DCL8 ENHANCED PULSED ND:YAG LASER SYSTEM

K964031 · Slt-J , Ltd. · GEX · Dec 23, 1996 · General, Plastic Surgery

Device Facts

Record IDK964031
Device NameDCL8 ENHANCED PULSED ND:YAG LASER SYSTEM
ApplicantSlt-J , Ltd.
Product CodeGEX · General, Plastic Surgery
Decision DateDec 23, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Intended Use

The DCL8 Enhanced Pulsed Nd:YAG laser system is substantially equivalent to other pulsed Nd:YAG laser systems currently in interstate commerce for use in procedures for incision, excision, coagulation or vaporization of soft tissues, such as: subcutaneous tissues, striated and smooth muscle, mucous membranes, along with organs and glands, in both open, endoscopic and microscopic procedures.

Device Story

Pulsed Nd:YAG laser system for soft tissue surgery; performs incision, excision, coagulation, vaporization. Input: 110 VAC power. Operation: user-controlled via touch pad; red diode aiming beam for targeting. Output: pulsed laser energy delivered via fiberoptic cables. Used in clinical settings (open, endoscopic, microscopic procedures) by physicians. Benefits: precise soft tissue management via established laser technology.

Clinical Evidence

Bench testing only; no clinical data provided. Equivalence established through comparison of operating characteristics and delivery system compatibility with predicate devices.

Technological Characteristics

Pulsed Nd:YAG laser; 110 VAC power source; red diode aiming beam; touch pad control interface; SMA 905 fiberoptic connector system for delivery.

Indications for Use

Indicated for incision, excision, coagulation, or vaporization of soft tissues, including subcutaneous tissues, striated and smooth muscle, mucous membranes, organs, and glands in open, endoscopic, and microscopic procedures.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} k964031 F. Summary of Safety and Effectiveness DEC 23 1996 The DCL8 Enhanced Pulsed Nd:YAG laser system is substantially equivalent to other pulsed Nd:YAG laser systems currently marketed by Sunrise Technologies (Fremont, CA) and American Dental Technologies (Troy, MI) This belief is based upon the following facts: 1. The DCL8 Enhanced Pulsed Nd:YAG laser system is substantially equivalent to other pulsed Nd:YAG laser systems currently in interstate commerce for use in procedures for incision, excision, coagulation or vaporization of soft tissues, such as: subcutaneous tissues, striated and smooth muscle, mucous membranes, along with organs and glands, in both open, endoscopic and microscopic procedures. The delivery systems which can be used in conjunction with the DCL8 Enhanced Pulsed Nd:YAG laser system have previously been cleared for marketing by the FDA. 2. The DCL8 Enhanced Pulsed Nd:YAG laser system has the same operating characteristics as the pulsed Nd:YAG laser systems offered by Sunrise Technologies and American Dental Technologies with respect to average power to tissue, pulse rates, operating controls and indicators. The DCL8 Enhanced Pulsed Nd:YAG laser system incorporates a red diode aiming beam, touch pad controls for setting of operating parameters, and uses the same input power (110 VAC) as the pulsed Nd:YAG laser systems offered by Sunrise Technologies and American Dental Technologies. 3. The DCL8 Enhanced Pulsed Nd:YAG laser system uses the industry standard SMA 905 fiberoptic connector system for its fiberoptic delivery systems. This is identical to the connector systems used by Sunrise Technologies and American Dental Technologies. The delivery systems which will be used with the DCL8 0021-a {1} Enhanced Pulsed Nd:YAG laser system have previously received market clearance for use with Nd:YAG laser systems to which the delivery systems can be attached. 0022-a
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...