------------------------------------------|--|
| 510(k) Number: | K081718-Special 510(k): Device Modification | |
| Device Name: | Modified Dornier Medilas H Laser | |
Indications for Use:
The Dornier Medilas H Laser is intended to be used in cutting, vaporization, ablation, and coagulation of soft tissue in conjunction with endoscopic equipment (including laparoscopes, hysteroscopes, bronchoscopes, gastroscopes, cystoscopes, and colonoscopes), or in incision/excision, vaporization, ablation and coagulation of soft tissue in contact or non-contact open surgery (with or without hand piece).
The Dornier Medilas H is indicated for use in medicine and surgery, in the following medical specialties:
- ◆ Urology
- ♦ Arthroscopy
- ◆ General Surgery
- ◆ Pulmonology
- · Gynecology
- · Gastroenterology
- ◆ ENT
- ◆ Lithotripsy
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | X |
|------------------|---|
|------------------|---|
or
| Over-the-Counter Use | |
|----------------------|--|
|----------------------|--|
| (Division Sign-Off) | |
|-----------------------------------------|---------|
| Division of General Restorative Devices | |
| 510(k) Number | R983963 |